K201092 · Imaging Biometrics, LLC · JAK · Oct 29, 2020 · Radiology
Device Facts
Record ID
K201092
Device Name
LSN
Applicant
Imaging Biometrics, LLC
Product Code
JAK · Radiology
Decision Date
Oct 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
LSN (Liver Surface Nodularity) is an image analysis software application intended to assist radiologists and other trained healthcare professionals in analyzing and reporting on the liver morphology depicted in computed tomography (CT) images for use in assessment of chronic liver disease. LSN is designed to assist the user in the evaluation and documentation of liver morphology, specifically liver surface nodularity, provided that the surface nodularity is adequately depicted on the CT images. LSN provides quantitative metrics related to liver fibrosis by automating segmentation of the liver surface within user-defined Regions of Interest (ROIs) and calculating distances and means related to the liver surface nodularity. LSN also offers reporting capabilities for documenting user-confirmed results, thereby facilitating communication with other trained healthcare professionals and assessment of changes over time. LSN is intended to provide image-related information that is interpreted by a trained professional, but it does not directly generate any diagnosis. The information provided by LSN should not be used in isolation when making patient management decisions. LSN is not intended for use with or for the diagnostic interpretation of mammography images.
Device Story
LSN is a post-processing software application for DICOM CT abdominal images. It assists radiologists in assessing chronic liver disease by quantifying liver surface nodularity. The user defines Regions of Interest (ROIs) on the liver edge; the software performs semi-automated segmentation of the liver surface within these ROIs. It fits a spline to the detected edge and calculates the mean distance between the edge and the spline. The output includes a quantitative 'LSN Score' (in tenths of a millimeter) and a report (PDF/DICOM) for PACS upload. Used in clinical settings, it does not provide a diagnosis; the clinician interprets the metrics to aid in patient management and longitudinal assessment. It is designed to standardize terminology and facilitate communication between clinicians.
Clinical Evidence
Bench testing only. Functional verification, integration testing, and validation were performed. Arithmetic and report accuracy were validated against alternative calculation mechanisms. Usability testing confirmed clinical operation. Results demonstrated repeatability across different CT imaging/reconstruction parameters and reproducibility across scanner vendors. Intra- and inter-observer measurement variability was low.
Technological Characteristics
Software-only application; platform-independent code integrated as a standalone PC app or Mac-based viewer plugin. Uses DICOM (NEMA PS 3.1-3.20) for image communication. Employs semi-automated edge detection and polynomial spline fitting for surface approximation. Complies with IEC 62304 (software lifecycle) and ISO 14971 (risk management).
Indications for Use
Indicated for radiologists and trained healthcare professionals to assist in the evaluation and documentation of liver surface nodularity in CT images for the assessment of chronic liver disease. Not for use with mammography images.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
October 29, 2020
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Imaging Biometrics, LLC % Mr. Timothy Dondlinger COO 13416 Watertown Plank Road, Suite 260 ELM GROVE WI 53122
Re: K201092
Trade/Device Name: LSN Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK, LLZ Dated: September 22, 2020 Received: September 24, 2020
Dear Mr. Dondlinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201092
Device Name LSN
## Indications for Use (Describe)
LSN (Liver Surface Nodularity) is an image analysis software application intended to assist radiologists and other trained healthcare professionals in analyzing and reporting on the liver morphology depicted in computed tomography (CT) images for use in assessment of chronic liver disease. LSN is designed to assist the user in the evaluation and documentation of liver morphology, specifically liver surface nodularity, provided that the surface nodularity is adequately depicted on the CT images.
LSN provides quantitative metrics related to liver fibrosis by automating segmentation of the liver surface within userdefined Regions of Interest (ROIs) and calculating distances and means related to the liver surface nodularity. LSN also offers reporting capabilities for documenting user-confirmed results, thereby facilitating communication with other trained healthcare professionals and assessment of changes over time.
LSN is intended to provide image-related information that is interpreted by a trained professional, but it does not directly generate any diagnosis. The information provided by LSN should not be used in isolation when making patient management decisions.
LSN is not intended for use with or for the diagnostic interpretation of mammography images.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## 510(k) Notification K201092
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Applicant /<br>Owner : | Imaging Biometrics, LLC<br>13416 Watertown Plank Road, Suite 260<br>Elm Grove, WI 53122 |
|----------------------------|-------------------------------------------------------------------------------------------------------|
| Contact Person: | Timothy Dondlinger<br>(262) 439-8252 (telephone)<br>(262) 439-8301 (fax)<br>tim@imagingbiometrics.com |
| Date Prepared: | April 17, 2020 |
| Common Name: | System, X-Ray, Tomography, Computed |
| Trade Name: | LSN |
| Classification<br>Name: | Computed Tomography X-Ray System |
| Review Panel: | Radiology |
| Regulation<br>Number: | 892.1750 |
| Device Class: | II |
| Product Code: | JAK |
| Subsequent<br>Product Code | LLZ (892.2050; System, Image Processing, Radiological; Picture Archiving and Communications System) |
## Predicate Device Information
| Type | 510(k) | Trade Name | Manufacturer | Regulatory<br>Citation<br>(21 CFR) | Regulation Name | Regulatory<br>Class | Primary<br>Product<br>Code |
|------------------------|---------|--------------|------------------------|------------------------------------|----------------------------------------|---------------------|----------------------------|
| Primary Predicate | K133649 | Hepatic VCAR | GE Medical Systems SCS | 892.1750 | System, X-Ray, Tomography, Computed | II | JAK |
| Secondary<br>Predicate | K101342 | OsiriX MD | Pixmeo SARL | 892.2050 | System, Image Processing, Radiological | II | LLZ |
## Device Description
LSN (Liver Surface Nodularity) is a post-processing software application which assists trained professionals in evaluating DICOM computed tomography image studies of patients with chronic liver disease. The software provides tools to enable the user to make quantitative measurements related to liver surface nodularity as depicted on CT images.
The generated information consists of a LSN Score (reported in tenths of a millimeter), a quantitative measure of the surface nodularity based on a set of user-defined ROIs sampling the liver surface. LSN calculates the distance between the detected liver edge and a smoothed polynomial line (spline) on a pixel-by-pixel basis inside ROIs and reports the mean of these distances on a per-slice basis as well as an overall LSN Score for the imaging series.
LSN provides the user with information the progression of chronic liver disease. LSN does not make clinical decisions and the information provided by LSN must not be used in isolation when making patient management decisions. The LSN Score may provide value by standardizing terminology used to describe surface neporting, thereby facilitating communications between radiologists and other clinicans invalient's treatment planning. In addition, standardized reporting metrics may also be helpful in assessing changes for the same patient over time.
LSN functions by displaying a DICOM CT abdominal series to the user paints a broad region of interest (RO)) delineating the liver edge on a subset of image slices. Then, for the painted region on each slice, the edge is detected using multiple algorithms. For each detected
{4}------------------------------------------------
edge, a spline is fit to the edge and the shortes from each edge pixel to the spline are calculated and averaged, resulting in a potential LSN value. The maximum LSN value calculated for an edge is reported as the LSN values for all slices on which ROIs have been painted are then averaged to determine the overall LSN score.
The core LSN algorithms are implemented in platform-independent code, and have been integrated into both a standalone PC research application and a Mac-based viewer plugin for clinical use. Both platforms produce an equivalent LSN score; the sthe algorithm to require less re-work by the user. The clinical version also produces a report containing images of the scores for each slice, and the overall LSN score. The report is produced in both PDF and DICOM formats and is ready for upload to PACS.
#### Indications for Use/Intended Use Statement
LSN (Liver Surface Nodularity) is an image analysis software application intended to assist radiologists and other trained healthicare professionals in analyzing and reporting on the liver morphology depicted in computed tomography (CT) images for use in assessment of chronic liver disease. LSN is designed to assist the user in the evaluation of liver morphology, specifically liver surface nodularity, provided that the surface nodularity is adequately depicted on the CT images.
LSN provides quantitative metrics related to liver fibrosis by automating segmentation of the liver surface within user-defined Regions of Interest (ROls) and calculating distances and means related to the liver surface nodularity. LSN also offers reporting userconfirmed results, thereby facilitating communication with other trained healthcare professionals and assessor time.
LSN is intended to provide information that is interpreted by a trained professional, but it does not directly generate any diagnoss. The information provided by LSN should not be used in isolation when making patient management decisions.
LSN is not intended for use with or for the diagnostic interpretation of mammography images.
#### Substantial Equivalence
The below table provides a comparison summary of the subject device's Indications for Use/Intended Use Statement to the primary and secondary predicate devices.
| Indications for Use /<br>Intended Use | Subject | Primary Predicate:<br>Hepatic VCAR | Secondary Predicate:<br>OsiriX MD | Rationale for<br>Substantial Equivalence |
|-------------------------------------------------------------------------------------------------------------------------------------|---------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| LSN (Liver Surface<br>Nodularity): | | | | |
| 1. is an image analysis<br>software application | Yes | Yes | Yes | Same |
| 2. intended to assist<br>radiologists and other<br>trained healthcare<br>professionals | Yes | Yes<br>From p.16 of the User Manual: "Intended<br>Operator Profile: Hepatic VCAR is<br>designed for clinicians, specifically,<br>radiologists, interventional radiologists,<br>surgeons, and other trained healthcare<br>professionals (3D technologists).<br>Clinicians and healthcare professionals<br>using this software must be educated in<br>the reading and quantitative and<br>qualitative analysis of CT liver exams." | Yes | Same |
| 3. in analyzing and<br>reporting on computed<br>tomography (CT) images | Yes | Yes | Yes (multimodality; CT included) | Same |
| 4. for assessment | Yes | Yes | Yes | Same |
| 5. in chronic liver disease. | Yes | Yes (Not explicitly stated, but implied.<br>Chronic liver disease is frequently a<br>precursor for liver lesions, e.g.<br>hepatocellular carcinoma). | Yes (Not specifically claimed in the<br>Intended Use statement, but implied:<br>General image viewing and analysis<br>application for all patients). | Same. Chronic liver<br>disease is a subset of the<br>intended patient<br>population for Hepatic<br>VCAR, as well as for<br>OsiriX MD. |
| 6. is designed to assist the<br>user | Yes | Yes | Not specifically claimed in the Intended<br>Use statement, but implicit in the<br>functionality. | Same |
| | | | | |
| 7. in the evaluation and<br>documentation of liver<br>morphology, | Yes | Yes | Yes (Not specifically claimed in the<br>Intended Use statement, but implied:<br>General image viewing and analysis<br>application for all patients). Yes (Not<br>specifically claimed in the Intended Use<br>statement, but implied: General image<br>viewing and analysis application for all<br>patients). | Same |
| 8. specifically liver surface<br>nodularity, provided that<br>the surface nodularity is<br>adequately depicted on<br>the CT images. | Yes | Not specifically claimed in the Intended<br>Use statement; Hepatic VCAR is<br>intended for evaluating liver morphology<br>which may include aspects of surface<br>nodularity, but features are primarily<br>focused on evaluation and quantitation<br>of liver lesions, provided they are visible<br>on the CT images. | Not specifically claimed in the Intended<br>Use statement; OsiriX MD may be used<br>to evaluate liver surface nodularity. | Same |
| 9. provides quantitative<br>metrics related to liver<br>fibrosis | Yes | Not specifically claimed in the Intended<br>Use statement; Hepatic VCAR is used to<br>assess liver volumes which are in turn<br>used to asses hepatic fibrosis. | Not specifically claimed in the Intended<br>Use statement, but these types of<br>measurement tools are included in the<br>device features. | Same for secondary<br>predicate device |
| 10. by automating<br>segmentation of the liver<br>surface within user-<br>defined Regions of<br>Interest (ROIs) | Yes | Yes | No | Same for primary<br>predicate device |
| 11. and calculating<br>distances and means<br>related to the liver surface<br>nodularity. | Yes | No | Not specifically claimed in the Intended<br>Use statement, but these types of<br>measurement tools are included in the<br>device features. | Same for secondary<br>predicate device |
| 12. offers reporting<br>capabilities for<br>documenting user-<br>confirmed results | Yes | Yes | Yes | Same |
| 13. facilitating<br>communication with other<br>trained healthcare<br>professionals | Yes | Yes | Not specifically claimed in the Intended<br>Use statement, but implied by "Images<br>and data can be communicated." | Same |
| 14. facilitating assessment<br>of changes over time | Yes | Similar - See #8. Hepatic VCAR is<br>intended to assist with evaluating<br>changes in liver lesions over time. LSN<br>is intended to assist with evaluating<br>changes in liver surface nodularity over<br>time. | Not specifically claimed in the Intended<br>Use statement. In general, quantitative<br>measurements facilitate assessment of<br>morphological changes over time, so<br>this is an implicit benefit of the device<br>features. | Same |
| 15. intended to provide<br>image-related information<br>that is interpreted by a<br>trained professional | Yes | Yes | Yes. (Not explicitly stated, but implied). | Same |
| 16. does not directly<br>generate any diagnosis | Yes | Yes. (Not explicitly stated, but implied). | Yes. (Not explicitly stated, but implied). | Same |
| 17. The information<br>provided should not be<br>used in isolation when<br>making patient<br>management decisions | Yes | Yes. (Not explicitly stated, but implied by<br>device classification). | Yes. (Not explicitly stated, but implied<br>by device classification). | Same |
| 18. not intended for use<br>with or for the diagnostic<br>interpretation of<br>mammography images | Yes | Yes. (Not explicitly stated for Hepatic<br>VCAR, but implied by Volume Viewer<br>Plus's exclusion of mammography<br>images). | No. (allows viewing of mammography<br>images with an appropriate monitor) | Same for primary<br>predicate device |
{5}------------------------------------------------
Subject Device Indications for Use/Intended Use Statement: LSN
{6}------------------------------------------------
LSN (Liver Surface Nodularity) is an image analysis software application intended to assist radiologists and other trained healthicare professionals in analyzing and reporting on the liver morphology depicted in computed tomography (CT) images for use in assessment of chronic liver disease. LSN is designed to assist the user in the evaluation of liver morphology, specifically liver surface nodularity, provided that the surface nodularity is adequately depicted on the CT images.
LSN provides quantitative metrics related to liver fibrosis by automating segmentation of the liver surface within user-defined Regions of Interest (RQIs) and calculating distances and means related to the liver surface nodularity. LSN also offers reporting capabilities for documenting user-confirmed results, thereby facilitating communication with other professionals and assessment of changes over time.
L SN is intended to provide image-related information that is interpreted by a trained professional, but it does not directly generate any diagnosis. The information provided by LSN should not be used in isolation when making patient management decisions.
LSN is not intended for use with or for the diagnostic interpretation of mammography images.
#### Primary Predicate Indications for Use/Intended Use Statement: Hepatic VCAR
Hepatic VCAR is a CT image analysis software package(1) that allows the analysis and visualization of Liver CT data(3.5) derived from DICOM 3.0 compliant CT scans. Hepatic VCAR is designed for the purpose of assessing liver morphology(5.7). including liver lesion, provided the lesion has different CT appearance from surrounding liver time114) through automated tools for liver. liver lobe, liver segments and liver lesion seqment(10). It is intended for use by clinicians(2) to process, review, archive, print and distribute(3,12,13) liver CT studies.
This software will assist the user(6) by providing initial 3D secmentation(10), visualization, and quantitative analysis of liver anatomy(7). The user has the ability to adjust the contour and confirm the final segmentation(2.6).
#### Secondary Predicate Indications for Use/Intended Use Statement: OsiriX MD
OsiriX MD TM is a software device intended for viewing of images acquired from CT(3), MR, CR, OR, US and other DICOM compliant medical imaging systems when installed on suitable commercial standard hardware. Images and data can be captured, stored, communicated(12,13), processed, and displayed(1) within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for image interpretation.
For primary diagnosis(4), post process DICOM for presentation" images must be used. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images.(18) It is the User's responsibility to ensure monitor quality, ambient light conditions, and image compression ratios are consistent with the clinical application(2).
A comparison of the subject device's technical characteristics compared to the predicate devices is summarized in the following table:
| Technical Characteristic | Subject | Primary Predicate: | Secondary Predicate: | OsiriX MD |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | | Hepatic VCAR | | |
| 1. Designed to work as a plug-in software<br>application that works with an external<br>software platform application by leveraging<br>the external platform's DICOM<br>communications, image display, and<br>general image processing capabilities | Yes | Similar - Works with Volume<br>Viewer Plus software application<br>from GE Medical Systems<br>(K041521) to provide liver-specific<br>functionality, including semi-<br>automated liver segmentation | No - OsiriX MD is the platform<br>application that provides DICOM<br>communications, image display,<br>and general image processing<br>capabilities which are leveraged<br>by the LSN application. | |
| 2. Works with DICOM-standard CT images | Yes | Yes | Multimodality including CT | |
| 3. No special CT acquisition required | Yes | Yes | Yes | |
| 4. User-defined ROIs | Yes (leverages OsiriX ROI tools<br>via published API) | Yes | Yes - offers support for a wide<br>range of ROI types and statistics | |
| 5. User may accept/reject any/all ROIs<br>before finalizing measurements and<br>creating a report | Yes | Yes | Yes | |
| 6. Semi-automated liver boundary detection<br>inside user-defined ROIs | Yes | Yes (ROIs may be complete<br>images that user has identified as<br>containing liver) | No | |
| 7. Calculates and displays smoothed liver<br>surface approximation using a polygonal<br>approximation to the surface | Creates and displays a<br>smoothed polynomial (spline)<br>from the bounding surface<br>determined by a semi-automated<br>edge detection based on<br>threshold processing of 2D data | Not verified how the liver surface is<br>extracted and and displayed, but<br>presumably using a smooth<br>polynomial mesh. | Yes - It is possible to perform this<br>step manually. | |
| 8. User-customizable colors for ROI and<br>text annotations on images | Yes | Yes | Yes | |
{7}------------------------------------------------
| 9. Average spline-to-surface distance for<br>each ROI and overall LSN Score calculation. | Yes | No | LSN measurements for a segment<br>of liver surface may be performed<br>manually using native OsiriX<br>measurement tools with off-line<br>calculation of the overall LSN<br>Score |
|-----------------------------------------------------------------------------------------------------------|-----------------------------------------|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 10. Display output of measurement tools for<br>user review and acceptance | Yes | Yes | Yes |
| 11. Report generation | Yes | Yes | Yes |
| 12. Runs on standard operating systems<br>and hardware | Yes: macOS, commercial<br>computers | Yes: Windows OS, commercial<br>computers | Yes: macOS, commercial<br>computers |
| Physical Characteristic | | | |
| 13. Post-processing application (non-real-<br>time, non-contacting, not life supporting or<br>sustaining) | Yes | Yes | Yes |
| 14. Sterilization and Shelf-life | Not applicable, software<br>application | Not applicable, software application | Not applicable, software<br>application |
| 15. Biocompatibility | Not applicable, software<br>application | Not applicable, software application | Not applicable, software<br>application |
The intended use and technical characteristics for LSN are similar to both predicate Device Information section above for a software application that provides a user interface to allow a clinician to load, display, measure, and identify regions of interests. The software algorithms calculate statistical information of the image pixels contained within user-identified regions of interest. The calculated values provide a clinician with relevant information for assessment of chronic liver disease.
LSN is a software-only device, is non-pating, and is used by highly-trained healthcare professionals such as radiologists and medical imaging technologists. LSN does not provide any interpretation of the statistical parameters displayed to the user. The trained professional is responsible for identifying, measuring, and interpreting the images and data being displayed.
LSN and both its predicate devices are substantially equivalent in the categories of technical characteristics and features. LSN does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices.
## Testing Information and Performance
All product specifications were verified and the product to meet user needs was validated. Testing was performed according to internal company procedures. Software testing and validation were conducted according to written testing was conducted. Test results were reviewed by designals before software proceeded to release. Validation test results support the conclusion that actual device performance satisfies the design intent. Testing was performed in accordance document, "General Principles of Software Validation," issued January 11, 2002, and documentation provided by FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued May 11, 2005. Bench testing (functional and integration) was conducted for LSN during product development.
Bench testing included functional verfication, icensing, labeling, and feature functionality all met design requirements. Arithmetic and report accuracy was veilied and validated by comparison to alternative calculation mechanisms. And clinical operation was validated through usability testing.
Test results demonstrated that LSN output is repeatable for different CT imaging and reconstruction parameters, reproducible across different CT scanner types and vendors, and that the intra- and inter-observer measurement variability is low.
To obtain consistent output, images should be obtained at resolutions with sub-mm pixel sizes in the axial plane depend on slice thickness, and to a lesser extent, other imaging parameters such as contrast agent presence, so the user is instructed that comparisons to previous results published in the scientific literant CT exams should be undertaken with a knowledge of these dependencies.
The LSN risk analysis was completed and risk control mere implemented to mitigate unacceptable hazards. LSN relies upon user expertise to determine the suitability of CT images with excessive noise, artifacts, and/or confounding anatomy or pathology obscuring any portion of the liver surface to be analyzed should not be used. Verfication testing results supported the claims of substantial equivalence.
LSN has not been evaluated with images from patients of all ethnicities. It has been primarily evaluated with White and Black racial backgrounds. LSN has not been evaluated with images from pediatric patients.
#### Design and Development Standards and Guidance
{8}------------------------------------------------
LSN was designed and developed using the following standards and guidance:
- · NEMA PS 3.1-3.20: Digital Imaging and Communications in Medicine (DICOM)
- ISO 14971: Medical devices Application of risk management to medical devices
- IEC 62304: Medical device software Software life cycle processes
- · ISO 13485: Medical devices Quality management systems
- FDA: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- · FDA: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
- FDA: Off-The-Shelf Software Use in Medical Devices
## Conclusion
LSN has the same intended use as the primary and secondary predicate devices as an image processing software application for DICOM-format CT images. There is significant overlap between LSN and the predicate devices in technological characteristics. Any differences between LSN and the predicate devices do not raise new or different questions of safety and effectiveness. The result of all testing conducted was found acceptable to support the claim of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.