K201031 · International Medical Industries, Inc. · QEH · Dec 16, 2020 · General Hospital
Device Facts
Record ID
K201031
Device Name
NRFit® Caps, Male and Female Neuraxial Tip Caps
Applicant
International Medical Industries, Inc.
Product Code
QEH · General Hospital
Decision Date
Dec 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The NRFit® cap is a sterile, single-use device intended to provide positive closure to prevent fluid loss and contamination in Neuraxial devices CADD Cassettes or Syringes utilizing NRFit® connections. The NRFit® cap is intended to be used for less than 24 hours on any single application.
Device Story
NRFit® Caps are sterile, single-use polypropylene caps designed to provide positive closure for neuraxial devices, CADD cassettes, or syringes featuring ISO 80369-6 compliant connections. Available in male and female configurations, the caps prevent fluid loss and contamination during clinical use. Operated by hospital or clinical personnel in general hospital settings, the caps are applied manually to the neuraxial device port. The device functions as a mechanical seal; it does not involve electronic processing or automated decision-making. By maintaining a secure connection, the caps support aseptic technique and prevent leakage, benefiting patients by ensuring the integrity of neuraxial delivery systems during short-term applications (less than 24 hours).
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 80369-6/20 performance testing, biocompatibility (cytotoxicity, sensitization, irritation, hemolysis, systemic toxicity, pyrogenicity), and packaging integrity (ASTM F1886).
Technological Characteristics
Medical grade polypropylene with yellow colorant. ISO 80369-6 compliant male/female neuraxial lock connections. Externally communicating, tissue/bone/dentin contacting. Sterilized via Ethylene Oxide. Single-use. Packaging integrity tested per ASTM F1886:2013.
Indications for Use
Indicated for use as a sterile, single-use closure for neuraxial devices, CADD cassettes, or syringes with NRFit® connections to prevent fluid loss and contamination for durations under 24 hours.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
December 16, 2020
International Medical Industries, Inc. David Meily Director of Quality & Regulatory Affairs 2981 Gateway Drive Pompano Beach, Florida 33069
Re: K201031
Trade/Device Name: NRFit® Caps, Male and Female Neuraxial Tip Caps Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QEH Dated: November 18, 2020 Received: November 19, 2020
Dear David Meily:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K201031
#### Device Name
NRFit® Caps, Male and Female Neuraxial Tip Caps
Indications for Use (Describe)
The NRFit® cap is a sterile, single-use device intended to provide positive closure to prevent fluid loss and contamination in Neuraxial devices CADD Cassettes or Syringes utilizing NRFit® connections. The NRFit® cap is intended to be used for less than 24 hours on any single application.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary NRFit® Caps, Male and Female Neuraxial Tip Caps
## K201031
#### 1. Submission Sponsor
International Medical Industries, Inc.
2981 Gateway Drive.
Pompano Beach
FL, 33069
USA
Contact: David Meily
Title: Director of Quality & Regulatory Affairs
### 2. Submission Correspondent
No applicable
#### 3. Date Prepared
December 16, 2020
### 4. Device Identification
| Trade/Proprietary Name: | NRFit® Caps, Male and Female Neuraxial Tip Caps |
|-------------------------|-------------------------------------------------|
| Common/Usual Name: | syringe, piston |
| Classification Name: | syringe, piston |
| Regulation Number: | 21 CFR§880.5860 |
| Product Code: | QEH |
| Device Class: | Class II |
| Classification Panel: | General Hospital |
## 5. Legally Marketed Predicate Device(s)
#### Predicate Device
{4}------------------------------------------------
#### K170900
| Trade/Proprietary Name: | DASH 6® NRFit Syringe Cap |
|-------------------------|---------------------------|
| Common/Usual Name: | syringe, Piston |
| Classification Name: | syringe, piston |
| Regulation Number: | 21 CFR 880.5860 |
| Product Code: | FMF |
| Device Class: | Class II |
| Classification Panel: | General Hospital |
## 6. Indication for Use Statement
The NRFit® cap is a sterile, single-use device intended to provide positive closure to prevent fluid loss and contamination in Neuraxial devices CADD Cassettes or Syringes utilizing NRFit® connections.
NRFit® cap is intended to be used for less than 24 hours on any single application.
### 7. Device Description
The NRFit® Caps (abbreviated as "NRFit") are ISO 80369-6 compliant caps designed to prevent fluid loss in Neuraxial devices. The NRFit cap is made from medical grade polypropylene with yellow colorant. The NRFit cap is available with either a male neuraxial lock connection or a female neuraxial lock connection. The NRFit cap is made of materials appropriate for a limited exposure duration, externally communicating, tissue/bone/dentin contacting medical device.
### 8. Substantial Equivalence Discussion
Based on the above information and FDA Guidance Document "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications", the proposed NRFit caps is concluded to be substantially equivalent to its predicate, DASH 6 NRFit syringe caps.
| Manufacturer | International Medical Industries,<br>Inc. | INTERVENE Group Limited | |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Proprietary Device<br>Name | NRFit® Caps, Male and Female<br>Neuraxial Tip Caps | DASH 6® NRFit™ Syringe Caps | Comparison |
| 510(k) # | K201031 | K170900 | N/A |
| Indications for Use | The NRFit® cap is a sterile, single-use device intended to provide positive closure to prevent fluid loss and contamination in Neuraxial devices CADD Cassettes | DASH 6® NRFit Syringe cap is intended to be used with ISO 80369-6 NRFit neuraxial syringes for sealing the tip of the syringe. | Similar, both are used with neuraxial devices |
| Table 5-A: NRFit Caps and Predicate Device Comparison | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | International Medical Industries,<br>Inc. | INTERVENE Group Limited | Comparison |
| Proprietary Device<br>Name | NRFit® Caps, Male and Female<br>Neuraxial Tip Caps | DASH 6® NRFit™ Syringe Caps | Comparison |
| Product Code | or Syringes utilizing NRFit®<br>connections. NRFit® cap is<br>intended to be used for less than<br>24 hours on any single<br>application. | FMF | New product<br>code for<br>neuraxial<br>devices |
| Regulation No. | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Classification | Class II | Class II | Same |
| Device Description | The NRFit® Caps are ISO 80369-6<br>compliant caps designed to<br>prevent fluid loss in Neuraxial<br>devices. The NRFit® cap is made<br>from medical grade<br>polypropylene with yellow<br>colorant. The NRFit® cap is<br>available with either a male<br>neuraxial lock connection or a<br>female neuraxial lock connection.<br>The NRFit® cap is made of<br>materials appropriate for a<br>limited exposure duration,<br>externally communicating,<br>tissue/bone/dentin contacting<br>medical device < 24 contact<br>duration | DASH 6® NRFit Syringe Caps are<br>Single Use, In-Hospital devices. It<br>incorporates a female ISO 80369-6<br>NRFit connector to be connected<br>on an ISO 80369-6 compliant male<br>port. The device will be supplied as<br>individually packed (sterile) and the<br>bulk packed (non-sterile). The<br>sterile packed items will be<br>supplied directly to the user and<br>the bulk packed will be supplied to<br>Anesthetic Conduction Kit<br>manufacturers to be packaged into<br>kit. The devices packaging will<br>indicate it is a syringe cap and the<br>sterility/sterilization method the<br>device has undergone. | Similar.<br>IMI's<br>description is<br>based on<br>function,<br>material<br>composition<br>and patient<br>contact.<br>INTERVENE's<br>description is<br>of packaging<br>and sterility.<br>Both<br>companies<br>agree on<br>compliance to<br>ISO 80369-6 |
| Environment | General Hospital | General Hospital | Same |
| User | Hospital/clinical personnel | Hospital/clinical personnel | Same |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same |
| Number of Uses | Single Use | Single Use | Same |
| Cap Material | Polypropylene (PP) | Polypropylene (PP) | Same |
| Biocompatibility<br>Contact and Duration | External communicating, blood<br>path (indirect) for a duration of<br><24 Hrs | External communicating, blood<br>path (indirect) for a duration of ≤24<br>Hrs | Same |
{5}------------------------------------------------
The predicate device is a Syringe Family and Accessories. The NRFit® caps are substantially equivalent to the accessory- DASH 6® NRFit Syringe caps.
Note: Based on the comparison above, there are no new or different questions of safety or effectiveness.
{6}------------------------------------------------
# 9. Non-Clinical Performance Data
| Table 5-B: Summary of NRFit® Caps Testing | | |
|-------------------------------------------|--|--|
|-------------------------------------------|--|--|
| Test Name | Test Description |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Falling Drop Liquid Leakage | NRFit® device shall show no signs of leakage sufficient to form a<br>falling drop of water |
| Sub-atmospheric-pressure air leakage | NRFit® Device shall not leak by more than 0.005 Pa (m^3)/s<br>during aspiration |
| Stress Cracking | NRFit® Device shall show no evidence of stress cracking (i.e.<br>visible cracks in the threads or taper of the connector) or<br>leakage after Stress Crack testing |
| Axial Separation | NRFit® Device shall not separate from a properly installed<br>reference connector while being subjected to at least 35N<br>disconnection tensile axial force |
| Resistance to Unscrewing | NRFit® Device shall not separate from a properly installed<br>reference connector |
| Resistance to Overriding | NRFit® device shall not override the threads or lugs of the<br>reference connector |
| Visual | NRFit® device shall be a similar yellow in color to other NRFit<br>products available on the open market |
| Visual Inspection | Device must be available in a primary package blister<br>configuration consisting of PETG film with Tyvek Lid. |
| Visual Inspection | Heat seals must meet ASTM F1886:2013 Standard test method<br>for determining integrity of seals for flexible packaging by visual<br>inspection |
| Wearing a set of disposable nitrile gloves,<br>run a finger along the outer edge of the<br>packaging and ensure glove does not catch<br>on any feature such that a hole or tear is<br>formed | Primary packaging must not expose the user to sharps that may<br>cause injury or a breach in aseptic technique |
| Visual Inspection | Product outer packaging must be free from gross cosmetic<br>defects, particulate, and foreign material |
| Transportation, Shelf Life | Simulated transportation conditions and shelf life followed by<br>design verification testing. |
| Product Validation | Human factors and Usability |
| Biocompatibility | Cytotoxicity, Intracutaneous Irritation testing, Sensitization<br>testing, Hemolysis testing direct/indirect, Acute systemic<br>toxicity, Pyrogenicity, Chemical Characterization, Extractables<br>and Leachables, Toxicological Risk assessment. |
| Performance | ISO 80369-6:2016 and ISO 80369-20:2015 |
{7}------------------------------------------------
## 10. Clinical Performance Data
There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device.
### 11. Statement of Substantial Equivalence
The male and female NRFit® caps have the same intended use, function, materials and patient contact as the predicate device. Both devices are compliant to ISO 80369-6. Minor technological differences do not raise new or different questions of safety and effectiveness.
The NRFit® caps, as designed and manufactured, are determined to be substantially equivalent to the referenced predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.