XBit Series Digital Color Doppler Ultrasound System
K200780 · CHISON Medical Technologies Co., Ltd. · IYN · Aug 5, 2020 · Radiology
Device Facts
Record ID
K200780
Device Name
XBit Series Digital Color Doppler Ultrasound System
Applicant
CHISON Medical Technologies Co., Ltd.
Product Code
IYN · Radiology
Decision Date
Aug 5, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The XBit Series Digital Color Doppler Ultrasound System is intended for diagnostic ultrasound imaging in B (2D/3D/4D),B/M,M,B+CFM,B+CPA (PD),B+DPD,B+PW,B+CW,B+CFM +D (PW)/CW, B+ CPA(PD) + D (PW)/CW, TDI and Fusion Harmonic Imaging modes. The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified clinician for evaluation of Fetal ,Abdominal,Pediatric,Small Organ (breast, thyroid,testes ), Neonatal Cephalic , Adult Cephalic, Cardiac (adult , pediatric), Musculo-skeletal (Conventional , Superficial) , Peripheral Vascular, Transesophageal, Trans-rectal, Trans-vaginal, OB/GYN and Urology.
Device Story
Mobile, software-controlled digital color Doppler ultrasound system; acquires ultrasound echo data via linear, convex, phased array, or volume probes. Transforms inputs using autocorrelation (color) and FFT (PW/CW Doppler) processing; displays images in B, M, CFM, Power Doppler, TDI, and 3D/4D modes. Used in hospitals/clinics by qualified clinicians. Features semi-automated functions (Auto Follicle, SonoBeam, SonoColor, SonoFusion) requiring user modification. Healthcare providers use real-time images and automated measurements to evaluate anatomy and fluid flow; aids clinical decision-making by providing diagnostic visualization of structures and blood flow; benefits patients through non-invasive diagnostic imaging.
Clinical Evidence
No clinical testing was required. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output (IEC 60601-2-37), and biocompatibility (ISO 10993-1).
Technological Characteristics
Mobile console with LED monitor, keyboard, and multiple probe ports. Supports linear, convex, phased, and volume arrays. Imaging modes: B, M, PW/CW Doppler, CFM, Power Doppler, TDI, 3D/4D. Connectivity: DICOM/standard image archiving. Software-controlled signal processing using autocorrelation and FFT. Safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1.
Indications for Use
Indicated for diagnostic ultrasound imaging in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal/adult cephalic, cardiac (adult/pediatric), musculoskeletal (conventional/superficial), peripheral vascular, transesophageal, trans-rectal, trans-vaginal, OB/GYN, and urology applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
CBit 9 Digital Color Doppler Ultrasound System (K180974)
Acuson x700 Diagnostic Ultrasound System (K141846)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
CHISON Medical Technologies Co., Ltd. % Liu Qifei Regulatory Affairs Manager No. 228, Changjiang East Road, Block 51 and 53 Phase 5, Shuofang Industrial Park Xinwu District, Wuxi, Jiangsu 214142 CHINA
August 5, 2020
### Re: K200780
Trade/Device Name: XBit Series Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: June 6, 2020 Received: June 6, 2020
#### Dear Liu Qifei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K200780
Device Name
XBit Series Digital Color Doppler Ultrasound System
#### Indications for Use (Describe)
The XBit Series Digital Color Doppler Ultrasound System is intended for diagnostic ultrasound imaging in B (2D/3D/4D),B/M,M,B+CFM,B+CPA (PD),B+DPD,B+PW,B+CW,B+CFM +D (PW)/CW, B+ CPA(PD) + D (PW)/CW, TDI and Fusion Harmonic Imaging modes. The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified clinician for evaluation of Fetal ,Abdominal,Pediatric,Small Organ (breast, thyroid,testes ), Neonatal Cephalic , Adult Cephalic, Cardiac (adult , pediatric), Musculo-skeletal (Conventional , Superficial) , Peripheral Vascular, Transesophageal, Trans-rectal, Trans-vaginal, OB/GYN and Urology.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 10px;"> <span style="border: 1px solid black; padding: 2px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="border: 1px solid black; padding: 2px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K200780
In accordance with 21 CFR 807.92 the following summary of information is provided:
## 1. Submitter:
| Submitter: | CHISON Medical Technologies Co., Ltd. |
|----------------|----------------------------------------------------------------------------------------------------------------------------------|
| Address: | No.228, Changjiang East Road, Block 51 and 53, Phase 5, Shuofang<br>Industrial Park, Xinwu District, Wuxi, Jiangsu, China 214142 |
| Contact: | Mr. Liu Qifei |
| Tel: | +86-510-85310019 |
| Fax: | +86-510-85310021 |
| Date Prepared: | March 12th, 2020 |
#### 2. Device :
Trade Name: XBit Series Digital Color Doppler Ultrasound System
Common Name: Diagnostic Ultrasound System with Transducers
Classification: Regulatory Class: II Review Category: Tier II
| Classfication Name | 21 CFR Section | Product Code |
|------------------------------------------|----------------|--------------|
| Ultrasonic pulsed doppler imaging system | 892.1550 | 90-IYN |
| Ultrasonic pulsed echo imaging system | 892.1560 | 90-IYO |
| Diagnostic ultrasonic transducer | 892.1570 | 90-ITX |
# 3. Predicate Device(s):
| Device | Model | Product Code | 510(k)Number |
|-------------------------|------------------------------------------------|--------------------|--------------|
| 1.Main predicate device | CBit 9 Digital Color Doppler Ultrasound System | IYN, IYO, ITX | K180974 |
| 2.Reference device | Resona 7 Diagnostic Ultrasound System | IYN, IYO, ITX | K171233 |
| 3.Reference device | Acuson x700 Diagnostic Ultrasound System | IYN, IYO, ITX, OBJ | K141846 |
### 4. Device Description:
The XBit Series Digital Color Doppler Ultrasound System is an integrated preprogrammed color doppler ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.
This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array. This system consists of a mobile console with keyboard control panel, power supply module, color LED monitor and optional probes.
{4}------------------------------------------------
This system is a mobile, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound echo data and to display the image B-Mode (including Fusion Harmonic Imaging), M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Power Doppler Mode, Directional Power Doppler Mode, TDI Mode or a combination of these modes, Elastography, contrast imaging, 3D/4D.
Auto Follicle, SonoBeam, SonoColor, and SonoFusion, these four features are all semi-automated functions, they need to be modified during processing.
### 5. Indications for Use:
The XBit Series Digital Color Doppler Ultrasound System is intended for diagnostic ultrasound imaging in B(2D/3D/4D), B/M, M, B+CFM,B+CPA (PD),B+DPD,B+PW,B+CW, B+ CFM + D (PW)/CW, B+ CPA(PD) + D (PW)/CW, TDI and Fusion Harmonic Imaging modes. The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified clinician for evaluation of Fetal,Abdominal,Pediatric,Small Organ (breast, thyroid,testes ), Neonatal Cephalic ,Adult Cephalic,Cardiac (adult , pediatric),Musculoskeletal (Conventional, Superficial), Peripheral Vascular,Transesophageal,Trans-rectal, Trans-vaginal, OB/GYN and Urology.
### 6. Summary of Non-Clinical Tests:
The XBit Series Digital Color Doppler Ultrasound System has been evaluated for electrical, mechanical, thermal and electromagnetic compatibility safety, biocompatibility and acoustic output.
The device has been found to conform to applicable medical device safety standards in regards to thermal, mechanical and electrical safety as well as biocompatibility.
IEC 60601-1: 2015 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
IEC 60601-1-2: 2014 Medical Electrical Equipment - Part 1-2: General Requirements For Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.
IEC 60601-2-37: 2015 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
Output Indices on Diagnostic Ultrasound Equipment
ISO 10993-1:2018 Biological Evaluation of Medical Devices -- Part 1: Evaluation And Testing Within A Risk Management Process
The device has been found to conform to applicable FDA medical device guidance documents titled as followings:
- . Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (Document issued on: June 27, 2019)
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005)
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (Document Issued on: October 2, 2014)
{5}------------------------------------------------
- Use of International Standard ISO 10993-1, "Biological evaluation of medical devices ● - Part 1: Evaluation and testing within a risk management process" (Document issued on: June 16, 2016)
## 7. Clinical Test:
No clinical testing was required.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on May 11, 2005", is also included as part of this submission.
#### 8. Comparison to Predicate Device:
#### Table 1 Substantial Equivalence Comparison
| | Main predicate<br>device | Reference<br>device | Reference<br>device | Submission<br>Device | |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Items | CBit 9 Digital<br>Color Doppler<br>Ultrasound<br>System<br>K180974 | Resona 7<br>Diagnostic<br>Ultrasound<br>System<br>K171233 | Acuson x700<br>Diagnostic<br>Ultrasound<br>System<br>K141846 | XBit Series<br>Digital Color<br>Doppler<br>Ultrasound<br>System | Remark |
| Indications for<br>Use | Fetal<br>Abdominal<br>Pediatric<br>Small Organ<br>(breast,<br>thyroid ,testes)<br>Neonatal<br>Cephalic ,Adult<br>Cephalic<br>Trans-rectal<br>Trans-vaginal<br>Musculo-skeletal<br>( Conventional,<br>Superficial)<br>Cardiac(adult ,p<br>ediatric)<br>Peripheral<br>Vascular<br>OB/GYN,Urology<br>Trans-esophage<br>al | Fetal<br>Abdominal<br>Intra-operative(a<br>bdominal,<br>thoracic, and<br>vascular)<br>Pediatric<br>Small<br>organ(breast,<br>thyroid ,testes)<br>Neonatal<br>Cephalic<br>Adult Cephalic<br>Trans-rectal<br>Trans-vaginal<br>Musculo-skeleta<br>I(conventional,<br>superficial),<br>Cardiac adult<br>Cardiac<br>pediatric<br>Trans-esoph.<br>(cardiac),<br>Peripheral<br>vessel<br>urology | Cardiac (Adult,<br>Pediatric)<br>Transesphageal<br>(Cardiac)<br>Intracardiac<br>Cerebrovascular<br>Peripheral<br>Vessel<br>Abdominal<br>Renal<br>Fetal<br>Abdominal<br>Intra-operative<br>Pediatric<br>Small Organ<br>Neonatal<br>Cepahalic<br>Adult Cephalic<br>Orthopedics<br>Musculo-skeleta<br> <br>Conventional<br>Musculo-skeleta<br>I Superficial<br>Pelvic<br>Obstetrical<br>Gynecological<br>Urological | Fetal<br>Abdominal<br>Pediatric<br>Small Organ<br>(breast,<br>thyroid ,testes)<br>Neonatal<br>Cephalic ,Adult<br>Cephalic<br>Trans-rectal<br>Trans-vaginal<br>Musculo-skeleta<br>I( Conventional,<br>Superficial)<br>Cardiac(adult ,p<br>ediatric)<br>Peripheral<br>Vascular<br>OB/GYN,Urology<br>Trans-esophage<br>al | Same |
| Items | Main predicate device | Reference device | Reference device | Submission Device | Remark |
| | CBit 9 Digital Color Doppler Ultrasound System K180974 | Resona 7 Diagnostic Ultrasound System K171233 | Acuson x700 Diagnostic Ultrasound System K141846 | XBit Series Digital Color Doppler Ultrasound System | |
| Design | Autocorrelation for color processing and FFT for pulse and CW Doppler processing. Supporting Linear, Curve, Phase array and Volume probes . Cine play back capability Image file archive | Transmit ultrasonic energy into patients, then perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. | The ACUSON X700 ultrasound system features a sophisticated imaging engine migrated from our premium imaging products as well as technology to bolster compatibility across systems. | Autocorrelation for color processing and FFT for pulse and CW Doppler processing. Supporting Linear, Curve, Phase array and Volume probes . Cine play back capability Image file archive | Same |
| | TGC 8 slider | TGC 8 slider | TGO: 8 controls | TGC 8 slider | Same |
| | Depth Range: 0 to 45 cm | Depth Range: 1.5 to 40 cm | Depth Range: 1 to 30 cm | Depth Range: 0 to 45 cm | Same |
| | 256 shades of gray | NA | NA | 256 shades of gray | Same |
| | B Dynamic range control: 20-280dB | B Dynamic range control: 30-260dB | B Dynamic range selection: 30-90dB | B Dynamic range control: 20-280dB | Same |
| | Gain:0-255,1/step | Gain:0-100,1/step | Gain:-30-+30,1/step | Gain:0-255,1/step | Same |
| | Focal Number: adjustable | No physical focus to adjust | Focal Number: adjustable | Focal Number: adjustable | Same |
| | Focus position: adjustable | No physical focus to adjust | Focus position: adjustable | Focus position: adjustable | Same |
| Operating Controls | B steer: available on linear transducers | B steer: available on linear transducers | B steer: available on linear transducers | B steer: available on linear transducers | Same |
| | B Persistence: 7 steps | B Persistence: 7 steps | B Persistence:5 levels | B Persistence: 7 steps | Same |
| | ROI size/position: adjustable | ROI size/position: adjustable | ROI size/position: adjustable | ROI size/position: adjustable | Same |
| | Color Wall Filter settings:8 steps | Color Wall Filter settings:8 steps | Wall filter : 4 selections | Color Wall Filter settings:8 steps | Same |
| | Color Baseline: 16 steps Color Maps: 21 maps Color Invert: | Color Baseline: 16 steps Color Maps: 21 maps Color Invert: | Color Baseline: 9 steps Color maps: 16 maps Color | Color Baseline: 16 steps Color Maps: 21 maps Color Invert: | Same |
| | Main predicate device | Reference device | Reference device | Submission Device | Remark |
| Items | CBit 9 Digital Color Doppler Ultrasound System K180974 | Resona 7 Diagnostic Ultrasound System K171233 | Acuson x700 Diagnostic Ultrasound System K141846 | XBit Series Digital Color Doppler Ultrasound System | |
| | on/off | on/off | invert :on/off | on/off | |
| | PW sweeping speed: 6 steps | PW sweeping speed: 6 steps | Sweep speed: 8 selections | PW sweeping speed: 6 steps | Same |
| | PW Wall Filter: 7 steps | PW Wall Filter: 8 or 9 steps | Wall filter: 25 – 3906 Hz, 8 steps | PW Wall Filter: 7 steps | Same |
| | PW sample volume:<br>0.5-30mm (PW only) | PW sample volume:<br>0.5-30mm (PW only) | Gate size: 0.2 - 20 mm | PW sample volume:<br>0.5-30mm (PW only) | Same |
| | PW angle correction:-89~8 9degrees, 1/step | PW angle correction:-89~8 9degrees, 1/step | Angle correction: 0 – 89° in 1° increments | PW angle correction:-89~8 9degrees, 1/step | Same |
| | Baseline: 8steps | Baseline: 9steps | Baseline shift: 17 levels | Baseline: 8steps | Same |
| | Cine control: step, play backward, play continuously | Cine control: step, play backward, play continuously | Cine control: step, play backward, play continuously | Cine control: step, play backward, play continuously | Same |
| | Doppler Auto Trace | Doppler Auto Trace | Autotrace Function | Doppler Auto Trace | Same |
| | Freeze control:Toggling freeze key | Freeze control:Toggling freeze key | Freeze control:Toggling freeze key | Freeze control:Toggling freeze key | Same |
| Safety Compliance | IEC60601-1<br>IEC60601-1-2<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>IEC 60601-2-37 | IEC60601-1<br>IEC60601-1-2<br>ISO 10993-1<br>ISO 9001<br>ISO 13485<br>IEC 60601-2-37 | EN/IEC 60601-1<br>EN/IEC 60601-1-1<br>EN/IEC 60601-1-2<br>EN/IEC 62304<br>EN/IEC 62366<br>EN/IEC 60601-2-18<br>EN/IEC 60601-2-25 | IEC60601-1<br>IEC60601-1-2<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>IEC 60601-2-37 | Same |
| Operation Mode | B mode | B mode | B mode | B mode | Same |
| | B/M mode | B/M mode | B/M mode | B/M mode | Same |
| | M mode | M mode | M mode | M mode | Same |
| | Dual mode | Dual mode | Dual mode | Dual mode | Same…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.