K201115 · Acclarent, Inc. · LRC · Aug 27, 2020 · Ear, Nose, Throat
Device Facts
Record ID
K201115
Device Name
Next Generation Balloon Dilation System
Applicant
Acclarent, Inc.
Product Code
LRC · Ear, Nose, Throat
Decision Date
Aug 27, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4420
Device Class
Class 1
Attributes
Therapeutic, Pediatric
Indications for Use
The NGB Balloon Dilation System is intended to provide a means to access the sinus space within and across nasal and sinus structures; dilate the sinus ostia and spaces associated with the paranasal sinus cavities for diagnostic and therapeutic procedures; irrigate from within a target sinus; suction throughout therapeutic procedures and to facilitate diagnostic procedures. For children aged 17 and under, the NGB Balloon Dilation System is intended to provide a means to access the sinus space within and across nasal and sinus structures; dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures; irrigate from within the maxillary sinus; suction throughout therapeutic procedures and to facilitate diagnostic procedures. The NGB Balloon Dilation System is intended to dilate the cartilaginous Eustachian tube for treatment of persistent Eustachian tube dysfunction in patients ages 18 and older. The NGB Balloon Dilation System with the navigation guidewire may be utilized in conjunction with the TruDi™ Navigation System, to help direct access to nasal and paranasal spaces, and to confirm placement in the targeted anatomy. NGB may be utilized in conjunction with the TruDi™ NAV Wire to confirm placement of the balloon in the Eustachian tube. The NGB Balloon Dilation System with the illumination guidewire may be utilized to illuminate within and provide transcutaneous illumination across nasal and sinus structures.
Device Story
Handheld integrated balloon sinuplasty and Eustachian tube dilation device; features adjustable grip, directable guide tip with four preset positions (Sphenoid, Eustachian tube, Frontal, Maxillary), wire slider, spinner, and balloon slider. Inputs include manual manipulation by physician; device supports optional TruDi Navigation System for EM tracking and illumination guidewire for transcutaneous visualization. Device provides mechanical dilation of sinus ostia or Eustachian tube; enables irrigation and suction. Used in clinical settings by physicians to treat sinus/Eustachian tube dysfunction. Benefits include minimally invasive access, real-time placement confirmation, and combined therapeutic/diagnostic capability.
Clinical Evidence
No clinical data provided; substantial equivalence determined via non-clinical bench testing including simulated use, dimensional attributes, cycle fatigue, balloon burst, bond separation, irrigation flow rate, and device accuracy.
Technological Characteristics
Handheld catheter with non-compliant balloon; materials include Pebax, PET, PEEK, stainless steel, polycarbonate, nylon, PTFE, silicone, and nitinol. Features EM navigation and illumination compatibility. Sterilized via Ethylene Oxide (EtO) per ISO 11135-1. Packaging per ASTM F1980-07.
Indications for Use
Indicated for sinus ostia/space dilation and irrigation/suction in paranasal sinus cavities (all ages) and maxillary sinus (children 17 and under); also indicated for dilation of the cartilaginous Eustachian tube in patients 18 and older with persistent Eustachian tube dysfunction.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
Predicate Devices
Acclarent AERA® Eustachian Tube Balloon Dilation System (K171761)
Relieva SpinPlus® Nav Balloon Sinuplasty System (K171687)
Relieva SpinPlus® Balloon Sinuplasty System (K143541)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 27, 2020
Acclarent Inc. David Locke Manager, Regulatory Affairs 31 Technology Dr., Suite 200 Irvine, California 92618
Re: K201115
Trade/Device Name: Next Generation Balloon Dilation System Regulation Number: 21 CFR 874.4180 Regulation Name: Eustachian Tube Balloon Dilation System Regulatory Class: Class II Product Code: PNZ, PGW, LRC Dated: July 25, 2020 Received: July 28, 2020
Dear David Locke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K201115
Device Name Next Generation Balloon Dilation System
#### Indications for Use (Describe)
The NGB Balloon Dilation System is intended to provide a means to access the sinus space within and across nasal and sinus structures; dilate the sinus ostia and spaces associated with the paranasal sinus cavities for diagnoutic procedures; irrigate from within a target sinus; suction throughout therapeutic procedures and to facilitate diagnostic procedures.
For children aged 17 and under, the NGB Balloon System is intended to provide a means to access the sinus space within and across nasal and sinus structures; dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures; irrigate from within the maxillary sinus; suction throughout therapeutic procedures and to facilitate diagnostic procedures.
The NGB Balloon Dilation System is intended to dilate the cartilaginous Eustachian tube for treatment of persistent Eustachian tube dysfunction in patients ages 18 and older.
The NGB Balloon Dilation System with the navigation guidewire may be utilized in conjunction with the TruDi™ Navigation System, to help direct access to nasal and paranasal spaces, and to confirm placement in the targeted anatomy. NGB may be utilized in conjunction with the TruDi™ NAV Wire to confirm placement of the balloon in the Eustachian tube.
The NGB Balloon Dilation System with the illumination guidewire may be utilized to illuminate within and provide transcutaneous illumination across nasal and sinus structures.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image displays the Acclarent logo, with the company name in bold, followed by a curved line above the 'ent'. Below the logo, the text "Next Generation Balloon Dilation System" is present. The text is in a smaller font size compared to the logo.
# K201115 - 510(K) SUMMARY
#### [807.92(a)(1)] Submitter Information
| Applicant: | Acclarent Inc.<br>31 Technology Drive, Suite 200<br>Irvine, CA 92618, USA<br>Tel.: 419-233-2611<br>Fax: 949-450-6886 |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | David Locke<br>Manager, Regulatory Affairs<br>Acclarent Inc.<br>Phone: 419-233-2611<br>Dlocke1 @its.jnj.com |
| Authored by: | David Locke<br>Manager, Regulatory Affairs<br>Acclarent Inc.<br>Phone: 419-233-2611 |
| Date Summary Prepared: | Aug 26, 2020 |
| [807.92(a)(2)] Name of Device | |
| Device Trade Name:<br>Device Common Name:<br>Classification Name: | Next Generation Balloon - Balloon Dilation System<br>Next Generation Balloon Dilation System<br>Eustachian tube balloon dilation system (21 CFR 874.4180) |
| Device Classification: | Class II |
| Product Code:<br>Secondary Codes:<br>Review Panel: | PNZ<br>PGW, LRC<br>Ear, Nose and Throat |
| [807.92(a)(3)] Legally Marketed Devices | |
| Primary Predicate Device: | Acclarent AERA® Eustachian Tube Balloon Dilation System<br>(K171761) |
| Secondary Predicate<br>Devices: | Relieva SpinPlus® Nav Balloon Sinuplasty System (K171687)<br>Relieva SpinPlus® Balloon Sinuplasty System (K143541) |
| Reference Device: | Acclarent TruDi NAV Wire (K190532) |
| Device Description: | The Next Generation Balloon Dilation System (NGB) is an integrated<br>balloon sinuplasty (BSP) and Eustachian tube (ET) dilation device, that<br>will be available in light fiber and navigation guidewire configurations for<br>compatibility with illumination and navigation technology. The system<br>includes a handle with several integrated features to allow for single-<br>handed use. The adjustable gripping feature allows for ergonomic<br>handling. The directable guide tip enables access (i.e.<br>manipulating/separating/dividing tissue) and placement near the targeted<br>anatomy. There are four preset approximate positions: Sphenoid<br>(straight), Eustachian tube (55°), Frontal (70°), Maxillary (110°). The<br>directable guide knob allows for positioning of the balloon tip towards the<br>target anatomy. The wire slider and spinner are used to advance, retract,<br>and rotate the guidewire. The balloon slider is used advance and retract<br>the balloon. The Next Generation Balloon Dilation System also allows for<br>suction and irrigation in and around the target anatomy. |
{4}------------------------------------------------
## [807.92(a)(4)] Device Description
## [807.92(a)(5)] Intended Use and Predicate Device Comparison
| Indications for Use: |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The NGB Balloon Dilation System is intended to provide a means to<br>access the sinus space within and across nasal and sinus structures; dilate<br>the sinus ostia and spaces associated with the paranasal sinus cavities for<br>diagnostic and therapeutic procedures; irrigate from within a target sinus;<br>suction throughout therapeutic procedures and to facilitate diagnostic<br>procedures. |
| For children aged 17 and under, the NGB Balloon Dilation System is<br>intended to provide a means to access the sinus space within and across<br>nasal and sinus structures; dilate sinus ostia and spaces associated with the<br>maxillary sinus for diagnostic and therapeutic procedures; irrigate from<br>within the maxillary sinus; suction throughout therapeutic procedures and<br>to facilitate diagnostic procedures. |
| The NGB Balloon Dilation System is intended to dilate the cartilaginous<br>Eustachian tube for treatment of persistent Eustachian tube dysfunction in<br>patients ages 18 and older. |
| The NGB Balloon Dilation System with the navigation guidewire may be<br>utilized in conjunction with the TruDiTM Navigation System, to help direct<br>access to nasal and paranasal spaces, and to confirm placement in the<br>targeted anatomy. NGB may be utilized in conjunction with the TruDiTM<br>NAV Wire to confirm placement of the balloon in the Eustachian tube. |
| The NGB Balloon Dilation System with the illumination guidewire may<br>be utilized to illuminate within and provide transcutaneous illumination<br>across nasal and sinus structures. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Acclarent" in bold, with a purple arc over the last few letters. Below that, the text "Next Generation Balloon Dilation System" is written in a smaller font. The word "Next" is capitalized, and the rest of the words are in lowercase.
# [807.92(a)(6)] Technical Characteristics
| Technological<br>Characteristics: | The Next Generation Balloon Dilation System combines Sinuplasty and<br>Eustachian Tube Dilation into a single handheld device. The device also<br>allows for the use of the Acclarent TruDiTM NAV Wire for real-time<br>tracking within nasal and Eustachian tube anatomy as well as an<br>illumination wire for transcutaneous illumination across, nasal and sinus<br>structures. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Non-clinical Performance<br>Data: | The Next Generation Balloon was tested to ensure that it functions in<br>accordance with the device design specifications related to substantial<br>equivalence in terms of device safety and effectiveness.<br>The following nonclinical tests were performed:<br><br>1. Bench testing has been performed and met all acceptance criteria for<br>attributes such as simulated use testing, dimensional attributes, cycle<br>fatigue, balloon burst, bond separation, irrigation flow rate and device<br>accuracy. Moreover, testing also showed that the Next Generation<br>Balloon Dilation System is biocompatible.<br><br>2. The sterilization process has been validated per AAMI/ANSI/ISO<br>11135-1: 2007 and demonstrated a sterility assurance level of 10-6.<br>The method used for sterilization validation is the overkill (half-cycle<br>approach) in a fixed chamber. Ethylene oxide residuals have been<br>tested and meet ISO 10993-7:2008 requirements. The subject device<br>is not tested nor labeled as “non-pyrogenic”.<br><br>3. Packaging shelf life has been established to be 3 months per ASTM<br>F1980-07.<br><br>4. Mechanical testing was performed, including tensile and flexural<br>testing of catheter joints and materials.<br><br>5. Durability testing was performed, including fatigue and burst pressure<br>testing of the balloon materials and components.<br><br>6. Inflation and deflation characterization testing was performed,<br>including time and pressure measurements, and leak testing of the<br>balloon.<br><br>7. Verification testing of safety features built into the device was<br>performed, including the characterization of catheter geometries and<br>distal tip insertion limitation mechanisms.<br><br>8. Simulated use testing in a clinically relevant model demonstrated the<br>reliability of the device to remain mechanically functional throughout<br>the anticipated conditions of use, and that the design features limit<br>access to only the cartilaginous portion of the Eustachian tube.<br><br>9. The Next Generation Balloon Dilation System passed all tests in<br>accordance with appropriate test criteria and standards, and the<br>modified device did not raise new questions of safety or effectiveness. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Acclarent logo, with the company name in bold black letters and a purple curved line above the "ent" portion of the name. Below the logo, the text "Next Generation Balloon Dilation System" is displayed in a smaller, non-bold font. The text is left-aligned and positioned directly beneath the Acclarent logo.
| Clinical Data: | Clinical data was not necessary to determine that the subject Next<br>Generation Balloon Dilation System performs as intended. |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | The modified Next Generation Balloon Dilation System device is<br>substantially equivalent to the currently cleared Acclarent AERA®<br>Eustachian Tube Balloon Dilation System and the secondary predicate<br>devices based on the completion of non-clinical bench testing as well as<br>similar principles of design, operation and indications for use. |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Acclarent" in black font. The "ent" portion of the word is in purple font. There is a purple arc above the "ent" portion of the word.
# Substantial Equivalence Table
| Attribute | Primary Predicate<br>Device:<br>ACCLARENT<br>AERA®<br>Eustachian Tube<br>Balloon Dilation<br>System | Secondary<br>Predicate<br>Device:<br>RELIEVA<br>SPINPLUS®<br>NAV Balloon<br>Sinuplasty<br>System | Secondary<br>Predicate<br>Device:<br>RELIEVA<br>SPINPLUS®<br>Balloon<br>Sinuplasty<br>System | Reference<br>Device:<br>TruDi™ NAV<br>Wire® | Subject<br>Device:<br>Next<br>Generation<br>Balloon<br>Dilation<br>System | Substantial<br>Equivalence<br>Rationale |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>number | K171761 | K171687 | K143541 | K190532 | K201115 | N/A |
| Manufacturer | Acclarent | Acclarent | Acclarent | Acclarent | Acclarent | N/A |
| Trade Name | ACCLARENT<br>AERA®<br>Eustachian<br>Balloon<br>Dilation<br>System | RELIEVA<br>SPINPLUS®<br>NAV<br>Balloon<br>Sinuplasty<br>System | Relieva<br>SpinPlus<br>Balloon<br>Sinuplasty<br>System | TruDi™ NAV Wire | Next<br>Generation<br>Balloon<br>Dilation<br>System | N/A |
| Common<br>Name | Eustachian<br>Tube Balloon<br>Dilation<br>Device | Sinus<br>Dilation<br>System | Sinus<br>Dilation<br>System | Image Guided<br>Surgery<br>System | Eustachian<br>Tube<br>Balloon<br>Dilation<br>Device | N/A |
| Class | II | I | I | II | II | Same as the<br>higher<br>classification<br>devices |
| Attribute | Primary Predicate<br>Device:<br>ACCLARENT<br>AERA®<br>Eustachian Tube<br>Balloon Dilation<br>System | Secondary<br>Predicate<br>Device:<br>RELIEVA<br>SPINPLUS®<br>NAV Balloon<br>Sinuplasty<br>System | Secondary<br>Predicate<br>Device:<br>RELIEVA<br>SPINPLUS®<br>Balloon<br>Sinuplasty<br>System | Reference<br>Device:<br>TruDi™ NAV<br>Wire® | Subject<br>Device:<br>Next<br>Generation<br>Balloon<br>Dilation<br>System | Substantial<br>Equivalence<br>Rationale |
| Classification<br>Product Code | PNZ | LRC | LRC | PGW | PNZ | Same as<br>primary<br>predicate<br>device |
| Subsequent<br>Product Code | N/A | PGW | N/A | N/A | PGW,<br>LRC | N/A |
| Classification<br>Section | 874.4180 | 874.4420 | 874.4420 | 882.4560 | 874.4180 | Same as<br>primary<br>predicate<br>device |
| Indications<br>for Use | The ACCLARENT<br>AERA® Eustachian<br>Tube Balloon Dilation<br>System is intended to<br>dilate the Eustachian<br>tube for treatment of<br>persistent Eustachian<br>tube dysfunction in<br>patients ages 18 and<br>older. | The RELIEVA<br>SPINPLUS® NAV<br>Balloon Sinuplasty<br>System is intended to<br>provide a means to<br>access the sinus space,<br>within and across<br>nasal and sinus<br>structures; dilate the<br>sinus ostia and spaces<br>associated with the<br>paranasal sinus<br>cavities for diagnostic<br>and therapeutic<br>procedures; and<br>irrigate from within a<br>target sinus for<br>therapeu…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.