K200970 · True Indicating, LLC · FRC · Dec 28, 2020 · General Hospital
Device Facts
Record ID
K200970
Device Name
Self-Contained Biological Indicator
Applicant
True Indicating, LLC
Product Code
FRC · General Hospital
Decision Date
Dec 28, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The True Indicating Self-Contained Biological Indicator is intended for monitoring the efficacy of saturated steam sterilization processes. The True Indicating Self-Contained Biological Indicator has a validated reduced incubation time of 10 hours and may be used in the following steam sterilization cycles: 121°C, 30 minutes (Gravity) 132°C, 4 minutes (Pre-Vacuum) 135°C, 3 minutes (Pre-Vacuum)
Device Story
Self-contained biological indicator (SCBI) for monitoring steam sterilization efficacy. Device consists of polypropylene vial containing a 6-mm filter paper disc inoculated with Geobacillus stearothermophilus (ATCC 7953) spores and a glass ampule of nutrient broth with Bromocresol Purple pH indicator. User places SCBI in the most difficult area to sterilize; after cycle completion, user activates device by crushing the ampule to submerge the disc in media. Device is incubated at 55-65°C for 10 hours. Growth is indicated by a color shift from purple to yellow and/or turbidity, signifying sterilization failure. Absence of growth (media remains purple) indicates effective sterilization. Used in clinical or healthcare settings to verify sterilization cycle performance.
Clinical Evidence
Bench testing only. Testing performed per ISO 11138-1:2017 and ISO 11138-3:2017 standards. Results: Viable spore population 1.0-4.8 x 10^5; D121 ≥ 1.7 min, D132 ≥ 18 s, D135 ≥ 12 s; z-value ≥ 11.6. Reduced incubation time of 10 hours demonstrated >97% alignment with 7-day conventional incubation. Simulated use testing confirmed growth in abbreviated cycles and no growth in half/full cycles.
Technological Characteristics
Materials: Polypropylene vial/cap, glass ampule, paper filter carrier. Sensing: Biological indicator (Geobacillus stearothermophilus). Energy: Thermal (incubation 55-65°C). Dimensions: Not specified. Connectivity: None. Sterilization: Not applicable (device is a monitor). Software: None.
Indications for Use
Indicated for monitoring the efficacy of saturated steam sterilization processes in gravity and pre-vacuum cycles.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 28, 2020
True Indicating LLC Thomas Riha CSO 946 Kane St Toledo, Ohio 43612
Re: K200970
Trade/Device Name: True Indicating Self-Contained Biological Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: December 02, 2020 Received: December 02, 2020
Dear Thomas Riha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200970
Device Name
True Indicating Self-Contained Biological Indicator
Indications for Use (Describe)
The True Indicating Self-Contained Biological Indicator is intended for monitoring the efficacy of saturated steam sterilization processes. The True Indicating Self-Contained Biological Indicator has a validated reduced incubation time of 10 hours and may be used in the following steam sterilization cycles:
121°C, 30 minutes (Gravity) 132°C, 4 minutes (Pre-Vacuum) 135°C, 3 minutes (Pre-Vacuum)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submitter: | True Indicating LLC<br>946 Kane Street<br>Suite A<br>Toledo, OH 43612<br>P: 419 476 7119<br>F: 419 470 8899<br>E: info@trueindicating.com |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contacts: | Julie Wheeler<br>President and Chief Operations Officer<br>P:419 304 6535<br>F: 419 470 8899<br>julie.wheeler@trueindicating.com |
| | Tom Riha<br>Chief Scientific Officer<br>P: 248-982-6492<br>F: 419 470 8899<br>E: tom.riha@trueindicating.com |
| Prepared on: | December 28, 2020 |
| Device Name: | True Indicating Self-Contained Biological Indicator |
| Classification: | 21 CFR 880.2800, Class II Medical Device, FDA Product Code FRC,<br>General Hospital |
| Predicate Devices:<br>(Legally Marketed) | SporView® 10 Steam Self Contained Biological Indicator (K122024) |
| Device Description: | The True Indicating Self-Contained Biological Indicator (SCBI) consists of a 6 -<br>mm filter paper disc inoculated with Geobacillus stearothermophilus ATCC®<br>7953, at a minimum of 105 bacterial spores encased within a polypropylene cap,<br>and a polypropylene vial which includes a glass ampule hermetically sealed<br>containing a nutrient broth culture medium modified with a pH indicator,<br>Bromocresol Purple. |
| Indications for Use: | The True Indicating Self Contained Biological Indicator is Intended formonitoring<br>the efficacy of saturated steam sterilization processes. The True Indicating Self-<br>Contained Biological Indicator has a validated reduced incubation time of 10<br>hours and may be used in the following steam sterilization cycles: 121°C, 30<br>minutes (Gravity), 132°C, 4 minutes (Pre-Vacuum), and 135°C, 3 minutes (Pre-<br>Vacuum). |
| Operational<br>Principles: | Place an SCBI in the most difficult area to sterilize in a load. When the cycle is<br>complete, the SCBI is retrieved and activated by crushing the growth medium<br>ampule to engulf the disc within the bottom of the SCBI vial. |
| | The activated SCBI should be incubated at 55-65°C for a minimum of 10 hours.<br>The SCBI should be monitored for visible signs of growth. Growth is indicated by<br>a visual shift in color of the growth medium from purple to yellow, and/or the<br>presence of turbidity. The absence of growth, resulting in growth medium<br>remaining purple in color and/or no turbidity being visually present, indicates the<br>cycle was effective. |
{4}------------------------------------------------
#### Subject Device Predicate Device True Indicating SCBI for SporView® 10 Steam Self Comparison Feature Contained Biological Indicator Steam (K200970) (K122024) Same Gravity 121°C 30 minutes Gravity 132°C 15 minutes Gravity 134°C 4 minutes Gravitv-Flash 132°C 3 minutes Intended Use: 121°C, 30 minutes (Gravity) Gravity-Flash 132°C 10 minutes 132°C, 4 minutes (Pre-Vac) Method of Dynamic Air 121°C 20 minutes Sterilization 135°C, 3 minutes (Pre-Vac) DynamicAir 121°C 30 minutes Dynamic Air 132°C 3 minutes Dynamic Air 132°C 4 minutes Dynamic Air 134°C 4 minutes Dynamic Air 135°C 3 minutes Product Code FRC FRC Same FDA Regulation 21 CFR§ 880.2800 21 CFR§ 880.2800 Same SporView® 10 Steam is a selfcontained biological indicator inoculated with viable Geobacillus Same stearothermophilus bacterial spores and is intended for monitoring the efficacy of saturated steam The True Indicating Self-Contained sterilization processes. Spor View® Biological Indicator is intended for 10 self-contained biological monitoring the efficacy of saturated indicators have a validated reduced steam sterilization processes. The incubation time of 10 hours and may True Indicating Self-Contained be used in the following steam Biological Indicator has a validated Indications for sterilization cycles: reduced incubation time of 10 Use (IFU) hours and may be used in the Gravity 121°C 30 minutes following steam sterilization cycles: Gravity 132°C 15 minutes Gravity 134°C 4 minutes 121°C, 30 minutes (Gravity) Gravity-Flash 132°C 3 minutes 132°C, 4 minutes (Pre-Vacuum) Gravity-Flash 132°C 10 minutes 135°C, 3 minutes (Pre-Vacuum) Dynamic Air 121°C 20 minutes DynamicAir 121°C 30 minutes Dynamic Air 132°C 3 minutes Dynamic Air 132°C 4 minutes Dynamic Air 134°C 4 minutes Dynamic Air 135°C 3 minutes Geobacillus stearothermophilus Geobacillus stearothermophilus Organism Same ATCC 7953 ATCC 7953 Biochemical activity of Biochemical activity of Geobacillus stearothermophilus Geobacillus stearothermophilus Mechanism of produces acid by-products that produces acid by-products that Same Action lower the pH of the media lower the pH of the media which changing its color from purple to changing its color from purple to yellow yellow
#### Technological Characteristic Comparison Table
{5}------------------------------------------------
| Feature | Subject Device<br>True Indicating SCBI for<br>Steam | Predicate<br>SporView® 10 Steam Self<br>Contained Biological Indicator<br>(K122024) | Comparison |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|------------|
| Device Design | Inoculated paper disc and glass<br>ampule with growth media<br>contained within a vial and<br>enclosed with a cap with liner | Inoculated paper disc and glass<br>ampule with growth media<br>contained within a vial and<br>enclosed with a cap with liner | Same |
| Growth Medium<br>Color Change | Purple to Yellow | Purple to Yellow | Same |
| Viable Spore<br>Population | ≥105 or greater | ≥105 or greater | Same |
| Carrier Material | Paper Filter Carrier | Paper Filter Carrier | Same |
| Incubation<br>Conditions | 55 - 65 °C | 55 - 60°C | Similar |
| SCBI Label | Film label with Lot Number and<br>expiration date | Film label with Lot Number and<br>expiration date and chemical<br>indicator that transitions from<br>blue to dark | Similar |
| D value | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s | Same |
| Reduced<br>Incubation Time | 10 hours | 10 hours | Same |
| Hold Time | 72 Hours | 48 Hours | Similar |
| z value | ≥ 10°C | ≥ 10°C | Same |
| Shelf Life | 24 Months | 24 Months | Same |
# Summary of
Nonclinical Testing: Per FDA recognized consensus standards and guidance documents testing was performed for steam sterilization processes using multiple lots of True Indicating SCBI for Steam over the range of the shelf life:
- Total Viable Spore Count was evaluated per manufacturer's methodology and in compliance with ISO 11138-1 Annex A
{6}------------------------------------------------
### 510(k) Summary K200970
- . Resistance Characteristic Studies were conducted including D value per ISO 11138-1 Annex D and ISO 11138-3 Annex A z value per ISO 11138-3 Annex B, and Survival/Kill Windows per ISO 11138-1 Annex E
- Carrier and Primary Packaging Materials were evaluated per Annex B . and ISO 11138-3 7.1
- Hold Time Assessment was evaluated for a period of 72-hours per Guidance for Industry and FDA Staff Biological Indicator (BI) Premarket Notification [510(k)] Submissions Section 7.D. Holding Time Assessment
- . and True Indicating Protocol
- Reduced Incubation Time Studies were conducted per Guidance for . Industry and FDA Staff Biological Indicator (BI) Premarket Notification [51-(k)] Submissions Section 7.E. and Attachment II and True Indicating standard operating procedure
- . Simulated Use - Full, Half, and Abbreviated Exposure Cycles were tested per Guidance for Industry and FDA Staff Biological Indicator (BI) Premarket Notification [51-(k)] Submissions Section 7 and True Indicating Protocols
#### Summary of Nonclinical Testing Table:
Testing was conducted on the Test Pack following the FDA guidance and the standards below:
- Guidance for Industry and FDA Staff, Biological Indicator Premarket Notification [510(k)] Submissions, October 4, 2007
- ISO 11138-1:2017 Sterilization of health care products Biological indicators, Part 1: General . requirements
- ISO 11138-3:2017 Sterilization of health care products Biological indicators, Part 3: Biological ● indicators for moist heat sterilization processes
| Name of Test | Purpose | Acceptance Criterial | Subject Device Result |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|----------------------------------------------------------------|
| Viable Spore<br>Population | Determine the spore population<br>per the manufacturer's<br>procedure following ISO 11138-1 | ≥105 or greater | PASS<br>1.0 - 4.8 x 105 Spore/SCB |
| D value | Determine the<br>resistance of the<br>BI following ISO<br>11138-1 and<br>11138-3 | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s | PASS<br>D121 ≥ 1.7 min<br>D132 ≥ 18 s<br>D135 ≥ 12 s |
| z value | Determine the z-value of the Bl<br>per ISO 11138-1 and 11138-3 | ≥10 | PASS<br>≥11.6 |
| Survival Time | Determine the exposure time for all<br>BI's to retain viable spores<br>(Survival Time) per ISO 11138-1,<br>11138-3 and Guidance for Industry<br>and Staff - Biological Indicator (BI)<br>Premarket Notification [510(k)]<br>Submission | Meets the longer of FDA<br>and ISO 11138-3<br>requirements | PASS<br>121°C ≥ 5.9 min<br>132°C ≥ 1.1 min<br>135°C ≥ 42 sec |
| Kill Time | Determine the exposure time for all<br>BI's to inactivate all spores (Kill<br>Time) per ISO 11138-1,11138-3<br>and Guidance for Industry and<br>Staff - Biological Indicator (BI)<br>Premarket Notification [510(k)]<br>Submission | Meets the shorter of FDA<br>and ISO 11138-3<br>requirements | PASS<br>121°C ≤ 28.0 min<br>132°C ≤ 2.8 min<br>135°C ≤ 1.8 min |
{7}------------------------------------------------
| Name of Test | Purpose | Acceptance Criterial | Subject Device Result |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Reduced<br>Incubation<br>Time (RIT) | Determine the Reduced<br>Incubation Time outlined in<br>Guidance for Industry and Staff -<br>Biological Indicator (BI)<br>Premarket Notification [510(k)]<br>Submission | Meets FDA's requirement<br>of > 97% alignment of the<br>10 hour<br>results with the<br>conventional incubation<br>time of 7 days | PASS<br>Minimum of 10 Hours |
| Carrier growth<br>inhibition /<br>media growth<br>promotion | Determine the carrier growth<br>inhibition/media growth<br>promotion per ISO 11138-1<br>and 11138-3 | Positive growth of less<br>than 100 spores after<br>primary packaging and<br>media are subject to<br>worst case steam<br>exposure | PASS |
| Hold Time | Determine the length of time<br>that an exposed BI can be held<br>before incubation (Hold Time)<br>per Guidance for Industry and<br>Staff - Biological Indicator (BI)<br>Premarket Notification [510(k)]<br>Submission | Performance not<br>affected if incubated<br>within 72 hours of<br>exposure to steam<br>sterilization | PASS |
| Simulated Use | Determine the simulated use<br>of the BI in a sterilizer per<br>Guidance for Industry and<br>Staff - Biological Indicator (BI)<br>Premarket Notification [510(k)]<br>Submission | Demonstrate growth when<br>exposed to abbreviated<br>cycle, all kill in a full cycle<br>and a half cycle. All cycles<br>utilized full loads using<br>porous, non-<br>porous, and mixed load<br>material | PASS<br>Abbreviated cycle - growth<br>Half cycle - no growth<br>Full cycle - no growth |
- Conclusion: Based on the non-clinical performance data, The True Indicating Self-Contained Biological Indicator is as safe, as effective, and performs as well as or better than the legally marketed predicate, SporView® 10 Steam Self Contained Biological Indicator cleared under K122024, Class II (21 CFR 880.2800), product code FRC.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.