The CHIRAVAC™ Blood Collection Needles are single use and are intended for venous blood collection in connection with blood collection tubes.
Device Story
Device is a double-ended hollow stainless steel cannula with a threaded polypropylene hub and a rubber sleeve; used for venous blood collection. Healthcare professional punctures vein with intravenous end; attaches evacuated blood collection tube to non-patient end. Rubber sleeve acts as a non-return valve, allowing multiple tube changes during a single venipuncture. Device is supplied sterile (ethylene oxide). No electronic components or software.
Clinical Evidence
Bench testing only. Testing included ISO 7864 (sterile hypodermic needles), ISO 6009 (color coding), ISO 9626 (stainless steel tubing), and internal procedures for vacuum function and sleeve integrity. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity), ASTM F756 (hemocompatibility), and European Pharmacopoeia 10.0 (pyrogenicity). Shelf-life validated to 5 years via ASTM F1980.
Technological Characteristics
Materials: 304 stainless steel cannula, polypropylene hub/caps, rubber sleeve, silicone lubricant. Dimensions: 18G-23G gauges; 1"-1.5" lengths. Principle: Manual venipuncture with vacuum-assisted blood draw. Sterilization: Ethylene oxide (SAL 10^-6). No software or energy source.
Indications for Use
Indicated for venous blood collection in patients requiring blood specimen collection via vacuum tubes. Intended for single use.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
July 21, 2021
CHIRANA T. Injecta % Nathan Wright Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K201009
Trade/Device Name: CHIRAVAC™ Blood Collection Needles Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA, FMI Dated: June 16, 2021 Received: June 21, 2021
Dear Nathan Wright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201009
Device Name CHIRAVAC™ Blood Collection Needles
Indications for Use (Describe)
The CHIRAVAC™ Blood Collection Needles are single use and are intended for venous blood collection in connection with blood collection tubes.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K201009 - 510K SUMMARY
| Submitter's Name: | CHIRANA T. Injecta |
|----------------------------|--------------------------------------------------------|
| Submitter's Address: | Nám. Dr. Schweitzera 194 |
| | Stará Turá 916 01 |
| | Slovak Republic |
| Submitter's Telephone: | 421-32-770-9972 |
| Contact Person: | Nathan Wright, MS |
| | Empirical Testing Corp. |
| | 719-351-0248 |
| | nwright@empiricaltech.com |
| Date Summary was Prepared: | July 14, 2021 |
| Trade or Proprietary Name: | CHIRAVACTM Blood Collection Needles |
| Common or Usual Name: | Blood specimen collection device |
| Classification: | Class II per 21 CFR §862.1675 |
| Product Code: | JKA, FMI |
| Classification Panel: | Office of In Vitro Diagnostics and Radiological Health |
#### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The CHIRAVAC™ Blood Collection Needles consists of a double-ended hollow stainless steel thin-walled cannula, a threaded polypropylene hub, a sleeve that interrupts blood flow between filling multiple tubes, and polypropylene caps. The device is single use and supplied sterile (ethylene oxide sterilized). The stainless steel cannula and silicone cannula lubricants are classified as externally communicating with limited (≤24 hours) duration contact; all other device components have no contact with patient blood, tissue, or skin. Once the vein has been accessed by puncturing with the intravenous end of the cannula, the healthcare professional will begin the blood collection process by placing evacuated blood collection tubes on the cannula's non-patient end, where the sleeve acts as a non-return valve and allows for more than one tube to be used to collect blood from the vein puncture.
| Size Metric<br>(mm) | Size (G x ") | Color Code | Pieces per Box<br>(pcs) | Pieces per Case<br>(pcs) | Item Code |
|---------------------|--------------|------------|-------------------------|--------------------------|------------|
| 0.6x25 | 23G x 1" | Blue | 100 | 1 000 | CHBCN23100 |
| 0.6x38 | 23G x 1.5" | Blue | 100 | 1 000 | CHBCN23112 |
| 0.7x25 | 22G x 1" | Black | 100 | 1 000 | CHBCN22100 |
| 0.7x38 | 22G x 1.5" | Black | 100 | 1 000 | CHBCN22112 |
| 0.8x25 | 21G x 1" | Green | 100 | 1 000 | CHBCN21100 |
| 0.8x38 | 21G x 1.5" | Green | 100 | 1 000 | CHBCN21112 |
| 0.9x25 | 20G x 1" | Yellow | 100 | 1 000 | CHBCN20100 |
| 0.9x38 | 20G x 1.5" | Yellow | 100 | 1 000 | CHBCN20112 |
| 1.2x25 | 18G x 1" | Pink | 100 | 1 000 | CHBCN18100 |
| 1.2x38 | 18G x 1.5" | Pink | 100 | 1 000 | CHBCN18112 |
#### Table 5-1 CHIRAVACTM Blood Collection Needles
#### INDICATIONS FOR USE
The CHIRAVAC™ Blood Collection Needles are single use and are intended for venous blood collection in connection with blood collection tubes.
{4}------------------------------------------------
The indications for use for the CHIRAVAC™ Blood Collection Needles are similar to the indications of the listed predicate devices.
| | Subject | Predicate |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | CHIRAVACTM Blood Collection<br>Needles<br>(Subject - K201009) | Sol-M Blood Collection Needles, Sol-Care Safety<br>Blood Collection Needles, Sol-Care Safety Blood<br>Collection Needle with Holder, Sol-M Blood<br>Collection Set<br>(K182146) |
| Indications<br>for Use | The CHIRAVACTM Blood<br>Collection Needles are single use<br>and are intended for venous blood<br>collection in connection with blood<br>collection tubes. | The Blood Collecting Needle is intended to be used<br>with vacuum blood collection tube for the collection of<br>venous blood.<br>The Safety Blood Collecting Needle is intended to be<br>used with vacuum blood collection tube for the<br>collection of venous blood. The safety shield is<br>intended to aid in the protection against accidental<br>needle stick injury. The Blood Collecting Set with<br>Holder is intended to be used with vacuum blood<br>collection tube for the collection of venous blood. |
## Discussion of differences in Indications for Use Statements
The subject and predicate device have indication for use in collection of blood from the veins. In addition to their blood collection needles, the predicate offers safety blood collecting needles and a blood collecting set with holder with particular indications for those devices that do not apply to the subject.
## TECHNICAL CHARACTERISTICS COMPARISON
The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are similar between the subject and predicates:
- Indications for Use .
- Sizes .
- Materials of manufacture
- Principles of Operations .
| Table 5-2: Predicate Devices | | | |
|------------------------------|--|--|--|
|------------------------------|--|--|--|
| 510k<br>Number | Trade or Proprietary or Model Name | Manufacturer | Predicate Type |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|----------------|
| K182146 | Sol-M Blood Collection Needles, Sol-Care Safety Blood Collection Needles, Sol-Care Safety Blood Collection Needle with Holder, Sol-M Blood Collection Set | Sol-Millennium Medical, Inc. | Primary |
{5}------------------------------------------------
| | CHIRAVACTM Blood Collection<br>Needles<br>(Subject - K201009) | Sol-M Blood Collection Needles,<br>Sol-Care Safety Blood Collection<br>Needles, Sol-Care Safety Blood<br>Collection Needle with Holder,<br>Sol-M Blood Collection Set<br>(K182146) | Discussion/<br>Comment |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation<br>Number | 21 CFR § 880.5570 | 21 CFR § 880.5570 | Same |
| Classification<br>Name | Needle, Hypodermic, Single-Lumen | Needle, Hypodermic, Single-Lumen | Same |
| Regulatory<br>Class | Class II | Class II | Same |
| Product Code | JKA, FMI | FMI | Same. JKA is current<br>product code for blood<br>collection needles. |
| Intended<br>Use/Indications<br>for Use | The CHIRAVACTM Blood<br>Collection Needles are single use<br>and are intended for venous blood<br>collection in connection with blood<br>collection tubes. | The Blood Collecting Needle is<br>intended to be used with vacuum<br>blood collection tube for the<br>collection of venous blood.<br>The Safety Blood Collecting Needle<br>is intended to be used with vacuum<br>blood collection tube for the<br>collection of venous blood. The<br>safety shield is intended to aid in the<br>protection against accidental needle<br>stick injury. The Blood Collecting<br>Set with Holder is intended to be<br>used with vacuum blood collection<br>tube for the collection of venous<br>blood. | Similar |
| Principles of<br>Operation | Needles accesses the vein by<br>puncturing with the intravenous end<br>of the cannula. With the non-patient<br>end of the cannula attached to the<br>blood collection tube, the blood is<br>drawn via the piston vacuum into<br>the syringe. The sleeve acts as a<br>non-return valve. | The blood collection devices form a<br>channel between patient's vein and<br>the vacuum blood collection tube<br>which draw the blood to the blood<br>collection tube. | Same |
| Color Coding | Pink - 18G<br>Yellow - 20G<br>Green - 21G<br>Black - 22G<br>Blue - 23G | Pink - 18G<br>Yellow - 20G<br>Green - 21G<br>Black - 22G<br>Blue - 23G | Same |
| Single Use? | Yes | Yes | Same |
| Cannula<br>Material | 304 Stainless Steel | 304 Stainless Steel | Same |
| Other Needle<br>Materials | Polypropylene (needle hub and<br>protective caps), rubber (sleeve),<br>glue. | Polypropylene; Dialyzing Paper;<br>Glue; Rubber | Same. Predicates include<br>safety shields with dialyzing<br>paper. The subject does not<br>have dialyzing paper and<br>does not require a safety<br>shield because it comes with<br>a protective cap which is<br>removed prior to use. |
| Cannula<br>Lubricant | Silicone Oil | Silicone Oil | Same |
| Cannula Length | 1", 1.5" | 5/8", ¾", 1", 1.25", 1.5", 1.75", 2" | Same; additional lengths<br>offered by predicate not<br>offered by subject. |
| | CHIRAVACTM Blood Collection<br>Needles<br>(Subject - K201009) | Sol-M Blood Collection Needles,<br>Sol-Care Safety Blood Collection<br>Needles, Sol-Care Safety Blood<br>Collection Needle with Holder.<br>Sol-M Blood Collection Set<br>(K182146) | Discussion/<br>Comment |
| Cannula Gauge | 18G, 20G, 21G, 22G, 23G | 18G, 19G, 20G, 21G, 22G, 23G,<br>25G | Same; additional gauges<br>offered by predicate not<br>offered by subject. |
| Biocompatibility | Per ISO 10993 | Per ISO 10993 | Same |
| Sterility | Ethylene Oxide to SAL of 10-6 | Ethylene Oxide to SAL of 10-6 | Same |
Table 5-3: Predicate Comparison (CHIRAVAC™ Blood Collection Needles)
{6}------------------------------------------------
Discussion of differences in Technological characteristics.
- The predicate device includes safety shields with dialyzing paper. The subject does ● not require a safety shield because it comes with a protective cap which is removed prior to use and therefore does not have dialyzing paper like the predicates.
- All subject sizes (cannula lengths and gauge options) are sizes available by the ● predicates; however, not all sizes offered by the predicates are available in the subject needles.
# PERFORMANCE TESTING SUMMARY
The CHIRAVAC™ Blood Collection Needles have been tested in the following test modes:
## A. Performance Testing - Bench
- ISO 7864:2016 Sterile hypodermic needles for single use-Requirements and test . methods.
- Color Coding o
- Size Designation O
- Cleanliness O
- pH Limits о
- Limits of Extractable Metals O
- Inspection of Needle Length O
- Inspection of Defects O
- Inspection of Lubrication O
- Needle/Hub Bond Strength O
- Lumen Patency O
- Fragmentation O
- Inspection of Needle Sharpness/Cleanness о
- o Penetration Force Test
- ISO 6009:2016 Hypodermic needles for single use. ●
- o Color coding
- ISO 9626: 2016 Stainless steel needle tubing for the manufacture of medical devices -. Requirements and test methods
- o pH Limits
{7}------------------------------------------------
- Inspection of Surface Appearance o
- O Inspection of Tubing Dimensions
- Stiffness Test O
- Breakage Resistance Test о
- Corrosion Resistance Test O
- Internal Procedure SOP 4113 ●
- Visual Inspection of Labeling о
- Internal Procedure SOP 4401 ●
- Inspection of Blister/Label Perforation, O
- Vacuum Function Evaluation o
- Test for Rubber Sleeve Returning to Original Position without Damage from o Repeated Use
- Internal Procedure SOP 4405
- Inspection of Cannula Transparency о
- o Cleanliness
# B. Biocompatibility
The stainless steel cannula and silicone cannula lubricants are classified as externally communicating with limited (<24 hours) duration contact; all other device components have limited (<24 hours) duration surface contact with intact skin of the health care professional user and the patient.
- Cytotoxicity per ISO 10993-5 Biological evaluation of medical devices Part 5: Tests for ● in vitro cytotoxicity.
- Sensitization per ISO 10993-10 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
- . Irritation per ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- Acute Systemic Toxicity per ISO 10993-11 Biological evaluation of medical devices -Part 11: Tests for systemic toxicity
- Hemocompatibility per ASTM F756 Standard Practice for Assessment of Hemolytic ● Properties of Materials
- Material Mediated pyrogenicity per European Pharmacopoeia 10.0 Article 2.6.8 Pyrogens .
C. Sterilization, Shipping and Shelf-life
- Packaging integrity test per internal procedures ●
- Sterile Barrier Packaging test done per ISO 7886-1 Sterile hypodermic syringes for single use ● – Part 1: Syringes for manual use
- . Shelf life of 5 years - validated by accelerated aging per ASTM F1980 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
# CONCLUSION
The overall technology characteristics and testing performance data lead to the conclusion that the CHIRAVAC™ Blood Collection Needles are substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.