Performance testing using test datasets of representative clinical exams
>1 (NM Physicists and Physicians)
Kidney segmentation
Deep Learning based algorithm
—
—
—
—
Performance testing using test datasets of representative clinical exams
>1 (NM Physicists and Physicians)
Indications for Use
The system is intended for use by Nuclear Medicine (NM) or Radiology practitioners and referring physicians for display, processing, archiving, printing, reporting and networking of NMI data, including planar scans (Static, Whole Body, Dynamic, Multi-Gated) and tomographic scans (SPECT, Gated SPECT, dedicated PET or Camera-Based-PET) acquired by gamma cameras or PET scanners. The system can run on dedicated workstation or in a server-client configuration. The NM or PET data can be coupled with registered and/or fused CT or MR scans, and with physiological signals in order to depict, localize, and/or quantify the distribution of radionuclical structures in scanned body tissue for clinical diagnostic purposes.
Device Story
Xeleris V is a software-based Nuclear Medicine processing/review system. Inputs include planar/tomographic NM data (SPECT, PET), registered CT/MR scans, and physiological signals. The system performs image reconstruction, reorientation, registration, segmentation, and quantification. It operates on dedicated workstations or server-client configurations. Physicians use the output for clinical diagnosis, treatment planning (e.g., SIRT), and monitoring disease progression. New features include DL-based lung fissure and kidney segmentation, and an ARCS algorithm for heart detection. The device aids in interpreting cognitive complaints, neuro-degenerative diseases, and lung/cardiac function, potentially improving diagnostic accuracy and workflow efficiency.
Clinical Evidence
No clinical studies required. Performance testing for DL-based algorithms (kidney/lung fissure segmentation) used representative clinical exam datasets. Ground truth established by manual segmentation from experienced NM Physicists/Physicians. Evaluation performed by clinicians using a 5-point Likert scale. Bench testing included verification of software, integration, and simulated use.
Technological Characteristics
Software-based system for NM image processing. Connectivity: standalone workstation or server-client. Standards: NEMA PS3.1-3.20, IEC62304, 21 CFR 820, ISO 13485. Algorithms: Rule-based and Deep Learning-based models for segmentation and detection. Deployment: workstation or remote server.
Indications for Use
Indicated for NM/Radiology practitioners and referring physicians to process/review NMI data (planar/tomographic) from gamma/PET scanners. Includes optional applications for: DaTscan visual/quantitative evaluation (Parkinson's); Q.Lung (PE, COPD, lung function assessment); Q.Brain (epilepsy, dementia, inflammation, brain death, stroke, TBI); Alcyone CFR (coronary vascular function/CAD); Exini Bone (prostate cancer bone metastases); Q.Liver (SIRT treatment planning).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Xeleris 4.0 Processing and Review Workstation (K153355)
{0}------------------------------------------------
September 25, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of a human figure, while the right side features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
GE Healthcare % Alexandra Lifshits Regulatory Affairs Program Manager 4 Hayozma Street Tirat Hacarmel, 30200 ISRAEL
Re: K201103
Trade/Device Name: Xeleris V Processing and Review Systems Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: August 13, 2020 Received: August 18, 2020
Dear Alexandra Lifshits:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K201103
Device Name Xeleris V Processing and Review System
#### Indications for Use (Describe)
The system is intended for use by Nuclear Medicine (NM) or Radiology practitioners and referring physicians for display, processing, archiving, printing, reporting and networking of NMI data, including planar scans (Static, Whole Body, Dynamic, Multi-Gated) and tomographic scans (SPECT, Gated SPECT, dedicated PET or Camera-Based-PET) acquired by gamma cameras or PET scanners. The system can run on dedicated workstation or in a server-client configuration. The NM or PET data can be coupled with registered and/or fused CT or MR scans, and with physiological signals in order to depict, localize, and/or quantify the distribution of radionuclical structures in scanned body tissue for clinical diagnostic purposes.
The DaTQUANT optional application enables visual evaluation of 1231-ioflupane (DaTscanTM) images. DaTQUANT Normal Database option enables quantification relative to normal population databases of 1231ioflupane (DaTscanTM) images. These applications may assist in detection of loss of functional dopaminergic neuron terminals in the striatum, which is correlated with Parkinson disease.
The O.Lung application may aid physicians in:
· Diagnosis of Pulmonary Embolism (PE), Chronic Obstructive Pulmonary Disease (COPD), Emphysema and other lung deficiencies.
• Assess the fraction of total lung function provided by a lobe or whole lung for Lung cancer resection requiring removal of an entire lobe, bilobectomy, or pneumonectomy.
The Q.Brain application allows the user to visualize and quantify relative changes in the brain's metabolic function or blood flow activity between a subject's images and controls, which may be resulting from brain functions in: · Epileptic seizures
· Dementia. Such as Alzheimer's disease, Lewy body dementia, Parkinson's disease with dementia, and frontotemporal dementia.
- · Inflammation
- · Brain death
- · Cerebrovascular disease such as Acute stroke, Chronic and acute ischemia
· Traumatic Brain Injury (TBI)
When integrated with the patient's clinical and diagnostic information may aid the physician in the interpretation of cognitive complaints, neuro-degenerative disease processes and brain injuries.
The Alcyone CFR application allows for the quantification of coronary vascular function by deriving Myocardial Blood Flow (MBF) and then calculating Coronary Flow Reserve (CFR) indices on data acquired on PET scanners and on stationary SPECT scanners with the capacity for dynamic SPECT imaging. These indices may add information to physicians using Myocardial Perfusion Imaging for the diagnosis of Coronary Artery Disease (CAD).
The Exini Bone application is intended to be used with NM bone scans for the evaluation of adult male patients with bone metastases from prostate cancer. Exini Bone quantifies the selected lesions and provides a Bone Scan Index value as adjunct information related to the progression of disease.
The Q.Liver application provides processing, quantification, and multidimensional review of Liver SPECT/PET and CT images for display, segmentation, and a calculation of the SPECT 'liver to lung' shunt value and the patient's Body Surface Area (BSA) for use in calculating a therapeutic dose for Selective Internal Radiation Therapy (SIRT) treatment
{3}------------------------------------------------
using a user defined formula.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "ge" in a cursive font, enclosed within a blue circle. The circle is surrounded by a decorative, swirling pattern, also in blue. The background is plain white.
K201103
# 510(k) SUMMARY OF SAFETY AND EFFECTIVNESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h):
| Date: | April 23rd, 2020 | | |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|--|
| Submitter: | GE Medical Systems Israel, Functional Imaging (GE Healthcare)<br>4 Hayozma Street<br>Tirat Hacarmel, 30200, Israel | | |
| Primary<br>Contact: | Alexandra Lifshits<br>Regulatory Affairs Program Manager<br>GE Medical Systems Israel, Functional Imaging<br>Tel: +972-4-8563666 ext. 805<br>Fax: +972-4-8577662<br>email: alexandra.lifshits@ge.com | | |
| Secondary<br>Contacts: | John Jaeckle<br>Chief Regulatory Affairs Strategist<br>GE Healthcare<br>Tel: 262-424-9547<br>email: john.jaeckle@ge.com | | |
| Device Trade<br>Name: | Xeleris V Processing and Review System | | |
| Common /<br>Usual Name: | System, Image Processing, Radiological | | |
| Classification<br>Names: | 21CFR 892.2050 | | |
| Product<br>Code: | LLZ | | |
| Predicate<br>Device | Device Name<br>510 (K) number<br>Regulation Nº<br>Product Code | Xeleris 4.0 Processing and Review Workstation<br>K153355<br>892.2050<br>LLZ | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are also some white swirls around the letters. The logo is simple and recognizable.
| Reference | Device Name | EXINI's aBSI | | |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--|--|
| Devices | 510 (K) number | K191262 | | |
| | Regulation Nº | 892.2050 | | |
| | Product Code | LLZ | | |
| | | | | |
| | Device Name | GE's Hepatic VCAR | | |
| | 510 (K) number | K133649 | | |
| | Regulation Nº | 892.1750 | | |
| | Product Code | JAK, LLZ | | |
| | | | | |
| Marketed<br>Devices | The Xeleris V Processing and Review System is a modification to the predicate<br>device, Xeleris 4.0 Processing and Review Workstation. It includes all of the<br>clinical applications and features in the current production version of Xeleris 4.0<br>and, introduces new clinical applications (i.e. Exini Bone, Q.Liver) and<br>enhancements to previously cleared clinical applications (i.e. Q.Lung,<br>Myovation, Q.Volumetrix MI, Q.Brain). | | | |
| | Xeleris V's Indications for Use remain the same as the predicate device's except<br>for additional indicates for the new clinical applications, Exini Bone and Q.Liver.<br>Xeleris V's Indications for Use do not create a new Intended Use. | | | |
| Device<br>Description: | The Xeleris V Processing and Review Workstation is a Nuclear Medicine<br>Software system that is designed for general nuclear medicine processing and<br>review procedures for detection of radioisotope tracer uptake in the patient's<br>body, using a variety of individual processing applications orientated to specific<br>clinical applications. | | | |
| | Xeleris V is a modification to the predicate device, Xeleris 4.0 Processing and<br>Review Workstation. It includes all the clinical applications and features in the<br>current production version Xeleris 4.0 and introduces new clinical applications<br>and enhancements to previously cleared clinical applications. | | | |
| | With Xeleris V, the customer now has an option where they can purchase the<br>Xeleris V software, and have it installed on a remote server. | | | |
| | New Clinical Applications | | | |
| | Exini Bone: The Exini Bone application is intended to be used with NM bone<br>scans for the evaluation of adult male patients with bone metastases from<br>prostate cancer. The application automatically segments bone lesions based on<br>2D whole-body planar bone scans and requires the user adjust and/or accept<br>the final segmentation before proceeding. Exini Bone quantifies the selected<br>lesions and provides a Bone Scan Index value as adjunct information related to<br>the progression of disease. | | | |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three water droplets surrounding the circle. The logo is simple and recognizable, and it is associated with a well-known and established company.
- · Q.Liver: The Q.Liver application is a comprehensive application that provides processing, quantification, and multidimensional review of liver SPECT/CT exams for display and segmentation. The application provides the user with tools to calculate a therapeutic dose for Selective Internal Radiation Therapy (SIRT) treatment.
# Enhancements to Clinical Applications
- Q.Lung: The Q.Lung application provides processing, quantification, and multidimensional review for pulmonary scintigraphy for display and quantification of global and regional ventilation and perfusion on SPECT and SPECT/CT studies. The change introduced in Xeleris V is the introduction of an optional lung fissure automatic detection using a Deep Learning based model.
- Myovation: The Myovation application is used for the evaluation of patients with suspected or known coronary artery disease. Myovation reconstructs and reformats the SPECT raw data and helps to analyze the myocardial perfusion and function in rest, stress, and viability studies. The change for Xeleris V is the introduction of a new optional, non-Al Automatic Recognition of Cardiac Structures (ARCS) algorithm for improved heart detection and orientation.
- Q.Volumetrix MI: The Q.Volumetrix MI application is a compressive tool for processing and reading non-cardiac volumetric data, including NM SPECT and hybrid SPECT-CT, PET-CT, external CT/MR (i.e. CT/MR from a separate nonhybrid scan). The enhancement for Xeleris V is a new, optional, DL-based automatic kidney segmentation algorithm.
- Q.Brain: The Q.Brain application is used to process, quantify, and review brain studies obtained that include SPECT , SPECT/CT , PET/CT, and MR datasets. Q.Brain processing includes reconstruction, reorientation, rigid registration, and non-rigid registration of single or sequential tomographic brain data. The enhancement introduced with Xeleris V is that the display of the CT images registered to the template is now included along with other currently available image types.
- The system is intended for use by Nuclear Medicine (NM) or Radiology Intended practitioners and referring physicians. The intended use of the system is to Use: provide digital processing, review and reporting of medical images, including data display, quality control, image manipulation and quantification analysis, transfer, storage and printing capabilities.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three water droplets surrounding the circle. The logo is simple and recognizable, and it is associated with a well-known company.
The system operates in a variety of configurations. The hardware components may include computer workstations, communications devices, video monitors, data storage and hardcopy devices.
Software components provide functions for performing operations related to image display, manipulation, enhancements, analysis and quantification and can operate on dedicated workstations and client-server architectures.
The system is intended for use by Nuclear Medicine (NM) or Radiology Indications practitioners and referring physicians for display, processing, archiving, for Use: printing, reporting and networking of NMI data, including planar scans (Static, Whole Body, Dynamic, Multi-Gated) and tomographic scans (SPECT, Gated SPECT, dedicated PET or Camera-Based-PET) acquired by gamma cameras or PET scanners. The system can run on dedicated workstation or in a server-client configuration.
> The NM or PET data can be coupled with registered and/or fused CT or MR scans, and with physiological signals in order to depict, localize, and/or quantify the distribution of radionuclide tracers and anatomical structures in scanned body tissue for clinical diagnostic purposes.
> The DaTQUANT optional application enables visual evaluation and quantification of 1231-ioflupane (DaTscanTM) images. DaTQUANT Normal Database option enables quantification relative to normal population databases of 123I-ioflupane (DaTscanTM) images. These applications may assist in detection of loss of functional dopaminergic neuron terminals in the striatum, which is correlated with Parkinson disease.
The Q.Lung application may aid physicians in:
- . Diagnosis of Pulmonary Embolism (PE), Chronic Obstructive Pulmonary Disease (COPD), Emphysema and other lung deficiencies.
- • Assess the fraction of total lung function provided by a lobe or whole lung for Lung cancer resection requiring removal of an entire lobe, bilobectomy, or pneumonectomy.
The Q.Brain application allows the user to visualize and quantify relative changes in the brain's metabolic function or blood flow activity between a subject's images and controls, which may be resulting from brain function alterations in:
- · Epileptic seizures
- . Dementia. Such as Alzheimer's disease, Lewy body dementia, Parkinson's disease with dementia, vascular dementia, and frontotemporal dementia.
- Inflammation .
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. There are three water droplet shapes surrounding the circle. The logo is blue.
- . Brain death
- · Cerebrovascular disease such as Acute stroke, Chronic and acute ischemia
- Traumatic Brain Injury (TBI)
When integrated with the patient's clinical and diagnostic information, Q.Brain application may aid the physician in the interpretation of cognitive complaints, neuro-degenerative disease processes and brain injuries.
The Alcyone CFR application allows for the quantification of coronary vascular function by deriving Myocardial Blood Flow (MBF) and then calculating Coronary Flow Reserve (CFR) indices on data acquired on PET scanners and on stationary SPECT scanners with the capacity for dynamic SPECT imaging. These indices may add information to physicians using Myocardial Perfusion Imaging for the diagnosis of Coronary Artery Disease (CAD).
The Exini Bone application is intended to be used with NM bone scans for the evaluation of adult male patients with bone metastases from prostate cancer. Exini Bone quantifies the selected lesions and provides a Bone Scan Index value as adjunct information related to the progression of disease.
Q.Liver application provides processing, quantification, The and multidimensional review of Liver SPECT/PET and CT images for display, segmentation, and a calculation of the SPECT 'liver to lung' shunt value and the patient's Body Surface Area (BSA) for use in calculating a therapeutic dose for Selective Internal Radiation Therapy (SIRT) treatment using a user defined formula.
- Xeleris V's basic functionality for processing and reviewing Nuclear Medicine Technology: and their associated CT images is not changed from its predicate. This functionality includes manual and automatic segmentation to depict, localize, and/or quantify the distribution of radionuclide tracers and anatomical structures in patients for diagnostic purposes. Much of this functionality is designed for streamlined user workflows.
Xeleris V introduces two new applications that are focused on specific clinical imaging scenarios, using this basic functionality Xeleris V also enhances three existing applications to improve their workflow efficiency using both DL-based and traditional algorithms.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized script in the center. There are three water droplets or stylized swirls around the letters. The logo is simple and recognizable, and it is associated with a well-known and established company.
| Attribute | Predicate Device: Xeleris 4.0 | Proposed Device: Xeleris V |
|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General<br>Workflows | Xeleris 4.0's workflows use<br>both manual and automated<br>processes. | Xeleris V's workflows continue<br>use both manual and automated<br>processes. Xeleris V also includes<br>two new, DL-based automatic<br>segmentation algorithms. |
| Use of<br>Deep<br>Learning | Xeleris 4.0 does not contain<br>any DL processes/algorithms. | Xeleris V introduces two new DL-<br>based algorithms for:<br>• Lung fissure segmentation as<br>an alternative to the manual<br>method.<br>• Kidney segmentation as an<br>enhancement to the non-DL<br>automatic segmentation. |
| Hardware<br>Needed to<br>Support<br>Xeleris V's<br>software | Xeleris 4.0 offered with a GE<br>workstation or as software<br>that can be loaded onto the<br>customer's workstation that<br>meets specification. | In addition to the two options for<br>the predicate, with Xeleris V the<br>software can also be installed on a<br>remote server. |
The Xeleris V Processing and Review System has identical or equivalent technological characteristics as its predicate and reference devices. The changes and the different technological characteristics do not raise new or different questions of safety and effectiveness. The software was developed, verified, and validated under GE Healthcare's QMS including software development lifecycle.
#### Design Control Testing: Testing
Summary
**Design Control Testing:**
Xeleris V and its clinical applications successfully completed all design control testing per GE Healthcare's quality system and also verified compliance with the relevant standards (i.e. NEMA PS3.1 - 3.20, IEC62304). The testing did not raise any new safety questions or identify any new risks. Xeleris V is designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. The following quality assurance measures were applied to the development of the system:
- Risk Analysis
- · Requirements Reviews
- · Design Reviews
- · SW Development Lifecycle
- Testing on Unit Level (Module Verification)
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a blue circle. There are three water droplets or splashes emanating from the top and sides of the circle, giving the logo a dynamic and fluid appearance. The color scheme is primarily blue and white.
- · Integration Testing (System Verification)
- · Performance Testing (Verification)
- Safety testing (Verification)
- · Simulated Use Testing (Validation)
The testing and their results do not raise different questions of safety and effectiveness than the predicate device. GE believes Xeleris V is of comparable type and substantially equivalent to the predicate and reference devices, and hence is safe and effective for its intended use. The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.
# Summary of Additional Testing:
In addition to the verification and validation testing successfully completed as required by GE Healthcare's quality system, additional engineering and clinical performance testing was performed to provide the requisite data to demonstrate performance and substantiate claims. This testing included:
Engineering testing including clinical datasets and clinician participation.
- · Demonstration of clinical outputs, workflow, and tools.
- Scientific testing for algorithm performance and claims substantiation.
- · Algorithm clinical performance testing using test datasets of representative clinical exams where the ground truth and evaluations were performed by clinicians.
# Clinical Evaluation
Xeleris V did not require clinical studies to support the determination of equivalence. However, the performance testing for the DL-based algorithms for kidney segmentation and lung fissure segmentation evaluates and demonstrates each algorithm's performance and uses test datasets of representative clinical exams.
The test datasets were comprised of representative clinical exams that were manually segmented by experienced NM Physicists and Physicians and were used as ground truth. The NM Physicists and Physicians also performed the algorithm evaluation. They reviewed the segmentation results and scored them using a 5-point Likert scale.
The scientific methods used to evaluate the effectiveness of Xeleris V are acceptable and support the determination of substantial equivalence.
Substantial The Substantial Equivalence of the proposed device has been demonstrated by:
- Equivalence
- Review of the proposed Indications for Use shows that they are substantially equivalent to the predicate device's, also with consideration of the reference devices. Xeleris V's Indications for Use do not create a new Intended Use.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The circle is surrounded by a swirling pattern, also in blue. The logo is simple and recognizable, and it is often used to represent the company's brand.
- The device description and the comparison of device characteristics show that Xeleris V's has identical or equivalent technological characteristics as its predicate and reference devices.
- The testing (engineering and clinical) demonstrates the effectivity of Xeleris V for its intended purpose.
- . The changes and the different technological characteristics do not raise new or different questions of safety and effectiveness. The proposed device is as safe and effective as the legally marketed predicate device as demonstrated by the:
- o Software verification and validation for a Moderate level of concern, without unexpected results;
- o Development under GE's quality management system, design control activities including risk management;
- o Device labeling; and
- o Engineering and clinical testing without unexpected results.
- Xeleris V's Indications for Use do not create a new Intended Use. Xeleris V has Conclusion: identical or equivalent technological characteristics as its predicate device.
Xeleris V was developed under GE Healthcare's quality system. Design verification testing, along with bench and clinical testing demonstrate that Xeleris V is substantially equivalent and hence as safe and as effective as the legally marketed predicate device. GE's quality system's design, verification, and risk management processes did not identify any new hazards, unexpected results, or adverse effects stemming from the changes to the predicate.
Based on development under GE Healthcare's quality system, the successful verification, engineering bench and clinical testing, GE Healthcare believes that Xeleris V is substantially equivalent to the predicate device and hence is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.