Halyard Pink Underguard Zero Nitrile Sterile Powder-Free Exam Gloves Tested with Chemotherapy Drugs are disposable device intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs and Fentanyl Citrate as per ASTM-D6978-05: Azacitidine (Vidaza) (25 mg/ml) No breakthrough up to 240 minutes Bendamustine (5 mg/ml) No breakthrough up to 240 minutes Bleomycin sulfate (15 mg/ml) No breakthrough up to 240 minutes Bortezomib (6 mg/ml) No breakthrough up to 240 minutes Busulfan (6 mg/ml) No breakthrough up to 240 minutes Capecitabine (26 mg/ml) No breakthrough up to 240 minutes Carboplatin (10 mg/ml) No breakthrough up to 240 minutes Carfilzomib (2 mg/ml) No breakthrough up to 240 minutes Carmustine (3.3 mg/ml) permeation occurred at 59.0 minutes Cetuximab (Erbitux) (2 mg/ml) No breakthrough up to 240 minutes Chloroquine (50 mg/ml) No breakthrough up to 240 minutes Cisplatin (1 mg/ml) No breakthrough up to 240 minutes Cladribine (1.0 mg/ml) No breakthrough up to 240 minutes Cyclophosphamide (20 mg/ml) No breakthrough up to 240 minutes Cyclosporine A (100 mg/ml) No breakthrough up to 240 minutes Cytarabine HCL (100 mg/ml) No breakthrough up to 240 minutes Cytovene (10 mg/ml) No breakthrough up to 240 minutes Dacarbazine (10 mg/ml) No breakthrough up to 240 minutes Dactinomycin (0.5 mg/ml) No breakthrough up to 240 minutes Daunorubicin HCL (5 mg/ml) No breakthrough up to 240 minutes Decitabine (5 mg/ml) No breakthrough up to 240 minutes Docetaxel (10 mg/ml) No breakthrough up to 240 minutes Doxorubicin HCL (2 mg/ml) No breakthrough up to 240 minutes Epirubicin (Ellence) (2 mg/ml) No breakthrough up to 240 minutes Etoposide (20 mg/ml) No breakthrough up to 240 minutes Fludarabine (25 mg/ml) No breakthrough up to 240 minutes 5-Fluorouracil (50 mg/ml) Nobreakthrough up to 240 minutes Fulvestrant (50 mg/ml) No breakthrough up to 240 minutes Gemcitabine (38 mg/ml) No breakthrough up to 240 minutes Idarubicin (1 mg/ml) No breakthrough up to 240 minutes Ifosfamide (50 mg/ml) No breakthrough up to 240 minutes Irinotecan (20 mg/ml) No breakthrough up to 240 minutes Leuprolide acetate salt (5mg/ml) No breakthrough up to 240 minutes Mechlorethamine HCL (1 mg/ml) No breakthrough up to 240 minutes Melphalan (5 mg/ml) No breakthrough up to 240 minutes Methotrexate (25 mg/ml) No breakthrough up to 240 minutes Mitomycin-C (0.5 mg/ml) No breakthrough up to 240 minutes Mitoxantrone (2 mg/ml) No breakthrough up to 240 minutes Oxaliplatin (2 mg/ml) No breakthrough up to 240 minutes Paclitaxel (6 mg/ml) No breakthrough up to 240 minutes Pemetrexed (25 mg/ml) No breakthrough up to 240 minutes Raltitrexed (0.5 mg/ml) No breakthrough up to 240 minutes Retrovir (10 mg/ml) No breakthrough up to 240 mimutes Rituximab (10 mg/ml) No breakthrough up to 240 minutes Temsirolimus (25 mg/ml) No breakthrough up to 240 minutes ThioTEPA (10 mg/ml) No breakthrough up to 240 minutes Topotecan HCL (1 mg/ml) No breakthrough up to 240 minutes Triclosan (1 mg/ml) No breakthrough up to 240 minutes Trisenox (0.1 mg/ml) No breakthrough up to 240 minutes Vinblastine (1 mg/ml) No breakthrough up to 240 minutes Vincrinstine Sulfate (1 mg/ml) No breakthrough up to 240 minutes Vinorelbine (10 mg/ml) No breakthrough up to 240 minutes Zoledronic Acid (0.8 mg/ml) No breakthrough up to 240 minutes Fentany I Citrate Injection, (100mcg /2mL). No breakthrough up to 240 minutes
Device Story
Disposable, pink-colored, chlorinated, nitrile, powder-free, textured fingertip, ambidextrous, sterile patient examination gloves. Used in clinical settings by healthcare providers to prevent cross-contamination between patient and examiner. Provides barrier protection against specific chemotherapy drugs and Fentanyl Citrate. Performance verified via ASTM D6978-05 permeation testing; most tested agents show no breakthrough up to 240 minutes, with the exception of Carmustine (59 minutes).
Clinical Evidence
Bench testing only. No clinical data. Testing included ASTM D6978-05 (chemotherapy drug permeation), ASTM D5151-06 (leakage/holes), ASTM D6124-06 (residual powder), ASTM D6319-10 (physical dimensions/properties), and ISO 10993 (biocompatibility: systemic toxicity, irritation, sensitization).
Indicated for use as a disposable medical examination glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for resistance to permeation by specific chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Halyard Pink Underguard Zero Nitrile Powder-Free Exam Glove Tested for Use with Chemotherapy Drugs (K183132)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 22, 2020
೦&M Halyard, Inc. Steven Dowdley Associate Director of Regulatory Affairs 9120 Lockwood Blvd Mechanicsville, Virginia 23116
Re: K200843
Trade/Device Name: Halyard Pink Underguard Zero Nitrile Sterile Powder-Free Exam Glove Tested with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO Dated: August 17, 2020 Received: August 21, 2020
Dear Steven Dowdley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K200843
#### Device Name
Halyard Pink Underguard Zero Nitrile Sterile Powder-Free Exam Gloves Tested with Chemotherapy Drues
#### Indications for Use (Describe)
Halyard Pink Underguard Zero Nitrile Sterile Powder-Free Exam Gloves Tested with Chemotherapy Drugs are disposable device intended for medical purposes that are worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with the following chemotherapy drugs and Fentanyl Citrate as per ASTM-D6978-05: Azacitidine (Vidaza) (25 mg/ml) No breakthrough up to 240 minutes Bendamustine (5 mg/ml) No breakthrough up to 240 minutes Bleomycin sulfate (15 mg/ml) No breakthrough up to 240 minutes Bortezomib (6 mg/ml) No breakthrough up to 240 minutes Busulfan (6 mg/ml) No breakthrough up to 240 minutes Capecitabine (26 mg/ml) No breakthrough up to 240 minutes Carboplatin (10 mg/ml) No breakthrough up to 240 minutes Carfilzomib (2 mg/ml) No breakthrough up to 240 minutes Carmustine (3.3 mg/ml) permeation occurred at 59.0 minutes Cetuximab (Erbitux) (2 mg/ml) No breakthrough up to 240 minutes Chloroquine (50 mg/ml) No breakthrough up to 240 minutes Cisplatin (1 mg/ml) No breakthrough up to 240 minutes Cladribine (1.0 mg/ml) No breakthrough up to 240 minutes Cyclophosphamide (20 mg/ml) No breakthrough up to 240 minutes Cyclosporine A (100 mg/ml) No breakthrough up to 240 minutes Cytarabine HCL (100 mg/ml) No breakthrough up to 240 minutes Cytovene (10 mg/ml) No breakthrough up to 240 minutes Dacarbazine (10 mg/ml) No breakthrough up to 240 minutes Dactinomycin (0.5 mg/ml) No breakthrough up to 240 minutes Daunorubicin HCL (5 mg/ml) No breakthrough up to 240 minutes Decitabine (5 mg/ml) No breakthrough up to 240 minutes Docetaxel (10 mg/ml) No breakthrough up to 240 minutes Doxorubicin HCL (2 mg/ml) No breakthrough up to 240 minutes Epirubicin (Ellence) (2 mg/ml) No breakthrough up to 240 minutes Etoposide (20 mg/ml) No breakthrough up to 240 minutes Fludarabine (25 mg/ml) No breakthrough up to 240 minutes 5-Fluorouracil (50 mg/ml) Nobreakthrough up to 240 minutes Fulvestrant (50 mg/ml) No breakthrough up to 240 minutes Gemcitabine (38 mg/ml) No breakthrough up to 240 minutes Idarubicin (1 mg/ml) No breakthrough up to 240 minutes Ifosfamide (50 mg/ml) No breakthrough up to 240 minutes Irinotecan (20 mg/ml) No breakthrough up to 240 minutes Leuprolide acetate salt (5mg/ml) No breakthrough up to 240 minutes Mechlorethamine HCL (1 mg/ml) No breakthrough up to 240 minutes
Melphalan (5 mg/ml) No breakthrough up to 240 minutes Methotrexate (25 mg/ml) No breakthrough up to 240 minutes
Mitomycin-C (0.5 mg/ml) No breakthrough up to 240 minutes
List of chemotherapy drugs continues on next page
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
{3}------------------------------------------------
K200843: Continuation of the list of chemotherapy drugs from previous page. These gloves were tested for use with the following chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05:
Mitoxantrone (2 mg/ml) No breakthrough up to 240 minutes Oxaliplatin (2 mg/ml) No breakthrough up to 240 minutes Paclitaxel (6 mg/ml) No breakthrough up to 240 minutes Pemetrexed (25 mg/ml) No breakthrough up to 240 minutes Raltitrexed (0.5 mg/ml) No breakthrough up to 240 minutes Retrovir (10 mg/ml) No breakthrough up to 240 mimutes Rituximab (10 mg/ml) No breakthrough up to 240 minutes Temsirolimus (25 mg/ml) No breakthrough up to 240 minutes ThioTEPA (10 mg/ml) No breakthrough up to 240 minutes Topotecan HCL (1 mg/ml) No breakthrough up to 240 minutes Triclosan (1 mg/ml) No breakthrough up to 240 minutes Trisenox (0.1 mg/ml) No breakthrough up to 240 minutes Vinblastine (1 mg/ml) No breakthrough up to 240 minutes Vincrinstine Sulfate (1 mg/ml) No breakthrough up to 240 minutes Vinorelbine (10 mg/ml) No breakthrough up to 240 minutes Zoledronic Acid (0.8 mg/ml) No breakthrough up to 240 minutes
Fentany I Citrate Injection, (100mcg /2mL). No breakthrough up to 240 minutes
CAUTION: The following chemotherapy drugs and concentration showed breakthrough detected in less than 60 minutes. Carmustine (BCNU): 59 minutes.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
| Date Summary<br>was Prepared | September 19, 2020 |
|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter | O & M Halyard, Inc.<br>5405 Windward Parkway<br>Alpharetta, GA 30004 |
| Primary Contact for<br>this 510(k) Submission | Steven Dowdley, RAC<br>Tel: 678-451-8062<br>Email: steven.dowdley@hyh.com |
| Device Trade Name | Halyard Pink Underguard Zero Nitrile Sterile Powder-Free Exam Gloves Tested with<br>Chemotherapy Drugs |
| Device Common Name | Medical Exam Gloves |
| Device Product Code<br>and Classification Name | LZC<br>Class I, 21 CFR §880.6250 Patient Examination Glove<br>LZC Class I, 21 CFR §880.6250 Patient Examination Glove, Specialty<br>OQD Class I, 21 CFR §880.6250 Fentanyl and other opioid protection glove |
| Predicate Device | Halyard Pink Underguard Zero Nitrile Powder-Free Exam Glove Tested for<br>Use with Chemotherapy Drugs cleared under K183132 |
| Subject Device Description | The Halyard Pink Underguard Zero Nitrile Sterile Powder-Free Exam<br>Gloves Tested with Chemotherapy Drugs are disposable, pink-colored,<br>chlorinated, nitrile, powder-free, textured fingertip, ambidextrous, sterile<br>patient examination gloves.<br>The device follows consensus standards:<br>• ASTM D5151-06 Standard Test Method for Detection of Holes in Medical<br>Gloves<br>• ASTM D6319-10 Standard Specification for Nitrile Examination Gloves<br>for Medical Applications<br>• ASTM D6124-06 Standard Test Method for Residual Powder on Medical<br>Gloves<br>• ASTM D6978-05 Standard Practice for Assessment of Resistance of<br>Medical Gloves to Permeation by Chemotherapy Drugs |
| | • ISO 10993-11:2017, Biological evaluation of medical devices - Part11:<br>Tests for Systemic Toxicity<br>• ISO 10993-10: 2010: Biological evaluation of medical devices - Part 10:<br>Tests for Irritation and Skin Sensitization |
| Indications for Use for the<br>Halyard Pink Underguard Zero,<br>Sterile Powder-Free Exam Glove<br>Tested with Chemotherapy | The Halyard Pink Underguard Zero Nitrile Sterile Powder-Free Exam Gloves<br>Tested with Chemotherapy Drugs are disposable devices intended for medical<br>purposes that is worn on the examiner's hand to prevent contamination between<br>patient and examiner. These gloves were tested for use with the following |
| Drugs | chemotherapy drugs and Fentanyl Citrate as per ASTM -D6978-05.: |
| | Azacitidine (Vidaza) (25 mg/ml) No breakthrough up to 240 minutes |
| | Bendamustine (5 mg/ml) No breakthrough up to 240 minutes |
| | Bleomycin sulfate (15 mg/ml) No breakthrough up to 240 minutes |
| | Bortezomib (6 mg/ml) No breakthrough up to 240 minutes |
| | Busulfan (6 mg/ml) No breakthrough up to 240 minutes |
| | Capecitabine (26 mg/ml) No breakthrough up to 240 minutes |
| | Carboplatin (10 mg/ml) No breakthrough up to 240 minutes |
| | Carfilzomib (2 mg/ml) No breakthrough up to 240 minutes |
| | Carmustine (3.3 mg/ml) permeation occurred at 59.0 minutes |
| | Cetuximab (Erbitux) (2 mg/ml) No breakthrough up to 240 minutes |
| | Chloroquine (50 mg/ml) No breakthrough up to 240 minutes |
| | Cisplatin (1 mg/ml) No breakthrough up to 240 minutes |
| | Cladribine (1.0 mg/ml) No breakthrough up to 240 minutes |
| | Cyclophosphamide (20 mg/ml) No breakthrough up to 240 minutes |
| | Cyclosporine A (100 mg/ml) No breakthrough up to 240 minutes |
| | Cytarabine HCL (100 mg/ml) No breakthrough up to 240 minutes |
| | Cytovene (10 mg/ml) No breakthrough up to 240 minutes |
| | Dacarbazine (10 mg/ml) No breakthrough up to 240 minutes |
| | Dactinomycin (0.5 mg/ml) No breakthrough up to 240 minutes |
| | Daunorubicin HCL (5 mg/ml) No breakthrough up to 240 minutes |
| | Decitabine (5 mg/ml) No breakthrough up to 240 minutes |
| | Docetaxel (10 mg/ml) No breakthrough up to 240 minutes |
| | Doxorubicin HCL (2 mg/ml) No breakthrough up to 240 minutes |
| | Epirubicin (Ellence) (2 mg/ml) No breakthrough up to 240 minutes |
| | Etoposide (20 mg/ml) No breakthrough up to 240 minutes |
| | Fludarabine (25 mg/ml) No breakthrough up to 240 minutes |
| | 5-Fluorouracil (50 mg/ml) No breakthrough up to 240 minutes |
| | Fulvestrant (50 mg/ml) No breakthrough up to 240 minutes |
| | Gemcitabine (38 mg/ml) No breakthrough up to 240 minutes |
| | Idarubicin (1 mg/ml) No breakthrough up to 240 minutes |
| | Ifosfamide (50 mg/ml) No breakthrough up to 240 minutes |
| | Irinotecan (20 mg/ml) No breakthrough up to 240 minutes |
| | Leuprolide acetate salt (5mg/ml) No breakthrough up to 240 minutes |
| | Mechlorethamine HCL (1 mg/ml) No breakthrough up to 240 minutes |
| | Melphalan (5 mg/ml) No breakthrough up to 240 minutes |
| | Methotrexate (25 mg/ml) No breakthrough up to 240 minutes |
| | Mitomycin-C (0.5 mg/ml) No breakthrough up to 240 minutes |
| | Mitoxantrone (2 mg/ml) No breakthrough up to 240 minutes |
| | Oxaliplatin (2 mg/ml) No breakthrough up to 240 minutes |
| | Paclitaxel (6 mg/ml) No breakthrough up to 240 minutes |
| | Pemetrexed (25 mg/ml) No breakthrough up to 240 minutes |
| | Raltitrexed (0.5 mg/ml) No breakthrough up to 240 minutes |
| | Retrovir (10 mg/ml) No breakthrough up to 240 minutes |
| | Rituximab (10 mg/ml) No breakthrough up to 240 minutes |
| | Temsirolimus (25 mg/ml) No breakthrough up to 240 minutes |
| | ThioTEPA (10 mg/ml) No breakthrough up to 240 minutes |
| | Topotecan HCL (1 mg/ml) No breakthrough up to 240 minutes |
| | Triclosan (1 mg/ml) No breakthrough up to 240 minutes |
| | Trisenox (0.1 mg/ml) No breakthrough up to 240 minutes |
| | Vinblastine (1 mg/ml) No breakthrough up to 240 minutes |
| | Vincrinstine Sulfate (1 mg/ml) No breakthrough up to 240 minutes |
| | Vinorelbine (10 mg/ml) No breakthrough up to 240 minutes |
| | Zoledronic Acid (0.8 mg/ml) No breakthrough up to 240 minutes |
| | Fentanyl Citrate (100mcg/2ml) No breakthrough up to 240 minutes. |
| | CAUTION: The following chemotherapy drugs and concentration showed<br>breakthrough detected in less than 60 minutes: Carmustine (BCNU): 59<br>minutes. |
{5}------------------------------------------------
{6}------------------------------------------------
| Summary of comparison of technological characteristics | The subject device and the predicate device are both sterile, powder-free nitrile exam gloves tested for resistance to permeation by chemotherapy drugs. |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological Characteristics Comparison Table | | | |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | Subject Device | Predicate Device K183132 | Comparison |
| FDA Product Code | LZA, LZC, QDO | LZA, LZC | Same |
| FDA Classification | Class I | Class I | Same |
| Regulation Number | 880.6250 | 880.6250 | Same |
| Common Name | Medical Exam Glove | Medical Exam Glove | Same |
| Device Trade Name | Halyard Pink Underguard Zero<br>Nitrile Sterile Powder-Free<br>Exam Gloves Tested with<br>Chemotherapy Drugs | Halyard Pink Underguard Nitrile<br>Powder-Free Exam Glove Tested for<br>Use with Chemotherapy Drugs | Similar |
| Intended Use | Halyard Pink Underguard Nitrile<br>Sterile Powder-Free Exam Gloves<br>Tested with Chemotherapy Drugs<br>are disposable device intended for<br>medical purposes that is worn on<br>the examiner's hand to prevent<br>contamination between patient<br>and examiner. These gloves were<br>tested for use with chemotherapy<br>drugs listed on the label. | The Halyard Pink Underguard<br>Nitrile Powder-Free Exam Glove<br>Tested for Use with Chemotherapy<br>Drugs is a disposable device<br>intended for medical purposes that<br>is worn on the examiner's hand to<br>prevent contamination between<br>patient and examiner.<br>These gloves were tested for use<br>with chemotherapy drugs listed on<br>the label. | Same |
| Technological<br>Characteristics | The glove is a colored, nitrile,<br>powder-free, textured fingertip,<br>ambidextrous, patient examination<br>glove. | The glove is a colored, nitrile,<br>powder-free, textured fingertip,<br>ambidextrous, patient examination<br>glove. | Same |
| Sizes of gloves | XS, S, M, L, XL(Sterile) | XS, S, M, L, XL | Different |
| Texture | Textured fingertips (Sterile) | Textured fingertips | Different |
{7}------------------------------------------------
| Sterility | Sterile. | Non Sterile | Different |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Biocompatibility | Based ISO 10993 Biological<br>evaluation of Medical devices -<br>Test for Systemic Injection, the<br>test article was considered non-<br>toxic. Meets the acceptance<br>criteria. | Based ISO 10993 Biological<br>evaluation of Medical devices -<br>Test for Systemic Injection, the test<br>article was considered non-toxic.<br>Meets the acceptance criteria. | Similar |
| | Based on ISO 10993- Biological<br>evaluation of Medical Devices -<br>Test for Skin Irritation, the device<br>extracts were not found to cause a<br>systemic response in the animal<br>model. Meets the acceptance<br>criteria. | Based on ISO 10993- Biological<br>evaluation of Medical Devices -<br>Test for Skin Irritation, the device<br>extracts were not found to cause a<br>systemic response in the animal<br>model. Meets the acceptance<br>criteria. | |
| | Based on ISO 10993- Biological<br>evaluation of Medical Device<br>Dermal Sensitization-<br>ISO Kligman Maximization Test -<br>NaCl and CSO extracts. The test<br>article were classified as a non-<br>sensitizer. Based on the criteria of<br>the protocol, the test article meets<br>the requirements of the ISO<br>10993-10 guidelines. | Based on ISO 10993- Biological<br>evaluation of Medical Device<br>Dermal Sensitization-<br>ISO Kligman Maximization Test -<br>NaCl and CSO extracts. The test<br>article were classified as a non-<br>sensitizer. Based on the criteria of<br>the protocol, the test article meets<br>the requirements of the ISO 10993-<br>10 guidelines. | |
| Performance Data for Chemotherapy Drugs | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.