← Product Code [HAE](/productcode/HAE) · K200768

# Rebellion, Phantom Multi-Bite Kerrison Rongeur (K200768)

_Morpheus AG · HAE · May 4, 2021 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200768

## Device Facts

- **Applicant:** Morpheus AG
- **Product Code:** [HAE](/productcode/HAE.md)
- **Decision Date:** May 4, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4840
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Rebellion is intended for cutting and removing bone, vertebral bodies and tissue in orthopedics, neurosurgery and spine surgery involving the skull or spinal column.

## Device Story

Rebellion is a manual bone punch (Kerrison rongeur) used in orthopedics, neurosurgery, and spine surgery. Device features a long barrel (200–250 mm) with a trigger mechanism for manual bone cutting. Integrated Luer Lock connection allows flushing with NaCl 0.9% to prevent tube clogging; posterior tube connects to a bone/tissue trap or standard suction system to remove bone fragments. Operated by surgeons in an OR setting. Suction-assisted bone removal distinguishes it from manual-only predicates. Device is sterile, single-use, and constructed from stainless steel and various polymers. Benefits include efficient bone removal and reduced risk of clogging during surgical procedures.

## Clinical Evidence

No clinical data submitted. Bench testing included biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemocompatibility), sterilization validation (ISO 11135:2014), packaging/shelf-life validation (ISO 11607:2017), and functional performance testing (cadaveric study in OR and simulated bone cutting to maximum life cycle).

## Technological Characteristics

Manual rongeur; stainless steel (420, 304, 302) and polymers (PVC, ABS, PA6 GF60, TPE, fluoropolymer). Features integrated Luer Lock for saline flushing and suction-assisted bone removal. 200-250 mm shaft length; 2-4 mm bite size; 90° up-biting. Sterile, single-use. Sterilization via ethylene oxide per ISO 11135:2014.

## Regulatory Identification

A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.

## Predicate Devices

- Fehling Punches ([K153243](/device/K153243.md))

## Reference Devices

- Steribite ([K180949](/device/K180949.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 4, 2021

Morpheus AG Timo Rack CEO Bahnhofstrasse 20 Spaichingen. 78549 DE

Re: K200768

Trade/Device Name: Rebellion, Phantom Multi-Bite Kerrison Rongeur Regulation Number: 21 CFR 882.4840 Regulation Name: Manual Rongeur Regulatory Class: Class II Product Code: HAE Dated: March 29, 2021 Received: April 2, 2021

Dear Timo Rack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200768

Device Name

Rebellion; Phantom Multi-Bite Kerrison Rongeur

Indications for Use (Describe)

The Rebellion is intended for cutting and removing bone, vertebral bodies and tissue in orthopedics, neurosurgery and spine surgery involving the skull or spinal column.

| Type of Use (Select one or both, as applicable) |                                              |
|-------------------------------------------------|----------------------------------------------|
| X Prescription Use (Part 21 CER 801 Subpart D)  | Over-The-Counter Use (21 C.FR 801 Subnart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

 Morpheus AG Rebellion 510(k) Premarket Notification

| DATE:           | 05/04/2021                                                                                                      |
|-----------------|-----------------------------------------------------------------------------------------------------------------|
| 510(k) Number:  | K200768                                                                                                         |
| APPLICANT:      | Morpheus AG Bahnhofstrasse 20 78549 Spaichingen Germany Tel: + 49 (0) 7461 70382-0 E-Mail: info@morpheus.com.ag |
| CONTACT PERSON: | Timo Rack CEO Tel.: + 49 (0) 7461 70382-11 E- Mail: t.rack@morpheus.com.ag                                      |

{4}------------------------------------------------

Morpheus AG Rebellion 510(k) Premarket Notification

### 1 Device Name

| Trade Name:                 | Rebellion;<br>Phantom Multi-Bite Kerrison Rongeur |
|-----------------------------|---------------------------------------------------|
| Common Name:                | Kerrison Rongeur                                  |
| Device Classification Name: | Manual rongeur                                    |

### Classification / Product Code 2

Rebellion can be classified according to following device name and product code:

| Device    | Regulation<br>Description | Regulation<br>Medical Specialty | Product Code | Regulation<br>Number | Device<br>Classification |
|-----------|---------------------------|---------------------------------|--------------|----------------------|--------------------------|
| Rebellion | Manual rongeur            | Neurology                       | HAE          | 21 CFR<br>882.4840   | II                       |

### Predicate Device / Reference Device 3

| Subject Device | Predicate Device | Reference Device | 510(k) number | 510(k) Holder                        |
|----------------|------------------|------------------|---------------|--------------------------------------|
| Rebellion      | Fehling Punches  | --               | K153243       | Fehling Instruments<br>GmbH & Co. KG |
|                | --               | Steribite        | K180949       | RJR Surgical, Inc.                   |

### Device Description 4

The Rebellion is a Bone Punch and it is available in two lengths (200 – 250 mm) with tree different tip sizes (2, 3 and 3 mm). The Rebellion can be connected to a rinsing solution (NaCl 0,9%) via integrated Luer Lock connection. The tubes are flushed with the rinsing solution, so a clogging of the tube is prevented. At posterior tube the Rebellion can be connected to the Morpheus Bone and Tissue Trap or to a standard suction system via sterile tubing.

Image /page/4/Figure/11 description: The image shows a tool labeled as "Figure 1: Rebellion". The tool has a long, white barrel with a black handle and trigger mechanism. A thin, flexible tube is connected to the back of the handle.

### 은 Indications for Use

The Rebellion is intended for cutting and removing bone, vertebral bodies and tissue in orthopedics, neurosurgery and spine surgery involving the skull or spinal column.

{5}------------------------------------------------

## Morpheus AG

Rebellion 510(k) Premarket Notification

# 6 Technological Characteristics

Morpheus Rebellion possesses similar technological characteristics as compared to the predicate device K153243 and reference device K180949.Different characteristics do not raise different questions of safety and effectiveness, and scientific methods were applied to evaluate different characteristics' effects on safety and effectiveness.

# 6.1 Device Characteristics Table

| Description            | Subject Device                                                                                                                                                               | Predicate Device<br>Fehling Instruments GmbH &<br>Co. KG<br>(Fehling-punches)                                                                                              | Reference Device<br>RJR Surgical, Inc.<br>(Steribite)                                                                                                             |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)                 | --                                                                                                                                                                           | K153243                                                                                                                                                                    | K180949                                                                                                                                                           |
| Product Code           | HAE                                                                                                                                                                          | HAE                                                                                                                                                                        | HAE                                                                                                                                                               |
| Class                  | II                                                                                                                                                                           | II                                                                                                                                                                         | II                                                                                                                                                                |
| Regulation #           | 882.4840                                                                                                                                                                     | 882.4840                                                                                                                                                                   | 882.4840                                                                                                                                                          |
| Classification<br>Name | Manual, Rongeur                                                                                                                                                              | Manual, Rongeur                                                                                                                                                            | Manual, Rongeur                                                                                                                                                   |
| Indication for Use     | The Rebellion is indicated to cut<br>and remove bone, vertebral<br>body and tissue in the<br>orthopedic, neuro and spine<br>surgery involving the skull or<br>spinal column. | Fehling rongeurs (bone<br>punches) are manually<br>operated instruments<br>indicated for cutting or biting<br>bone during surgery involving<br>the skull or spinal column. | The Steribite® Rongeur is<br>a manually operated<br>instrument indicated for<br>cutting or biting bone<br>during surgery involving<br>the skull or spinal column. |
| Anatomical<br>location | Skull and spine                                                                                                                                                              | Skull and spine                                                                                                                                                            | Skull and spine                                                                                                                                                   |
| Sterility              | Sterile                                                                                                                                                                      | Non-Sterile                                                                                                                                                                | Sterile                                                                                                                                                           |
| Re-Use                 | No                                                                                                                                                                           | Yes                                                                                                                                                                        | No                                                                                                                                                                |
| Material               | Stainless Steel: 420, 304 and<br>302<br>Polymer: PVC, ABS, PA6 GF60<br>(Grivory GV-6H); TPE<br>(TM5MED) and Fluoropolymer<br>(Altera MT1000A)                                | 420 and 304 Stainless Steels;<br>Coating: CERAMO® (TiAIN)<br>Polymer: Silicone                                                                                             | 420, 17- 4PH, 302 and<br>316 stainless steel;<br>Polymer: polyarylamide<br>resin                                                                                  |
| Patient<br>Contacting  | Stainless Steel: 420, 304<br>Polymer: Fluoropolymer (Altera<br>MT1000A)                                                                                                      | 420 and 304 Stainless Steels;<br>CERAMO® (TiAIN)                                                                                                                           | 420 and 17- 4PH stainless<br>steel                                                                                                                                |
| Design features        | Manual Rongeur                                                                                                                                                               | Manual Rongeur                                                                                                                                                             | Manual Rongeur                                                                                                                                                    |
| Bone Removal           | Suction                                                                                                                                                                      | Manual                                                                                                                                                                     | Manual                                                                                                                                                            |
| Shaft-length           | 200 - 250 mm                                                                                                                                                                 | 110 - 400 mm                                                                                                                                                               | 200 - 280 mm                                                                                                                                                      |
| Jaw opening            | 14 mm                                                                                                                                                                        | 9 - 19 mm                                                                                                                                                                  | Not known                                                                                                                                                         |
| Bite size              | 2 - 4 mm                                                                                                                                                                     | 0.8 - 8 mm                                                                                                                                                                 | 1 - 5 mm                                                                                                                                                          |
| Cutting<br>angulation  | 90° up-biting                                                                                                                                                                | 40° and 90° up/down biting                                                                                                                                                 | 40° up-biting                                                                                                                                                     |

{6}------------------------------------------------

Morpheus AG Rebellion 510(k) Premarket Notification

### 7 Performance Data

# 7.1 Clinical performance testing

Clinical performance testing was not submitted in this 510(k).

# 7.2 Biocompatibility

The Morpheus Rebellion is categorized as an external communicating device with limited contact (≤24 h) with tissue/bone. The appropriate biocompatibility endpoints that were assessed include:

- Cytotoxicity
- Sensitization
- . Irritation or Intracutaneous Reactivity
- . Acute Systematic Toxicity
- Material-Mediated Pyrogenicity
- . Hemocompatibility

# 7.3 Sterilization

Sterilization validation was conducted in accordance with ISO 11135:2014, Sterilization of health-care products — Ethylene oxide — Requirements for the development, validation and routine control of a sterilization process for medical devices.

# 7.4 Packaging and Shelf Life

Packaging and Shelf Life validation was conducted in accordance with ISO 11607:2017, Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems.

# 7.5 Functional performance data

To verify function and performance under simulated environmental and application conditions, a cadaver test was performed in an operating room with two potential users.

A product validation was also performed in which the product was tested up to the maximum life cycle by cutting simulated bone.

#### 8 Conclusion

Morpheus Rebellion possesses indications for use and technological characteristics similar to the predicate devices. The performance testing demonstrated that the device is as safe and effective as the predicate and has equivalent performance to the predicate. Therefore, Morpheus Rebellion is substantially equivalent to the predicate.

---

**Source:** [https://fda-staging.innolitics.com/device/K200768](https://fda-staging.innolitics.com/device/K200768)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
