DB-Orthopedics Bone Void Filler is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. DB-Orthopedics is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
Device Story
DB-Orthopedics is a moldable, injectable, biocompatible calcium phosphate bone void filler; supplied as a two-component kit (calcium-phosphate powder and mixing solution). Components are mixed by the clinician prior to implantation to fill non-structural bony voids or gaps in the skeletal system (extremities, spine, pelvis). The material resorbs and is replaced by bone during the healing process. Used in clinical/surgical settings; intended for single use.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including working/setting time, Ca/P ratio (1.5), heavy metal analysis (ASTM F1185-03), pH profile, FTIR, XRD, temperature profile, solubility/dissolution, tensile testing, dimensional stability, and SEM microstructure analysis.
Technological Characteristics
Calcium phosphate bone void filler; supplied in 3cc, 5cc, and 10cc kits. Sterilized via gamma radiation (ISO 11137-2:2013, VDMax25). Heavy metal analysis per ASTM F1185-03. Material composition and manufacturing processes identical to reference device K182742.
Indications for Use
Indicated for patients with bony voids or gaps in the skeletal system (extremities, spine, pelvis) resulting from surgical procedures or traumatic injury. Contraindicated for voids or gaps intrinsic to the stability of the bony structure.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Dimensional Bioceramics, LLC % Patsy Trisler Regulatory Consultant Trisler Consulting 7949 Beaumont Green East Dr Indianapolis, Indiana 46250
Re: K200752
Trade/Device Name: DB-Orthopedics Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: March 17, 2020 Received: March 23, 2020
Dear Patsy Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
April 22, 2020
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jesse Muir, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200752
Device Name
DB-Orthopedics
Indications for Use (Describe)
DB-Orthopedics Bone Void Filler is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects created from traumatic injury to the bone. DB-Orthopedics is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary – K200752
| I. SUBMITTER | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Submitter Name: | Dimensional Bioceramics, LLC | |
| Submitter<br>Address: | 2161 Delaware Ave., Suite A<br>Santa Cruz, CA 95060 | |
| Contact Person:<br>Phone Number: | Duran N. Yetkinler, M.D., Ph.D.<br>408-757-6603 | |
| Date Prepared: | March 17, 2020 | |
| II. DEVICE | | |
| Trade Name: | DB-Orthopedics Bone Void Filler | |
| Common Name | Bone Void Filler | |
| Classification<br>Name Number<br>Product Code<br>Regulatory<br>Class | Resorbable calcium salt bone void filler device<br>21 CFR 888.3045<br>MQV<br>Class 2 | |
| III. PREDICATE DEVICE | | |
| Primary<br>Predicate<br>Device: | K051123, Callos™ Bone Void Filler, Skeletal Kinetics | |
| Reference<br>Device: | K182742, DB-Cranial Bone Void Filler, Dimensional Bioceramics, LLC | |
| IV. INDICATIONS FOR USE<br>This submission expands the indications for use of the DB-Cranial 510(k) (Reference<br>Device K182742) clearance to orthopedics. | | |
| Indications for<br>Use Statement | DB-Orthopedics Bone Void Filler is indicated to fill bony voids or<br>gaps of the skeletal system (i.e. extremities, spine, pelvis). These<br>defects may be surgically created osseous defects or osseous<br>defects created from traumatic injury to the bone. DB-Orthopedics<br>is indicated only for bony voids or gaps that are not intrinsic to the<br>stability of the bony structure. The product provides a bone void<br>filler that resorbs and is replaced by bone during the healing<br>process. | |
| V. DEVICE DESCRIPTION | | |
| Device<br>Identification | DB-Orthopedics Bone Void Filler is a moldable, injectable and<br>biocompatible calcium phosphate bone void filler. The DB-<br>Orthopedics kit is comprised of two components: a calcium-<br>phosphate powder and a mixing solution in premeasured quantities,<br>which will be mixed together prior to implantation. | |
| Technological<br>Characteristics | The 3 cc, 5 cc, and 10 cc DB-Orthopedics Bone Void Filler kits are<br>provided sterile to SAL of 10-6 and are for single use only. | |
| VI. PERFORMANCE TESTING | | |
| Test | Test Method Summary | |
| Working Time<br>In Vitro | Ensures sufficient manipulation time is provided while also ensuring<br>cement setting-times are met in the operative theater. | |
| Setting Time | This test measures the time for a bone void filler to set in simulated<br>physiologic conditions. | |
| Ca to P Ratio | This test determines CA/P ratios via ICP-MS. DB-Orthopedics has a<br>ratio of 1.5. | |
| Heavy Metal<br>Analysis | Samples are analyzed for trace heavy metal content using ICP-MS,<br>according to ASTM F1185-03 | |
| pH Profile | Examines effects of the device on pH surrounding the implanted<br>device. pH is measured in physiologic buffer solutions surrounding<br>curing cements. | |
| FTIR Analysis | This test identifies the chemical composition of subject and predicate<br>device following curing in simulated physiologic conditions. | |
| Crystallographic<br>Analysis | XRD analysis is performed with samples set in simulated physiologic<br>conditions for specified times and evaluated using powder x-ray<br>diffraction and compared against known mineralogic standards. | |
| Temperature<br>Profile | Device samples are tested in simulated physiologic solutions to<br>measure temperature of curing cement. | |
| Solubility and<br>Dissolution | Test samples are cured and incubated at simulated physiological<br>conditions for a specified time and measured for solubility and<br>dissolution. | |
| Tensile Testing | Test samples were mixed and cured for 24 hours at simulated<br>temperature and pH. Tensile testing was performed using a<br>mechanical tester and load at sample breakage was recorded and<br>compared. | |
| Dimensional<br>Stability | Dimensional stability is measured to establish that the bone void<br>fillers maintain shape and do not dissolve in an untimely manner. | |
| Physical Form | Test samples were imaged by SEM to determine microstructure.<br>Testing confirmed hydroxyapatite crystal formation. | |
| Biocompatibility | No biocompatibility studies were needed to demonstrate substantial<br>equivalence. No changes to the product were made from the<br>original submission. | |
| Pyrogenicity | Pyrogenicity (LAL) testing is routinely performed on the product. | |
| Sterilization and<br>Shelf Life | The sterilization method is gamma radiation. Sterilization validation<br>is based on ISO 11137-2:2013 (VDMax25).<br>DB-Orthopedics Bone Void Filler will be labeled with a shelf life of 30<br>months. | |
| Animal Testing | No animal studies were needed to demonstration substantial equivalence. | |
| Clinical Testing | This product category does not require human clinical testing. | |
{4}------------------------------------------------
I
{5}------------------------------------------------
### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
DB-Orthopedics Bone Void Filler's intended use and critical specifications are identical to the predicate device, Skeletal Kinetics' Callos™ Bone Void Filler, K051123.
Further, this product is composed of the identical material, and is manufactured and packaged by the identical processes and in the identical facility, as the recently cleared Reference Device, K182742.
There are no notable differences in comparison to the predicate device, therefore, no questions related to safety and effectiveness were raised.
## VIII. CONCLUSIONS
Based on the comparisons provided and the data submitted in this 510(k), it can be concluded the DB-Orthopedics is substantially equivalence to Callos™ predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.