The Apple iMac 27" 5K Retina with PL4.0 is intended to be used for displaying and viewing medical images, for review and analysis by trained medical practitioners. The device can NOT be used in mammography. The device can NOT be used for a life-support system. The device is intended for prescription use.
Device Story
Apple iMac 27" 5K Retina integrated with QUBYX PerfectLum 4.0 (PL 4.0) software; functions as medical image display system. Software suite enables DICOM calibration, AAPM TG18 verification, and remote management. Used in clinical settings by trained medical practitioners for review/analysis of medical images. System transforms display output to meet DICOM Part 14 GSDF and AAPM TG18 standards via software-driven calibration and test pattern generation. Healthcare providers use calibrated display to interpret medical images, supporting clinical decision-making. Benefits include high-resolution (5120 x 2880) and high-luminance (500 cd/m2) visualization of diagnostic data.
Clinical Evidence
Bench testing only. Compliance verified via DICOM conformance testing and AAPM TG18 acceptance testing using QUBYX PL 4.0 software and X-Rite i1 Display Pro measurement device. Results confirmed device meets DICOM Part 14 GSDF and AAPM TG18 standards for primary category medical image interpretation.
Technological Characteristics
Color IPS Retina display; 27" diagonal; 5120 x 2880 resolution; 0.1167 mm pixel pitch; 500 cd/m2 max luminance; 1000:1 contrast ratio. Software: QUBYX PerfectLum 4.0 for DICOM calibration and AAPM TG18 verification. Connectivity: Standard computer display interface. Power: Standard AC. Sterilization: Not applicable (non-contact device).
Indications for Use
Indicated for trained medical practitioners to display and view medical images for review and analysis. Not for use in mammography or life-support systems. Prescription use only.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
DELL UP3017 With QUBYX PerfectLum Bundle (K171229)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is placed inside a blue square. The text is stacked in three lines, with "ADMINISTRATION" on the bottom line.
% Marc Leppla Director Quantum Spectra LLC 501, Silverside Road St 105 Wilmington, Delaware 19809
June 2, 2020
Re: K200889
Trade/Device Name: Apple iMac 27" 5K Retina with PL4.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: PGY Dated: January 21, 2020 Received: April 3, 2020
Dear Marc Leppla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200889
Device Name Apple iMac 27" 5K Retina with PL4.0
Indications for Use (Describe)
The Apple iMac 27" 5K Retina with PL4.0 is intended to be used for displaying and viewing medical images, for review and analysis by trained medical practitioners.
The device can NOT be used in mammography.
The device can NOT be used for a life-support system.
The device is intended for prescription use.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|--|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | |
| <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary as required by 807.92
K200889
1. Company Identification
QUANTUM SPECTRA LLC 501 SILVERSIDE RD STE 105 WILMINGTON DE 19809
2. Official Correspondent
Mr. Marc Leppla President and CTO (Chief Technical Officer) marc@qs-group.net
3. Date of Submission 01/21/2020
4. Device Trade name Apple iMac 27" 5k Retina with PL 4.0
5. Common/Usual Name Image display system, Color LCD Monitor, image monitor/display
6, Classification Number Medical displays classified in Class II per 21 CFR 892.2050 Classification Name: Picture Archiving and Communications System
7.Predicate device Name: DELL UP3017 With QUBYX PerfectLum Bundle Manufacturer: DELL and QUBYX 510(k) number: K171229 Classification Number: Medical displays classified in Class II per 21 CFR 892.2050 Classification Name: Picture Archiving and Communications System
# 8. Device description
The Apple iMac 27″ 5k Retina with PerfectLum 4.0 (PL 4.0) is a computer with color display for medical viewing.
It is combined with QUBYX PerfectLum 4.0 and PerfectLum remote management, a user- friendly DICOM calibration and AAPM TG18 verification software suite. The software allows setting the display function to DICOM, displaying test pattern and performing acceptance and constancy tests.
Image /page/3/Picture/16 description: The image is a logo featuring a light blue circle with a stylized design inside. The design appears to be a combination of geometric shapes and lines, possibly representing letters or abstract forms. The overall impression is modern and minimalist, with the light blue color giving it a clean and fresh look. The logo is simple, yet visually striking, making it easily recognizable.
{4}------------------------------------------------
#### 9. Indications for use
The Apple iMac 27″ 5k Retina with PerfectLum 4.0 (PL 4.0) is intended to be used for displaying and viewing medical images, for review and analysis by trained medical practitioners.
The device can NOT be used in mammography. The device can NOT be used for a life-support system. The device is intended for prescription use.
### 10. Comparison of technological characteristics
| Specification | Apple iMac 27" 5k Retina<br>with PL 4.0 | DELL UP3017<br>with PerfectLum |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Screen technology | Color IPS Retina | Color TFT LCD Panel (IPS) |
| Active screen size (diagonal) | 68 cm / 27" | 30" |
| Resolution | 5120 x 2880 | 2560 x 1600 |
| Pixel pitch | 0.1167 mm | 0.25 mm |
| Color imaging | Yes | Yes |
| Gray imaging | Yes | Yes |
| Maximum luminance | 500 cd/m2 | 350 cd/m2 |
| DICOM calibrated luminance | 350 cd/m2 | 300 cd/m2 |
| Contrast ratio (typical) | 1000:1 | 1000:1 |
| Intended use | The Apple iMac 27″ 5k Retina with<br>PL 4.0 is intended to be used for<br>displaying and viewing medical<br>images, for review and analysis by<br>trained medical practitioners.<br>The device can NOT be used in<br>mammography.<br>The device can NOT be used for a<br>life-support system.<br>The device does<br>not contact with the patient | The DELL U3017 with<br>QUBYX PerfectLum is<br>intended to be used for<br>displaying and viewing<br>medical images, for review<br>and analysis by trained<br>medical practitioners. The<br>DELL U3017 can be used<br>only in conjunction with<br>QUBYX PerfectLum.<br>The device can not be used<br>in primary image diagnosis in<br>mammography.<br>The device can not be<br>used for a life-support<br>system. The device does<br>not contact with the patient |
{5}------------------------------------------------
## Details of testing:
To verify DICOM conformance, a DICOM conformance test was performed, using QUBYX PL 4.0 software and an X-Rite i1 Display Pro measurement device. The test procedure was generated by the software in accordance with the requirements of the DICOM standard. It consisted of measurement steps, where the meter measured display's characteristics and the software recorded them. Then the software analyzed the results in comparison with target values, defined by DICOM standard, and generated the report, stating that the display is DICOM-conformant.
The display device has successfully passed DICOM conformance test, so it is compliant with DICOM Part 14 GSDF standard. So is the predicate device, so the two devices are substantially equivalent in this regard.
To verify AAPM TG18 conformance, an acceptance test was performed, using QUBYX PerfectLum software and an X-Rite i1 Display Pro measurement device. The test procedure was generated by the software in accordance with the requirements of the AAPM TG18 standard and consisted of measurement and visual parts.
During the measurement steps, the meter measured display's characteristics and the software recorded them. During the visual steps, the user analyzed test patterns, generated by the software in accordance with AAPM standard. The software recorded the user's answers. Then the software analyzed the results in comparison with target values, defined by AAPM standard, and generated the report, stating that the display passes AAPM TG18 acceptance test.
The display device has successfully passed AAPM TG18 acceptance test, so it is compliant with AAPM TG18 standard and can be used as a primary category display for interpretation of medical images. The same is true for the predicate device, so the two devices are substantially equivalent in this regard.
Both devices have the same indications for use, except for predicate device it is not specified that it will not contact with the patient.
We can conclude that the new and predicate devices are substantially equivalent in terms of performance, indications for use, and principles of operation.
International standards: Directives: 2014/53/EU 2009/125/EC 2011/65/EU Safety and Health: EN 60950-1:2006+A1:2010+A11:2009+A12:2011+A2:2013 EN 50665:2017 2000-12-21 489-1 V2.2.0 EN 301 489-17 V3.2.0
RF Spectrum Efficiency: EN 300 328 V2.1.1 , EN 301 893 V2.1.1 , EN 300 440 V2.1.1
Additional Compliance: RoHS: EN50581:2012 , Ene
{6}------------------------------------------------
## Conclusion:
The comparison table shows that the subject device (Apple iMac 27" 5k Retina with PerfectLum 4.0) has the same intended use as the predicate. Although the devices have some different technological characteristics, as a higher resolution and higher maximum luminance, these differences do not make the subject device less safe and reliable, so the subject device fits for diagnostic use as the predicate device does.
Both devices are compliant with DICOM Part 14 GSDF and AAPM TG18 standards. To verify DICOM and AAPM compliance for the subject device, AAPM acceptance test and DICOM conformance test were also performed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.