K200925 · Clarus Medical, LLC · GWG · Mar 18, 2021 · Neurology
Device Facts
Record ID
K200925
Device Name
Clarus Peel-Away Introducer Sheath
Applicant
Clarus Medical, LLC
Product Code
GWG · Neurology
Decision Date
Mar 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1480
Device Class
Class 2
Indications for Use
The Clarus Peel-Away Introducer Sheath is intended to be used in indications requiring access through the brain into the ventricular system.
Device Story
The Clarus Peel-Away Introducer Sheath is a thin-walled, cylindrical device used to create a communicating passageway through brain tissue. It consists of a peel-away sheath and an inner obturator. The device is advanced as a unit; the obturator features a distal tip hole allowing pressurized cerebrospinal fluid to egress upon reaching the ventricular system. Once positioned, the obturator is unlocked via a quarter-turn rotation and removed, leaving the sheath in place to maintain the access site. The sheath includes tab handles for lengthwise tearing to facilitate removal or adjustment. Used in healthcare facilities by physicians, the device serves as a conduit for the subsequent insertion of other cleared medical instruments. It benefits patients by providing a stable, temporary access channel for neurosurgical procedures.
Clinical Evidence
Bench testing only. No clinical or animal studies were required. Performance verification included biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, material-mediated endotoxin) and mechanical testing (fluid patency, prolapse force, peak tensile force, peel force). All test results were acceptable and comparable to predicate devices.
Technological Characteristics
Two-part assembly: peel-away sheath and obturator. Materials: HDPE (obturator hub/tubing with 15% barium sulfate) and PTFE (sheath tubing with 9% bismuth oxide). Dimensions: 12F, 14F, 15F diameters; 8-13cm lengths. Sterilization: Ethylene Oxide. Single-use. Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring access through the brain into the ventricular system.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
Martech Medical Products Super Sheath Introducer (K120617)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 18, 2021
Clarus Medical, LLC Mark Brown Director, Operations 13355 10th Ave N Suite 110 Plymouth, Minnesota 55441
Re: K200925
Trade/Device Name: Peel-Away Introducer Sheath Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG Dated: February 11, 2021 Received: February 16, 2021
Dear Mark Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200925
Device Name Clarus Peel-Away Introducer Sheath
Indications for Use (Describe)
The Clarus Peel-Away Introducer Sheath is intended to be used in indications requiring access through the brain into the ventricular system.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> |
|----------------------------------------------|-------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) Summarv
#### A. General Information:
| 510(K) Submission Number: | K200925 |
|---------------------------|--------------------------------------------------------------------------------------|
| Submitter: | Clarus Medical, LLC<br>13355 10th Ave North<br>Minneapolis, MN 55441<br>763-525-8403 |
| Establishment Number: | 2183911 |
| Contact: | Mark F. Brown<br>Director of Operations<br>mbrown@clarus-medical.com |
| Date Prepared: | April 3, 2020 |
**B. Trade Name:**
| Common Name: | Clarus Peel-Away Introducer Sheath |
|----------------|------------------------------------|
| Product Code: | Endoscope Introducer |
| C.F.R Section: | GWG |
| Class: | 882.1480 |
| | II |
**C. Predicate Device:**
| 510(k) Number | Description |
|---------------|--------------------------------------|
| K990333 | Medtronic Peelaway Introducer Sheath |
| <b>D. Reference Devices:</b> | |
|------------------------------|--------------------------------------------------|
| 510(k) Number | Description |
| K883607 | Codman® Peel-Away Catheter Introducer |
| K120617 | Martech Medical Products Super Sheath Introducer |
# E. Device Description:
The Clarus Peel-Away Introducer Sheath is a thin-walled, cylindrical device that is placed in position so that it provides a communicating passageway through brain tissue. This is accomplished by fitting the sheath securely over an obturator, advancing both devices through the tissue together as a unit. A hole through the distal tip of the obturator aids in placement by allowing pressurized cerebrospinal fluid to egress once the ventricular system is reached. The proximal hub on the obturator is rotated, unlocking it from the sheath, and the obturator is removed from the sheath, leaving the sheath alone in the desired position acting to hold the penetrated site in an open condition. The peel away sheath is capable of being pulled apart lengthwise to allow varying insertion depths into brain tissue. The proximal end of the sheath has tab handles that are provided to facilitate grasping and tearing. At this point, a regulatorily cleared device can be inserted through the sheath for patient procedures.
{4}------------------------------------------------
The Peel-Away Introducer Sheath has the following parts:
Image /page/4/Figure/2 description: The image contains two figures, labeled as Figure 1 and Figure 2, which depict medical instruments. Figure 1 shows a detailed view of an obturator hub with labels pointing to the conical luer fitting, peel-away sheath tab, peel-away sheath tubing, and obturator tubing. Figure 2 presents a side-by-side comparison of a peel-away sheath and an obturator.
{5}------------------------------------------------
## Component description:
Obturator hub: The obturator hub is injection molded directly onto the obturator tubing with high-density polyethylene (HDPE) resin. The hubs are color coded to indicate the French size. Also, the French size number is molded onto the side of the hub
Obturator tubing: The tubing for the obturator is extruded using high-density polyethylene (HDPE) resin with 15% barium sulfate to aid in visualization under fluoroscopy.
Peel-away Sheath tabs: The peel-away sheath tabs are injection molded directly onto the sheath tubing with high-density polyethylene (HDPE) resin. The molded tabs incorporate a slot to allow the obturator hub to lock onto the sheath.
Peel-away sheath: The sheath tubing is extruded using Polytetrafluoroethylene PTFE resin with 9% Bismuth oxide to aid in visualization under fluoroscopy.
## F. Intended Use:
The Clarus Peel-Away Introducer Sheath is intended to be used in indications requiring access through the brain into the ventricular system.
## G. Comparison to Predicate and reference Devices:
The Clarus Peel-Away Introducer Sheath is substantially equivalent to the predicate and reference devices based on the following technological elements:
- . Design: The predicate and reference devices all have a two-part design, a peelaway sheath and an inner obturator introducer. Both parts lock together.
- . Materials: The materials of the subject device and the Medtronic predicate device are very similar, and the Martech Medical Products PTFE Super Sheath reference device are identical in terms of materials.
- Construction: The construction of the subject device and predicate Medtronic . Peelaway Introducer Sheath are very similar. They both include two parts, a peelaway sheath and a blunt distal tip obturator, which fits inside the sheath.
- Environment of Use: The Clarus Peel-Away Introducer Sheath and the Medtronic . Peelaway Introducer Sheath must be used by a physician in a healthcare facility.
The following technological differences exist between the subject and predicate devices:
- . The way in which the outer sheath and obturator lock together is different. The subject device locks to the sheath by rotating the obturator a quarter turn engaging two side tabs into the sheath. The predicate device snaps together.
- The way in which fluid can be drained is different. The subject device has a small hole at ● the distal tip of the obturator to aid in placement by allowing pressurized cerebrospinal fluid to egress once the ventricular system is reached. The predicate device does not have this feature which requires the user to completely remove the obturator to relieve pressurized fluid. The Codman reference device obturator distal tip designs are identical. They both have a hole through the side to aids in placement by allowing pressurized cerebrospinal fluid to egress once the ventricular system is reached.
{6}------------------------------------------------
| | Subject Device | Predicate Devices | Reference Device | Reference Device |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Clarus Peel-Away<br>Introducer Sheath | Medtronic<br>Peelaway<br>Introducer Sheath | Codman® Peel-<br>Away Catheter<br>Introducer | Martech Medical<br>Products PTFE Super<br>Sheath Introducer |
| 510(k) | K200925 | K990333 | K883607 | K120617 |
| Indications<br>for Use | The Clarus Peel-<br>Away Introducer<br>Sheath is a device<br>used to make a<br>channel through the<br>brain into the<br>ventricular system. | The Introducer<br>Sheath is a device<br>used to<br>make a channel<br>through the brain<br>into the ventricular<br>system. | The Peel-Away<br>Catheter Introducer<br>assists in placing a<br>catheter in the<br>ventricle when<br>hydrocephalus<br>shunting is<br>indicated and when<br>using an<br>endoscope. | The PTFE Super<br>Sheath Introducers are<br>intended to obtain<br>central venous access<br>to facilitate catheter<br>insertion or placing<br>pacing leads into the<br>central venous system. |
| Image | | | | |
| Product<br>Code | GWG | GWG | GYK | DYB |
| Enviornment<br>of use | Rx Only | Rx Only | Rx Only | Rx Only |
| Sheath<br>Diameter<br>and Length | 12F, 14F, 15F<br>8-13cm | 12F, 14F, 15F<br>9.5cm | 9F, 11F, 12.5F,<br>14F<br>14.7cm | 5F through 18F<br>10cm and 13.5cm |
| Sterilization<br>method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Conditions<br>supplied | Sterile, single use | Sterile, single use | Sterile, single use | Sterile, single use |
| Device<br>components | Peel-away Sheath,<br>obturator | Peel-away Sheath,<br>obturator | Peel-away Sheath,<br>obturator | Peel-away Sheath,<br>dilator |
{7}------------------------------------------------
# H. Performance Data:
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the Clarus Peel-Away Introducer Sheath was conducted with the following FDA quidance, "Use of International Standards ISO 10993-1,"Biological evaluation of medical devices - Part 1: Evaluation and testing within risk management process". The battery of testing included the following tests:
- . Cytotoxicity
- Sensitization ●
- Irritation ●
- . Systemic Toxicity
- Hemocompatibility ●
- Material-Mediated Endotoxin ●
The nature of body contact and the related device category: Category: External Communicating device Contact: Circulating Blood Contact Duration: limited (<24 hours)
The selected tests are appropriate based on the intended use for the Clarus Peel-Away Introducer Sheath. All tests results are acceptable demonstrating the Clarus Peel-Away Introducer Sheath is safe for its intended use.
## Mechanical Testing:
Components of the Clarus Peel-Away Introducer Sheath and predicate devices were tested for:
| Test performed | Description | Test results |
|-----------------------|-----------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Fluid Patency | Visually inspect inside diameter for<br>cleanliness and quality of manufacturing. | Pass. All devices had<br>comparable results. |
| Prolapse Force | Resistance to deformation of device while<br>advancing distal end into flat plate. Force<br>was measured. | Pass. All devices had<br>comparable results. |
| Peek Tensile<br>Force | Measure peak tensile force of inner<br>obturator/dilator and sheath hub per ISO<br>11070. | Pass. All devices had<br>comparable results. |
| Peel Force | Each sheath is gripped and force measure to<br>break hub, and peel sheath. | Pass. All devices had<br>comparable results. |
The subject device and predicate devices were found to be substantially equivalent in performance.
# Clinical and animal Studies:
No animal studies were required (other than biocompatibility testing) for the Clarus Peel-Away Introducer Sheath.
{8}------------------------------------------------
No clinical studies were required for the Clarus Peel-Away Introducer Sheath to support substantial equivalence.
#### l. Conclusion
The non-clinical data supports the safety of the device and performance verification & validation testing demonstrate that the Clarus Peel-Away Introducer Sheath should perform as intended in the specified use conditions. Testing demonstrates that the Clarus Peel-Away Introducer Sheath is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.