← Product Code [KWQ](/productcode/KWQ) · K200995

# Insignia™ Anterior Cervical Plate System (K200995)

_Alphatec Spine, Inc. · KWQ · May 21, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200995

## Device Facts

- **Applicant:** Alphatec Spine, Inc.
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** May 21, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Insignia™ Anterior Cervical Plate System is intended for anterior screw fixation to the (C2-T1) for the following indications: degenerative disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis, lordosis or scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, and spinal stenosis.

## Device Story

Insignia Anterior Cervical Plate System provides anterior spinal fixation; consists of titanium alloy plates and screws. Used by surgeons in clinical settings to stabilize cervical spine (C2-T1) segments. Device components are implanted during surgical procedures to address degenerative disease, trauma, tumors, deformity, pseudarthrosis, failed fusions, spondylolisthesis, or spinal stenosis. System includes associated surgical instruments. Output is mechanical stabilization of the cervical spine; benefits patient by facilitating fusion and correcting spinal alignment.

## Clinical Evidence

No clinical data. Substantial equivalence determination based on bench testing only, including static compression, static torsion, and dynamic compression testing per ASTM F1717.

## Technological Characteristics

Titanium alloy (ASTM F136) plates and screws. Mechanical fixation device. Non-sterile. No software or electronic components.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Tempus Anterior Cervical Plate System ([K131374](/device/K131374.md))
- Trestle Luxe Anterior Cervical Plating System ([K102820](/device/K102820.md))
- MaxAn Anterior Cervical Plate System ([K133518](/device/K133518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

May 21, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Alphatec Spine, Inc. Sandy Gill Regulatory Affairs Associate 5818 El Camino Real Carlsbad, California 92008

Re: K200995

Trade/Device Name: Insignia™ Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: April 15, 2020 Received: April 16, 2020

Dear Sandy Gill:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K200995

Device Name InsigniaTM Anterior Cervical Plate System

### Indications for Use (Describe)

The Insignia™ Anterior Cervical Plate System is intended for anterior screw fixation to the (C2-T1) for the following indications: degenerative disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis, lordosis or scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, and spinal stenosis.

Type of Use (Select one or both, as applicable)

| <div> <span style="font-size:10pt">       Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> |
|-----------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:10pt">       Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div>  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.

| I. | SUBMITTER:             | Alphatec Spine, Inc.<br>5818 El Camino Real<br>Carlsbad, CA 92008<br>Phone: (760) 431-9286<br>Fax: (760) 431-0289 |
|----|------------------------|-------------------------------------------------------------------------------------------------------------------|
|    | Contact Person:        | Sandy Gill<br>Regulatory Affairs Associate<br>Contact Phone: (760) 431-9286                                       |
|    | Date Summary Prepared: | April 14, 2020                                                                                                    |

#### DEVICE II.

| Name of Device:       | Insignia™ Anterior Cervical Plate System                          |
|-----------------------|-------------------------------------------------------------------|
| Common or Usual Name: | Appliance, Fixation, Spinal Intervertebral Body                   |
| Classification Name:  | Spinal Intervertebral Body Fixation Orthosis<br>(21 CFR 888.3060) |
| Regulatory Class:     | Class II                                                          |
| Product Code:         | KWQ                                                               |

#### LEGALLY MARKETED PREDICATE DEVICES III.

| 510(k)                      | Product Code | Trade Name                                    | Manufacturer         |
|-----------------------------|--------------|-----------------------------------------------|----------------------|
| Primary Predicate Device    |              |                                               |                      |
| K131374                     | KWQ          | Tempus Anterior Cervical Plate System         | NeuroStructure, LLC  |
| Additional Predicate Device |              |                                               |                      |
| K102820                     | KWQ          | Trestle Luxe Anterior Cervical Plating System | Alphatec Spine, Inc. |
| K133518                     | KWQ          | MaxAn Anterior Cervical Plate System          | Zimmer Biomet        |

#### IV. DEVICE DESCRIPTION

The Insignia Anterior Cervical Plate System is intended for anterior fixation to the cervical spine. The Insignia Anterior Cervical Plate System consists of a variety of sizes of plates and screws that are manufactured from titanium alloy conforming to ASTM F136 and are offered non-sterile. The instruments in this system are intended for use in surgical procedures.

{4}------------------------------------------------

#### V. INDICATIONS FOR USE

The Insignia™ Anterior Cervical Plate System is intended for anterior screw fixation to the cervical spine (C2-T1) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (kyphosis, lordosis or scoliosis), pseudarthrosis, failed previous fusions, spondylolisthesis, and spinal stenosis.

#### VI. TECHNOLOGICAL COMPARISON TO PREDICATES

The technological design features of the subject implants were compared to the predicates in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent.

#### VII. PERFORMANCE DATA

Nonclinical testing was presented on the Insignia Anterior Cervical Plate System using the performance criteria proposed in Draft Guidance for Industry and Food and Drug Administration Staff, Spinal Plating Systems- Performance Criteria for Safety and Performance Based Pathway, issued on September 20, 2019. The following testing was performed:

- o Static compression testing per ASTM F1717
- Static torsion testing per ASTM F1717 0
- Dynamic compression testing per ASTM F1717 ●

The results demonstrate that the subject Insignia Anterior Cervical Plate System is substantially equivalent to other predicate devices for nonclinical testing.

# Clinical Information

Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.

#### CONCLUSION VIII.

Based upon the information provided in this 510(k) submission, it has been determined that the subject devices are substantially equivalent to legally marketed devices in regards to indications for use, intended use, design, technology, and performance.

---

**Source:** [https://fda-staging.innolitics.com/device/K200995](https://fda-staging.innolitics.com/device/K200995)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).