BD Insyte Autoguard Shielded IV Catheter, BD Insyte Autoguard BC Shielded IV Catheter, BD Insyte Autoguard BC Pro Shielded IV Catheter
K201075 · Becton, Dickinson and Company · FOZ · Dec 18, 2020 · General Hospital
Device Facts
Record ID
K201075
Device Name
BD Insyte Autoguard Shielded IV Catheter, BD Insyte Autoguard BC Shielded IV Catheter, BD Insyte Autoguard BC Pro Shielded IV Catheter
Applicant
Becton, Dickinson and Company
Product Code
FOZ · General Hospital
Decision Date
Dec 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BD Insyte™ Autoguard™ Shielded IV Catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may also be used to facilitate the placement of guidewires and other vascular access devices without pre-attached hubs. These for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068kPa). BD Insyte™ Autoguard™ BC shielded IV catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa). BD Insyte™ Autoguard™ BC Pro shielded IV catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
Device Story
Over-the-needle peripheral IV catheters; include radiopaque polyurethane catheter, needle, flash chamber, and spring-loaded push-button shielding mechanism. BC/BC Pro models include blood control septum to limit blood exposure during Luer connection. Instaflash™ technology provides visual confirmation of vessel entry. Used by clinicians for vascular access; facilitates guidewire placement. Needle retracts into barrel upon button activation to prevent needlestick injuries. Power injection compatible (22-18 GA).
Clinical Evidence
Bench testing only. Evaluated per ISO 10555-1, ISO 10555-5, ISO 594-1/2, ISO 23908, and ISO 11070. Biocompatibility testing conducted per ISO 10993-1, -3, -4, -5, -6, -7, -10, -11, and -12. Sterilization and packaging validated per ISO 11135, ISO 11607-1/2, and USP <788> for particulate matter.
Technological Characteristics
Single lumen peripheral IV catheter; materials: Polypropylene (cover, adapter, barrel), Polycarbonate (grip, hub, button), Stainless Steel (needle, wedge, spring), Polyurethane (tubing), Silicone (lubricant, septum), Porous Polyethylene (plug). Active push-button needle shielding; blood control septum (BC/BC Pro). Sterilized via Ethylene Oxide.
Indications for Use
Indicated for patients requiring peripheral vascular access for short-term blood sampling, blood pressure monitoring, or fluid administration. Suitable for any patient population based on vascular anatomy, procedure, and therapy duration. 22-18 GA models compatible with power injectors up to 300 psi.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
BD Insyte™ Autoguard™ Shielded IV Catheter (K952861)
BD Insyte™ Autoguard™ BC Shielded IV Catheter (K110443)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' and 'ADMINISTRATION' in blue text.
December 18, 2020
Kimberly Geisler Sr. Manager, Regulatory Affairs 9450 South State Street Sandy, Utah 84070
Re: K201075
Trade/Device Name: BD Insyte™ Autoguard ™Shielded IV Catheter, BD Insyte™ Autoguard™ BC Shielded IV Catheter, BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ, DYB Dated: November 17, 2020 Received: November 19, 2020
Dear Kimberly Geisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan M. Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201075
#### Device Name
BD Insyte™ Autoguard™ Shielded IV Catheter BD Insyte™ Autoguard™ BC Shielded IV Catheter BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter
Indications for Use (Describe)
BD Insyte™ Autoguard™ Shielded IV Catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These devices may also be used to facilitate the placement of guidewires and other vascular access devices without pre-attached hubs. These for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068kPa).
BD Insyte™ Autoguard™ BC shielded IV catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
BD Insyte™ Autoguard™ BC Pro shielded IV catheters are intended to be inserted into a patient's peripheral vascular system for short term use to sample blood pressure, or administer fluids. These devices may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
bd.com
Image /page/3/Picture/2 description: The image shows the logo for BD, a global medical technology company. The logo consists of two parts: a circular symbol on the left and the letters "BD" on the right. The circular symbol is orange and features a stylized image of a sun or star. The letters "BD" are in blue and have a rounded, sans-serif font.
#### Submitter Name: Becton Dickinson Infusion Therapy Systems Inc. Submitter Information Submitter Address: 9450 South State Street Sandy, Utah 84070 Contact Person: Paul Holman Regulatory Affairs Specialist Email Address: paul.holman@bd.com Phone Number: (801) 565-2838 Date of Preparation: December 17, 2020 BD Insyte™ Autoguard™ Shielded IV Catheter Subject Device Trade Name: BD Insyte™ Autoguard™ BC Shielded IV Catheter BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter 510(k) Reference: K201075 Peripheral Intravascular or IV Catheter Common Name: Regulation Number: 21 CFR §880.5200 (primary) 21 CFR §870.1340 (secondary for BD Insyte™ Autoguard™ Shielded IV Catheters only) Regulation Name: Catheter, intravascular, therapeutic, short-term less than 30 days Regulatory Class: II FOZ (primary) Product Code: DYB (secondary for BD Insyte™ Autoguard™ Shielded IV Catheter only) Classification Panel: General Hospital BD Insyte™ Autoguard™ Shielded IV Catheter Predicate Trade Name: Device 510(k) Reference: K952861 Common Name: Peripheral Intravascular or IV Catheter Regulation Number: 21 CFR §880.5200 Regulation Name: Catheter, intravascular, therapeutic, short-term less than 30 days Regulatory Class: II Product Code: FOZ Classification Panel: General Hospital
# 510(k) Summary: K201075
{4}------------------------------------------------
| | bd.com | |
|--|--------|--|
| Predicate<br>Device | Trade Name: | BD Insyte™ Autoguard™ BC Shielded IV Catheter |
|-----------------------|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 510(k) Reference: | K110443 |
| | Common Name: | Peripheral Intravascular or IV Catheter |
| | Regulation Number: | 21 CFR §880.5200 |
| | Regulation Name: | Catheter, intravascular, therapeutic, short-term less than<br>30 days |
| | Regulatory Class: | II |
| | Product Code: | FOZ |
| | Classification Panel: | General Hospital |
| Predicate<br>Device | Trade Name: | Introcan® Safety™ IV Catheter |
| | 510(k) Reference: | K021094 |
| | Common Name: | Peripheral Intravascular or IV Catheter/<br>Catheter Introducer |
| | Regulation Number: | 21 CFR §880.5200 and §870.1340 |
| | Regulation Name: | Catheter, intravascular, therapeutic, short-term less than 30<br>days Catheter Introducer |
| | Regulatory Class: | II |
| | Product Code: | FOZ/DYB |
| | Classification Panel: | Cardiovascular |
| Device<br>Description | | BD Insyte™ Autoguard™ Shielded IV Catheter |
| | | BD Insyte™ Autoguard™ shielded IV catheters are over-the-needle, intravascular (IV)<br>catheters. These devices include a radiopaque BD Vialon™ catheter, needle, grip with<br>button, flash chamber with vent plug, and barrel. The needle and catheter are protected<br>by a needle cover. |
| | | The 24-20 GA (0.7-1.1 mm) devices have BD Instaflash™ needle technology, allowing<br>for immediate visualization of blood along the catheter. The flash chamber provides<br>confirmation that the device has entered the vessel. These devices incorporate BD<br>Autoguard™ push-button shielding technology which is activated when the button is<br>depressed. A spring mechanism retracts the needle and flash chamber into the barrel,<br>fully encasing the needle and reducing the risk of accidental needlestick injury. |
| | | These devices are available with or without wings. The catheter hub and wings are<br>color coded to indicate outside catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA<br>(0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm)<br>= Grey, 14 GA (2.1 mm) = Orange). |
| | | These devices facilitate the introduction of guidewires and other vascular access<br>devices. The following labeled gauge sizes correspond to the listed inside diameter: 24<br>GA (0.40 mm), 22 GA (0.55 mm), 20 GA (0.7 mm), 18 GA (0.8 mm), 16 GA (1.2 mm),<br>and 14 GA (1.6 mm). |
| | | The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a<br>maximum pressure of 300 psi (2068 kPa) |
{5}------------------------------------------------
#### bd.com
# BD Insyte™ Autoguard™ BC Shielded IV Catheter
BD Insyte™ Autoguard™ BC shielded IV catheters are over-the-needle, intravascular (IV) catheters with blood control technology. These devices include a radiopaque BD Vialon™ catheter, needle, grip with button, flash chamber with vent plug, and barrel. The needle and catheter are protected by a needle cover.
The blood control technology is designed to stop the flow of blood from the catheter hub until the initial Luer connection is made. Once the connection is made, fluids or blood can flow through the catheter hub in either direction. The blood control technology limits blood exposure to the device user.
The 24-20 GA (0.7-1.1 mm) devices have BD Instaflash™ needle technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the device has entered the vessel.
These devices incorporate BD Autoguard™ push-button shielding technology which is activated when the button is depressed. A spring mechanism retracts the needle and flash chamber into the barrel, fully encasing the needle and reducing the risk of accidental needlestick injury.
These devices are available with or without wings. The catheter hub and wings are color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9 mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) = Grev).
The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa).
# BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter
BD Insyte™ Autoguard™ BC Pro shielded IV catheters are over-the-needle, intravascular (IV) catheters with blood control technology. These devices include a radiopaque BD Vialon™ catheter, needle, grip with button, flash chamber with vent plug, and barrel. The needle and catheter are protected by a needle cover.
The blood control technology is designed to stop the flow of blood from the catheter hub until the initial Luer connection is made. Once the connection is made, fluids or blood can flow through the catheter hub in either direction. The blood control technology limits blood exposure to the device user.
The 24-18 GA (0.7-1.1 mm) devices have BD Instaflash™ needle technology, allowing for immediate visualization of blood along the catheter. The flash chamber provides confirmation that the device has entered the vessel.
These devices incorporate BD Autoguard™ push-button shielding technology which is activated when the button is depressed. A spring mechanism retracts the needle and
{6}------------------------------------------------
| flash chamber into the barrel, fully encasing the needle and reducing the risk of<br>accidental needlestick injury. | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| These devices are available with or without wings. The catheter hub and wings are<br>color coded to indicate the catheter gauge size (24 GA (0.7 mm) = Yellow, 22 GA (0.9<br>mm) = Blue, 20 GA (1.1 mm) = Pink, 18 GA (1.3 mm) = Green, 16 GA (1.7 mm) =<br>Grey).<br>The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a<br>maximum pressure of 300 psi (2068 kPa). | |
| Indications for<br>Use<br>(21 CFR §<br>807.92(a)(5)) | BD Insyte™ Autoguard™ Shielded IV Catheter<br><br>BD Insyte™ Autoguard™ shielded IV catheters are intended to be inserted into a<br>patient's peripheral vascular system for short term use to sample blood, monitor blood<br>pressure, or administer fluids. These devices may also be used to facilitate the<br>placement of guidewires and other vascular access devices without pre-attached hubs.<br>These devices may be used for any patient population with consideration given to<br>adequacy of vascular anatomy, procedure being performed, fluids being infused, and<br>duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power<br>injectors set to a maximum pressure of 300 psi (2068 kPa). |
| | BD Insyte™ Autoguard™ BC Shielded IV Catheter<br><br>BD Insyte™ Autoguard™ BC shielded IV catheters are intended to be inserted into a<br>patient's peripheral vascular system for short term use to sample blood, monitor blood<br>pressure, or administer fluids. These devices may be used for any patient population<br>with consideration given to adequacy of vascular anatomy, procedure being performed,<br>fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are<br>suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa). |
| | BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter<br><br>BD Insyte™ Autoguard™ BC Pro shielded IV catheters are intended to be inserted into<br>a patient's peripheral vascular system for short term use to sample blood, monitor<br>blood pressure, or administer fluids. These devices may be used for any patient<br>population with consideration given to adequacy of vascular anatomy, procedure being<br>performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm)<br>devices are suitable for use with power injectors set to a maximum pressure of 300 psi<br>(2068 kPa). |
| Technological<br>Characteristics | Technological characteristics of the subject devices are substantially equivalent to the<br>predicate devices. The subject BD Insyte™ Autoguard™ Shielded IV Catheter, BD<br>Insyte™ Autoguard™ BC Shielded IV Catheter, and BD Insyte™ Autoguard™ BC Pro<br>Shielded IV Catheter achieve their intended use based on the same technology and<br>principles of operation as the predicate devices. |
{7}------------------------------------------------
#### bd.com
Comparisons of the subject and predicate devices technological characteristics are provided in the tables below.
{8}------------------------------------------------
#### bd.com
# Subject / Predicate Device Comparison of Intended Use, Materials and Technological Characteristics Subject Device: BD Insyte™ Autoguard™ Shielded IV Catheter
| Attribute | SUBJECT<br>BD Insyte™ Autoguard™ Shielded IV<br>Catheter | PREDICATE<br>BD Insyte™ Autoguard™ Shielded IV<br>Catheter<br>(K952861) | PREDICATE<br>BBraun Introcan® Safety™ IV<br>Catheter<br>(K021094) | Substantial Equivalence |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR §880.5200<br>21 CFR §870.1340<br>Class II<br>FOZ - Intravascular Catheter (primary)<br>DYB - Catheter Introducer (secondary) | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter | 21 CFR §880.5200<br>21 CFR §870.1340<br>Class II<br>FOZ – Intravascular Catheter<br>DYB – Catheter Introducer | Same |
| Indications for<br>Use | BD Insyte™ Autoguard™ shielded IV<br>catheters are intended to be inserted into<br>a patient's peripheral vascular system for<br>short term use to sample blood, monitor<br>blood pressure, or administer fluids.<br>These devices may also be used to<br>facilitate the placement of guidewires and<br>other vascular access devices without<br>pre-attached hubs. These devices may be<br>used for any patient population with<br>consideration given to adequacy of<br>vascular anatomy, procedure being<br>performed, fluids being infused, and<br>duration of therapy. The 22-18 GA (0.9-<br>1.3 mm) devices are suitable for use with<br>power injectors set to a maximum<br>pressure of 300 psi (2068 kPa). | An intravascular catheter is intended to<br>be inserted into the patient's vascular<br>system for short term use to sample<br>blood, monitor blood pressure, or<br>administer fluids intravenously. | The BBraun Introcan® Safety IV<br>Catheter is a passive anti-needle<br>stick device to provide venous or<br>arterial access for the infusion of<br>fluids, drugs, and/or blood<br>components, or to facilitate the<br>placement of Vascular Access<br>devices such as guidewires,<br>indwelling central venous<br>catheters, peripherally inserted<br>central catheters, and midline<br>catheters into the vascular system. | Substantially Equivalent<br>The subject device Indication for Use has<br>been modified compared to the BD Insyte™<br>Autoguard™ Shielded IV Catheter<br>predicate to clarify vascular insertion<br>location, patient population, and power<br>injection capabilities. These revisions serve<br>to clarify product use and do not change the<br>intended use of the product or introduce<br>new features or functionality for vascular<br>access; therefore, there are no new<br>questions of safety or effectiveness.<br>The subject device Indication for Use<br>includes the facilitation of guidewire and<br>other vascular access device placement<br>which does not alter the overall intended<br>use of the product with respect to the<br>BBraun predicate device. |
| Attribute | SUBJECT<br>BD Insyte ™ Autoguard™ Shielded IV<br>Catheter | PREDICATE<br>BD Insyte ™ Autoguard™ Shielded IV<br>Catheter<br>(K952861) | PREDICATE<br>BBraun Introcan® Safety™ IV<br>Catheter<br>(K021094) | Substantial Equivalence |
| Fundamental<br>Scientific<br>Technology | Single lumen with tapered tip peripheral<br>intravascular catheter designed with an<br>active needlestick safety mechanism.<br>Incorporates BD Instaflash™ technology<br>to assist with flashback visualization.<br>Facilitates placement of guidewires and<br>other vascular access devices. | Single lumen with tapered tip peripheral<br>intravascular catheter designed with an<br>active needlestick safety mechanism.<br>Incorporates BD Instaflash™ technology<br>to assist with flashback visualization. | Single lumen with tapered tip<br>peripheral intravascular catheter<br>designed with a passive needlestick<br>safety* mechanism. Facilitates<br>placement of guidewires and other<br>vascular access devices. | Substantially Equivalent<br>The subject and predicate BD Insyte™<br>Autoguard™ Safety IV Catheter active<br>needlestick safety mechanisms are identical.<br>*Passive and active needle safety<br>mechanisms are well accepted technology<br>for needlestick injury prevention in the<br>clinical setting. |
| Primary Device<br>Components /<br>Materials | Needle Cover: Polypropylene<br>Barrel: Polypropylene<br>Grip: Polycarbonate<br>Needle Hub: Polycarbonate<br>Safety Activation Button: Polycarbonate<br>Catheter Adapter: Polypropylene<br>Catheter Wedge: Stainless Steel<br>Needle: Stainless Steel<br>Catheter Tubing: Polyurethane with<br>radiopaque barium sulfate<br>Lubricants: Silicone<br>Spring: Stainless Steel<br>Porous Flow Plug: Porous Polyethylene<br>with Carboxymethyl Cellulose (CMC) | Needle Cover: Polypropylene<br>Barrel: Polycarbonate<br>Grip: Polycarbonate<br>Needle Hub: Polycarbonate<br>Safety Activation Button: Polycarbonate<br>Catheter Adapter: Polypropylene<br>Catheter Wedge: Stainless Steel<br>Needle: Stainless Steel<br>Catheter Tubing: Polyurethane with<br>radiopaque barium sulfate<br>Lubricants: Silicone<br>Spring: Stainless Steel<br>Porous Flow Plug: Porous Polyethylene<br>with Carboxymethyl Cellulose (CMC) | Needle Cover: Polypropylene<br>Catheter Tubing: Radiopaque<br>Polyurethane (PUR) or Teflon<br>(FEP)<br>Catheter Adapter: Polypropylene<br>Needle: Chrome-Nickel steel<br>Needle Hub: ABS<br>Catheter Wedge: Stainless Steel<br>Safety Clip: Stainless steel<br>Lubricants: Silicone | Substantially Equivalent<br>The subject and predicate BD Insyte™<br>Autoguard™ Shielded IV Catheter<br>materials are the same except the barrel<br>material. The change in barrel material<br>(non-patient contacting) does not alter or<br>raise new questions of safety and efficacy.<br>Subject device material differences<br>compared to the predicate BBraun<br>Introcan® Safety™ IV Catheter (K021094)<br>are non-significant relative to the catheter<br>introducer indication. Materials contacting<br>and interfacing with the guidewire or other<br>vascular access device include catheter<br>tubing, catheter adapter and catheter wedge.<br>These materials are substantially equivalent<br>to the predicate BBraun Introcan® Safety™<br>IV Catheter. |
| Attribute | SUBJECT<br>BD Insyte™ Autoguard™ Shielded IV<br>Catheter | PREDICATE<br>BD Insyte™ Autoguard™ Shielded IV<br>Catheter<br>(K952861) | PREDICATE<br>BBraun Introcan® Safety™ IV<br>Catheter<br>(K021094) | Substantial Equivalence |
| Catheter<br>Dimensions | Catheter Diameters<br>14 GA, 16 GA, 18 GA, 20 GA, 22 GA,<br>24 GA<br>Catheter Lengths<br>0.56 IN, 0.75 IN, 1.00 IN, 1.16 IN,<br>1.75 IN, 1.77 IN, 1.88 IN<br>Catheter ID<br>0.0190 IN - 0.0705 IN | Catheter Diameters<br>14 GA, 16 GA, 18 GA, 20 GA, 22 GA,<br>24 GA<br>Catheter Lengths<br>0.56 IN, 0.75 IN, 1.00 IN, 1.16 IN,<br>1.75 IN, 1.77 IN, 1.88 IN<br>Catheter ID<br>0.0190 IN - 0.0705 IN | Catheter Diameters<br>14 GA, 16 GA, 18 GA, 20 GA,<br>22 GA, 24 GA<br>Catheter Lengths<br>0.55 IN, 0.75 IN, 1.00 IN, 1.25 IN,<br>1.75 IN, 2.00 IN, 2.50 IN<br>Catheter ID<br>0.018 IN - 0.064 IN | Substantially Equivalent<br>The subject device and predicate BD<br>Insyte™ Autoguard™ Shielded IV Catheter<br>dimensions are identical. The subject and<br>predicate devices are available in<br>winged/non-winged configurations with<br>similar gauges, lengths, and sizes.<br>Catheter ID dimensions allow passage of<br>vascular access devices, including a<br>standard appropriately sized guidewire or<br>catheter through the lumen of the<br>subject/predicate devices. |
| Performance | Power Injection<br>The 22-18 GA (0.9-1.3 mm) devices are<br>suitable for use with power injectors set<br>to a maximum pressure of 300 psi (2068<br>kPa)<br>Flashback Chamber / Technology<br>Yes<br>Safety Design<br>Yes<br>Radiopaque<br>Yes<br>Must allow passage of appropriate size<br>guide wire or catheter<br>Yes | Power Injection<br>The 22-18 GA (0.9-1.3 mm) devices are<br>suitable for use with power injectors set<br>to a maximum pressure of 300 psi (2068<br>kPa)<br>Flashback Chamber / Technology<br>Yes<br>Safety Design<br>Yes<br>Radiopaque<br>Yes | Power Injection<br>The 22-18 GA power injection<br>max pressure 300 psi (2068 kPa)<br>Flashback Chamber / Technology<br>Yes<br>Safety Design<br>Yes<br>Radiopaque<br>Yes<br>Must allow passage of appropriate<br>size guide wire or catheter<br>Yes | Same |
| Attribute | SUBJECT<br>BD Insyte™ Autoguard™ Shielded IV<br>Catheter | PREDICATE<br>BD Insyte™ Autoguard™ Shielded IV<br>Catheter<br>(K952861) | PREDICATE<br>BBraun Introcan® Safety™ IV<br>Catheter<br>(K021094) | Substantial Equivalence |
| Standards<br>Compliance | ISO 594-1<br>ISO 594-2<br>ISO 10555-1<br>ISO 10555-5<br>ISO 23908<br>ISO 11070 | ISO 594-1<br>ISO 594-2<br>ISO 10555-1<br>ISO 10555-5<br>ISO 23908 | Unknown | Substantially Equivalent<br>The subject device was evaluated to current<br>FDA recognized and industry consensus<br>standards applicable to the device<br>classification and FDA product codes. |
| Attribute | SUBJECT<br>BD Insyte™ Autoguard™ BC Shielded IV Catheter<br>BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter | PREDICATE<br>BD Insyte™ Autoguard™ BC Shielded IV<br>Catheter<br>(K110443) | Substantial Equivalence | |
| Classification | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter | Same | |
| Indications for Use | BD Insyte™ Autoguard™ BC Shielded IV Catheter:<br>BD Insyte™ Autoguard™ BC shielded IV catheters are<br>intended to be inserted into a patient's peripheral vascular<br>system for short term use to sample blood, monitor blood<br>pressure, or administer fluids. These devices may be used for<br>any patient population with consideration given to adequacy of<br>vascular anatomy, procedure being performed, fluids being<br>infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm)<br>devices are suitable for use with power injectors set to a<br>maximum pressure of 300 psi (2068 kPa).<br>BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter:<br>BD Insyte™ Autoguard™ BC Pro shielded IV catheters are<br>intended to be inserted into a patient's peripheral vascular<br>system for short term use to sample blood, monitor blood<br>pressure, or administer fluids. These devices may be used for<br>any patient population with consideration given to adequacy of<br>vascular anatomy, procedure being performed, fluids being<br>infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm)<br>devices are suitable for use with power injectors set to a<br>maximum pressure of 300 psi (2068 kPa). | The BD Insyte™ Autoguard™ BC catheter is<br>inserted into a patient's vascular system to sample<br>blood, monitor blood pressure, or administer<br>fluids. | Substantially Equivalent<br>The subject device Indication for Use has been<br>modified to clarify vascular insertion location,<br>duration of use, patient population, power<br>injection capabilities and to align with current<br>industry standards and FDA expectations<br>These revisions serve to clarify product use and<br>do not change the intended use of the product or<br>introduce new features or functionality;<br>therefore, there are no new questions of safety<br>or effectiveness when considering a claim of<br>substantial equivalence. | |
| Attribute | SUBJECT<br>BD Insyte™ Autoguard™ BC Shielded IV Catheter<br>BD Insyte™ Autoguard™ BC Pro Shielded IV Catheter…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.