Rhythmlink International Subdermal Needle Electrodes are intended for use with recording, monitoring and stimulation equipment for the purpose of recording of biopotential signals. Examples include: Electromyography (EMG), Electroencephalography (EEG), and Nerve potential signals. The electrodes are sterile and for single patient use.
Device Story
Guardian Needle Electrodes are sterile, single-use subdermal needle electrodes used for recording biopotential signals (EMG, EEG, nerve potentials) and stimulation. The device consists of a 304VM stainless steel needle, PVC-coated leadwire, and a 1.5mm DIN 42802 touch-proof connector. The subject device includes a thermoplastic polyester elastomer needle sheath and adhesive tape for securing the electrode to the patient, features not present in the predicate. Used in hospital settings, the device is inserted subcutaneously by a physician or practitioner. The electrodes interface with neurophysiology monitoring equipment to capture electrical signals from muscle or nerve tissue. The output is displayed on the monitoring equipment, assisting clinicians in diagnostic or intraoperative monitoring. Benefits include reliable signal acquisition for neurophysiological assessment.
Clinical Evidence
No clinical data. Bench testing only, including electrical continuity, adhesion testing, and biocompatibility evaluation per ISO 10993.
Technological Characteristics
Needle: 304VM stainless steel (0.4mm diameter, 7-22mm length). Leadwire: PVC-coated. Connector: 1.5mm DIN 42802. Sheath: Thermoplastic polyester elastomer. Sterilization: EtO. Standards: IEC 60601-1, DIN 42802, ISO 14971, ISO 15223-1, ISO 10993.
Indications for Use
Indicated for patients requiring biopotential signal recording, monitoring, or stimulation during procedures such as EMG, EEG, and nerve potential studies. For use by licensed physicians or practitioners.
Regulatory Classification
Identification
A needle electrode is a device which is placed subcutaneously to stimulate or to record electrical signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 27, 2020
Rhythmlink International, LLC Gabriel Orsinger Vice President of Engineering and R&D 1140 First Street South Columbia, South Carolina 29209
Re: K200984
Trade/Device Name: Guardian Needle Electrode Regulation Number: 21 CFR 882.1350 Regulation Name: Needle Electrode Regulatory Class: Class II Product Code: GXZ Dated: April 9, 2020 Received: April 14, 2020
Dear Gabriel Orsinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name Guardian Needle™ Electrode
#### Indications for Use (Describe)
Rhythmlink International Subdermal Needle Electrodes are intended for use with recording, monitoring and stimulation equipment for the purpose of recording of biopotential signals. Examples include: Electromyography (EMG), Electroencephalography (EEG), and Nerve potential signals. The electrodes are sterile and for single patient use.
Type of Use (*Select one or both, as applicable*)
| <span> <span style="text-decoration:overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Guardian Needle™ Electrode Rhythmlink International, LLC
{3}------------------------------------------------
# 510(k) Summary – K200984
| 807.92(a)(1)<br>Submitter<br>Information | Rhythmlink International, LLC<br>1140 First Street South<br>Columbia, SC 29209<br>Phone: 803-252-1222<br>FDA Registration #: 1067162 | |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official<br>Correspondent | Gabriel Orsinger, PhD<br>Vice President of Engineering and R&D<br>Email: gorsinger@rhythmlink.com<br>Phone: 803-365-9664 | |
| Summary Date | May 27, 2020 | |
| 807.92(a)(2)<br>Device<br>Identification | Device Trade Name:<br>Classification Name:<br>Product Code:<br>Classification:<br>Classification Panel: | Guardian Needle™ Electrode<br>Needle Electrode<br>GXZ<br>21 CFR 882.1350 Class II<br>Neurology |
| 807.92(a)(3)<br>Predicate Device | Device Trade Name:<br>Classification Name:<br>510(k) Number:<br>Product Code:<br>Classification:<br>Classification Panel:<br>Manufacturer: | Subdermal Needle Electrodes<br>Needle Electrode<br>K022914<br>GXZ<br>21 CFR 882.1350 Class II<br>Neurology<br>Rhythmlink International, LLC<br>1140 First Street South<br>Columbia, SC 29209 |
| 807.92(a)(4)<br>Device Description | Guardian Needle™ Electrodes are a sterile, single-use device. The<br>electrodes are applied during the study of biopotentials such as<br>electromyography (EMG), electroencephalography (EEG), nerve conduction<br>and stimulation/response. The electrodes are invasive as they are placed<br>subcutaneously or in contact with nerve or muscle tissue. The needle is<br>housed inside a sheath until application and secured to the patient during use<br>with adhesive tape. | |
| 807.92(a)(5)<br>Intended Use | Rhythmlink International Guardian Needle™ Electrodes are intended<br>for use with recording, monitoring and stimulation equipment for the<br>purpose of recording of biopotential signals. Examples include:<br>Electromyography (EMG), Electroencephalography (EEG) and Nerve<br>potential signals. The electrodes are sterile and for single patient use<br>only | |
| 807.92(a)(6)<br>Technological<br>Characteristics | The technological characteristics of Guardian Needle™ Electrodes are<br>identical to the predicate device (K022914), with the addition of a needle<br>sheath and adhesive tape, both of which have been assessed to not change the<br>fundamental scientific technology, intended use, performance, or safety and<br>effectiveness of the predicate device (reference Substantial Equivalence of | |
{4}------------------------------------------------
| Technological Characteristics table, below). The test methods were identical<br>to those used to assess the predicate device. |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>807.92(b)(1)<br/>Summary of Non-<br/>Clinical Tests</b> |
| Guardian Needle™ Electrodes are substantially equivalent in technology,<br>safety, and effectiveness as the predicate device (K022914), as demonstrated<br>by the test results. |
| The Guardian Needle™ Electrode was assessed by applying the TIR 28:2009<br>Guidance, from which it was determined that the subject device does not<br>challenge the Submitter's Master Challenge Device and can thus fall under<br>the previously validated EtO sterilization cycle. Reevaluation of the EtO<br>residuals was performed to confirm low levels of EtO and Ethylene<br>Chlorohydrin (ECH) after a 24-hour aeration time below the maximum limits<br>described in ANSI/AAMI/ISO 10993-7:2008/(R) 2012. |
| The biocompatibility evaluation was conducted within the risk management<br>framework and in compliance with ISO 10993 standards. This<br>biocompatibility evaluation establishes the biological safety for the Guardian<br>Needle™ Electrode with intact skin contact for a limited (≤24 hours)<br>duration |
| Functional performance equivalency was determined by electrical and<br>adhesive benchtop testing, as follows: |
| Electrical Continuity Adhesion Testing |
| Benchtop performance testing passed predetermined acceptance criteria,<br>demonstrating Guardian Needle™ Electrodes are equivalent to the predicate<br>device in functionality, safety, and effectiveness. |
| <b>807.92(b)(2)<br/>Clinical Tests</b> |
| No Clinical Tests were conducted as referenced in 21 CFR 807.92(b)(2). |
| <b>807.92(b)(3)<br/>Clinical Summary</b> |
| No Clinical Tests were conducted as referenced in 21 CFR 807.92(b)(3). |
{5}------------------------------------------------
### Substantial Equivalence Table
| Characteristic | Subject Device:<br>Guardian Needle™<br>Electrodes | Predicate Device:<br>Subdermal Needle<br>Electrodes | Substantial<br>Equivalence<br>(SE)1 |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K200984 | K022914 Subdermal Needle<br>Electrodes | |
| Manufacturer | Rhythmlink International,<br>LLC | Rhythmlink International,<br>LLC | SE |
| Device Class | Class II | Class II | SE |
| Product Code | GXZ | GXZ | SE |
| Regulatory<br>Name | Subdermal Needle | Subdermal Needle | SE |
| Device Type | Subdermal Needle Electrodes | Subdermal Needle Electrodes | SE |
| Regulation # | 21 CFR 882.1350 | 21 CFR 882.1350 | SE |
| Intended Use | Rhythmlink International<br>Subdermal Needle Electrodes<br>are intended for use with<br>recording, monitoring and<br>stimulation equipment for the<br>purpose of recording of<br>biopotential signals.<br>Examples include:<br>Electromyography (EMG),<br>Electroencephalography<br>(EEG), and Nerve potential<br>signals. The electrodes are<br>sterile and for single patient<br>use. | Rhythmlink International<br>Subdermal Needle Electrodes<br>are intended for use with<br>recording, monitoring and<br>stimulation equipment for the<br>purpose of recording of<br>biopotential signals.<br>Examples include:<br>Electromyography (EMG),<br>Electroencephalography<br>(EEG), and Nerve potential<br>signals. The electrodes are<br>sterile and for single patient<br>use. | SE |
| Anatomical<br>Site(s) | Subdermal muscle or nerve<br>tissue | Subdermal muscle or nerve<br>tissue | SE |
| Environment<br>usage | Hospital | Hospital | SE |
| Electrode<br>Material | Medical Grade 304VM<br>(vacuum melt) Stainless Steel | Medical Grade 304VM<br>(vacuum melt) Stainless Steel | SE |
| Electrode<br>Length | 7mm to 22 mm | 7mm to 22 mm | SE |
| Electrode<br>Diameter | 0.4mm | 0.4mm | SE |
| Leadwire<br>Material | PVC-coated Electrical Wire | PVC-coated Electrical Wire | SE |
| Leadwire<br>Length | 1.0m to 3.0m | 1.0m to 3.0m | SE |
| Connector | 1.5mm DIN 42 802 pin touch<br>proof connector | 1.5mm DIN 42 802 pin touch<br>proof connector | SE |
| Characteristic | Subject Device:<br>Guardian Needle™<br>Electrodes | Predicate Device:<br>Subdermal Needle<br>Electrodes | Substantial<br>Equivalence |
| 510(k)<br>Number | K200984 | K022914 Subdermal Needle<br>Electrodes | (SE)1 |
| Use of<br>Adhesive Tape | YES | NO | SE; Material<br>change has<br>been validated<br>by testing and<br>does not<br>adversely<br>impact safety<br>Or<br>effectiveness |
| Needle Cover | Thermoplastic polyester<br>elastomer sheath | PVC tube | SE; Material<br>and<br>dimensional<br>change has<br>been validated<br>by testing and<br>does not<br>adversely<br>impact safety<br>Or<br>effectiveness |
| Operation of<br>Applying the<br>Device | Applied subcutaneously to the<br>skin | Applied subcutaneously to the<br>skin | SE |
| Prescription<br>Use | YES - Inserted by a licensed<br>physician or practitioner | YES - Inserted by a licensed<br>physician or practitioner | SE |
| Targeted<br>Procedures | IONM, EMG, EP, EEG | IONM, EMG, EP, EEG | SE |
| Compatibility<br>with other<br>devices | Neurophysiology Monitors<br>with a 1.5mm DIN 42 802 pin<br>touch proof connector | Neurophysiology Monitors<br>with a 1.5mm DIN 42 802 pin<br>touch proof connector | SE |
| Packaging | 24 electrodes sealed in a<br>Tyvek pouch and placed<br>inside a sealed, labeled box | 24 electrodes sealed in a<br>Tyvek pouch and placed<br>inside a sealed, labeled box | SE |
| Sterilization<br>Method | Supplied EtO Sterile | Supplied EtO Sterile | SE |
| Single Patient<br>Use | YES - disposable | YES - disposable | SE |
| Electrical<br>Safety | Connectors comply with IEC<br>60601-1 (1988) sub clause<br>56.3(c) per CFR 898.12 | Connectors comply with IEC<br>60601-1 (1988) sub clause<br>56.3(c) per CFR 898.12 | SE |
| Characteristic | Subject Device:<br>Guardian Needle™<br>Electrodes | Predicate Device:<br>Subdermal Needle<br>Electrodes | Substantial<br>Equivalence<br>(SE)1 |
| 510(k)<br>Number | K200984 | K022914 Subdermal Needle<br>Electrodes | |
| Mechanical<br>Safety | Leadwires are soldered to<br>electrode using Tin/Silver<br>solder and covered with heat<br>shrink | Leadwires are soldered to<br>electrode using Tin/Silver<br>solder and covered with heat<br>shrink | SE |
| Duration of<br>use | ≤24 hours | ≤24 hours | SE |
| Standards | • IEC 60601-1-1: 1988/a1:<br>1991/A2 :1995 § 56.3(c)<br>per CFR 898.12<br>• DIN 42802<br>• ISO 14971<br>• ISO 15223-1 | • IEC 60601-1-1: 1988/a1:<br>1991/A2 : 1995 § 56.3(c)<br>per CFR 898.12<br>• DIN 42802<br>• ISO 14971<br>• ISO 15223-1 | SE |
| Biocompatible | Yes | Yes | SE |
{6}------------------------------------------------
{7}------------------------------------------------
¹Only differences are indicated in this column.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.