← Product Code [NUC](/productcode/NUC) · K200894

# Agape Cooling Personal Lubricant (K200894)

_CC Wellness, LLC · NUC · Jul 2, 2020 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200894

## Device Facts

- **Applicant:** CC Wellness, LLC
- **Product Code:** [NUC](/productcode/NUC.md)
- **Decision Date:** Jul 2, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Agape Cooling Personal Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

## Device Story

Agape Cooling Personal Lubricant is a clear, semi-viscous, non-sterile personal lubricant; intended for OTC use to enhance comfort during intimate sexual activity. Formulated with water, propanediol, gluconolactone, hydroxyethylcellulose, sodium benzoate, polysorbate 20, citric acid, and peppermint leaf extract. Packaged in PET cylinder bottles. Device provides lubrication and moisture; supplements natural body lubrication. Compatible with natural rubber latex and polyisoprene condoms; incompatible with polyurethane condoms. Users apply topically as needed. Benefits include reduced friction and increased comfort during sexual activity.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, vaginal irritation, acute systemic toxicity) confirmed the device is non-cytotoxic, non-irritating, non-sensitizing, and non-systemically toxic. Shelf-life validated via real-time aging study (3 years). Condom compatibility tested per ASTM D7661-10.

## Technological Characteristics

Water-based personal lubricant. Ingredients: Water, Propanediol, Gluconolactone, Hydroxyethylcellulose, Sodium Benzoate, Polysorbate 20, Citric Acid, Mentha Piperita Leaf Extract. Specifications: Viscosity 2150-4000 cps (USP <911>), pH 3.50-4.50 (USP <971>), Specific Gravity 1.010-1.030 (USP <841>), Osmolality 500-750 mOsm/kg (USP <785>). Non-sterile. Packaged in PET bottles. Tested for antimicrobial effectiveness per USP <51>.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Premium Cooling Personal Lubricant ([K200602](/device/K200602.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 2, 2020

CC Wellness LLC Marlent Perez Quality Specialist 29000 N. Hancock Pkwy. Valencia, CA 91355

K200894 Re: Trade/Device Name: Agape Cooling Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: April 2, 2020 Received: April 3, 2020

Dear Marlent Perez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K200894

Device Name Agape Cooling Personal Lubricant

Indications for Use (Describe)

Agape Cooling Personal Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

|                           |                                                                                    | 510(k) Summary – K200894<br>Agape Cooling Personal Lubricant                                              |  | Page 1 of |
|---------------------------|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--|-----------|
| 510(k) Owner:             | CC Wellness LLC                                                                    |                                                                                                           |  |           |
| Street Address:           | 29000 N. Hancock Parkway<br>Valencia, CA 91355                                     |                                                                                                           |  |           |
| Contact Person:           | Marlent Perez<br>Quality Specialist                                                |                                                                                                           |  |           |
|                           | Bruce Albert<br>Head of Technical Services                                         |                                                                                                           |  |           |
| Contact Numbers:          | Phone: (661) 295-1700, ext. 1007<br>Phone: (661) 295-1700, ext. 231                |                                                                                                           |  |           |
| Summary Preparation Date: | June 29, 2020                                                                      |                                                                                                           |  |           |
| Trade Name:               | Agape Cooling Personal Lubricant                                                   |                                                                                                           |  |           |
| Common Name:              | Personal Lubricant                                                                 |                                                                                                           |  |           |
| Device Classification:    | Regulation Name:<br>Regulation Number:<br>Product Code:<br>Device Class:           | Condom<br>21 CFR § 884.5300<br>NUC (lubricant, personal)<br>Class II                                      |  |           |
| Predicate Device:         | Product Name:<br>510(k) Number:<br>Manufacturer:<br>Product Code:<br>Device Class: | Premium Cooling Personal Lubricant<br>K200602<br>CC Wellness LLC<br>NUC (lubricant, personal)<br>Class II |  |           |

The predicate device has not been subject to a design-related recall.

#### Device Description:

Agape Cooling Personal Lubricant is a clear, semi-viscous personal lubricant that is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

This device is a non-sterile personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity supplement the body's natural lubrication.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo for a wellness company. The logo features two stylized, cursive letters that appear to be "CC". Below the letters, the word "WELLNESS" is written in a smaller, sans-serif font. The overall design is simple and elegant, with a focus on the company's name and its association with wellness.

# 510(k) Summary - K200894 Agape Cooling Personal Lubricant

oz./60 mL sizes provided in clear polyethylene terephthalate (PET) cylinder bottles. The 1 fl. oz./30 mL size bottles are capped with natural disc tops. The 2 fl. oz./60 mL size bottles are capped with silver disc tops. The individual bottles are hermetically sealed during the production process.

This device is composed of Water, Propanediol, Gluconolactone, Hydroxyethylcellulose, Sodium Benzoate, Polysorbate 20, Citric Acid and Mentha Piperita (Peppermint) Leaf Extract.

The device specifications are listed in the table below:

| Property                                                        | Specification                                                  |
|-----------------------------------------------------------------|----------------------------------------------------------------|
| Appearance                                                      | Semi-viscous liquid                                            |
| Color                                                           | Colorless                                                      |
| Odor                                                            | Peppermint                                                     |
| Viscosity (cps) per USP <911>                                   | 2150 to 4000                                                   |
| pH per USP <971>                                                | 3.50 to 4.50                                                   |
| Specific Gravity per USP <841>                                  | 1.010 to 1.030                                                 |
| Osmolality per USP <785> (1:2 dilution)                         | 500 to 750 mOsm/kg                                             |
| Antimicrobial effectiveness per USP <51>                        | Meets USP <51> acceptance criteria for Category 2<br>products. |
| Total aerobic microbial count (TAMC) per USP<br><61> and <1111> | Less than 100 cfu/g                                            |
| Total yeast and mold count (TYMC) per USP <61><br>and <1111>    | Less than 10 cfu/g                                             |
| Presence of Pathogens per USP <62>                              | Specification                                                  |
| Pseudomonas aeruginosa                                          | Absent                                                         |
| Staphylococcus aureus                                           | Absent                                                         |
| Salmonella/Shigella                                             | Absent                                                         |
| Escherichia coli                                                | Absent                                                         |
| Candida albicans                                                | Absent                                                         |

### Table 1: Device Specifications for Agape Cooling Personal Lubricant

## Indications for Use:

Agape Cooling Personal Lubricant is a personal lubricant for penile, anal and/or vaginal application, intended to lubricate and moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

# Predicate Device Comparison:

The table below lists the comparative indications for use and technological characteristics of the subject and predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo for a business called "Wellness". The logo features a stylized, cursive font. The word "Wellness" is written in smaller letters to the right of the cursive letters. The logo is simple and elegant, and it is likely used to represent a business that focuses on health and well-being.

# Table 2: Comparator Table for Subject Device – Agape Cooling Personal Lubricant and Predicate Device – Premium Cooling Personal Lubricant

| Feature                    | Agape Cooling Personal<br>Lubricant                                                                                                                                                                                                                                                                                                                                                                                              | Premium Cooling Personal<br>Lubricant (K200602)                                                                                                                                                                                                                                                                                                                                     |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification Name | Lubricant, Personal                                                                                                                                                                                                                                                                                                                                                                                                              | Lubricant, Personal                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code               | NUC                                                                                                                                                                                                                                                                                                                                                                                                                              | NUC                                                                                                                                                                                                                                                                                                                                                                                 |
| Indications for Use        | Agape Cooling Personal Lubricant is<br>a personal lubricant for penile, anal<br>and/or vaginal application,<br>intended to lubricate and<br>moisturize, to enhance the ease<br>and comfort of intimate sexual<br>activity and supplement the body's<br>natural lubrication. This product is<br>compatible with natural rubber<br>latex and polyisoprene condoms.<br>This product is not compatible with<br>polyurethane condoms. | Premium Cooling Personal<br>Lubricant is a personal lubricant for<br>penile, anal and/or vaginal<br>application, intended to lubricate<br>and moisturize, to enhance the<br>ease and comfort of intimate sexual<br>activity and supplement the body's<br>natural lubrication. This product is<br>compatible with natural rubber<br>latex, polyisoprene and<br>polyurethane condoms. |
| Water soluble              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                              | No                                                                                                                                                                                                                                                                                                                                                                                  |
| Contains water             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                              | No                                                                                                                                                                                                                                                                                                                                                                                  |
| Ingredients                | Water, Propanediol, Gluconolactone,<br>Hydroxyethylcellulose, Sodium<br>Benzoate, Polysorbate 20, Citric Acid<br>and Mentha Piperita (Peppermint)<br>Leaf Extract                                                                                                                                                                                                                                                                | Dimethicone, Dimethiconol<br>and Menthol                                                                                                                                                                                                                                                                                                                                            |
| Over the counter use       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                 |
| Sterile                    | No                                                                                                                                                                                                                                                                                                                                                                                                                               | No                                                                                                                                                                                                                                                                                                                                                                                  |
| Condom Compatibility       | Natural Rubber Latex and<br>Polyisoprene                                                                                                                                                                                                                                                                                                                                                                                         | Natural Rubber Latex, Polyisoprene<br>and Polyurethane                                                                                                                                                                                                                                                                                                                              |
| Biocompatibility Tested    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                 |
| Antimicrobial Tested       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                 |
| Shelf life                 | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                          | 3 years                                                                                                                                                                                                                                                                                                                                                                             |

The subject and predicate device indications for use are not identical as the predicate device indications for use also states that it is compatible with polyurethane condoms. This difference does not represent a new intended use as the intended use of the subject device is the same as the predicate device, i.e., provides lubrication during intimate sexual activity. In addition, the subject and predicate devices have different technological characteristics; for example, different formulations, water solubility, condom compatibility, etc. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness.

## Summary of Performance Data:

## Biocompatibility

Biocompatibility studies were performed in accordance with the 2016 FDA Guidance document Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process" and ISO 10993-1:2018 as follows:

{6}------------------------------------------------

- Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010)
- Vaginal Irritation (ISO 10993-10:2010)
- Acute Systemic Toxicity (ISO 10993-11:2006) ●

The results of this testing demonstrated that the subject lubricant is non-cytotoxic, non-irritating, non-sensitizing and non-systemically toxic.

# Shelf-Life:

The subject device is a non-sterile personal lubricant with a three-year shelf-life in accordance with the results of a real time aging study. All device specifications listed in Table 1 were tested at Time 0 and at the end of the three-year shelf-life. The subject device met the device specifications at all time points assessed.

# Condom Compatibility:

The compatibility of the subject device with natural rubber latex, polyisoprene and polyurethane condoms was evaluated in accordance with ASTM D7661-10 (R)2017 - Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. The results of this test indicated that Agape Cooling Personal Lubricant is compatible with natural rubber latex and polyisoprene condoms. This product is not compatible with polyurethane condoms.

## Conclusion:

The results of the performance testing described above demonstrate that Agape Cooling Personal Lubricant is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K200894](https://fda-staging.innolitics.com/device/K200894)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
