# HEALIX ADVANCE Anchor with DYNATAPE Suture (K200949)

_Medos International SARL · MBI · Jun 4, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200949

## Device Facts

- **Applicant:** Medos International SARL
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Jun 4, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The HEALIX ADVANCE Anchor is indicated for use in soft-tissue-to-bone fixation in association with postoperative immobilization as follows: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromioclavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis. Elbow: Biceps Tendon Reattachment, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction.

## Device Story

Threaded suture anchor preloaded on disposable inserter; available in absorbable BR or non-absorbable PEEK materials. Suture features silicone core with salt particles; when placed in aqueous environment, salt elutes, creating micro-porous structure; core hydrates and expands radially; radial expansion causes axial shortening of suture to resist laxity and maintain approximation force. Used by surgeons in clinical settings for soft-tissue-to-bone fixation. Provides mechanical fixation to facilitate healing; minimizes gap formation at repair site.

## Clinical Evidence

Bench testing only. Performance testing included straight and knot tensile strength, suture approximation force, anchor fixation post cyclic loading, and sterility. Biocompatibility, chemical characterization, and in-vivo testing were included by reference to predicates.

## Technological Characteristics

Threaded suture anchor; materials: absorbable BR or PEEK; preloaded on disposable inserter. Suture: silicone core with salt particles for radial expansion. Sterilization: Ethylene Oxide per ANSI/AAMI/ISO 11135:2014 (SAL 10^-6). Non-pyrogenic per ANSI/AAMI ST-72:2011.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- HEALIX ADVANCE Anchor with DYNACORD Suture ([K173859](/device/K173859.md))
- HEALIX ADVANCE Anchor with DYNACORD Suture ([K183506](/device/K183506.md))

## Reference Devices

- HEALIX ADVANCE Anchor with PERMATAPE Suture ([K170639](/device/K170639.md))
- HEALIX ADVANCE Anchor with DYNA+TAPE Sutures ([K191483](/device/K191483.md))
- FiberWire (Arthrex) ([K021434](/device/K021434.md))
- FiberWire (Arthrex) ([K041553](/device/K041553.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 4, 2020

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of this is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

Medos International SARL % Elizabeth Messana Regulatory Affairs Specialist II DePuy Synthes Mitek, a Johnson and Johnson Company 325 Paramount Drive Raynham, Massachusetts 02767

Re: K200949

Trade/Device Name: HEALIX ADVANCETM Anchor with DYNATAPE™ Suture Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: April 8, 2020 Received: April 9, 2020

Dear Ms. Messana:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jesse Muir, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200949

#### Device Name

HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures

Indications for Use (Describe)

The HEALIX ADVANCE Anchor is indicated for use in soft-tissue-to-bone fixation in association with postoperative immobilization as follows:

| Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromioclavicular |
|--------------------------------------------------------------------------------------------------------|
| Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction                      |

- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair
- Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Elbow: Biceps Tendon Reattachment, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction

| Type of Use (Select one or both, as applicable)  |                                             |  |  |  |
|--------------------------------------------------|---------------------------------------------|--|--|--|
| X   Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |  |  |  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY

## HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures Date Prepared: April 3, 2020

| Submitter's<br>Name and<br>Address | DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br><br>On behalf of:<br>Medos International SARL<br>Chemin-Blanc 38, Le Locle Neuchatel<br>CH 2400, Switzerland                                                                                                                                                                                                              |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person                     | Elizabeth Messana<br>Regulatory Affairs Specialist II<br><br>Telephone: 508-828-3150<br>Email: emessan1@its.jnj.com<br><br>DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767                                                                                                                                                                                                   |
| Name of Medical<br>Device          | Proprietary Name: HEALIX ADVANCETM Anchor with<br>DYNATAPETM Sutures<br>A) HEALIX ADVANCE™ BR Anchor with DYNATAPE™ Suture<br>B) HEALIX ADVANCE™ PEEK Anchor with DYNATAPE™<br>Suture<br><br>Classification Name:<br>A) Single/multiple component metallic bone fixation appliances and<br>accessories<br>B) Smooth or threaded metallic bone fixation fasteners<br><br>Product Code:<br>A) MAI<br>B) MBI<br><br>Common Name: Suture Anchor |

{4}------------------------------------------------

| Substantial<br>Equivalence            | The HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is<br>substantially equivalent to:<br>K173859, K183506 - HEALIX ADVANCE™ Anchor with DYNACORD™ Suture              Reference Devices:  K170639 - HEALIX ADVANCE™ Anchor with PERMATAPE™ Suture K191483 - HEALIX ADVANCE™ Anchor with DYNA+TAPE™ Sutures K021434, K041553- FiberWire® (Arthrex)                                                                                                                                                                                                                                                    |                                                                                                                                                                                         |  |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Device<br>Classification              | The HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is classified as:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                         |  |
|                                       | A) HEALIX ADVANCE BR Anchor with DYNATAPE Suture is classified as: Single/multiple component metallic bone fixation             appliances and accessories, classified as Class II, product code MAI, regulated under 21 CFR 888.3030.                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                         |  |
|                                       | B) HEALIX ADVANCE PEEK Anchor with DYNATAPE Suture is classified as: Smooth or threaded metallic bone fixation             fasteners, classified as Class II, product code MBI, regulated under 21 CFR 888.3040.                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                         |  |
| Device Panel                          | Orthopedic Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                         |  |
| Device<br>Description                 | The proposed HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is a line extension to the currently marketed HEALIX             ADVANCE™ Anchor family. HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is a threaded suture anchor preloaded on a             disposable inserter assembly. HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is available in absorbable BR and non-             absorbable PEEK materials. HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is provided sterile and is for single use only.                                                                                    |                                                                                                                                                                                         |  |
| Technological<br>Characteristics      | The HEALIX ADVANCETM Anchor with DYNATAPETM Sutures is<br>intended for soft-tissue-to-bone fixation in association with<br>postoperative mobilization.                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                         |  |
|                                       | When DYNATAPET™ Suture is placed in an aqueous environment, the<br>salt particles within the silicone core elute out, leaving behind a micro-<br>porous structure within the silicone core. These small voids are<br>consequently filled with surrounding fluid as the core hydrates,<br>resulting in a radial expansion of the suture. If laxity is present, this<br>radial expansion of the braid causes an axial shortening of the total<br>suture length. The DYNATAPETM Suture is designed to resist laxity<br>and minimize gap formation by maintaining approximation<br>(compression) force. |                                                                                                                                                                                         |  |
| Indications for<br>Use                | The HEALIX ADVANCE Anchor is indicated for use in soft-tissue-to-<br>bone fixation in association with postoperative immobilization as<br>follows:                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                         |  |
|                                       | Shoulder:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Rotator Cuff Repair, Bankart Repair, SLAP Lesion<br>Repair, Biceps Tenodesis, Acromioclavicular<br>Separation Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction |  |
|                                       | Foot/Ankle:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Lateral Stabilization, Medial Stabilization, Achilles<br>Tendon Repair                                                                                                                  |  |
|                                       | Knee:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Medial Collateral Ligament Repair, Lateral Collateral<br>Ligament Repair, Posterior Oblique Ligament Repair,<br>Iliotibial Band Tenodesis                                               |  |
|                                       | Elbow:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Biceps Tendon Reattachment, Ulnar Collateral<br>Ligament Reconstruction, Radial Collateral Ligament<br>Reconstruction                                                                   |  |
| Comparison to<br>Predicate<br>Devices | The proposed HEALIX ADVANCE™ Anchors with DYNATAPE™<br>Sutures is a line extension to the HEALIX ADVANCE™ Anchor<br>family, inclusive of the HEALIX ADVANCE™ Anchors with<br>DYNACORD™ Suture (K173859, K183506). The anchors, suture<br>materials, intended use, packaging and sterilization are identical to the<br>currently marketed HEALIX ADVANCE™ Anchors with<br>DYNACORD™ Suture (K173859, K183506).                                                                                                                                                                                       |                                                                                                                                                                                         |  |

{5}------------------------------------------------

{6}------------------------------------------------

|                           | The inner and outer sheath of the DYNATAPE™ Suture has been<br>redesigned with thicker fibers and a unique braiding pattern to create<br>the flattened structure specific to DYNATAPE ™ Suture. The proposed<br>HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures will be<br>preloaded with one strand of 2.5 mm DYNATAPE ™ Suture and one<br>strand of #2 DYNACORD ™ Suture.                                                                                                                                |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Non-clinical<br>Testing   | Device safety and effectiveness is supported by non-clinical testing<br>performed on the proposed device and / or its predicate. Performance<br>testing included straight and knot tensile strength, suture approximation<br>force, anchor fixation post cyclic loading and sterility. In-vitro anchor<br>fixation, torque testing, chemical characterization, biocompatibility, in-<br>vivo testing, packaging and shelf-life of the predicate device were<br>included by reference (K173859, K183506). |
|                           | Ethylene Oxide Sterilization was validated according to<br>ANSI/AAMI/ISO 11135: 2014 to a SAL of 1 x 10-6.                                                                                                                                                                                                                                                                                                                                                                                               |
|                           | EO residuals were tested per AAMI/ANSI/ISO 10993-7:2008                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                           | The proposed device has been determined to be non-pyrogenic per the<br>requirements set forth in ANSI/AAMI ST-72:2011, United States<br>Pharmacopeia (USP), and European Pharmacopeia (EP) using the<br>bacterial endotoxin testing (BET) method.                                                                                                                                                                                                                                                        |
| Safety and<br>Performance | Results of performance testing have demonstrated that the proposed<br>devices are suitable for their intended use. Based on similarities in the<br>indications for use, technological characteristics, and performance in<br>comparison to the predicate devices, the proposed HEALIX<br>ADVANCE™ Anchor with DYNATAPE™ Sutures has shown to be<br>substantially equivalent to the predicate devices under the Federal Food,<br>Drug and Cosmetic Act.                                                   |

---

**Source:** [https://fda-staging.innolitics.com/device/K200949](https://fda-staging.innolitics.com/device/K200949)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).