RX-1 Rhythm Express Remote Cardiac Monitoring System
Applicant
VivaQuant, Inc.
Product Code
MLO · Cardiovascular
Decision Date
Sep 2, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Attributes
AI/ML, Pediatric
Indications for Use
The Rhythm Express remote cardiac monitoring system is intended for use by patients greater than 10 kg who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers the recorded cardiac activity to the server where it is presented for review by a medical professional. The data received from the Rhythm Express device can be used by another device for arrhythmia analysis, reporting and signal measurements. The Rhythm Express device is not intended to sound any alarms. The device does not deliver any therapy, administer any drugs, provide interpretive or diagnostic statements or provide for any life support. The Rhythm Express system communicates events from the monitoring center within one to seven minutes (assuming cell service is available) and hence is not suitable for use as a real-time arrhythmia event monitor.
Device Story
RX-1 Rhythm Express is a wearable ambulatory cardiac monitor for patients >10 kg; used in Mobile Cardiac Telemetry (MCT), Event Recorder (ER), or Wireless Holter (WH) modes. Device captures 2-channel ECG via standard electrodes; embedded algorithm processes signals to detect arrhythmias, compress data, and remove in-band noise. Device uses cellular modem to transmit data to RS-1 Web Service for review by medical professionals. Operated by patient; worn for 1 day to 4 weeks. Output is raw/processed ECG data for clinical review; no alarms, therapy, or diagnostic statements provided. Benefits include remote monitoring of cardiac activity for patients with suspected or known arrhythmias. Not for real-time event monitoring.
Ambulatory cardiac monitor; 2-channel ECG; rechargeable Li-Ion battery (7-day life); cellular (LTE) connectivity; USB charging port. Functional blocks: ECG front end, DSP, MCU, flash storage, RF modem, LCD, keypad. Software: embedded algorithm for arrhythmia detection and noise removal. Sterilization: not specified. Standards: IEC 60601 series, ISO 10993-1, ANSI/AAMI TIR57.
Indications for Use
Indicated for patients >10 kg with or at risk of cardiac disease, or those with intermittent symptoms requiring continuous cardiac monitoring.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Predicate Devices
RX-1 Rhythm Express Remote Cardiac Monitoring System (K183704)
PocketECG III - Medicalgorithmics Unified Arrhythmia Diagnostic System (K152550)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
September 2, 2020
VivaQuant Inc. % Brodie Pedersen Consultant Borderless Compliance, LLC 7118 Teakwood Cir Maple Grove, Minnesota 55369
Re: K200833
Trade/Device Name: RX-1 Rhythm Express Remote Cardiac Monitoring System Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MLO, DSI Dated: July 31, 2020 Received: August 3, 2020
Dear Brodie Pedersen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200883
Device Name
RX-1 Rhythm Express Remote Cardiac Monitoring System
### Indications for Use (Describe)
The Rhythm Express remote cardiac monitoring system is intended for use by patients greater than 10 kg who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers the recorded cardiac activity to the server where it is presented for review by a medical professional.
The data received from the Rhythm Express device can be used by another device for arrhythmia analysis, reporting and signal measurements. The Rhythm Express device is not intended to sound any alarms.
The device does not deliver any therapy, administer any drugs, provide interpretive or diagnostic statements or provide for any life support. The Rhythm Express system communicates events from the monitoring center within one to seven minutes (assuming cell service is available) and hence is not suitable for use as a real-time arrhythmia event monitor.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <path d="M2 2 L14 14 M14 2 L2 14" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
(as required by 21CFR 807.92)
#### SUBMITTER 1.
VivaQuant Inc. 4339 Nancy Place, Suite 100, St. Paul, MN 55126 Contact Person: Brian Brockway bbrockway@vivaquant.com Phone: 651-217-2176 Date Prepared: July 31, 2020
#### II. DEVICE
RX-1 Rhythm Express Remote Cardiac Monitoring System Name of Device: Classification Name: 870.2800. 870.1025 Electrocardiograph, Ambulatory, With Analysis Algorithm Detector and Alarm, Arrhythmia, Common or Usual Name: Ambulatory Cardiac Monitor Device Panel: Cardiovascular Regulatory Class: Class 2 Product Code: MLO, DSI
#### PREDICATE DEVICE III.
The RX-1 system is substantially equivalent in intended use and similar technological characteristics to the following devices: RX-1 Rhythm Express Remote Cardiac Monitoring System cleared under K183704, and PocketECG III - Medicalgorithmics Unified Arrhythmia Diagnostic System, cleared under K152550.
{4}------------------------------------------------
#### DEVICE DESCRIPTION IV.
The Rhythm Express RX-1 will be worn by patients for a period of time as prescribed by a physician, typically 1 day to 4 weeks, and will continuously monitor ECG. RX-1 will function in one of three modes: a) Mobile Cardiac Telemetry (MCT), b) Event Recorder (ER), and Wireless Holter (WH). The device will connect to standard ECG electrodes to capture 2 channel ECGs. An embedded algorithm processes the acquired ECG to detect arrhythmias, compress the ECG, and remove most in-band noise without distorting ECG morphology. RX-1 incorporates a cellular modem to communicate with the RS-1 Web Service.
The RX-1 device is not a life-supporting or life-sustaining system. Clinical judgment and experience are used to check and interpret the data.
#### V. INTENDED USE
The Rhythm Express remote cardiac monitoring system is intended for use by patients greater than 10 kg who either have or are at risk of having cardiac disease and those that demonstrate intermittent symptoms indicative of cardiac disease and require cardiac monitoring on a continuing basis. The device continuously records ECG data and upon detection by an ECG analysis algorithm or manually initiated by the patient, automatically delivers the recorded cardiac activity to the server where it is presented for review by a medical professional.
The data received from the Rhythm Express device can be used by another device for arrhythmia analysis. reporting and signal measurements. The Rhythm Express device is not intended to sound any alarms.
The device does not deliver any therapy, administer any drugs, provide interpretive or diagnostic statements or provide for any life support. The Rhythm Express system communicates events from the patient to the monitoring center within one to seven minutes (assuming cell service is available) and hence is not suitable for use as a real-time arrhythmia event monitor.
{5}------------------------------------------------
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The RX-1 system is substantially equivalent in intended use and similar technological characteristics of RX-1 system cleared under K183704.
| Category | Identical/<br>Different | RX-1 | RX-1 predicate | PocketECG III |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | | Pending | K183704 | K152550 |
| Classification<br>Name | Identical | Electrocardiograph,<br>Ambulatory, With<br>Analysis Algorithm | Medical<br>Mobile Cardiac<br>Monitor | Arrhythmia Detector<br>and Alarm |
| Product Code | Identical | MLO, DSI | DXH | DSI |
| Intended Use | Similar | The Rhythm Express<br>remote cardiac<br>monitoring system is<br>intended for use by<br>patients greater than 10<br>kg who either have or<br>are at risk of having<br>cardiac disease and<br>those that demonstrate<br>intermittent symptoms<br>indicative of cardiac<br>disease and require<br>cardiac monitoring on a<br>continuing basis. The<br>device continuously<br>records ECG data and<br>upon detection by an<br>ECG analysis algorithm or<br>manually initiated by the<br>patient, automatically<br>delivers the recorded<br>cardiac activity to the<br>server where it is<br>presented for review by | The Rhythm Express<br>remote cardiac<br>monitoring system is<br>intended for use by<br>adult patients who<br>either have or are at<br>risk of having cardiac<br>disease and those that<br>demonstrate<br>intermittent symptoms<br>indicative of cardiac<br>disease and require<br>cardiac monitoring on<br>a continuing basis. The<br>device continuously<br>records ECG data and<br>upon detection by an<br>ECG analysis algorithm<br>or manually initiated<br>by the patient,<br>automatically delivers<br>the recorded cardiac<br>activity to the server<br>where it is presented | 1. Patients who have a<br>demonstrated need for<br>cardiac monitoring. These<br>may include but are not<br>limited to patients who<br>require monitoring for: a)<br>non-life threatening<br>arrhythmias such as<br>supraventricular<br>tachycardias (e.g. atrial<br>fibrillation, atrial flutter,<br>PACS, PSVT) and<br>ventricular ectopy; b)<br>evaluation of brady<br>arrhythmias and<br>intermittent bundle<br>branch block, including<br>after cardiovascular<br>surgery and myocardial<br>infarction; and c)<br>arrhythmias associated<br>with co-morbid conditions<br>such as hyperthyroidism<br>or chronic lung disease; |
| Category | Identical/<br>Different | RX-1 | RX-1 predicate | PocketECG III |
| | | a medical professional. | and can be reviewed<br>by a medical<br>professional. | 2. Patients with symptoms<br>that may be due to cardiac<br>arrhythmias. These may<br>include but are not limited<br>to symptoms such as: a)<br>dizziness or<br>lightheadedness; b)<br>syncope of unknown<br>etiology in which<br>arrhythmias are suspected<br>or need to be excluded;<br>and c) dyspnea (shortness<br>of breath). |
| | | The data received from<br>the Rhythm Express<br>device can be used by<br>another device for<br>arrhythmia analysis,<br>reporting and signal<br>measurements. The<br>Rhythm Express device is<br>not intended to sound<br>any alarms. | The data received from<br>the Rhythm Express<br>device can be used by<br>another device for<br>arrhythmia analysis,<br>reporting and signal<br>measurements. The<br>Rhythm Express device<br>is not intended to<br>sound any alarms. | 3. Patients with<br>palpitations with or<br>without known<br>arrhythmias to obtain<br>correlation of rhythm with<br>symptoms. |
| | | The device does not<br>deliver any therapy,<br>administer any drugs,<br>provide interpretive or<br>diagnostic statements or<br>provide for any life<br>support. The Rhythm<br>Express system<br>communicates events<br>from the patient to the<br>monitoring center within<br>one to seven minutes<br>(assuming cell service is<br>available) and hence is<br>not suitable for use as a<br>real-time arrhythmia<br>event monitor. | The device does not<br>deliver any therapy,<br>administer any drugs,<br>provide interpretive or<br>diagnostic statements<br>or provide for any life<br>support. The Rhythm<br>Express system<br>communicates events<br>from the patient to the<br>monitoring center<br>within one to seven<br>minutes (assuming cell<br>service is available) and<br>hence is not suitable<br>for use as a real-time<br>arrhythmia event<br>monitor. | 4. Patients who require<br>monitoring of effect of<br>drugs to control<br>ventricular rate in various<br>atrial arrhythmias (e.g.<br>atrial fibrillation).<br>5. Patients recovering<br>from cardiac surgery who<br>are indicated for<br>outpatient arrhythmia<br>monitoring.<br>6. Patients with diagnosed<br>sleep disordered<br>breathing including sleep<br>apnea (obstructive,<br>central) to evaluate<br>possible nocturnal<br>arrhythmias.<br>7. Patients requiring<br>arrhythmia evaluation of<br>etiology of stroke or<br>transient cerebral<br>ischemia, possibly<br>secondary to atrial<br>fibrillation or atrial flutter. |
| Category | Identical/<br>Different | RX-1 | RX-1 predicate | PocketECG III |
| Delivered device<br>includes, patient<br>ECG cable<br>- rechargeable<br>battery<br>-Wall Battery<br>charger | Similar | Yes | Yes | Yes |
| Monitor<br>functional<br>blocks: ECG<br>front end,<br>DSP<br>MCU,<br>flash data<br>storage,<br>RF modem for<br>data<br>transmission,<br>LCD screen, and<br>keypad, | Similar | Yes | Yes | Yes |
| The server<br>facilitates data<br>communication<br>with the device,<br>provide data<br>storage, and<br>present the data<br>for evaluation<br>by a medical<br>professional: | Similar | Yes | Yes | Yes |
| | | | | 8. Data from the device<br>may be used by another<br>device to analyze,<br>measure or report QT<br>interval. The device is not<br>intended to sound any<br>alarms for QT interval<br>changes. |
| Category | Identical/<br>Different | RX-1 | RX-1 predicate | PocketECG III |
| Device form<br>factor: small,<br>lightweight<br>ambulatory<br>cardiac<br>monitors. | Similar | Yes | Yes | Yes |
| Wireless<br>technology used<br>to transmit data<br>to server | Similar | LTE | LTE | GSM |
| Device is battery<br>powered by a<br>rechargeable Li-<br>Ion battery | Similar | For 7 Days | For 7 Days | For 24 hours |
| using a server,<br>can adjust<br>device<br>programming<br>parameters<br>such as pre-post<br>recording times<br>and auto-<br>triggering<br>configuration. | Similar | Yes | Yes | Yes |
| devices have<br>keypad for<br>manual event<br>recordings and a<br>user interface to<br>indicate device<br>status and mode<br>of operation. | Similar | Yes | Yes | Yes |
| Device<br>incorporate<br>embedded ECG | Similar | Yes | Yes | Yes |
| Category | Identical/<br>Different | RX-1 | RX-1 predicate | PocketECG III |
| analysis<br>algorithm to<br>auto-capture<br>arrhythmia<br>events between<br>the signal<br>acquisition point<br>and the server. | | | | |
| device has at<br>least 2 ECG<br>channels and 3-<br>lead electrodes | Identical | Yes | Yes | Yes |
| Functional,<br>Environmental<br>and Electrical<br>characteristics | Similar | Ambulatory use, charge<br>from ac adapter<br>0-45C | Ambulatory use, charge<br>from ac adapter<br>0-45C | Ambulatory use, charge<br>from ac adapter<br>0-43C |
| Power Port | Different | Yes, USB port to charge<br>the battery, cannot be<br>connected during ECG<br>recording, not used for<br>data download. | Yes, USB port to charge<br>the battery, cannot be<br>connected during ECG<br>recording, not used for<br>data download. | No, Separate Battery<br>charger to recharge<br>removable battery pack. |
| Storage<br>conditions | Similar | -25 to 45 C<br>10 to 95% RH | -25 to 45 C<br>10 to 95% RH | -20 to 60 C<br>15 to 93% RH |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
#### PERFORMANCE TESTING VII.
The following performance and safety tests have been passed successfully:
- IEC 60601-2-47:2012 Medical electrical equipment Part 2-47: . Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.
- IEC 60601-1-1:2012 Ed. 3.1 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance
- IEC 62304:2015 Ed. 1.1 Medical Device Software Software Life Cvcle ● Processes.
- IEC 60601-1-2:2014 4th Edition, Medical Electrical Equipment Part 1-● 2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60601-1-11:2015 Edition 1.1, Medical electrical equipment Part 1-● 11: General requirements for basic safety and essential performance -Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- ANSI/AAMI TIR57:2015, Principles for medical device security-Risk ● management
- IEC 62366-1:2015 Edition 1.0. Medical devices - Part 1: Application of usability engineering to medical devices
- ANSI IEEE C63.27-2017 Evaluation of Wireless Coexistence ●
- Biocompatibility testing of patient contacting materials according to ISO ● 10993-1.
- Bench test results verify that RX-1 Monitor system can continuously record ECG signal, store ECG data in the device memory, and transmit manual or auto activated event recordings to the server via mobile network connection for evaluation by a medical professional. Test results verify that all requirements were met and that the RX-1 Monitor performs as designed.
#### VIII. SUBSTANTIAL EQUIVALENCE RATIONALE
The intended use, performance and technological characterics of the RX-1 Monitor system compared to the named predicate device demonstrates that the RX-1 Monitor is substantially equivalent to the predicate.
#### CONCLUSIONS IX.
The analysis of the differences between RX-1 Monitor and the predicate device does not raise new questions of safety and effectiveness. Based on device performance test results, VivaQuant determines that the RX-1 Monitor system performs within its design specifications and is
{11}------------------------------------------------
substantially equivalent to the predicate devices.
The information in this 510(k), submission demonstrates that the RX-1 Monitor system is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.