K200849 · Remote Diagnostic Technologies, Ltd. A Philips Company · LDD · Jul 24, 2020 · Cardiovascular
Device Facts
Record ID
K200849
Device Name
Tempus LS - Manual
Applicant
Remote Diagnostic Technologies, Ltd. A Philips Company
Product Code
LDD · Cardiovascular
Decision Date
Jul 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5300
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The TEMPUS LS - MANUAL is a multi-function, portable of providing defibrillation therapy, noninvasive external pacing, displaying ECG, and CPR feedback in hospital settings. It is to be used by qualified medical health care professionals trained in the use and operation of the device and qualified by training in advanced cardiovascular life (ACLS) support. In Manual Defibrillation mode, the TEMPUS LS - MANUAL is indicated for victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions: - Unconsciousness - Absence of breathing, and - Absence of pulse. This product should be used only by qualified medical personnel for terminating ventricular fibrillation and ventricular tachycardia. Defibrillation is indicated for adult and pediatic patients. When a victim is less than 8 years of age, or weighs less than 55 lbs. (25kg), the TEMPUS LS - MANUAL should be used with the Tempus Pediatric Electrodes or in the device's pediatic mode with the Tempus Adult Electrodes. Therapy should not be delayed to determine the patient's exact age or weight. The TEMPUS LS - MANUAL may be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardia (VT) by using the R-wave of the patient's ECG as a timing reference. A qualified user trained in the operation of the device must decide when synchronized cardioversion is appropriate. Cardioversion is indicated for adult and pediatric patients. The TEMPUS LS - MANUAL is indicated for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include: - Resuscitation from standstill or bradycardia of any etiology. - As a standby when standstill or bradycardia might be expected. External pacing is indicated for adult and pediatric patients. The TEMPUS LS - MANUAL is intended to display a patient's ECG from a 3- or 4- lead wire ECG patient cable. It is indicated for use in adult or pediatric for attended monitoring of the patient's ECG signal (e.g., before starting demand pacing or after successful defibrillation). The TEMPUS LS - MANUAL may be supplied with an optional CPR feedback sensor, which is intended to provide visual and audio feedback to encourage rescuers to perform CPR in accordance with AHA/ERC guidelines for chest compression rate and depth. It is indicated for use on adult and pediatric patients ≥ 8 years of age or weighing ≥ 25 kg / 55 lbs. CPR Feedback is not indicated for use on neonatal and pediatric patients under 8 years of age or weighing less than 25 kg/55 lbs.
Device Story
Compact, portable DC defibrillator; delivers up to 200J via self-adhesive electrodes. Inputs: ECG signals (3/4-lead), CPR feedback sensor data. Operation: Manual defibrillation (asynchronous/synchronized), non-invasive pacing (fixed/demand), ECG monitoring, CPR metronome/feedback. Used in hospital/pre-hospital settings by ACLS-trained professionals. Output: Visual/audio CPR guidance, ECG display, shock delivery. Benefits: Rapid intervention for cardiac arrest/arrhythmias. Data export via Bluetooth/USB (Tempus Data Link).
Clinical Evidence
Bench testing only. Evaluated defibrillation, cardioversion, pacing, ECG display, and CPR feedback performance. Environmental testing (vibration, shock, altitude, climate) and electrical safety/EMC testing performed. Human factors validation study with 15 participants confirmed usability in intended environments. No clinical data required.
Indicated for adult and pediatric patients experiencing cardiac arrest (unconsciousness, apnea, pulselessness) requiring defibrillation, or patients requiring synchronized cardioversion for AF/VT, or temporary external pacing for bradycardia/standby. Also indicated for ECG monitoring and CPR feedback (patients ≥ 8 years or ≥ 25 kg). Contraindicated for CPR feedback in neonates/pediatrics < 8 years or < 25 kg.
Regulatory Classification
Identification
Low-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of a maximum of 360 joules of energy used for defibrillating (restoring normal heart rhythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. This generic type of device includes low energy defibrillators with a maximum electrical output of less than 360 joules of energy that are used in pediatric defibrillation or in cardiac surgery. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body. High-energy DC-defibrillator: A device that delivers into a 50 ohm test load an electrical shock of greater than 360 joules of energy used for defibrillating the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The device may either synchronize the shock with the proper phase of the electrocardiogram or may operate asynchronously. The device delivers the electrical shock through paddles placed either directly across the heart or on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 24, 2020
Remote Diagnostic Technologies Ltd., a Philips Company % Meaghan Bailey Executive Director NSF Health Sciences, LLC 2001 Pennsylvania Avenue NW, Suite 950 Washington, District of Columbia 20006
Re: K200849
Trade/Device Name: TEMPUS LS - MANUAL Regulation Number: 21 CFR 870.5300 Regulation Name: DC-Defibrillator (Including Paddles) Regulatory Class: Class II Product Code: LDD, DRO, DPS, LIX Dated: June 26, 2020 Received: June 26, 2020
Dear Meaghan Bailey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200849
Device Name TEMPUS LS - MANUAL
### Indications for Use (Describe)
The TEMPUS LS - MANUAL is a multi-function, portable of providing defibrillation therapy, noninvasive external pacing, displaying ECG, and CPR feedback in hospital settings. It is to be used by qualified medical health care professionals trained in the use and operation of the device and qualified by training in advanced cardiovascular life (ACLS) support.
### Manual Defibrillation
In Manual Defibrillation mode, the TEMPUS LS - MANUAL is indicated for victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions:
- · Unconsciousness
- · Absence of breathing, and
- · Absence of pulse.
This product should be used only by qualified medical personnel for terminating ventricular fibrillation and ventricular tachycardia.
Defibrillation is indicated for adult and pediatic patients. When a victim is less than 8 years of age, or weighs less than 55 lbs. (25kg), the TEMPUS LS - MANUAL should be used with the Tempus Pediatric Electrodes or in the device's pediatic mode with the Tempus Adult Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.
### Cardioversion
The TEMPUS LS - MANUAL may be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardia (VT) by using the R-wave of the patient's ECG as a timing reference. A qualified user trained in the operation of the device must decide when synchronized cardioversion is appropriate. Cardioversion is indicated for adult and pediatric patients.
### External Pacing
The TEMPUS LS - MANUAL is indicated for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include:
- · Resuscitation from standstill or bradycardia of any etiology.
- · As a standby when standstill or bradycardia might be expected.
External pacing is indicated for adult and pediatric patients.
### ECG Display
The TEMPUS LS - MANUAL is intended to display a patient's ECG from a 3- or 4- lead wire ECG patient cable. It is indicated for use in adult or pediatric for attended monitoring of the patient's ECG signal (e.g., before starting demand pacing or after successful defibrillation).
### CPR Feedback The TEMPUS LS - MANUAL may be supplied with an optional CPR feedback sensor, which is intended to provide
FORM FDA 3881 (7/17)
{3}------------------------------------------------
visual and audio feedback to encourage rescuers to perform CPR in accordance with AHA/ERC guidelines for chest compression rate and depth. It is indicated for use on adult and pediatric patients ≥ 8 years of age or weighing ≥ 25 kg / 55 lbs.
CPR Feedback is not indicated for use on neonatal and pediatric patients under 8 years of age or weighing less than 25 kg/55 lbs.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY
### Submitter Information
- Applicant
| Name: | Remote Diagnostic Technologies Ltd, a Philips company |
|--------------------------|----------------------------------------------------------------------------------|
| Address: | Pavilion C2, Ashwood Park, Ashwood Way, Basingstoke,<br>RG23 8BG, United Kingdom |
| | Attn: Martin Newman, Director, Quality Assurance +<br>Regulatory Affairs |
| Phone No: | +44 (0) 1256 362 400 |
| Submitter | |
| Name: | NSF Health Sciences |
| Address: | 2001 Pennsylvania Ave. NW, Suite 950, Washington DC<br>20006 |
| Contact Person: | Meaghan Bailey, Executive Director |
| Phone No: | 202-445-0577 |
| Email: | mbailey@nsf.org |
| Date Summary Prepared: | July 22, 2020 |
| Device Identification | |
| Device Name: | TEMPUS LS – MANUAL |
| Common Name: | Low-Energy Defibrillator |
| Classification: | II |
| Product Code: | LDD |
| Secondary Product Codes: | DRO, LIX, DPS |
| Regulation Number: | 21 CFR 870. 5300 |
| Predicate Device(s) | |
| Primary Predicate: | ZOLL E-Series ALS (K140502) |
| Secondary Predicate: | Philips HeartStart XL (K021453) |
| | Philips HeartStart XL+ (K133659) |
### Device Description
The TEMPUS LS – MANUAL is a compact, lightweight, portable electromedical device that contains a direct current (DC) defibrillator capable of delivering up to 200 joules of electrical energy to a patient through selfadhesive external electrode pads. The device is intended for use by advanced cardiac life support (ACLS) trained health care professionals in hospital settings to provide defibrillation therapy, non-
| Traditional 510(k) – TEMPUS LS – MANUAL Defibrillator |
|-------------------------------------------------------|
| 510(k) Summary |
{5}------------------------------------------------
invasive external pacing, cardiopulmonary resuscitation (CPR) feedback, and non-alarming electrocardiograph (ECG) monitoring. The device is intended for prescription-use only. The unit is powered by a rechargeable lithium-ion battery which is recharged when connected to a mains power supply. The unit incorporates a color LCD display that displays information related to a clinical intervention. The trained user may select an energy level between 1 - 200 ioules and deliver a shock in either an asynchronous mode of delivery or through synchronized cardioversion using the patient's R-wave as a timing reference for the shock.
The device provides non-invasive transcutaneous pacing in either fixed or demand modes of operation through the same self-adhesive pads used in manual defibrillation. In addition, the device incorporates an audible metronome to quide a user as to the correct rate at which chest compressions should be administered in accordance with current American Heart Association (AHA) resuscitation guidelines. The device may be supplied with an optional cardiopulmonary resuscitation (CPR) feedback sensor which assists caregivers during CPR in delivering chest compressions at a rate and depth in compliance with current AHA guidelines. The device may also be paired to a compatible external device via a secure wireless communications protocol called Tempus Data Link (TDL) to export event data (without patient identifiers) at the convenience of the user.
TEMPUS LS - MANUAL features list:
- Manual mode defibrillation delivered in a Bi-phasic Truncated Exponential (BTE) waveform ●
- 1 200 joule user-configurable energy levels 0
- Synchronized cardioversion 0
- External pacer (fixed & demand modes) .
- CPR metronome & optional sensor providing feedback on CPR rate and CPR depth ●
- 3-lead wire and 4-lead wire ECG monitoring (non-alarming) ●
- Export of event data via Bluetooth and USB cable
The following accessories are provided with the TEMPUS LS – MANUAL:
- External electrode pads (adult/child, pediatric) ●
- . 10.8 V Lithium-ion battery
- Mains power supply unit with mains cable ●
- 3-lead or 4-lead wire ECG cable (optional)
- CPR feedback sensor & disposable adhesive securing pads (optional) 0
## Indications for Use
The TEMPUS LS – MANUAL is a multi-function, portable device that is capable of providing defibrillation therapy, non-invasive external pacing, displaying ECG, and CPR feedback in hospital and pre-hospital settings. It is to be used by qualified medical health care professionals trained in the use and operation of the device and qualified by training in advanced cardiovascular life (ACLS) support.
## Manual Defibrillation
In Manual Defibrillation mode, the TEMPUS LS - MANUAL is indicated for victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions:
- Unconsciousness .
- . Absence of breathing, and
- o Absence of pulse.
This product should be used only by qualified medical personnel for terminating ventricular fibrillation and ventricular tachycardia.
| Traditional 510(k) – TEMPUS LS – MANUAL Defibrillator |
|-------------------------------------------------------|
| 510(k) Summary |
{6}------------------------------------------------
Defibrillation is indicated for adult and pediatric patients. When a victim is less than 8 years of age, or weighs less than 55 lbs. (25kg), the TEMPUS LS-MANUAL should be used with the Tempus Pediatric Electrodes or in the device's pediatric mode with the Tempus Adult Electrodes. Therapy should not be delayed to determine the patient's exact age or weight.
# Cardioversion
The TEMPUS LS - MANUAL may be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardia (VT) by using the R-wave of the patient's ECG as a timing reference. A qualified user trained in the operation of the device when synchronized cardioversion is appropriate. Cardioversion is indicated for adult and pediatric patients.
## External Pacing
The TEMPUS LS - MANUAL is indicated for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include:
- Resuscitation from standstill or bradycardia of any etiology. .
- . As a standby when standstill or bradycardia might be expected.
External Pacing is indicated for adult and pediatric patients.
## ECG Display
The TEMPUS LS-MANUAL is intended to display a patient's ECG from a 3- or 4- lead wire ECG patient cable. It is indicated for use in adult or pediatric patients for attended monitoring of the patient's ECG signal (e.g., before starting demand pacing or after successful defibrillation).
## CPR Feedback
The TEMPUS LS - MANUAL may be supplied with an optional CPR feedback sensor, which is intended to provide visual and audio feedback to encourage rescuers to perform CPR in accordance with AHA/ERC quidelines for chest compression rate and depth. It is indicated for use on adult and pediatric patients ≥ 8 years of age or weighing ≥ 25 kg / 55 lbs.
CPR Feedback is not indicated for use on neonatal and pediatric patients under 8 years of age or weighing less than 25 kg/55lbs.
## Summary of Technological Characteristics
The TEMPUS LS - MANUAL utilizes features and technologies substantially equivalent to the primary (K140502) and secondary (K021453 and K133659) predicate devices.
The intended use, indications for use, and technological features provided by the TEMPUS LS -MANUAL (subject device) are substantially equivalent to those provided by the primary predicate, the ZOLL E-Series ALS (K140502). The fundamental technological characteristics of the TEMPUS LS -MANUAL and ZOLL E-Series ALS are also the same in that both devices use a biphasic waveform for defibrillation therapy.
The electrical waveform delivered in manual defibrillation is substantially equivalent to the biphasic truncated exponential (BTE) waveform utilized by the Philips HeartStart XL (K021453). The electrical waveform delivered in non-invasive pacing therapy is substantially equivalent to the monophasic rectilinear waveform utilized by the Philips HeartStart XL+ (K133659) device.
| Traditional 510(k) – TEMPUS LS – MANUAL Defibrillator |
|-------------------------------------------------------|
| 510(k) Summary |
{7}------------------------------------------------
Based on the information provided in this premarket notification, Remote Diagnostic Technologies Limited concludes that the TEMPUS LS – MANUAL defibrillator is substantially equivalent to the predicate devices. See Tables 1 and 2 below for a more detailed summary of substantial equivalence between the TEMPUS LS – MANUAL subject device and the predicate devices.
| Table 1: Comparison of Technological Characteristics with ZOLL E-Series ALS (K140502) | |
|---------------------------------------------------------------------------------------|--------------------|
| ZOLL E-Series ALS (K140502) | TEMPUS LS - MANUAL |
| | ZOLL E-Series ALS (K140502) | TEMPUS LS - MANUAL |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| TECHNOLOGICAL CHARACTERISTICS - GENERAL | | |
| Indications for Use | Multifunctional, portable<br>electronic device – low<br>energy defibrillator | Identical |
| Size | 13.1" (W)<br>10.5" (D)<br>5.75" (H) | 6.4" (W)<br>3.7" (D)<br>7.8" (H) |
| Weight | 13.2 lbs | 4.5 lbs |
| Ingress Protection | IP34 | IP65 |
| Operating<br>Temperature | 0° C to 55° C | 0°C to 50°C |
| Operating Pressure | 594 to 1060 mBar | 540 to 1060 mBar |
| Shock & Vibration | MIL-STD-810 | Identical |
| Display Type | Color Liquid Crystal Display (LCD) | Identical |
| Screen Size | 5.6 inches diagonally | 5.7 inches diagonally |
| Electrical Safety<br>Standard | IEC 60601-1 | Identical |
| EMC Standard | IEC 60601-1-2<br>RTCA DO-160 | IEC 60601-1-2<br>RTCA DO-160G<br>Cat. M and R |
| Defibrillator Standard | IEC/EN 60601-2-4 | Identical |
| ECG Standard | IEC 60601-2-27 | Identical |
| TECHNOLOGICAL CHARACTERISTICS - DEFIBRILLATION AND CARDIOVERSION | | |
| Manual Defibrillation<br>Modes | Asynchronous manual<br>Synchronous manual<br>(cardioversion) | Identical |
| Output Energy | 1 - 200 Joules into a 50 Ohm load | Identical |
| Defibrillation Waveform | Biphasic Rectilinear | Biphasic Truncated Exponential |
| TECHNOLOGICAL CHARACTERISTICS - PACING | | |
| Modes | Fixed<br>Demand | Identical |
| Fixed-mode principle | Use configurable; pre-set intervals;<br>independent of intrinsic cardiac<br>activity | Identical |
| Demand-mode<br>principle | Use configurable; triggers when<br>heart rate falls below target | Identical |
| Pacing delivery | Multifunction electrode pads | Identical |
| Pacing waveform | Monophasic Rectilinear, constant<br>current, 40 milliseconds | Monophasic Rectilinear, constant<br>current, 20 milliseconds |
| Pulse Amplitude | 5 mA to 140 mA ±5% or 5 mA,<br>whichever is greater | 10 mA to 200 mA ± 10% or 5 mA,<br>whichever is greatest |
| Pacing Rate | 30 ppm to 180 ppm ± 1.5% | 40 ppm to 180 ppm ± 1.5% |
| TECHNOLOGICAL CHARACTERISTICS – ECG MONITORING | | |
| Standard | IEC 60601-2-27:2011 | Substantially Equivalent<br>For attended monitoring only, only<br>technical alerts |
| Patient Connection | ECG cable | Identical |
| Traditional 510(k) – TEMPUS LS – MANUAL Defibrillator |
|-------------------------------------------------------|
| 510(k) Summary |
{8}------------------------------------------------
| | ZOLL E-Series ALS (K140502) | TEMPUS LS – MANUAL |
|------------------------------|--------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Protection | Multifunction electrode pads<br>Fully isolated, defibrillation-protected | Identical |
| Heart Rate Display | Digital readout on display | Identical |
| Heart Rate Range | 0 to 300 bpm with accuracy ± 5% bpm | Substantially Equivalent<br>15 to 350 bpm with accuracy ± 10% bpm or 5 bpm, whichever is greater |
| Heart rate averaging | Averages the interval between the last 5 detected beats | Substantially Equivalent<br>Averages the interval between the last 4-16 detected beats (user configurable) |
| Bandwidth | 0.05 Hz – 150 Hz (-3 dB) | Identical |
| Tall T-wave rejection | < 1.0 mV | Substantially Equivalent<br>< 0.8 mV |
| Pacemaker detection | ± 2 mV to ± 700 mV<br>0.1 ms to 2 ms width | Identical |
| Reaction to irregular rhythm | A1: 40 bpm<br>A2: 60 bpm<br>A3: 120 bpm<br>A4: 90 bpm | Substantially Equivalent per 60601-2-27<br>A1: 80 bpm<br>A2: 60 bpm<br>A3: 120 bpm<br>A4: 90 bpm |
| Alarms | Yes | Yes, for technical alerts |
### Table 2: Comparison of waveform characteristics with Philips HeartStart XL (K021453) and Philips HeartStart XL+ (K133659)
| | Philips HeartStart XL | HeartStart XL+ | TEMPUS LS - MANUAL |
|----------------------------|-----------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|
| Defibrillation<br>Waveform | Biphasic truncated<br>exponential | N/A | Biphasic truncated<br>exponential |
| Pacing<br>Waveform | N/A | Monophasic Rectilinear, constant current, 20 milliseconds duration | Monophasic Rectilinear, constant current, 20 milliseconds duration |
### Safety and Performance Data
The Tempus LS - MANUAL was evaluated via non-clinical safety and performance testing to demonstrate that the device met pre-established acceptance criteria and/or the requirements of FDA recognized consensus standards, and testing demonstrates substantially equivalence to the predicate devices.
| Table 3: Summary of Performance Testing | | | |
|-----------------------------------------|--|--|--|
|-----------------------------------------|--|--|--|
| Performance Testing | Summary |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization and shelf-life | Non-sterile device and accessories |
| | 36-month shelf-life for defibrillator electrode pads, meeting<br>requirements for electrical testing per 60601-2-4 |
| | 24-month shelf-life for CPR feedback sensor securing pads |
| Biocompatibility | IEC 10993 testing met requirements for defibrillator electrode pads<br>and CPR feedback sensor securing pads |
| Electrical safety | IEC 60601-1 testing met requirements of the standard |
| Electromagnetic compatibility | IEC 60601-1-2, RTCA DO 160G and AIM 7351731 and wireless<br>coexistence testing met requirements |
| Software | Appropriate documentation for Major level of concern SW, including<br>verification and validation testing at the unit, integration and system |
| Traditional 510(k) – TEMPUS LS – MANUAL Defibrillator |
|-------------------------------------------------------|
| 510(k) Summary |
{9}------------------------------------------------
| Performance Testing – Bench | Bench testing was performed to evaluate the performance of the<br>TEMPUS LS-MANUAL, inclusive of defibrillation, synchronized<br>cardioversion, CPR metronome, pacing, and ECG display. The<br>performance of the accessories including the CPR feedback sensor,<br>electrode pads, ECG leads, power management was also evaluated.<br>Furthermore, environmental performance, including vibration and<br>shock, altitude, decompression and pressure, and climate, was<br>evaluated. The tests showed passing results and the device is<br>considered compliant with applicable standards (IEC 60601-2-4, IEC<br>60601-2-27; IEC 60601-1-8, AAMI EC53; IEC 60601-1-12; MIL-STD-<br>810G; IEC 62133; UN 38.3; ASTM D4169-16) |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Testing –<br>Human Factors | A human factors validation testing involving 15 participants<br>representing intended users was performed to evaluate the usability<br>of the TEMPUS – LS MANUAL in its intended use environment (pre-<br>hospital; in-hospital). The study demonstrated that users were able<br>to successfully and safely use the device. |
Clinical evidence was not necessary to demonstrate substantial equivalence.
### Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, Remote Diagnostic Technologies Limited conclude that the TEMPUS LS - MANUAL defibrillator described in this pre-market notification is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.