K200736 · I.T.S. Group S.R.L. · GEX · Jun 26, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200736
Device Name
Stardust Med
Applicant
I.T.S. Group S.R.L.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
STARDUST MED is indicated for use in aesthetic, surgical and cosmetic applications and in selective treatments required in the medical of dermatology and general and plastic surgery. STARDUST MED is indicated for hair removal and permanent hair reduction. Permanent hair reduction is defined as the long term, stable reductions in the number of hairs when measured at 6, 9 and 12 months after the completion of a treatment regime
Device Story
STARDUST MED is a laser-emitting device for aesthetic and surgical dermatology/plastic surgery. It consists of a main console and an 808 nm diode laser handpiece. Operated by a clinician, the handpiece is applied in contact with the skin and triggered via footswitch or finger switch. A microprocessor-based system controller monitors and directs all system functions and the graphic user interface. The device delivers laser energy to the target area to achieve hair removal and permanent hair reduction. Clinical benefit is realized through long-term, stable reduction in hair count.
Clinical Evidence
Bench testing only. Provided electrical safety and EMC testing per IEC 60601-2-22 and IEC 60601-1-2, and software verification and validation per FDA guidance.
Technological Characteristics
Laser surgical instrument (21 CFR 878.4810). Energy source: 808 nm diode laser. Output: up to 40 J/cm² fluence, 10-300 ms pulse duration, 1-10 Hz repetition rate. Spot size: 1.0x1.0 cm. Operation: CW/pulsed burst. Microprocessor-based system controller with GUI. Power: 110-240 Vac, 50/60Hz. Contact-based handpiece.
Indications for Use
Indicated for hair removal and permanent hair reduction in patients requiring aesthetic, surgical, or cosmetic dermatological and plastic surgery treatments.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
June 26, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
I.T.S. Group S.R.L. % Chiara Violini Consultant Endo Engineering SRL Via Del Consorzio, 41 Falconara Marittima, Ancona 60015 Italy
Re: K200736
Trade/Device Name: Stardust Med Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 11, 2020 Received: March 23, 2020
Dear Chiara Violini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post-marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K200736
Device Name STARDUST MED
#### Indications for Use (Describe)
STARDUST MED is indicated for use in aesthetic, surgical and cosmetic applications and in selective treatments required in the medical of dermatology and general and plastic surgery.
STARDUST MED is indicated for hair removal and permanent hair reduction.
Permanent hair reduction is defined as the long term, stable reductions in the number of hairs when measured at 6, 9 and 12 months after the completion of a treatment regime
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for NTS Group. The logo consists of a blue circle with the letters 'nts' in white, followed by the word 'group' in black. The 'nts' is written in a stylized font, with the 't' and 's' connected.
#### 510(k) Summary ഗ
### Introduction:
This document contains the 510(k) Summary for the STARDUST MED device. The content of this summary is based on the requirements of 21 CFR 807.92(c).
| Applicant/<br>Manufacturer<br>Name and Address: | I.T.S. GROUP s.r.l.<br>Via del Cerchio, 1<br>40012 Calderara di Reno (BO)<br>Italy |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Chiara Violini<br>Consultant<br>Email: c.violini@endoengineering.it<br>Phone: +39-071-9156048<br>Fax: +39-071-0971883 |
| Date Prepared: | 11/03/2020 |
| Device Name: | STARDUST MED |
| Classification: | Class II |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery |
| Regulation number: | 21 CFR 878.4810 |
| Classification product code: | GEX |
| Predicate Device: | MT ONE DIAMOND- M&T SRL - K191942 |
# Description of the device:
STARDUST MED is a laser emitting device that is operated with handpiece in contact with the skin.
STARSDUST MED comprises a main console unit and a handpiece that is triggered by means of footswitch or a finger switch.
A microprocessor based system controller is used to monitor and direct all the system function and the graphic user interface.
The main console can be connected to the following handpiece:
- -808 nm Laser diode.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for HTS Group. The logo features a blue circle on the left side with the letters 'HTS' in white. To the right of the circle, the word 'group' is written in a thin, black sans-serif font. The logo is simple and modern, with a clean design.
# STARDUST MED
# Intended Use:
STARDUST MED is indicated for use in aesthetic, surgical and cosmetic applications and in selective treatments required in the medical of dermatology and general and plastic surgery.
STARDUST MED is indicated for hair removal and permanent hair reduction.
Permanent hair reduction is defined as the long term, stable reductions in the number of hairs when measured at 6, 9 and 12 months after the completion of a treatment regime
# Comparison of Technological Characteristics:
STARDUST MED utilizes technological characteristics (energy source, control mechanisms) and specifications that are similar to the Predicate Device.
{5}------------------------------------------------
| Image: ITsgroup logo | STARDUST MED | I.T.S. GROUP s.r.l.<br>Via del Cerchio, 1<br>40012 Calderara di Reno<br>(BO) - Italy |
|----------------------|--------------|--------------------------------------------------------------------------------------|
|----------------------|--------------|--------------------------------------------------------------------------------------|
| Specifications | Predicate Device | Device object of 510(k) |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name<br>(K number) | MT ONE DIAMOND<br>K191942 | STARDUST MED |
| Submitter | M&T S.r.l. | ITS GROUP S.r.l. |
| Product Code | GEX | GEX |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Regulatory Class | II | II |
| Product Picture | Image: MT ONE DIAMOND | Image: STARDUST MED |
| Mains | 90-110 Vac / 190-230 Vac, 50 Hz/ 60 Hz | 110-240 Vac, 50/60Hz |
| Intended use | MT ONE DIAMOND and its handpiece is indicated<br>for use in aesthetic, surgical and cosmetic<br>applications and in selective treatments required<br>in the medical of dermatology and general and<br>plastic surgery. MT ONE DIAMOND with HR808<br>nm laser handpiece is indicated for hair removal<br>and permanent hair reduction. Permanent hair<br>reduction is defined as the long term, stable<br>reductions in the number of hairs when<br>measured at 6, 9 and 12 months after the<br>completion of a treatment regime | STARDUST MED is indicated for use in aesthetic,<br>surgical and cosmetic applications and in<br>selective treatments required in the medical of<br>dermatology and general and plastic surgery.<br>STARDUST MED is indicated for hair removal and<br>permanent hair reduction. Permanent hair<br>reduction is defined as the long term, stable<br>reductions in the number of hairs when<br>measured at 6, 9 and 12 months after the<br>completion of a treatment regim |
| Wavelength | 808 nm | 808 nm |
| Fluence | Up to 90 J/cm² | Up to 40 J/cm² |
| Pulse duration | 1 to 200 ms | 10 to 300 ms |
| Mode | Cw/pulsed burst | Cw/pulsed burst |
| Repetition rate | Up to 10 Hz | 1 to 10 Hz |
| Spot size | 1.68 cm2 | 1.0x1.0 cm |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image contains the logo for NTS Group. The logo features a blue circle on the left side with the letters "NTS" in white. To the right of the circle, the word "group" is written in a thin, sans-serif font. The overall design is clean and modern.
# Performance data:
The following performance data are provided in support of the substantial equivalence determination:
# Safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the STARDUST MED device.
The system complies with the IEC 60601-2-22 standards for safety and the IEC 60601-1-2 standard for EMC.
# Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
# Comparison of Intended Use:
STARDUST MED device's Intended Use is the same Intended Use of its predicate device.
# Substantial Equivalence:
The STARDUST MED device has the same intended use, and similar technological characteristics and specifications, compared to the predicate device. Non-clinical testing supports that the device's technological characteristics and performance do not raise new questions regarding the device's safety and efficacy for its intended use. The STARDUST MED device is considered substantially equivalent to the predicate K191942 device..
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.