K200852 · GE Medical Systems Ultrasound and Primary Care Diagnostics · LLZ · Sep 18, 2020 · Radiology
Device Facts
Record ID
K200852
Device Name
EchoPAC Software Only, EchoPAC Plug-In
Applicant
GE Medical Systems Ultrasound and Primary Care Diagnostics
Product Code
LLZ · Radiology
Decision Date
Sep 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Cardiac 2D measurements
AI-based feature
—
—
—
—
—
—
Doppler spectra recognition
AI-based feature
—
—
—
—
—
—
Left ventricle strain analysis
—
—
—
—
—
—
—
Left ventricular function assessment
—
—
—
—
—
—
—
Right ventricular strain analysis
—
—
—
—
—
—
—
Left atrial strain and volume analysis
—
—
—
—
—
—
—
Tricuspid annulus and valve quantification
—
—
—
—
—
—
—
Indications for Use
EchoPAC Software Only / EchoPAC Plug-in is intended for diagnostic review and analysis of ultrasound images, patient record management and reporting, for use by, or on the order of a licensed physician. EchoPAC Software Only / EchoPAC Plug-in allows post-processing of raw data images from GE ultrasound scanners and DICOM ultrasound images.
Device Story
EchoPAC Software Only/Plug-in provides post-acquisition image processing, annotation, analysis, measurement, reporting, and storage for ultrasound images acquired via GE Vivid scanners. Software operates as a standalone application on customer PC hardware or as a plug-in for image management workstations. It processes raw data and DICOM images; supports B, M, Color M, Color, Power, Pulsed/CW Doppler, Coded Pulse, Harmonic, 3D, and 4D modes. Clinicians use the software for diagnostic review and clinical decision-making across various anatomical applications. Features include AI-based automated measurements (Auto Measure, Spectrum Recognition, AFI, Auto EF, RV/LA strain analysis) and visualization tools (FlexiLight, HD Color). The device facilitates DICOM-compliant data transfer via LAN. It benefits patients by enabling advanced quantitative analysis and standardized reporting of ultrasound data, aiding in accurate diagnosis and longitudinal monitoring.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including risk analysis, requirements/design reviews, unit-level verification, integration testing, performance testing, and safety testing. Compliance with ISO 14971, NEMA PS 3.1-3.20 (DICOM), IEC 62304, and IEC 62366-1.
Technological Characteristics
Software-only workstation application for ultrasound image post-processing. DICOM-compliant; supports LAN connectivity. Implements AI/ML-based automated measurement and recognition algorithms. Complies with ISO 14971 (risk management), IEC 62304 (software lifecycle), and IEC 62366-1 (usability).
Indications for Use
Indicated for diagnostic review and analysis of ultrasound images, patient record management, and reporting. Patient populations include fetal/obstetrics, abdominal (renal, GYN), urology (prostate), pediatric, small organs (breast, testes, thyroid), neonatal, adult cephalic, cardiac (adult/pediatric), peripheral vascular, transesophageal, musculoskeletal, transrectal, transvaginal, intraoperative, intra-cardiac, and intra-luminal applications.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC % Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226
## Re: K200852
Trade/Device Name: EchoPAC Software Only, EchoPAC Plug-In Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: August 21, 2020 Received: August 25, 2020
Dear Tracey Ortiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
September 18, 2020
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
2
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200852
Device Name
EchoPAC Software Only, EchoPAC Plug-in
#### Indications for Use (Describe)
EchoPAC Software Only / EchoPAC Plug-in is intended for diagnostic review and analysis of ultrasound images, patient record management and reporting, for use by, or on the order of a licensed physician. EchoPAC Software Only / EchoPAC Plug-in allows post-processing of raw data images from GE ultrasound scanners and DICOM ultrasound images.
Ultrasound images are acquired via B (2D), M, Color M modes, Color, Power, Pulsed and CW Doppler modes, Coded Pulse, Harmonic, 3D, and Real time (RT) 3D Mode (4D).
Clinical applications include: Fetal/Obstetrics; Abdominal (including renal and GYN), Urology (including prostate), Pediatric; Small organs (breast, testes, thyroid), Neonatal and Adult Cephalic, Cardiac (adult and pediatric), Peripheral Vascular, Transesophageal (TEE), Musculo-skeletal Conventional, Musculo-skeletal Superficial, Transrectal (TR), Transvaginal (TV), Intraoperative (vascular), Intra-cardiac, and Intra-luminal.
Type of Use (Select one or both, as applicable):
| <div> <span> <b> × Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> |
|------------------------------------------------------------------------------------------------------------|
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized, cursive font. The letters are white and are set against a blue circular background. The blue background has a textured appearance, resembling a swirling pattern.
K200852 GE Healthcare 510(k) Premarket Notification Submission
## 510(k) Summary
| In accordance with 21 CFR 807.92 the following summary of information is provided: | |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Date: | March 27, 2020 |
| Submitter: | GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
| Primary Contact Person: | Tracey Ortiz<br>Regulatory Affairs Director<br>GE Healthcare T:(262)676-<br>6120 |
| Secondary Contact Person: | Charlotte K. Munthe Jørgensen<br>Senior Regulatory Affairs Leader<br>GE Vingmed Ultrasound AS |
| Device Trade Name: | EchoPAC Software Only / EchoPAC Plug-in |
| Common/Usual Name: | Workstation Software for ultrasound image review, analysis and<br>reporting |
| Classification Names: | Class II |
| Product Code: | Picture archiving and communications system, 21 CFR 892.2050<br>LLZ |
| Primary Predicate<br>Device(s): | EchoPAC Software Only / EchoPAC Plug-in (K170847) |
| Reference Predicate<br>Device(s): | Vivid S70N (K182450)<br>Vivid E95 (K181685)<br>Vivid E95 (K200743) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The logo is enclosed within a circular border, and the color of the logo and border is a light blue.
# GE Healthcare 510(k) Premarket Notification Submission
EchoPAC Software Only / EchoPAC Plug-in provides image Device Description: processing, annotation, analysis, measurement, report generation, communication, storage and retrieval functionality to ultrasound images that are acquired via the Vivid family of ultrasound scanners by GE Healthcare. The EchoPAC Software Only / EchoPAC Plug-in software is an integral component of each Vivid system, providing the post-acquisition image management and reporting functions of the scanner. EchoPAC Software Only will be offered as SW-only to be installed directly on customer PC hardware, and EchoPAC Plug-in will be offered as an accessory to image management workstations. EchoPAC Software Only / EchoPAC Plug-in is DICOM compliant, transferring images and data via LAN between scanners, hard copy devices, file servers and other workstations. EchoPAC Software Only / EchoPAC Plug-in is intended for Intended Use/ Indication for Use: diagnostic review and analysis of ultrasound images, patient record management and reporting, for use by, or on the order of a licensed physician. EchoPAC Software Only / EchoPAC Plug-in allows post-processing of raw data images from GE ultrasound scanners and DICOM ultrasound images. Ultrasound images are acquired via B (2D), M, Color M modes, Color, Power, Pulsed and CW Doppler modes, Coded Pulse, Harmonic, 3D, and Real time (RT) 3D Mode (4D). Clinical applications include: Fetal/Obstetrics; Abdominal (including renal and GYN), Urology (including prostate), Pediatric; Small organs (breast, testes, thyroid), Neonatal and Adult Cephalic, Cardiac (adult and pediatric), Peripheral Vascular, Transesophageal (TEE), Musculo-skeletal Conventional, Musculoskeletal Superficial, Transrectal (TR), Transvaginal (TV), Intraoperative (vascular), Intra-cardiac, and Intra-luminal. Technology: EchoPAC Software Only / EchoPAC Plug-in employs the same fundamental scientific technology as its predicate device. Determination of Comparison to Predicates Substantial Equivalence: EchoPAC Software Only / EchoPAC Plug-in is substantially equivalent to the predicate device with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script. The logo is enclosed in a circular frame with decorative swirls around the perimeter.
## GE Healthcare 510(k) Premarket Notification Submission
The following is an overview of the difference between the proposed EchoPAC Software Only / EchoPAC Plug-in and its predicates.
Features/Functionality additions:
- . Al Auto Measure – 2D: same feature as cleared on Vivid E95 (K200743). It is Al (Artificial Intelligence) based Cardiac Auto 2D feature that enables semi-automated measurements on a PLAX image.
- Al Auto Measure – Spectrum Recognition: same feature as cleared on Vivid E95 (K200743). It is Al based Spectrum Recognition feature that enables automated recognition of common Doppler spectra and automatically starts the Auto Doppler measurement or opens the appropriate manual measurement folder.
- . AFI 3.0: same feature as cleared on Vivid E95 (K200743), based on AFI 2.0 (cleared in Vivid E95, K181685), adds the ability to analyze the left ventricle on both GEHC raw data images and DICOM images from 3rd party ultrasound scanners.
- Auto EF 3.0: same feature as cleared on Vivid E95 ● (K200743), based on Auto EF 2.0 (cleared in Vivid E95, K181685), adds the ability to assess LV function on raw data images acquired with GEHC scanners as well as on DICOM images from other vendors systems.
- . AFI RV: same feature as cleared on Vivid E95 (K200743), based on AFI 2.0 (cleared in Vivid E95, K181685) but modified for the right ventricle (RV). It is a parametric tool giving quantitative data for right ventricular longitudinal global strain, free wall strain and segmental strain derived from the apical 4-chamber RV focused view.
- AFI LA: same feature as cleared on Vivid E95 (K200743), ● based on AFI 2.0 (cleared in Vivid E95, K181685) but modified for the left atrium. It provides quantitative data for left atrial (LA) global strain. The tool also supports measurements of LA volumes and emptying fraction (EF).
- Launchpad: same feature as cleared on Vivid E95 ● (K200743), allows the display and launch/starting of thirdparty software apps. The offered apps are verified to be compatible with EchoPAC Software Only / EchoPAC Plugin.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle is surrounded by swirling, wave-like shapes, also in blue, giving the logo a dynamic and recognizable appearance. The logo is simple, yet distinctive, and has been used by GE for many years.
# GE Healthcare
510(k) Premarket Notification Submission
- FlexiLight and HD Color: same features as cleared on ● Vivid E95 (K200743), are visualization tools enhancing depth and color perception respectively of 3D objects on a 2D monitor
- 4D Auto TVQ: same feature as cleared on Vivid E95 ● (K200743), based on 4D Auto MVQ (cleared in EchoPAC Software Only/EchoPAC Plug-in, K170847), but modified for quantification of tricuspid annulus and valve in 4D echocardiographic data.
- DICOM PDF Read: same feature as cleared on Vivid E95 (K200743). The system supports read-only access to DICOM PDF reports created on a DICOM server
- The new feature "Open 4D" loads 3D or 4D ultrasound ● images from 300 party vendors and allows user to do measurements and analysis.
- . Z-score enhancement: same enhancements as cleared on Vivid E95 (K200743). More Z-score sets of values added based on published literature - are being added to those previously cleared in predicate EchoPAC Software Only / EchoPAC Plug-in.
- 4D marker enhancement: same feature as cleared on Vivid ● E95 (K200743). Feature cleared in predicate Vivid E95 (K181685), is modified to allow user to modify individual markers.
- Adding clinical applications Obstetrics, Intra-luminal and ● Indications for use: Intra-Cardiac which were cleared with Vivid S70N (K182450).
- Clarified Real time 3D as 4D and listed 3D.
The EchoPAC Software Only / EchoPAC Plug-in and its applications Summary of Non-Clinical Tests: comply with voluntary standards: 1. ISO 14971:2007 - Medical devices - Application of risk management to medical devices 2. NEMA PS 3.1 -3.20 Digital Imaging and Communications in Medicine (DICOM) Set, (Radiology), 2016 3. IEC 62304:2006 A1 2015 - Medical device software -Software life cycle process 4. IEC 62366-1: 2015 – Medical Device-Part 1: Application of Usability.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The letters are also blue, matching the color of the circle that surrounds them.
GE Healthcare 510(k) Premarket Notification Submission
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews ●
- Design Reviews
- Testing on unit level (Module verification) ●
- Integration testing (System verification) ●
- Performance testing (Verification)
- Safety testing (Verification ●
The subject of this premarket submission, EchoPAC Software Conclusion: Only / EchoPAC Plug-in, did not require clinical studies to support substantial equivalence.
> GE Healthcare considers the proposed EchoPAC Software Only / EchoPAC Plug-in to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
Summary of Clinical Tests:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.