K200949 · Medos International SARL · MBI · Jun 4, 2020 · Orthopedic
Device Facts
Record ID
K200949
Device Name
HEALIX ADVANCE Anchor with DYNATAPE Suture
Applicant
Medos International SARL
Product Code
MBI · Orthopedic
Decision Date
Jun 4, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HEALIX ADVANCE Anchor is indicated for use in soft-tissue-to-bone fixation in association with postoperative immobilization as follows: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromioclavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair. Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis. Elbow: Biceps Tendon Reattachment, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction.
Device Story
Threaded suture anchor preloaded on disposable inserter; available in absorbable BR or non-absorbable PEEK materials. Suture features silicone core with salt particles; when placed in aqueous environment, salt elutes, creating micro-porous structure; core hydrates and expands radially; radial expansion causes axial shortening of suture to resist laxity and maintain approximation force. Used by surgeons in clinical settings for soft-tissue-to-bone fixation. Provides mechanical fixation to facilitate healing; minimizes gap formation at repair site.
Clinical Evidence
Bench testing only. Performance testing included straight and knot tensile strength, suture approximation force, anchor fixation post cyclic loading, and sterility. Biocompatibility, chemical characterization, and in-vivo testing were included by reference to predicates.
Technological Characteristics
Threaded suture anchor; materials: absorbable BR or PEEK; preloaded on disposable inserter. Suture: silicone core with salt particles for radial expansion. Sterilization: Ethylene Oxide per ANSI/AAMI/ISO 11135:2014 (SAL 10^-6). Non-pyrogenic per ANSI/AAMI ST-72:2011.
Indications for Use
Indicated for soft-tissue-to-bone fixation in shoulder, foot/ankle, knee, and elbow procedures requiring postoperative immobilization.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
HEALIX ADVANCE Anchor with DYNACORD Suture (K173859)
HEALIX ADVANCE Anchor with DYNACORD Suture (K183506)
Reference Devices
HEALIX ADVANCE Anchor with PERMATAPE Suture (K170639)
HEALIX ADVANCE Anchor with DYNA+TAPE Sutures (K191483)
{0}------------------------------------------------
June 4, 2020
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of this is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Medos International SARL % Elizabeth Messana Regulatory Affairs Specialist II DePuy Synthes Mitek, a Johnson and Johnson Company 325 Paramount Drive Raynham, Massachusetts 02767
Re: K200949
Trade/Device Name: HEALIX ADVANCETM Anchor with DYNATAPE™ Suture Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: April 8, 2020 Received: April 9, 2020
Dear Ms. Messana:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jesse Muir, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200949
#### Device Name
HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures
Indications for Use (Describe)
The HEALIX ADVANCE Anchor is indicated for use in soft-tissue-to-bone fixation in association with postoperative immobilization as follows:
| Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromioclavicular |
|--------------------------------------------------------------------------------------------------------|
| Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction |
- Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair
- Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis
- Elbow: Biceps Tendon Reattachment, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction
| Type of Use (Select one or both, as applicable) | | | | |
|--------------------------------------------------|---------------------------------------------|--|--|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
## HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures Date Prepared: April 3, 2020
| Submitter's<br>Name and<br>Address | DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br><br>On behalf of:<br>Medos International SARL<br>Chemin-Blanc 38, Le Locle Neuchatel<br>CH 2400, Switzerland |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Elizabeth Messana<br>Regulatory Affairs Specialist II<br><br>Telephone: 508-828-3150<br>Email: emessan1@its.jnj.com<br><br>DePuy Synthes Mitek Sports Medicine<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767 |
| Name of Medical<br>Device | Proprietary Name: HEALIX ADVANCETM Anchor with<br>DYNATAPETM Sutures<br>A) HEALIX ADVANCE™ BR Anchor with DYNATAPE™ Suture<br>B) HEALIX ADVANCE™ PEEK Anchor with DYNATAPE™<br>Suture<br><br>Classification Name:<br>A) Single/multiple component metallic bone fixation appliances and<br>accessories<br>B) Smooth or threaded metallic bone fixation fasteners<br><br>Product Code:<br>A) MAI<br>B) MBI<br><br>Common Name: Suture Anchor |
{4}------------------------------------------------
| Substantial<br>Equivalence | The HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is<br>substantially equivalent to:<br>K173859, K183506 - HEALIX ADVANCE™ Anchor with DYNACORD™ Suture Reference Devices: K170639 - HEALIX ADVANCE™ Anchor with PERMATAPE™ Suture K191483 - HEALIX ADVANCE™ Anchor with DYNA+TAPE™ Sutures K021434, K041553- FiberWire® (Arthrex) | | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Device<br>Classification | The HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is classified as: | | |
| | A) HEALIX ADVANCE BR Anchor with DYNATAPE Suture is classified as: Single/multiple component metallic bone fixation appliances and accessories, classified as Class II, product code MAI, regulated under 21 CFR 888.3030. | | |
| | B) HEALIX ADVANCE PEEK Anchor with DYNATAPE Suture is classified as: Smooth or threaded metallic bone fixation fasteners, classified as Class II, product code MBI, regulated under 21 CFR 888.3040. | | |
| Device Panel | Orthopedic Devices | | |
| Device<br>Description | The proposed HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is a line extension to the currently marketed HEALIX ADVANCE™ Anchor family. HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is a threaded suture anchor preloaded on a disposable inserter assembly. HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is available in absorbable BR and non- absorbable PEEK materials. HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures is provided sterile and is for single use only. | | |
| Technological<br>Characteristics | The HEALIX ADVANCETM Anchor with DYNATAPETM Sutures is<br>intended for soft-tissue-to-bone fixation in association with<br>postoperative mobilization. | | |
| | When DYNATAPET™ Suture is placed in an aqueous environment, the<br>salt particles within the silicone core elute out, leaving behind a micro-<br>porous structure within the silicone core. These small voids are<br>consequently filled with surrounding fluid as the core hydrates,<br>resulting in a radial expansion of the suture. If laxity is present, this<br>radial expansion of the braid causes an axial shortening of the total<br>suture length. The DYNATAPETM Suture is designed to resist laxity<br>and minimize gap formation by maintaining approximation<br>(compression) force. | | |
| Indications for<br>Use | The HEALIX ADVANCE Anchor is indicated for use in soft-tissue-to-<br>bone fixation in association with postoperative immobilization as<br>follows: | | |
| | Shoulder: | Rotator Cuff Repair, Bankart Repair, SLAP Lesion<br>Repair, Biceps Tenodesis, Acromioclavicular<br>Separation Repair, Deltoid Repair, Capsular Shift or<br>Capsulolabral Reconstruction | |
| | Foot/Ankle: | Lateral Stabilization, Medial Stabilization, Achilles<br>Tendon Repair | |
| | Knee: | Medial Collateral Ligament Repair, Lateral Collateral<br>Ligament Repair, Posterior Oblique Ligament Repair,<br>Iliotibial Band Tenodesis | |
| | Elbow: | Biceps Tendon Reattachment, Ulnar Collateral<br>Ligament Reconstruction, Radial Collateral Ligament<br>Reconstruction | |
| Comparison to<br>Predicate<br>Devices | The proposed HEALIX ADVANCE™ Anchors with DYNATAPE™<br>Sutures is a line extension to the HEALIX ADVANCE™ Anchor<br>family, inclusive of the HEALIX ADVANCE™ Anchors with<br>DYNACORD™ Suture (K173859, K183506). The anchors, suture<br>materials, intended use, packaging and sterilization are identical to the<br>currently marketed HEALIX ADVANCE™ Anchors with<br>DYNACORD™ Suture (K173859, K183506). | | |
{5}------------------------------------------------
{6}------------------------------------------------
| | The inner and outer sheath of the DYNATAPE™ Suture has been<br>redesigned with thicker fibers and a unique braiding pattern to create<br>the flattened structure specific to DYNATAPE ™ Suture. The proposed<br>HEALIX ADVANCE™ Anchor with DYNATAPE™ Sutures will be<br>preloaded with one strand of 2.5 mm DYNATAPE ™ Suture and one<br>strand of #2 DYNACORD ™ Suture. |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Non-clinical<br>Testing | Device safety and effectiveness is supported by non-clinical testing<br>performed on the proposed device and / or its predicate. Performance<br>testing included straight and knot tensile strength, suture approximation<br>force, anchor fixation post cyclic loading and sterility. In-vitro anchor<br>fixation, torque testing, chemical characterization, biocompatibility, in-<br>vivo testing, packaging and shelf-life of the predicate device were<br>included by reference (K173859, K183506). |
| | Ethylene Oxide Sterilization was validated according to<br>ANSI/AAMI/ISO 11135: 2014 to a SAL of 1 x 10-6. |
| | EO residuals were tested per AAMI/ANSI/ISO 10993-7:2008 |
| | The proposed device has been determined to be non-pyrogenic per the<br>requirements set forth in ANSI/AAMI ST-72:2011, United States<br>Pharmacopeia (USP), and European Pharmacopeia (EP) using the<br>bacterial endotoxin testing (BET) method. |
| Safety and<br>Performance | Results of performance testing have demonstrated that the proposed<br>devices are suitable for their intended use. Based on similarities in the<br>indications for use, technological characteristics, and performance in<br>comparison to the predicate devices, the proposed HEALIX<br>ADVANCE™ Anchor with DYNATAPE™ Sutures has shown to be<br>substantially equivalent to the predicate devices under the Federal Food,<br>Drug and Cosmetic Act. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.