K200828 · Athelas, Inc. · GKZ · Mar 22, 2022 · Hematology
Device Facts
Record ID
K200828
Device Name
Athelas Home
Applicant
Athelas, Inc.
Product Code
GKZ · Hematology
Decision Date
Mar 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K200828 · Mar 22, 2022
Athelas Home
Athelas, Inc.
Patient-generated health data from in-home testing
The sponsor conducted in-home studies to evaluate the precision and performance of the Athelas Home device when used by patients in their home environment, comparing results to the predicate device and assessing usability.
Studies from the Athelas One (K181288) were used to support the Athelas Home.
—
At-Home Method Comparison: 165 samples taken in at-home settings in the US. Method Comparison: 312 patient samples taken at 3 point of care sites in the US.
—
Neutrophil percentage
Computer vision based image analysis
Neutrophil % Precision: 5% SD OR 15% CV. Neutrophil % Bias: ±10% bias or ±5% Neut% total error (whichever larger).
Studies from the Athelas One (K181288) were used to support the Athelas Home.
—
At-Home Method Comparison: 165 samples taken in at-home settings in the US. Method Comparison: 312 patient samples taken at 3 point of care sites in the US.
—
Indications for Use
The Athelas Home is indicated for the quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary whole blood. The Athelas Home system is for In Vitro Diagnostic use and for prescription use only. The Athelas Home is only to be used with Athelas Home Test Strips. The Athelas Home is indicated for use by patients and caregivers within home settings with results viewable by health care professionals. The Athelas Home is intended for patients at risk of neutropenia. For patients with psychiatric conditions, clinical judgement should be exercised when deciding the end-user, based on the instructions for use (IFU) of Athelas Home, the treating physician should determine which patients are competent to perform the test by themselves. The Athelas Home is indicated for adult populations only (aged 21 and older). Results obtained with the Athelas Home should not be the sole basis for patient diagnosis, treatment or management of leukopenia and neutropenia. All results should be evaluated by a healthcare provider. The Athelas Home is intended to be used by a single person and should not be shared. Prescription Use Only.
Device Story
Athelas Home is an automated differential cell counter for home use. Input: 3.5 µL capillary whole blood via microfluidic test strip. Operation: Test strip creates stained monolayer; device captures multiple images; computer vision algorithm counts/classifies WBCs. Output: WBC and NEUT% values transmitted via cloud to healthcare provider. Used by patients/caregivers at home; physician-prescribed. Features remote lockout capability for safety. Benefits: Enables remote monitoring of patients at risk of neutropenia without requiring frequent lab visits; supports clinical decision-making by providing quantitative hematology data to physicians.
Clinical Evidence
Clinical performance supported by method comparison (n=312) against Sysmex XE-5000 and at-home method comparison (n=165) against Athelas One. Results showed high correlation (WBC r=0.97-0.99; NEUT% r=0.92-0.96). Precision studies (bench and at-home) met predefined criteria (WBC CV <7.5%; NEUT% SD <5% or CV <15%). Flagging accuracy for morphological abnormalities was 94.87% overall. Bench testing confirmed linearity, LoD (0.079 K/µL), and robustness to environmental factors (tilt, drop, temperature, humidity).
Technological Characteristics
Microfluidic test strip with pre-loaded Cresyl Violet and Methylene Blue stain. Sensing principle: Computer vision-based image analysis of stained cell monolayer. Energy source: Electronic device. Connectivity: Internet-connected for cloud processing. Dimensions/form factor: Handheld. Software: Automated image analysis algorithm. Sterilization: Not specified.
Indications for Use
Indicated for quantitative determination of WBC and NEUT% in capillary whole blood for adults (aged 21+) at risk of neutropenia. For use by patients/caregivers in home settings. Contraindicated for patients deemed incompetent by a physician; not for sole diagnostic use.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K200828
B Applicant
Athelas Inc.
C Proprietary and Established Names
Athelas Home
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GKZ | Class II | 21 CFR 864.5220 - Automated Differential Cell Counter | HE - Hematology |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
White blood cell count (WBC) and percent neutrophil count (NEUT%)
C Type of Test:
Quantitative
## III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
B Indication(s) for Use:
The Athelas Home is indicated for the quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary whole blood. The Athelas Home system is for In Vitro Diagnostic use and for prescription use only. The Athelas Home is only to be used with Athelas Home Test Strips. The Athelas Home is indicated for use by patients and caregivers within home settings with results viewable by health care professionals. The Athelas Home is intended for patients at risk of neutropenia. For patients with psychiatric conditions, clinical judgement should be exercised when deciding the end-user, based on the instructions for use (IFU) of Athelas Home, the treating physician should determine which patients are competent to perform the test by themselves. The Athelas Home is indicated for adult populations only (aged 21 and older).
Results obtained with the Athelas Home should not be the sole basis for patient diagnosis, treatment or management of leukopenia and neutropenia. All results should be evaluated by a healthcare provider.
The Athelas Home is intended to be used by a single person and should not be shared.
Prescription Use Only.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
Athelas Home
IV Device/System Characteristics:
A Device Description:
The Athelas Home is an automated cell counter system which consists of the Athelas Home analyzer and the Athelas Home Test Strips. The Athelas Home is identical to the Athelas One (K181288) in its technology, all software, the analytes measured, all components, and usage process. The Athelas Home has additional software and quality control safeguards to enable at-home usage of the device. Additionally, results generated from the Athelas Home are not shown to end-patients, rather are transmitted via the software directly to their healthcare provider. The Athelas Home also has Remote Lockout capabilities, such that a healthcare provider or manufacturer can remotely lock-out a device for safety in usage.
The Athelas Home Tests Strips collects a capillary whole blood sample to generate a layer of cells for counting and image analysis. The Athelas Home Test Strips are comprised of an upper optical panel, lower optical panel and a stain coated region containing methylene blue and cresyl violet stains. The test strip channel is optically clear for the camera module to take pictures of the cells in the blood sample.
K200828 - Page 2 of 14
{2}
A smartphone/tablet with the Athelas controlling mobile application is required to initiate a test with the Athelas Home System. The smartphone/tablet models compatible to initiate testing are those devices supporting iOS 9 and up, or Android 8 and up.
## B Principle of Operation:
The Athelas Home uses image processing and microfluidics in order to measure WBC and NEUT% values from whole blood in home settings. The Athelas Home system consists of the Athelas Home analyzer and Athelas Home test strips. The Athelas Home test strips serve as both a sample container and a reaction chamber. A capillary whole blood sample is directly transferred via fingerstick to the Athelas Home test strip which automatically spreads the sample into a monolayer. The pre-coated stain within the strip chamber interacts with the monolayer of blood and stains the WBCs. The Athelas Home test strip is inserted into the Athelas Home analyzer which utilizes a proximity sensor to lock in place. A servo stabilizes and auto-focuses the blood sample, and then a stage actuator scans the strip across various fields while the optical module takes multiple images of the cells across the monolayer. The images of the blood sample are transmitted to the server where they are analyzed using a locked down image processing algorithm. The algorithm recognizes the nucleation and WBCs to generate a WBC count and NEUT% result based on the concentrations and types of cells present.
## Instrument Description Information:
1. Instrument Name:
Athelas Home
2. Specimen Identification:
Capillary whole blood
3. Specimen Sampling and Handling:
A capillary whole blood sample is directly transferred via fingerstick to the Athelas Home test strip which is inserted into the Athelas Home analyzer.
4. Calibration:
Factory calibrated and automatic calibration at the beginning of each test. No manual calibration by the end user.
5. Quality Control:
Auto-Check 3-Level Gridded Strip
OR ATH-CHECK (3 level control)
## V Substantial Equivalence Information:
## A Predicate Device Name(s):
Athelas One
K200828 - Page 3 of 14
{3}
B Predicate 510(k) Number(s):
K181288
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K200828 | K181288 |
| --- | --- | --- |
| Device Trade Name | Athelas Home | Athelas One |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The Athelas Home is indicated for the quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary whole blood. The Athelas Home system is for In Vitro Diagnostic use and for prescription use only. The Athelas Home is only to be used with Athelas Home Test Strips. The Athelas Home is indicated for use by patients and caregivers within home settings with results viewable by health care professionals. The Athelas Home is intended for patients at risk of neutropenia. For patients with psychiatric conditions, clinical judgement should be exercised when deciding the end-user, based on the instructions for use (IFU) of Athelas Home, the treating physician should determine which patients are competent to perform the test by themselves. The Athelas Home is indicated for adult | Athelas One is indicated for use for quantitative determination of white blood cells (WBC) and Neutrophil percentages (NEUT%) in capillary or K2EDTA venous whole blood. The Athelas One system is for In Vitro Diagnostic use only. The Athelas One is only to be used with Athelas One Test Strips. The Athelas One is indicated for use in clinical laboratories and for point of care settings. The Athelas One is only indicated for use in adult populations (aged 21 and older). |
K200828 - Page 4 of 14
{4}
| Device & Predicate Device(s): | K200828 | K181288 |
| --- | --- | --- |
| Device Trade Name | Athelas Home | Athelas One |
| General Device Characteristic Similarities | | |
| | populations only (aged 21 and older).
Results obtained with the Athelas Home should not be the sole basis for patient diagnosis, treatment or management of leukopenia and neutropenia. All results should be evaluated by a healthcare provider.
The Athelas Home is intended to be used by a single person and should not be shared.
Prescription Use Only. | |
| Modes of Operation | Single mode of operation for capillary samples | Single mode of operation for both venous and capillary samples |
| Software/Hardware | Internet connected device for processing results on cloud server | Same |
| Sample Volume | 3.5 μL | Same |
| Measurement Range | WBC: 1–25 x10^{3}/μL
NEUT%: 0–100% | Same |
| Calibration | Factory calibrated and automatic calibration at the beginning of each test. No manual calibration by the end user | Same |
K200828 - Page 5 of 14
{5}
| Device & Predicate Device(s): | K200828 | K181288 |
| --- | --- | --- |
| Device Trade Name | Athelas Home | Athelas One |
| General Device Characteristic Similarities | | |
| Test Principle | A microfluidic test strip channel creates a stained monolayer of white blood cells. Multiple images are taken of the monolayer and the cells are counted and classified by computer vision-based image analysis | Same |
| Parameters | WBC, NEUT% | Same |
| Target Population | Adults (aged 21 and older) | Same |
| Reagents | Cresyl Violet, Methylene Blue Stain (pre-loaded/coated dry and contained in test strip) | Same |
| Controls/ Calibrators | Auto-Check 3-Level Gridded Strip OR ATH-CHECK (3 level control) | Same |
| General Device Characteristic Differences | | |
| Intended Use Settings | Home | Point of Care, Clinical Laboratory |
| Specimen Type | Capillary whole blood | Capillary whole blood and K_{2}EDTA venous whole blood |
| Result Viewing | Results are remotely and securely transmitted to the prescribing physician or healthcare provider | Results are shown on the mobile/tablet application used for system operation |
| Remote Lockout Capability | Yes | No |
VI Standards/Guidance Documents Referenced:
CLSI H26-A2: Validation, Verification and Quality Assurance of Automated Hematology Analyzers; Approved Guideline – Second Edition
K200828 - Page 6 of 14
{6}
CLSI EP07: Interference Testing in Clinical Chemistry, Third Edition
CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures; Approved Guideline
CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
CLSI EP35: Assessment of Equivalency or Suitability of Specimen Types for Medical Laboratory Measurement Procedures; First Edition
CLSI EP09c-A3: Measurement Procedure Comparison and Bias Estimation Using Patient Samples. Third Edition
CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition
CLSI EP25-A: Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
CLSI EP28-A3c: Defining, Establishing, and Verifying Reference Intervals in Clinical Laboratory; Approved Guideline – Third Edition
## VII Performance Characteristics (if/when applicable):
## A Analytical Performance:
### 1. Precision:
Two separate precision studies were performed to evaluate the performance of the Athelas Home by home users. The precision studies were conducted in the home environment with an Athelas Home device system.
a. A group of 24 patients were instructed to conduct tests on the Athelas Home following the instructions for use. Each patient collected two capillary whole blood samples from finger pricks (from different fingers) within five minutes of each other, filling two unique test strips for measurements. The precision profile approach was used to estimate repeatability as shown in the table below.
| WBC Mean (x 10³/μL) | Interval | SD | %CV |
| --- | --- | --- | --- |
| 2.53 | 1.0–4.3 | 0.071 | 2.8% |
| 6.12 | 4.7–7.6 | 0.379 | 6.2% |
| 11.78 | 9.8–13.9 | 0.89 | 7.6% |
| NEUT% Mean | Interval | SD | %CV |
| --- | --- | --- | --- |
| 36% | 27–48% | 2.08 | 5.8% |
| 56% | 51–59% | 2.03 | 3.7% |
| 66% | 55–80% | 3.16 | 4.8% |
K200828 - Page 7 of 14
{7}
b. A separate group of 10 patients were provided with Athelas Home Quality Control materials (low, moderate, and high levels). They used the dropper on the Quality Control material vial to conduct triplicate measurements for the assigned sample by the given patient. The precision profile approach was used to estimate repeatability for the home users using the Quality Control materials. The result was presented in the table below.
| Sample | N | WBC Mean (x 10³/μL) | SD | %CV |
| --- | --- | --- | --- | --- |
| Low QC | 15 | 2.79 | 0.288 | 10.3% |
| Medium QC | 9 | 7.76 | 0.271 | 3.5% |
| High QC | 6 | 22.78 | 1.087 | 4.8% |
2. Linearity:
Refer to K181288
3. Analytical Specificity/Interference:
Refer to K181288
4. Assay Reportable Range:
The reportable range for white blood cells (WBC) is 1.0–25.0 x 10³/μL.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Reference range from K181288:
| Parameter | Male (N = 60) | | Female (N = 60) | |
| --- | --- | --- | --- | --- |
| | Lower Limit | Upper Limit | Lower Limit | Upper Limit |
| WBC (x10³/μL) | 3.91 | 10.90 | 4.49 | 12.68 |
| NEUT% | 41.0 | 70.7 | 42.9 | 74.3 |
6. Detection Limit:
Refer to K181288
7. Assay Cut-Off:
Not applicable
8. Accuracy (Instrument):
Not applicable
9. Carry-Over:
K200828 - Page 8 of 14
{8}
Not applicable
## B Comparison Studies:
1. Method Comparison with Predicate Device (At-Home Usability Study):
The study was performed to evaluate the performance of the Athelas Home analyzer in the home setting when compared to the predicate device Athelas One. A total of 61 patients (including patients with neutropenia, patients with autoimmune diseases, patients on immunosuppressive oncology therapy and patients on clozapine) were enrolled in the study, 1–4 visits per patient. Patients (untrained operators) were instructed to conduct tests on the Athelas Home following the labeling process. Fifteen (15) separate lots of test strips were used across all patients in the study. Tests were conducted in a single measurement by home users on the Athelas Home. The trained operators performed triplicate measurements on the predicate device. The average of the triplicates on the predicate performed by trained operator was used for comparison against the single measurement by self-testers and caregivers on the Athelas Home device in the home settings. Passing-Bablok regression was used to evaluate the performance of the Athelas Home. The tables presented below demonstrate the correlation and estimated bias.
| Parameter | Interval | N | (r) | Slope (95% CI) | Intercept (95% CI) |
| --- | --- | --- | --- | --- | --- |
| WBC | 1.3–14.4 | 117 | 0.97 | 1.00 (0.97, 1.05) | 0.00 (-0.33, 0.16) |
| NEUT % | 21-78 | 117 | 0.95 | 1.00 (0.93, 1.05) | 0.00 (-2.25, 5.00) |
Bias at Medical Decision Levels of WBC
| WBC Level (x10³/μL) | Bias (95% CI) | %Bias (95% CI) |
| --- | --- | --- |
| 3.9 | 0.00 (-0.18, 0.10) | 0.00% (-4.50%, 2.56%) |
| 10.4 | 0.00 (-0.19, 0.28) | 0.00% (-1.85%, 2.65%) |
Bias at Medical Decision Levels of NEUT%
| NEUT% | Bias (95% CI) | %Bias (95% CI) |
| --- | --- | --- |
| 46.4 | 1.00 (-0.16, 2.00) | 2.16% (-0.34%, 4.31%) |
| 76.9 | 1.00 (-0.66, 2.00) | 1.30% (-0.86%, 2.60%) |
2. Matrix Comparison:
Not applicable
## C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
K200828 - Page 9 of 14
{9}
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
Refer to K181288
F Other Supportive Instrument Performance Characteristics Data:
1. At-Home Usability Analysis
During the first At-Home visits, study investigators were asked to conduct usability surveys to evaluate home users in setting up the device system for testing and their understanding of how to run the test. The results of this study are described in the table below.
| Survey Question | Yes | No |
| --- | --- | --- |
| Was the user able to use the lancet correctly? Mark yes if the user was able to open the lancet, puncture the correct site on their finger, and dispose of the lancet correctly. | 100%
(32/32) | 0%
(0/32) |
| Did the user fill up the test-strip fully? Mark yes if the user placed enough blood such that the blood reached the end of the channel on the test-strip. | 94%
(30/32) | 6%
(2/30) |
| Did the user face any errors when running the test (not error codes)? | 97%
(31/32) | 3%
(1/32) |
| Did the user correctly know how to run the test using the application? | 100%
(32/32) | 0%
(0/32) |
| Was this test performed by a caregiver or the patient?
(‘Yes’ if conducted by patient, ‘No’ if conducted by caregiver) | 31%
(10/32) | 69%
(22/32) |
2. At-Home Error Analysis
During At-Home testing, flex studies were conducted to verify the usability and ease of use of the Athelas Home when used by patients in the intended use settings. Over 94% of tests were run successfully without errors. Errors were resolved after running a single additional test strip. The results were presented in the table below:
| Error Type | Occurrence # | Percent of Tests |
| --- | --- | --- |
| No Error | 76 | 94 |
| Test-Strip/Optical Error | 3 | 3.7 |
| Incorrect Insertion Direction | 2 | 2.5 |
| HH! (shows as success to the patient) | 1 | 1.2 |
| LL! (shows as success to the patient) | 1 | 1.2 |
K200828 - Page 10 of 14
{10}
K200828 - Page 11 of 14
## 3. Flex Studies
### Delay in Testing After Sample Collection
A study was conducted to evaluate whether there was measurand drift in the analyte results if there was a delay between sampling and running the test strip on device. The study was to test for the case where a user fills the strip and waits for up to 30 minutes before inserting it into the device for analysis. Five different timepoints (5 minutes, 15 minutes, 25 minutes, 30 minutes and 35 minutes) were evaluated across two lots of test strips with two Athelas Home devices. Nine whole blood samples covering the measuring range were used. The linear regression analysis showed no significant measurand drift for tested samples. Results validated the claimed stability up to 30 minutes.
### Sample Strip Fill-Volume Testing
To determine whether the Athelas Home correctly handles variable amounts of blood inserted into the test-strip, a study was conducted with variable amounts of blood inserted into the test-strip using three K2EDTA whole blood samples with known WBC and NEUT% values. Each volume sample (0 μL, 1.4 μL, 3.4 μL, and 5.4 μL) was tested in two replicates by using one lot of test strips. Results were found acceptable and met the defined acceptance criteria.
### Tilt Testing
A study was conducted to determine whether the Athelas Home is robust against being placed on tilted surfaces in different orientations. Three K2EDTA whole blood samples were tested in two replicates with one lot of test strips on one device. One level of tilt angle was tested (30 degrees) with the analyzer tested in four orientations (front to back, back to front, left to right and right to left). Samples representing a low, normal, and abnormal (high) level of WBC and Neut% were analyzed in triplicates in each test position. Results demonstrate that the Athelas Home is not affected by tilted placement of the device.
### Drop Testing
A study was conducted to determine whether the Athelas Home continues to return accurate results after being dropped from multiple orientations and heights. 12 Athelas Home analyzers were tested in accordance with ASTM 1469 - D03. The test method included manual handling, vehicle stacking, loose load vibrations, vehicle vibrations. The Athelas Home were assessed in the following ways: Verify no visible physical damage or degradation; Verify the following functional characteristics (LED function, device stage flatness, ABS surface damage, motor function, liner actuator function and camera function); Verify a single run of the Auto-Check 3-Level Gridded Strip returns results within the specification range. Results demonstrate that all devices continued to pass functional tests after being subjected to the full testing schedules.
{11}
K200828 - Page 12 of 14
# Temperature Testing
A study was conducted to verify the performance of the Athelas Home device and test strips after extreme and varying temperature. The study was divided into two parts, one for the analyzer, and another for the test strips. For the device, three Athelas Home devices were placed in the given investigated temperature (-20°C (extreme low), 20°C (room temp), 40°C (extreme high)) for 12 hours and then allowed to return to room temperature for 4 hours. All test strips came from a single lot stored only at room temperature. Auto-Check 3-Level Gridded Strip were tested in three replicates on each device. For the test strips, three lots of Athelas Home test strips were placed in the given temperature ((-20°C (extreme low), 20°C (room temp), 40°C (extreme high)) for 12 hours. The test strips were allowed to return at room temperature for 4 hours. Auto-Check 3-Level Gridded Strip were tested in three replicates on one device only stored at room temperature for each test strip lot. Results demonstrate that both the Athelas Home device and Athelas Home test-strip meet performance specifications after exposure to different temperatures.
# Humidity Testing
A study was conducted to verify the performance of the Athelas Home under extreme humidity conditions. The study evaluated the performance of the Athelas Home device after exposure to RH levels (from 30% to 90% RH) outside of the recommended 60% RH. Three Athelas Home analyzers were exposed to three separate simulated conditions for 72 hours. For each simulated condition, tests were performed on the Athelas Home using Auto-Check 3-Level Gridded Strip (Low, Normal, and High) in three replicates. Results demonstrate that the Athelas Home is functional even after exposure to relative humidity levels outside of the specified operational range. Exposure to relative humidity levels within the range tested (up to 72 hours) shows no impact on performance.
# Vibration Testing
A study was conducted to ensure that the Athelas Home device can continue to function or return appropriate error codes after exposure to extensive vibration in handling by a patient or shipping. The analyzer was subjected to three levels of vibration: 4.8 millimeters/second (mm/s), 10 mm/s, and 15 mm/s. Samples with different levels (low, medium and high) were tested in triplicates at each vibration level. Results were compared to the same sample levels tested without vibration. All results were within the acceptance criteria and no erroneous results were produced. All low and abnormal (high) samples flagged and/or suppressed as expected.
# Ambient Room Light
A study was conducted to verify the performance of the Athelas Home device and test strips for exposure in various ambient lighting conditions during operation. Three devices were used in this study with one device in a fully dark room (0-20 lumens), second device in ambient room light (400-600 lumens), and third device in bright sunlight (3000-6000 lumens). Each device was tested with three runs on an Auto-Check 3-Level Gridded Strip. Results demonstrate that the Athelas Home is capable of performing in a range of ambient lighting conditions from complete darkness to extreme external brightness.
{12}
K200828 - Page 13 of 14
# 4. Flagging Comparison Study
This study was conducted on Athelas One (identical to Athelas Home) compared to the Sysmex XE-5000 to assess the flagging capabilities (distributional and morphological). This study was performed with 312 patient samples from either capillary whole blood or venous whole blood collected in $\mathrm{K}_2\mathrm{EDTA}$ anticoagulant. Results met the pre-defined acceptance criteria. Summarized data is presented below for both distributional flags as well as morphological flags.
## Distributional Flags
The results of the Athelas One WBC distributional flagging (leukocytosis, leukocytopenia) compared to the Sysmex XE-5000 were divided into two categories: 1) No flags, negative judgement and 2) patients with positive distributional abnormalities with flags present, positive judgement.
| | Sysmex XE-5000 | | | |
| --- | --- | --- | --- | --- |
| | | Positive (Abnormal) | Negative (Normal) | Total |
| Athelas One | Positive (Abnormal) | 34 | 4 | 38 |
| | Negative (Normal) | 5 | 269 | 274 |
| | Total | 39 | 273 | 312 |
%Positive Agreement (Sensitivity) = 87.2%; 95% CI: 72.57, 95.70
%Negative Agreement (Specificity) = 98.5%; 95% CI: 96.29, 99.60
%Overall Agreement = 97.12%; 95% CI: 94.59, 98.67
## Morphological Flags
The results of the Athelas One WBC morphological flagging (nucleated RBCs, platelet clumps, etc.) compared to the Sysmex XE-5000 were divided into two categories: 1) No flags, negative judgement and 2) patients with positive distributional abnormalities with flags present, positive judgement.
| | Sysmex XE-5000 | | | |
| --- | --- | --- | --- | --- |
| | | Positive (Abnormal) | Negative (Normal) | Total |
| Athelas One | Positive (Abnormal) | 90 | 7 | 97 |
| | Negative (Normal) | 9 | 206 | 215 |
| | Total | 99 | 213 | 312 |
%Positive Agreement (Sensitivity) = 90.91%; 95% CI: 83.44, 95.76
%Negative Agreement (Specificity) = 96.71%; 95% CI: 93.35, 98.67
%Overall Agreement = 94.87%; 95% CI: 91.81, 97.04
{13}
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K200828 - Page 14 of 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.