← Product Code [QAS](/productcode/QAS) · K200855

# CINA (K200855)

_Avicenna.Ai · QAS · Jun 24, 2020 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200855

## Device Facts

- **Applicant:** Avicenna.Ai
- **Product Code:** [QAS](/productcode/QAS.md)
- **Decision Date:** Jun 24, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2080
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K200855 · Jun 24, 2020 | CINA | Avicenna.Ai | Retrospective clinical anonymized cases (NCCT and CTA images) | The sponsor conducted a retrospective, multicenter, multinational study to evaluate the sensitivity, specificity, and time-to-notification of the CINA software using clinical images collected from routine practice. | Retrospective study; Multicenter; Clinical images; ICH; LVO |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective, blinded, multicenter, multinational study; Retrospective, blinded, multicenter, multinational study | Patients with suspected intracranial hemorrhage (ICH) and large vessel occlusion (LVO); Sample Size: 1,290 total (814 NCCT cases; 476 CTA cases); Number of Sites: 3 clinical sources | Not applicable for this study | Sensitivity, specificity, and time-to-notification |

## Indications for Use

CINA is a radiological computer aided triage and notification software indicated for use in the analysis of (1) non-enhanced head CT images and (2) CT angiographies of the head. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communicating suspected positive findings of (1) head CT images for Intracranial Hemorrhage (ICH) and (2) CT angiographies of the head for large vessel occlusion (LVO). CINA uses an artificial intelligence algorithm to analyze images and highlight cases with detected (1) ICH or (2) LVO on a standalone Web application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH or LVO findings. Notifications include compressed preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device. The results of CINA are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.

## Device Story

CINA is a radiological computer-assisted triage and notification software; operates in parallel to standard-of-care image interpretation. Inputs: non-enhanced head CT images and head CT angiographies. Processing: AI algorithms analyze DICOM images to detect suspected ICH or LVO; system flags positive cases. Outputs: active/passive notifications on a standalone web-based worklist application; includes compressed, non-diagnostic preview images. Usage: hospital networks; operated by radiologists. Impact: facilitates workflow triage/prioritization by alerting clinicians to suspected findings; enables earlier review of images in PACS. Benefit: potential for faster clinical intervention for time-sensitive conditions like ICH and LVO.

## Clinical Evidence

Retrospective, blinded, multicenter study (814 NCCT cases for ICH; 476 CTA cases for LVO). Ground truth established by three US-board-certified neuroradiologists. ICH application: 91.4% sensitivity, 97.5% specificity (AUC 0.94). LVO application: 97.9% sensitivity, 97.6% specificity (AUC 0.98). Mean time-to-notification: 21.6s (ICH) and 34.7s (LVO).

## Technological Characteristics

Runs on off-the-shelf server/workstation. Components: CINA Platform (interoperability/DICOM integration), ICH application, LVO application. Uses AI algorithms for image analysis. Connectivity: DICOM-compliant integration with hospital PACS/modalities. Software-based; no hardware components.

## Regulatory Identification

Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.

## Special Controls

Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

## Predicate Devices

- BriefCase ([K180647](/device/K180647.md))

## Reference Devices

- ContaCT ([DEN170073](/device/DEN170073.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 24, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is the Department of Health & Human Services logo, which features a stylized eagle. To the right of the eagle, there is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

AVICENNA.AI % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street, NW WASHINGTON DC 20004

Re: K200855

Trade/Device Name: CINA Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: March 31, 2020 Received: March 31, 2020

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

K200855

Device Name

CINA

Indications for Use (Describe)

CINA is a radiological computer aided triage and notification software indicated for use in the analysis of (1) nonenhanced head CT images and (2) CT angiographies of the head. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communicating suspected positive findings of (1) head CT images for Intracranial Hemorrhage (ICH) and (2) CT angiographies of the head for large vessel occlusion (LVO).

CINA uses an artificial intelligence algorithm to analyze images and highlight cases with detected (1) ICH or (2) LVO on a standalone Web application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH or LVO findings. Notifications include compressed preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.

The results of CINA are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D)  |
| Over-The-Counter Use (21 CFR 801 Subpart C)     |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

## AVICENNA.AI's CINA

# I. Submitter

## Applicant:

AVICENNA.AI 93 avenue des Sorbiers, Zone Athelia IV 13600 La Ciotat France

#### Contact Person:

Cyril Di Grandi CEO of Avicenna.Al Phone: +33 6 60 06 80 92 E-mail: cyril.di-grandi@avicenna.ai

Date prepared: March 31, 2020

## II. Device Identification

| Name of Device:       | CINA                                                               |
|-----------------------|--------------------------------------------------------------------|
| Classification Name:  | Radiological Computer-Assisted Triage And<br>Notification Software |
| Regulation No:        | 21 CFR § 892.2080                                                  |
| Product Code:         | QAS                                                                |
| Regulatory Class:     | Class II                                                           |
| Classification Panel: | Radiology devices                                                  |

#### III. Predicate Device

The CINA device is substantially equivalent to the following FDA cleared predicate device with regard to indications for use, performance, and technological characteristics:

| 510(k):              | K180647                                                            |
|----------------------|--------------------------------------------------------------------|
| Trade Name:          | BriefCase                                                          |
| Manufacturer:        | AiDoc Medical, Ltd                                                 |
| Classification Name: | Radiological Computer-Assisted Triage And<br>Notification Software |
| Regulation No:       | 21 CFR § 892.2080                                                  |
| Product Code:        | QAS                                                                |
| Regulatory Class:    | Class II                                                           |

{4}------------------------------------------------

A reference device is Viz.ai's ContaCT (DEN170073), which is a Class II device under the same requlation and product code as above.

# IV. Device Description

CINA is a radiological computer-assisted triage and notification software device.

The software system is based on algorithm-programmed components and is comprised of a standard off-the-shelf operating system and additional image processing applications.

DICOM images are received, recorded and filtered before processing. The series are processed chronologically by running algorithms on each series to detect suspicious results of an intracranial hemorrhage (ICH) or a large vessel occlusion (LVO), then notifications on the flagged series are sent to the Worklist Application.

The Worklist Application (on premise) displays the pop-up notifications of new studies with suspected findings when they come in, and provides both active and passive notifications. Active notifications are in the form of a small pop-up containing patient name, accession number and the type of suspected findings (ICH or LVO). All the non-enhanced head CT images and head CT angiographies studies received by CINA device are displayed in the worklist and those on which the algorithms have detected a suspected finding (ICH or LVO) are marked with an icon (i.e., passive notification). In addition, a compressed, small black and white image that is marked "not for diagnostic use" is displayed as a preview function. This compressed preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification. Presenting the radiologist with notification facilitates earlier triage by allowing one to prioritize images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.

# V. Intended Use / Indications for Use

CINA is a radiological computer aided triage and notification software indicated for use in the analysis of (1) non-enhanced head CT images and (2) CT angiographies of the head. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communicating suspected positive findings of (1) head CT images for Intracranial Hemorrhage (ICH) and (2) CT angiographies of the head for large vessel occlusion (LVO).

CINA uses an artificial intelligence algorithm to analyze images and highlight cases with detected (1) ICH or (2) LVO on a standalone Web application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH or LVO findings. Notifications include compressed preview images that are meant for informational purposes only. and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a diagnostic device.

The results of CINA are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.

{5}------------------------------------------------

# VI. Summary of Technological Characteristics

CINA runs on a standard "off the shelf" server/workstation and is made of the following software components:

- . CINA Image Processing Applications including two applications: ICH and LVO;
- CINA Platform server that includes the Worklist client application in which notifications from . the CINA Image Processing applications (ICH and LVO) are received.

Each of these components is briefly described below.

## VI.1. CINA Platform

The CINA platform is the technical platform for integrating and deploying CINA image processing applications. It provides the necessary requirements for interoperability with existing systems in the hospital radiology department such as CT modalities or other DICOM nodes (DICOM router or PACS for example). It is responsible for transferring, converting formats, notifying of suspected findings and display medical device data such as radiological data.

# VI.2. ICH Application

The ICH application includes the software algorithm responsible for identifying image characteristics that are consistent with an ICH. This application reads provided DICOM files, checks the DICOM properties to verify the compatibility with the recommended acquisition protocol, launches the algorithm and provides notification results (when an ICH is detected) compatible with the CINA Platform and with DICOM format.

## VI.3. LVO Application

The LVO application includes the software algorithm responsible for identifying and quantifying image characteristics that are consistent with an LVO. This application reads provided DICOM files, checks the DICOM properties to verify the compatibility with recommended acquisition protocol, launches the algorithm part and provides notification results (when a LVO is detected) compatible with the CINA Platform and with DICOM format.

## VII. Summary of Performance Data

The following performance data were provided in support of the substantial equivalence determination.

## VII.1. Software Verification and Validation Testing

{6}------------------------------------------------

The CINA device has been evaluated and verified in accordance with software specifications and applicable performance standards through Software Development and Validation Process to ensure performance according to specifications, User Requirements and Federal Requlations and Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

## VII.2. Performance Testing

Avicenna.Al conducted a retrospective, blinded, multicenter, multinational study with the CINA software with the primary endpoint to evaluate the software's performance in 1) non-contrast CT (NCCT) head images pertaining to patient with suspected intracranial hemorrhage (ICH) findings and 2) CT angiography (CTA) head series pertaining to patient with suspected large vessel occlusion (LVQ) findings, in 814 and 476 clinical anonymized cases, respectively. The device's sensitivity and specificity were analyzed, in addition to time-to-notification.

The data was provided from 3 clinical sources (2 US and 1 OUS). There were 255 (31.3%) positive ICH (images with ICH) and 188 (39.5%) positive LVO (images with LVO) cases included in the analysis.

Device sensitivities and specificities were compared to ground truth established by concurrence of three US-board-certified neuroradiologist readers.

Sensitivity and specificity for the "ICH" prioritization and triage application were observed to be 91.4% (95% Cl: 87.2% – 94.5%) and 97.5% (95% Cl: 95.8% – 98.6%), respectively. These findings achieved the 80% performance goal and are similar to the results reported for the predicate device BriefCase (Aidoc Medical): Sensitivity and specificity of 93.6% (95% Cl: 86.6% - 97.6%) and 92.3% (95% CI: 85.4% - 96.6%), respectively. ROC curve showed an AUC of 0.94.

Regarding the "LVO" prioritization and triage application, Sensitivity of 97.9% (95% CI: 94.6% - 99.4%) and 97.6% (95% Cl: 95.1% - 99%), respectively were observed. These results achieved the 80% performance goal and are similar to the ones reported by the reference device ContaCT (Viz.Al): Sensitivity was 87.8% (95% Cl: 81.2% – 92.5%) and specificity was 89.6% (95% CI: 83.7% – 93.9%). ROC curve showed an AUC of 0.98.

The results of the standalone assessment study demonstrated an overall agreement (accuracy) of 95.6% and 97.7% for the "ICH" and "LVO" tested cases, respectively, when compared to the ground truth (operators' visual assessments).

Positive predictive value (PPV) and negative predictive value (NPV) with varying prevalence, for both applications, are presented in Table 1 below:

{7}------------------------------------------------

|            | CINA - ICH triage application |         | CINA - LVO triage application |         |
|------------|-------------------------------|---------|-------------------------------|---------|
| Prevalence | PPV (%)                       | NPV (%) | PPV (%)                       | NPV (%) |
| 10%        | 80.2                          | 99.0    | 81.7                          | 99.8    |
| 15%        | 86.6                          | 98.5    | 87.7                          | 99.6    |
| 20%        | 90.1                          | 97.8    | 91.0                          | 99.5    |
| 25%        | 92.4                          | 97.1    | 93.1                          | 99.3    |
| 30%        | 94.0                          | 96.3    | 94.5                          | 99.1    |
| 35%        | 95.2                          | 95.5    | 95.6                          | 98.8    |
| 40%        | 96.1                          | 94.4    | 96.4                          | 98.6    |
| 45%        | 96.8                          | 93.2    | 97.1                          | 98.2    |
| 50%        | 97.3                          | 91.9    | 97.6                          | 97.9    |

## Table 1: PPV and NVP values for ICH and LVO image processing applications

Additionally, both "ICH" and "LVO" prioritization and triage effectiveness were evaluated by the standalone per-case processing time of the device (time-to-notification), the results are presented in Table 2 below:

| Time-to-Notification | MEAN ± SD   | MEDIAN    | MIN       | MAX       |
|----------------------|-------------|-----------|-----------|-----------|
|                      | (seconds)   | (seconds) | (seconds) | (seconds) |
| CINA - ICH           | 21.6 ± 4.4  | 20.4      | 14.4      | 53.3      |
| CINA - LVO           | 34.7 ± 10.7 | 33.4      | 14.3      | 63.3      |

The standalone effectiveness assessment demonstrated a substantial equivalence of the CINA -ICH triage application when compared to the predicate device (BriefCase). Specifically, the CINA's "ICH" triage mean "time-to-notification" was estimated to be 21.6 seconds. This is similar to the one reported by the predicate BriefCase device (mean: 4.46 minutes).

{8}------------------------------------------------

Regarding CINA - "LVO" triage application, the mean "time-to-notification" was estimated to be 34.7 seconds. This demonstrated a substantial equivalence of the CINA - LVO triage application effectiveness when compared to the reference ContaCT device (mean: 7.32 minutes).

The performance testing of the CINA device establishes that the subject device is as safe and effective as the predicate and reference devices. This established that the CINA device meets its intended use statement and is substantially equivalent to the predicate and reference devices.

#### VIII. Substantial Equivalence

The subject CINA for ICH and LVO prioritization and triage, the predicate device BriefCase device for ICH triage and the reference Viz.Al for LVO triage are all intended to aid in prioritization and triage of radiological images for ICH and/or LVO. The labeling of the subject and the predicate devices clearly states that the devices are not for diagnostic use. All devices are software packages with similar technological characteristics and principles of operation, and incorporate deep learning Al algorithms that process images, and software to send notifications and to display unannotated preview images. In all three devices, the labeling instructs the user to further evaluate and diagnose based only on the original images in the local PACS.

All three devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, and do not remove images from the standard of care FIFO queue, thus not disturbing standard interpretation of the images by the attending radiologists. The subject, predicate and reference devices achieve performance of the time-to-notification metric in similar ranges of time, and thus contribute similarly to effective triage and early involvement of the radiologist in evaluating suspected images of ICH and/or LVO.

The standalone performance and effectiveness assessment studies demonstrated that the CINA device performs as intended is therefore substantially equivalent to the BriefCase and Viz.Al reference devices.

{9}------------------------------------------------

Table 3 compares the key features of the subject and the predicate and reference devices.

| Table 3: Comparison of key features between CNA and predicate device (Aidoc BriefCase) and reference device (Viz.Al ContaCT) |  |  |  |  |  |
|------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|
|------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|

|                                                                | Subject device: CINA Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device: Aidoc BriefCase<br>Software (K180647)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Reference device: Viz.AI ContaCT<br>software (DEN170073)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for Use                          | CINA is a radiological computer<br>aided triage and notification<br>software indicated for use in the<br>analysis of (1) non-enhanced head<br>CT images and (2) CT<br>angiographies of the head.<br><br>The device is intended to assist<br>hospital networks and trained<br>radiologists in workflow triage by<br>flagging and communicating<br>suspected positive findings of (1)<br>head CT images for Intracranial<br>Hemorrhage (ICH) and (2) CT<br>angiographies of the head for large<br>vessel occlusion (LVO).<br><br>CINA uses an artificial intelligence<br>algorithm to analyze images and<br>highlight cases with detected (1)<br>ICH or (2) LVO on a standalone<br>Web application in parallel to the<br>ongoing standard of care image<br>interpretation. The user is presented<br>with notifications for cases with | BriefCase is a radiological computer<br>aided triage and notification software<br>indicated for use in the analysis of<br>non-enhanced head CT images.<br><br>The device is intended to assist<br>hospital networks and trained<br>radiologists in workflow triage by<br>flagging and communication of<br>suspected positive findings of<br>pathologies in head CT images,<br>namely Intracranial Hemorrhage<br>(ICH).<br><br>BriefCase uses an artificial<br>intelligence algorithm to analyze<br>images and highlight cases with<br>detected ICH on a standalone<br>desktop application in parallel to the<br>ongoing standard of care image<br>interpretation. The user is presented<br>with notifications for cases with<br>suspected ICH findings.<br><br>Notifications include compressed | ContaCT is a notification-only,<br>parallel workflow tool for use by<br>hospital networks and trained<br>clinicians to identify and<br>communicate images of specific<br>patients to a specialist, independent<br>of standard of care workflow.<br><br>ContaCT uses an artificial intelligence<br>algorithm to analyze images for<br>findings suggestive of a pre-specified<br>clinical condition and to notify an<br>appropriate medical specialist of<br>these findings in parallel to standard<br>of care image interpretation.<br>Identification of suspected findings is<br>not for diagnostic use beyond<br>notification. Specifically, the device<br>analyzes CT angiogram images of<br>the brain acquired in the acute setting<br>and sends notifications to a<br>neurovascular specialist that a<br>suspected large vessel occlusion has<br>been identified and recommends<br>review of those images. Images can |
|                                                                | Subject device: CINA Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device: Aidoc BriefCase<br>Software (K180647)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Reference device: Viz.Al ContaCT<br>software (DEN170073)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                                                | suspected ICH or LVO findings.<br>Notifications include compressed<br>preview images that are meant for<br>informational purposes only, and are<br>not intended for diagnostic use<br>beyond notification. The device does<br>not alter the original medical image,<br>and it is not intended to be used as<br>a diagnostic device.<br>The results of CINA are intended to<br>be used in conjunction with other<br>patient information and based on<br>professional judgment to assist with<br>triage/prioritization of medical<br>images. Notified clinicians are<br>ultimately responsible for reviewing<br>full images per the standard of care.                                                                                                                                                                                 | preview images that are meant for<br>informational purposes only and not<br>intended for diagnostic use beyond<br>notification. The device does not alter<br>the original medical image and is not<br>intended to be used as a diagnostic<br>device.<br>The results of BriefCase are intended<br>to be used in conjunction with other<br>patient information and based on<br>professional judgment, to assist with<br>triage/prioritization of medical<br>images. Notified clinicians are<br>responsible for viewing full images<br>per the standard of care.                                                                                                                                                                                                                                   | be previewed through a mobile<br>application.<br>Images that are previewed through<br>the mobile application are<br>compressed and are for informational<br>purposes only and not intended for<br>diagnostic use beyond notification.<br>Notified clinicians are responsible for<br>viewing non-compressed images on<br>a diagnostic viewer and engaging in<br>appropriate patient evaluation and<br>relevant discussion with a treating<br>physician before making care-related<br>decisions or requests. ContaCT is<br>limited to analysis of imaging data<br>and should not be used in-lieu of full<br>patient evaluation or relied upon to<br>make or confirm diagnosis.                                                                                                                                                                                                                                              |
| User population                                                | Radiologist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Radiologist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Clinician (e.g., neurovascular<br>specialist)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Anatomical region<br>of interest                               | Head                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Head                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Head                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Data acquisition<br>protocol                                   | Non contrast CT scan of the head or<br>neck and CT angiogram images of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Non contrast CT scan of the head or<br>neck                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | CT angiogram images of the brain                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                                | Subject device: CINA Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate device: Aidoc BriefCase Software (K180647)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Reference device: Viz.Al ContaCT software (DEN170073)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                | the brain                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| View DICOM data                                                | DICOM information about the patient, study and current image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | DICOM information about the patient, study and current image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | DICOM information about the patient, study and current image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Segmentation of region of interest                             | No; device does not mark, highlight, or direct users' attention to a specific location in the original image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | No; device does not mark, highlight, or direct users' attention to a specific location in the original image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | No; device does not mark, highlight, or direct users' attention to a specific location in the original image                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Algorithm                                                      | Artificial intelligence algorithm with database of images                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Artificial intelligence algorithm with database of images                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Artificial intelligence algorithm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Notification / Prioritization                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Preview images                                                 | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes.<br>The device operates in parallel with the standard of care, which remains the default option for all cases.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes.<br>The device operates in parallel with the standard of care, which remains the default option for all cases.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Presentation of a small, compressed, black and white preview image that is labeled "Not for diagnostic use".                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Alteration of original image                                   | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Removal of cases from worklist queue                           | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Structure                                                      | - LVO and ICH image processing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | - AHS module (image acquisition),                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | - Image Forwarding Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Subject device: CINA Software                                  | Predicate device: Aidoc BriefCase<br>Software (K180647)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Reference device: Viz.Al ContaCT<br>software (DEN170073)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| applications<br>- CINA Platform (worklist and Image<br>Viewer) | - ACS module (image processing),<br>- Aidoc Worklist application for<br>workflow integration (worklist and<br>Image Viewer).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | - Image Processing and Analysis<br>Software<br>- Non-diagnostic DICOM viewing<br>mobile application                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K200855](https://fda-staging.innolitics.com/device/K200855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).