The FIREBIRD SI Fusion System is intended for fixation of sacroiliac joint disruptions, and intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.
Device Story
FIREBIRD SI Fusion System is a temporary, multi-component orthopedic fixation system used for sacroiliac joint fusion and stabilization. The system consists of cannulated titanium alloy (Ti-6Al-4V ELI) screws featuring fenestrations for bone graft packing; 11mm and 12mm screws include 3D-printed porous mid-shaft regions for osseointegration. The system is used by surgeons in a clinical/OR setting to treat sacroiliac joint disruptions and degenerative sacroiliitis. The device is implanted to provide mechanical stability to the joint, facilitating fusion. The addition of 9mm screws expands the available sizing options for the previously cleared system.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical performance was validated via static and dynamic cantilever beam testing, static torsion testing, and static axial pull-out testing in accordance with ASTM F2193 and ASTM F543 standards.
Indicated for patients requiring fixation of sacroiliac joint disruptions or fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 2, 2020
Orthofix Inc. Jacki Koch Senior Regulatory Affairs Specialist 3451 Plano Parkway Lewisville, Texas 75056
Re: K200696
Trade/Device Name: FIREBIRD SI Fusion System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR Dated: March 16, 2020 Received: March 17, 2020
Dear Ms. Koch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'Neill, MBE Assistant Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200696
Device Name FIREBIRD SI Fusion System
Indications for Use (Describe)
The FIREBIRD SI Fusion System is intended for fixation of sacroiliac joint disruptions, and intended for sacrolliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.
Type of Use (Select one or both, as applicable)
| <div><span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CER 801 Subpart D)</div> |
|-----------------------------------------------------------------------------------------------------------|
| <div>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in bold, black letters. A small registration mark is visible to the upper right of the word "ORTHOFIX".
FIREBIRD SI Fusion System
# 510(k) SUMMARY
# FIREBIRD SI Fusion System (Formally SambaScrew 3D SI Fusion System)
| 510(k) Owner Information | |
|-----------------------------------------------------|-------------------------------------------------------------|
| Name:<br>Address: | Orthofix Inc.<br>3451 Plano Parkway<br>Lewisville, TX 75056 |
| Telephone Number:<br>Fax Number:<br>Email: | 214-937-2100<br>214-937-3322<br>jackikoch@orthofix.com |
| Registration Number: | 2183449 |
| Contact Person: | Jacki Koch, Senior Regulatory Affairs Specialist |
| Date Prepared: | March 16, 2020 |
| Name of Device<br>Trade Name / Proprietary<br>Name: | FIREBIRD SI Fusion System |
| Common Name: | Sacroiliac Joint Fixation Bone Screw |
| Product Code: | OUR |
| Regulatory Classification: | Class II – 21 CFR § 888.3040 |
| Review Panel: | Orthopedic Device Panel |
| Primary Predicate: | K183342 - SambaScrew 3D SI Fusion System |
| Reference Device: | K121148 - SAMBA Screw System |
Reason for 510(k) Submission: Due to advancements in surgeon requests, Orthofix is submitting this Special 510(k) premarket notification for the addition of the following device below to the previously cleared SambaScrew 3D SI Fusion System (K183342) now branded as the FIREBIRD SI Fusion System:
- . Addition of new 9mm FIREBIRD SI Screws
## Device Description
The FIREBIRD SI Fusion System (formally SambaScrew 3D SI Fusion System) is a temporary multiple component system consisting of non-sterile instruments, sterile 11mm and 12mm FIREBIRD SI Screws and non-sterile 9mm FIREBIRD SI Screws. The FIREBIRD SI Screws are manufactured from medical-grade titanium alloy (Ti-6AI-4V ELI) with lengths ranging from 25mm to 70mm.
{4}------------------------------------------------
K200696 - Page 2 of 3
Image /page/4/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, bold letters. The "®" symbol is located to the upper right of the "X" in "ORTHOFIX".
FIREBIRD SI Fusion System
The 11mm FIREBIRD SI Screw is a cannulated screw featuring multiple fenestrations along the shaft, with a tapered proximal end and dual-pitch threads. The 12mm FIREBIRD SI Scew is a cannulated screw featuring multiple fenestrations along the shaft, with a single pitch thread on the proximal and distal ends. The 11mm and 12mm FIREBIRD SI Screws are 3D printed with a mid-shaft porous region. The porous titanium region has open macroscopic 3D pores with a microscopic roughened surface. The 9mm FIREBIRD SI Screw features multiple fenestrations along the shaft and maintains a single pitch thread along the proximal and distal ends of the screw.
The FIREBIRD SI Fusion System allows for packing of autograft and allograft materials.
### Intended Use / Indications for Use
The FIREBIRD SI Fusion System is intended for fixation of sacroiliac joint disruptions, and intended for sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.
### Summary of the Technological Characteristics of the Device Compared to the Selected Predicate Devices
The technological characteristics of the FIREBIRD SI Fusion System are similar to the predicate (K183342 SambaScrew 3D SI Fusion System) in terms of intended use, indication for use and patient population. The subject 9mm FIREBIRD SI Screws are similar to the reference device (K121148 SAMBA Screw System) in terms of design and performance characteristics.
The technical characteristics of the subject device present no significant differences between the predicate and reference devices which would adversely affect the use of the product.
#### PERFORMANCE DATA - Summary of Non-Clinical Test Conducted for Determination of Substantial Equivalence
Mechanical testing consisting of Static and Dynamic cantilever beam Test. Static Torsion Test and Static Axial Pull-Off Test were conducted in accordance to ASTM F2193 Standard Specifications and Test Methods for Components Used in the Surgical Fixation of the Spinal Skeletal System and in accordance to ASTM F543 Standard Specification and Test Methods for Metallic Medical Bone Screws.
| Subject<br>Device | Mechanical Testing | Appendix |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| | Static cantilever beam test per<br>ASTM F2193-18a Standard<br>Specifications and Test Methods<br>for Components Used in the<br>Surgical Fixation of the Spinal<br>Skeletal System | Appendix B: Static Cantilever Beam Testing<br>10117508AA |
| 9mm<br>FIREBIRD<br>SI Screws | Dynamic cantilever beam test per<br>ASTM F2193-18a Standard<br>Specifications and Test Methods<br>for Components Used in the<br>Surgical Fixation of the Spinal<br>Skeletal System | Appendix C: Dynamic Cantilever Beam<br>Testing 10117509AA |
| | Static torsion test per ASTM F543-<br>17 Standard Specification and<br>Test Methods for Metallic Medical<br>Bone Screws | Appendix D: Static Torsion Testing<br>10117507AA |
#### Table 1: Mechanical Performance Testing
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, sans-serif font. The "R" in "ORTHOFIX" has a registered trademark symbol next to it.
FIREBIRD SI Fusion System
| | Static axial pullout test per ASTM<br>F543-17 Standard Specification<br>and Test Methods for Metallic<br>Medical Bone Screws | Appendix E: Static Axial Pullout Testing<br>10117506AA |
|--|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
|--|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
## Conclusions
The addition of the 9mm FIREBIRD SI Screws to the previously cleared FIREBIRD SI Fusion System (formally known as SambaScrew 3D SI Fusion System) does not change the previously cleared (K183342) indications for use, intended use, contraindications, warnings or precautions. Furthermore, the surgical approach and implantation technique remains the same as previously cleared (K183342).
The performance and technological characteristics of the subject device are similar to the reference device (K121148) in terms of design, materials and performance characteristics.
There are no significant differences between the subject device, predicate device and reference device, which would adversely affect the use of the product.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.