K200761 · Bryan Medical, Inc. · BTR · Nov 25, 2020 · Anesthesiology
Device Facts
Record ID
K200761
Device Name
Tenax Laser Resistant Endotracheal Tube
Applicant
Bryan Medical, Inc.
Product Code
BTR · Anesthesiology
Decision Date
Nov 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tenax® Laser Resistant Endotracheal Tube is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns) or KTP (532 nm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthetic gases or to overcome emergency obstruction of an airway.
Device Story
Tenax Laser Resistant Endotracheal Tube is a cuffed, single-use airway device designed for use during CO2 or KTP laser surgeries. It features an inner aluminum wrapping for laser resistance and an outer atraumatic silicone covering. The device includes dual inflatable cuffs and a pilot balloon containing blue dye to alert clinicians to potential punctures or damage. It is used by physicians in clinical settings to maintain an airway and deliver anesthetic gases. The device is inserted into the trachea; the laser-resistant construction protects the airway from accidental laser strikes during surgery. It is discarded after a single use.
Clinical Evidence
Bench testing only. Performance data included dimensional testing, bend radius, cuff inflation/deflation/burst testing, dye visibility, laser resistance, and connector leakage. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity.
Technological Characteristics
Endotracheal tube with aluminum scattering layer for laser resistance and silicone outer covering. Features dual inflatable cuffs and blue dye pilot balloon indicator. Sterile (EO). Dimensions: 34cm length, 5.0mm-7.5mm ID. Connects to standard anesthesia gas supply. Biocompatibility: Surface device, mucosal membrane, <24h duration.
Indications for Use
Indicated for adults (≥ 22 years) requiring endotracheal intubation during laser surgery (CO2 or KTP) in the upper gastro-respiratory tract. Contraindicated in patients with narrow airways or conditions precluding intubation. Do not exceed specified power densities (10,394 W/cm² for CO2; 5,305 W/cm² for KTP) or use with continuous irradiation > 30 seconds.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 25, 2020
Bryan Medical, Inc. % Ian Broome Regulatory Affairs Consultant AlvaMed, Inc. 1116 Great Plain Avenue, Suite 1 Needham, Massachusetts 02492
Re: K200761
Trade/Device Name: Tenax Laser Resistant Endotracheal Tube Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: Class II Product Code: BTR, BSK Dated: October 23, 2020 Received: October 26, 2020
Dear Ian Broome:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200761
Device Name Tenax® Laser Resistant Endotracheal Tube
Indications for Use (Describe)
The Tenax® Laser Resistant Endotracheal Tube is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns) or KTP (532 nm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthetic gases or to overcome emergency obstruction of an airway.
Type of Use (Select one or both, as applicable)
| <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------|
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### _ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5.0 TRADITIONAL 510(K) SUMMARY FOR TENAX® LASER RESISTANT ENDOTRACHEAL TUBE
#### K200761
#### I. SUBMITTER
Bryan Medical, Inc. 5725 Dragon Way, Suite 300 Cincinnati, OH 45227 Phone: (888) 331-3485, x 700 Fax: (617) 249-0955 Contact Person: lan Broome, AlvaMed, Inc. Date Prepared: November 23, 2020
### II. DEVICE
Name of Device: Tenax® Laser Resistant Endotracheal Tube Common or Usual Name: Endotracheal tube (cuffed) Classification Name: Tracheal tube (21 CFR 868.5730) Requlatory Class: II Product Code: BTR, BSK
#### III. PREDICATE DEVICE
Laser-Shield® II, K993582
### IV. DEVICE DESCRIPTION
The Tenax® Laser Resistant Endotracheal Tube is an endotracheal tube with two inflatable cuffs. It is indicated for use in surgical procedures where intubation is required in the presence of KTP or CO2 lasers.
The device is provided sterile (EO). After use, the device is discarded and disposed of in accordance with local regulations. There are no associated device accessories.
{4}------------------------------------------------
### V. INDICATION FOR USE
The Tenax® Laser Resistant Endotracheal Tube is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns) or KTP (532 nm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthetic gases or to overcome emergency obstruction of an airway.
| Intended Population: | Adults (≥ 22 years old) requiring endotracheal intubation during<br>laser surgery procedures in the upper gastro-respiratory tract. |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Environment of Use: | Trachea/Upper GI Tract |
| Contraindications: | The Tenax® Laser Resistant Endotracheal Tube should not be used in patients with narrow airways, which could restrict ventilation, resulting in excessive elevation of intratracheal pressure. Do not use in any patient suffering from conditions that may preclude endotracheal intubation. Do not use this device for any purpose other than its intended use. To avoid device damage, do not exceed a power density of 10,394 W/cm² when used with a CO2 laser. To avoid device damage, do not exceed a power density of 5,305 W/cm² when used with a KTP laser. Do not use with continuous irradiation lasting longer than 30 seconds. |
{5}------------------------------------------------
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| | Subject:<br>Tenax® Laser Resistant Endotracheal<br>Tube | Predicate Device:<br>Laser-Shield® II |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Bryan Medical, Inc. | Medtronic/Xomed |
| 510(k) Number | K200761 | K993582 |
| Product Code | BTR, BSK | BTR |
| Regulation | 868.5730 | 868.5730 |
| Regulation<br>Description | Tracheal tube | Tracheal tube |
| Common Name | Endotracheal tube | Endotracheal tube |
| Indications for<br>Use | The Tenax® Laser Resistant Endotracheal<br>Tube is intended for endotracheal<br>intubation. It is indicated for use for all<br>types of surgical procedures involving<br>carbon dioxide (10.60 microns) or KTP (532<br>nm) laser use (normal pulsed or continuous<br>beam delivery in the non-contact mode),<br>when endotracheal intubation is required to<br>administer anesthetic gases or to overcome<br>emergency obstruction of an airway. | Xomed Laser-Shield® II is intended for<br>endotracheal intubation. It is indicated for<br>use for all types of surgical procedures<br>involving carbon dioxide (10.60 microns) or<br>KTP (532 nm) laser use (normal pulsed or<br>continuous beam delivery in the non-<br>contact mode), when endotracheal<br>intubation is required to administer<br>anesthesia gases or to overcome<br>emergency obstruction of an airway. |
| Sterility | Sterile (ethylene oxide)<br>SAL 10-6 | Sterile (method unknown) |
| Biocompatibility<br>Classification | Surface Device, Mucosal Membrane,<br>Limited Duration (< 24h) | Surface Device, Mucosal Membrane,<br>Limited Duration (< 24h) |
| Design Features | ● Endotracheal tube with cuffs at distal<br>end<br>● Aluminum wrapping for laser<br>resistance<br>● Outer atraumatic silicone covering<br>● Blue dye within pilot balloons for<br>leak/damage indication | ● Endotracheal tube with cuffs at distal<br>end<br>● Aluminum wrapping for laser<br>resistance<br>● Outer atraumatic polymer covering<br>● Blue dye within pilot balloons for<br>leak/damage indication |
| | Subject: | Predicate Device: |
| | Tenax® Laser Resistant Endotracheal<br>Tube | Laser-Shield® II |
| Connection to<br>Ventilation<br>Source | Standard straight connector to anesthesia<br>gas supply | Standard straight connector to anesthesia<br>gas supply |
| Dimensions | Length: 34cm<br>ID: (6 sizes)<br>5.0mm (model TG-0050) 5.5mm (TG-0055) 6.0mm (TG-0060) 6.5mm (TG-0065) 7.0mm (TG-0070) 7.5mm (TG-0075) | Length: 33cm<br>ID: (9 sizes)<br>4.0mm (model 7060100) 4.5mm (7060150) 5.0mm (7060200) 5.5mm (7060250) 6.0mm (7060300) 6.5mm (7060350) 7.0mm (7060400) 7.5mm (7060450) 8mm (7060500)<br>OD:<br>6.6mm (7060100) |
| | OD:<br>7.9mm (TG-0050) 8.6mm (TG-0055) 9.3mm (TG-0060) 10.0mm (TG-0065) 10.6 (TG-0070) 11.0mm (TG-0075) | 7.3mm (7060150)8.0mm (7060200)8.6mm (7060250)9.0mm (7060300)10.0mm (7060350)10.5mm (7060400)11.0mm (7060450)11.5mm (7060500) |
| Single Use? | Yes | Yes |
{6}------------------------------------------------
Both devices are laser resistant endotracheal tubes. They share an identical indication for use and are intended for use in the same manner. The subject and predicate devices also share the following technological characteristics:
- Laser resistance using an inner aluminum scattering layer, ●
- . Atraumatic outer layer for protection of patient tissue,
- Inflatable cuffs,
- . Use of internal dye as a warning indicator to alert practitioners of device puncture or damage.
{7}------------------------------------------------
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
- Dimensional testing, ●
- Bend radius,
- . Cuff inflation/deflation/burst testing (time and pressure),
- Dye visibility.
- . Laser resistance,
- Tracheal tube dimensions, ●
- . Connector leakage.
- ●
#### Biocompatibility Testing
The biocompatibility evaluation for the Tenax® Laser Resistant Endotracheal Tube was conducted in accordance with the FDA's guidance "International Standard ISO 10993-1 'Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process." The biocompatibility testing included the following tests:
- Cytotoxicity,
- Sensitization, ●
- Irritation, ●
- Acute Systemic Toxicity, ●
- Material-Mediated Pyrogenicity. .
The Tenax® Laser Resistant Endotracheal Tube is considered tissue-contacting for a duration of less than 24 hours. The device's materials are well-characterized for medical use.
#### VIII. CONCLUSIONS
The non-clinical test data support the safety of the device and the verification and validation testing data demonstrate that the Tenax® Laser Resistant Endotracheal Tube shall perform as intended in the specified use conditions. The subject device and the predicate device have the same intended use, and the technological differences do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.