K200710 · Andon Health Co, Ltd. · FLL · Oct 1, 2020 · General Hospital
Device Facts
Record ID
K200710
Device Name
Infrared PT2L Digital No-Touch Thermometer
Applicant
Andon Health Co, Ltd.
Product Code
FLL · General Hospital
Decision Date
Oct 1, 2020
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infrared Digital No-Touch Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. It can be used by consumers in the household environment and by healthcare providers.
Device Story
Hand-held, battery-operated, non-contact infrared thermometer; measures natural thermal radiation from central forehead. User positions probe ≤ 3cm from forehead; thermopile sensor detects infrared radiation; electrical signal amplified and processed to calculate oral-equivalent temperature. Output displayed on LCD; device provides audible beep and colored screen alerts for temperature ranges. Used by consumers or healthcare providers in home or clinical settings. Benefits include non-invasive, rapid temperature assessment without skin contact.
Clinical Evidence
Clinical accuracy validation conducted per ASTM E1965-98 with 130 subjects: 40 infants (0-1 year), 40 children (1-5 years), and 50 subjects (>5 years). Results confirmed compliance with ASTM E1965-98 (2016) requirements.
Technological Characteristics
Thermopile sensor; infrared radiation detection; 2x1.5V AAA battery power; LCD display; ABS and PMMA materials; non-contact measurement distance ≤ 3cm. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 80601-2-56 (clinical thermometers), and ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for intermittent forehead body temperature measurement in all ages; suitable for household and clinical use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are placed side by side.
October 1, 2020
Andon Health Co., Ltd. Liu Yi President No. 3 Jin Ping Street, Ya An Road, Nankai District Tianjin, 300190 China
Re: K200710
Trade/Device Name: Infrared PT2L Digital No-Touch Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: Class II Product Code: FLL Dated: August 24, 2020 Received: September 1, 2020
Dear Liu Yi:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sapana Patel -S
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200710
#### Device Name
iHealth PT2L Infrared Digital No-Touch Thermometer
#### Indications for Use (Describe)
The Infrared Digital No-Touch Thermometer is intennittent measurement of body temperature from the forehead on people of all ages. It can be used by consumers in the household environment and by healthcare providers.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary-K200710
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
#### 1.0 submitter's information
| Name: | Andon Health Co., Ltd. |
|----------------------|-------------------------------------------------------------------------|
| Address: | No 3, Jinping Street Ya An Road, Nankai District, Tianjin<br>P.R. China |
| Phone number: | 86-22-87611660 |
| Fax number: | 86-22-6052 6162 |
| Contact: | Liu Yi |
| Date of Preparation: | 8/21/2020 |
#### 2.0 Device information
Trade name: iHealth PT2L Infrared Digital No-Touch Thermometer Infrared forehead thermometer Common name: Regulation name: Clinical Electronic Thermometer
#### 3.0 Classification
Production code: FLL Regulation number: 880.2910 Classification: II Panel: General Hospital
#### 4.0 Predicate device information
Manufacturer: Guangzhou Berrcom Medical Device Co., Ltd. Device: Infrared Thermometer Model MD-H30 510(k) number: K191570
#### 5.0 Indications for Use
The Infrared Digital No-Touch Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. It can be used by consumers in the household environment and by healthcare providers.
{4}------------------------------------------------
### 6.0 Device description
The iHealth PT2L Infrared Digital No-Touch Thermometer is hand-held, reusable, battery operated device, which can measure human body forehead skin temperature without touching patient's skin. Its operation is based on measuring the natural thermal radiation from the central forehead. Put the probe of the thermometer at a measurement distance less than or equal to 1.18 inch (3cm) without touching the skin at the center of the patient's forehead. It uses a thermopile sensor to measure the patient's forehead temperature and convert it to the oral equivalent. Pressing the activation button to start the measurement of proposed infrared radiation. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The final measured temperature will be appeared on a LCD display.
| Item | Subject Device<br>(K200710) | Predicate Device<br>(K191570) | Comparison |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Models | PT2L | MD-H30 | Different |
| Product code | FLL | FLL | Same |
| Regulation number | 21 CFR 880.2910 | 21 CFR 880.2910 | Same |
| Manufacturer | Andon Health Co., Ltd | Guangzhou Berrcom<br>Medical Device Co., Ltd. | Different |
| Indications for Use | The Infrared Digital No-Touch<br>Thermometer is intended for<br>the intermittent measurement<br>of body temperature from the<br>forehead on people of all ages.<br>It can be used by consumers<br>in the household environment<br>and by healthcare providers. | The infrared thermometer is<br>intended for the intermittent<br>measurement of body<br>temperature from the<br>auditory canal or central<br>forehead skin surface on<br>people of all ages. It can be<br>used by consumers in the<br>household environment and<br>by healthcare providers. | Different |
| Probe contact type | Non-contact;<br>measurement distance: ≤ 1.18<br>in(3 cm) | Skin contacting | Different |
| Sensor | Thermopile | Thermopile | Same |
| Measurement method | Infrared radiation detection | Infrared radiation detection | Same |
| Measurement Site | Forehead | Forehead or Ear | Different |
| Reference site | Oral | Oral | Same |
# 7.0 Summary comparing technological characteristics with predicate device
{5}------------------------------------------------
| Display type | LCD | LCD | Same |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|------------|
| Dimension | 141mm×32mm×48mm | Not known | different |
| Other feature | Beep when get measurement<br>colored screen indicates the<br>temperature is outside of<br>'normal' or 'high' temperature<br>ranges | Beep when get<br>measurement<br>Probe cover provided | different |
| Measuring range | 34-43° C (93.2-109.4° F) | 32- 43° C (89.6- 109.4° F) | different |
| Accuracy | ±0.5°F (±0.3°C) for 34-<br>43° C (93.2- 109.4° F) | ±0.4° F (±0.2° C) within<br>95~107.6°F (35~42° C),<br>±0.5°F (±0.3° C) for<br>other range. | Different |
| Temperature unit | °C or °F | °C or °F | Same |
| Display resolution | 0.1°F (0.1°C) | 0.1 °F (0.1 °C) | Same |
| Power<br>requirements | 2X1.5V AAA battery | 2X1.5V AAA battery | Same |
| Operation<br>environment | 15-40°C, ≤95%RH,<br>70-106kPa | 10°C~40°C; ≤95% RH | Different, |
| Storage<br>environment | -25-55°C, ≤95%RH,<br>70-106kPa | -20~55°C; ≤95%RH | Different, |
| Materials of<br>skin-contacting<br>components | ABS and PMMA | ABS | Different |
| biocompatibility | Comply with ISO 10993-1 | Comply with ISO 10993-1 | Same |
| electrical safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
The manufacturer of the subject device is different from the predicate device, so the model is also different, there is slightly difference in the indication for use because the subject device only take measurement from the forehead skin while the predicate device take measurement from both forehead skin and auditory canal. Moreover, the predicate device is skin contacting when taking measurement while the subject device is non-contact device. The software validation and clinical accuracy test demonstrated the difference does not raise any new performance questions.
Dimension of the subject device is different from the predicate device, but the electrical safety and EMC test and the performance test demonstrated the difference does not raise any new performance questions.
The subject device has no probe cover provided, while the predicate device will provide probe cover, and the subject device has function of flashing/colored screen that indicates the temperature is outside of 'normal' or
{6}------------------------------------------------
'high' temperature ranges, but the performance test of the subject device demonstrated the difference does not raise any new performance questions.
The Measuring range and accuracy is different from the predicate device, the performance test and the clinical accuracy test demonstrated the difference does not raise any new performance questions. The measuring range and accuracy is within the ASTM 1965-98 standard for Skin IR thermometers.
The operation temperature range of the proposed device PT2L is narrow than the predicate device, and the storage humidity range of the proposed device PT2L is also wide than the predicate device, but the performance test demonstrated the difference does not raise any new performance questions.
The material of skin-contacting components is different, the biocompatibility report confirms that there is no new risk for the proposed device.
### 8.0 Discussion of non-clinical and clinical test performed
Non-clinical tests were conducted to verify that the proposed device meet all design specifications in order to demonstrate that it is Substantially Equivalent to the predicate device.
### Non-clinical Tests
Electrical Safety test was performed according to IEC 60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
EMC test was performed according to 60601-1-2:2014. Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility requirements and tests
Performance test was performed according to ISO 80601-2-56 , Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. [Including: Amendment 1 (2018)]. is the applicable standards for this device.
The software verification and validation were conducted according to FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", dated May 11, 2005.
{7}------------------------------------------------
Biocompatibility test (including cytotoxicity, sensitization, irritation) was leveraged from their own reference device K 2 0 0 5 3 1 . The tests performed on the reference device according to (1) ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, (2)ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity and (3) ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
### Reprocessing: Cleaning and Disinfection
The subject device is a non-sterile device. The cleaning validation was performed according to the Instructions for Use.
# Clinical Accuracy Validation Test
Clinical investigation and data analysis have performed according to ASTM F1965-98. The test report shows result of 40 subjects in infants group (0-1 year), 40 subjects in children group (1-5 years) and 50 subjects in group (>5 years) (Total 130 subjects) are within acceptable range. Based on the test results, it demonstrates the iHealth PT2L Infrared Digital No-Touch Thermometer complies with the requirement of ASTM E1965-98 (2016).
### 9.0 Comparison to the predicate device and the conclusion
Basing on the performance testing, comparison and analysis above, the subject device iHealth PT2L Infrared Digital No-Touch Thermometer is substantially equivalent to the predicate device .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.