The Feelix Stethoscope is intended for the detection, amplification and recording of sounds from the heart, lungs, anterior and posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs with selective frequency ranges. It can be used on any person undergoing a physical examination.
Device Story
Electronic stethoscope; microphone-based signal acquisition; captures heart, lung, and other internal organ sounds. Features active noise cancellation; up to 25X amplification; real-time audio output via Bluetooth or wired headset. Operates as standalone device; stores up to 50 recordings (10 seconds each). Used in clinical, transport, and home environments by healthcare professionals. Provides digital audio for physical examination; aids clinicians in auscultation; improves sound quality and consistency via uniform face sensitivity and noise suppression.
Clinical Evidence
Bench testing only. No clinical or animal data required. Bench testing included software verification/validation, electrical safety (IEC 60601-1, -1-2, -1-11, -1-12), usability testing, and evaluation of signal acquisition and noise suppression functions. Results demonstrate acoustic performance equivalent to other legally marketed electronic stethoscopes.
Technological Characteristics
Microphone-based sensing; lithium power source; Bluetooth and wired connectivity; active noise cancellation; 25X amplification. Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-1-12. Software compliant with IEC 62304:2015. Cleaning via isopropyl alcohol.
Indications for Use
Indicated for detection, amplification, and recording of body sounds (heart, lungs, abdomen, etc.) in any person undergoing a physical examination.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 14, 2020
Sonavi Labs, Inc. Ian Mclane Chief Technology Officer 1100 Wicomico St., Suite 600 Baltimore, Maryland 21230
Re: K200862
Trade/Device Name: Feelix Stethoscope Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD Dated: August 11, 2020 Received: August 12, 2020
Dear Ian Mclane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200862
Device Name Feelix Stethoscope
Indications for Use (Describe)
The Feelix Stethoscope is intended for the detection, amplification and recording of sounds from the heart, lungs, anterior and posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs with selective frequency ranges. It can be used on any person undergoing a physical examination.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|------------------------------------------------------------------------------------|
| <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Sonavi Labs. The logo consists of a stylized graphic of connected circles in purple and yellow above the company name in a simple sans-serif font. Below the company name is the address "1100 Wicomico St., Suite 600 Baltimore, MD 21230".
# 510(k) Summary - K200862
(per 21 CFR 807.92)
## Submitter Information
| Name | Sonavi Labs, Inc. |
|----------------|-----------------------------------------------------|
| Address | 1100 Wicomico St., Suite 600<br>Baltimore, MD 21230 |
| Phone Number | (410) 417-8167 |
| Contact Person | lan McLane<br>Chief Technology Officer |
| Date Prepared | March 30, 2020 |
### Device Information
| Trade Name | Feelix Stethoscope |
|----------------|------------------------------------|
| Common Name | Electronic Stethoscope |
| Classification | Stethoscope |
| | 21 CFR 870.1875 (Product Code DQD) |
### Predicate Device Information
| Device Name | M3DICINE's Stethee Pro 1 |
|-------------|--------------------------|
| K Number | K172296 |
#### Device Description
Sonavi Labs, in collaboration with Johns Hopkins University, has developed a next generation stethoscope with pristine digital audio to help clinicians and healthcare workers detect, amplify, and record sounds during physical examination. The Feelix Stethoscope (Feelix) re-engineers the classic stethoscope and offers three major improvements: (1) more uniform sensitivity over its entire face so that sound quality is less dependent on placement on the body (2) leading noise suppression and sound volume control, and (3) onboard real-time algorithms so the device can operate as a stand-alone rather than requiring connection to a computer.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Sonavi Labs. The logo consists of a stylized molecular structure above the company name, which is written in a simple, sans-serif font. Below the company name is the address: 1100 Wicomico St., Suite 600, Baltimore, MD 21230.
### Intended Use
To detect, amplify, and record sounds during a physical examination
### Indications for Use
The Feelix Stethoscope is intended for the detection, amplification and recording of sounds from the heart, lungs, anterior and posterior chest, abdomen, neck, limbs, arteries, veins and other internal organs with selective frequency ranges. It can be used on any person undergoing a physical examination.
#### Comparison to Predicate
The Feelix Stethoscope's indications for use statement includes a couple minor differences from the predicate device. These changes were implemented to better represent expected use, without impacting the underlying intended use.
### Technological Characteristics
As shown in Table 1, many technological similarities exist between the subject and predicate devices. The operating mechanism utilized in the Feelix Stethoscope is microphone-based, which is the same as the predicate device. Both devices demonstrate both safety and effectiveness by the ability to obtain a sound (signal acquisition) and to suppress surrounding noise without impacting the sound of interest. This is proven through scientific evidence presented in the premarket notification. Other technological characteristics of the Feelix Stethoscope, including software, EMC and ES testing, are similar to the predicate device.
| Characteristic | Subject Device<br>Feelix Electronic Stethoscope | Predicate Device (K172296)<br>M3DICINE StetheeTM Pro 1 | Comparison |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Feelix (or Feelix Pro)<br>Stethoscope | StetheeTM Pro 1 | Different<br>Trade name differences have<br>no impact on substantial<br>equivalence. |
| Common Name | Electronic Stethoscope | Electronic Stethoscope | Same |
| 510(K) Number | Unknown | K172296 | Different<br>510(k) numbers have no<br>impact on substantial<br>equivalence. |
| Regulation Classification | 21 CFR 870.1875(b) | 21 CFR 870.1875(b) | Same |
| Product Code | DQD | DQD | Same |
| Intended Use | Detection, amplification and<br>recording of sounds | Detection, amplification and<br>recording of sounds | Same<br>Note that the recording of<br>sounds is present in<br>M3DICINE's marketing<br>materials (e.g., website). |
| Characteristic | Subject Device<br>Feelix Electronic Stethoscope | Predicate Device (K172296)<br>M3DICINE Stethee™ Pro 1 | Comparison |
| Indications for Use | The Feelix Stethoscope is<br>intended for the detection,<br>amplification and recording of<br>sounds from the heart, lungs,<br>anterior and posterior chest,<br>abdomen, neck, limbs,<br>arteries, veins and other<br>internal organs with selective<br>frequency ranges. It can be<br>used on any person<br>undergoing a physical<br>examination. | The Stethee™ Pro 1 is an<br>electronic stethoscope<br>intended for medical<br>diagnostic purposes only. It<br>may be used for the detection<br>and amplification of sounds<br>from the heart, lungs,<br>arteries, veins, and other<br>internal organs. It can be used<br>on any person undergoing a<br>physical assessment. | Equivalent<br>Minor differences in language<br>add clarity and do not change<br>the overall intended use of<br>the device. |
| Anatomical Use Areas | Heart, lungs, anterior and<br>posterior chest, abdomen,<br>neck, limbs, arteries, veins<br>and other internal organs | Heart, lungs, arteries, veins,<br>and other internal organs | Equivalent<br>Minor differences in language<br>add clarity and do not change<br>the overall intended use of<br>the device. |
| Use Environment | Clinical, Transport (e.g.,<br>Ambulance), and Home | unknown | Undetermined<br>Predicate device marketing<br>materials state that device is<br>"telehealth ready" which<br>implies the same use<br>environment as Feelix.<br>Clinical and transport<br>environments are<br>commonplace for<br>stethoscope use. |
| Rx or OTC | Rx | Rx | Same |
| Power Source | Lithium | Lithium | Same |
| Battery Life | < 3 hours (continuous use) | unknown | Undetermined<br>Differences in battery life do<br>not have an impact on safety<br>or effectiveness. |
| Wire/Wireless Charging | Both | Wired | Different<br>The additional capability for<br>Feelix to be charged<br>wirelessly does not have an<br>impact on safety or<br>effectiveness. |
| Direct Listening | Sounds can be heard in real<br>time using a Bluetooth<br>enabled headset; sounds can<br>also be heard in real time by<br>wired headset | Sounds can be heard in real<br>time using a Bluetooth<br>enabled headset | Equivalent<br>The additional capability for<br>Feelix to have direct listening<br>by headset does not have an<br>impact on safety or<br>effectiveness. This function is<br>common to other electronic<br>stethoscopes. |
| Characteristic | Subject Device<br>Feelix Electronic Stethoscope | Predicate Device (K172296)<br>M3DICINE Stethee™ Pro 1 | Comparison |
| Noise Reduction Type | Proprietary Noise<br>Cancellation (Active) | Proprietary Noise<br>Cancellation | Equivalent<br>Effectiveness is proven out in<br>testing for both devices. |
| Amplification | Up to 25X | Up to 24X | Equivalent<br>Both devices have a ceiling to<br>prevent excessively loud<br>outputs that could cause<br>hearing damage. |
| Number of Recordings<br>(Capacity) | 50 recordings | None on device itself | Different<br>The additional capability for<br>Feelix is a non-medical<br>feature that does not impact<br>safety or effectiveness. |
| Length of Recordings | 10 seconds | 20 seconds | Different<br>Differences in recording<br>length do not have an impact<br>on safety or effectiveness. |
| Software | Proprietary Software<br>Compliant to IEC 62304:2015 | Proprietary Software<br>Compliant to IEC 62304:2015 | Same |
| Data Transfer | Bluetooth or Wired | Bluetooth | Equivalent<br>The additional capability for<br>Feelix to have a wired<br>connection does not have an<br>impact on safety or<br>effectiveness. This connection<br>type is common to other<br>electronic stethoscopes. |
| Cleaning | lsopropyl alcohol | Isopropyl alcohol | Same |
| ES & EMC | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-1-12 | IEC 60601-1<br>IEC 60601-1-2<br>[unknown if additional<br>collateral tests were<br>conducted] | Different<br>Additional tests performed by<br>Sonavi Labs do not raise new<br>or different questions of<br>safety or effectiveness. |
#### Table 1 - Technological Comparison
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Sonavi Labs. The logo consists of the word "SONAVI" in large, sans-serif font, with the word "LABS" in a smaller font below it. Above the word "SONAVI" is a graphic of three connected circles, with a color gradient from purple to yellow.
1100 Wicomico St., Suite 600
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Sonavi Labs. The logo features a stylized molecular structure in shades of purple, orange, and yellow above the word "SONAVI" in a sans-serif font. Below "SONAVI" is the word "LABS" in a smaller, sans-serif font.
1100 Wicomico St., Suite 600
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Sonavi Labs. The logo consists of the word "SONAVI" in large, sans-serif font, with the word "LABS" in a smaller font underneath. Above the word "SONAVI" is a graphic of three connected circles that transition in color from purple to yellow.
1100 Wicomico St., Suite 600 Baltimore, MD 21230
| Characteristic | Subject Device<br>Feelix Electronic Stethoscope | Predicate Device (K172296)<br>M3DICINE Stethee™ Pro 1 | Comparison |
|---------------------|-------------------------------------------------|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing | Usability Testing<br>Acoustic Performance | Performance (over range of<br>operation) | Different<br>The additional usability test<br>performed by Sonavi Labs<br>does not raise new or<br>different questions of safety<br>or effectiveness.<br>It is likely that the<br>performance testing was<br>different, as no information is<br>available on the predicate<br>device's testing regimen. The<br>subject device's acoustic<br>performance test<br>demonstrated that the device<br>was as effective as other<br>legally marketed electronic<br>stethoscopes. |
### Performance Data
Bench performance data were collected to support a substantial equivalence determination. Animal and clinical testing were not necessary. Bench performance testing included software verification and validation, electrical safety and EMC testing (including home and EMT collaterals), usability testing, and evaluation of signal acquisition and noise suppression functions. Performance test results demonstrate reasonable assurance that the Feelix Stethoscope can safely and effectively detect, amplify, and record sounds during a physical examination. These data suggest the Feelix Stethoscope is as safe and as effective as the identified predicate device.
#### Conclusions
The Feelix Stethoscope and its predicate device have intended use and similar indications, both serving as medical devices to detect, amplify, and record sounds during a physical examination. As described above, the minor technological differences between the Feelix Stethoscope and its predicate do not present any new or different issues of safety or effectiveness. Performance testing further confirms that these differences do not raise different questions of safety or effectiveness. Based upon this analysis and valid scientific evidence it can be concluded that the Feelix Stethoscope is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.