← Product Code [GEX](/productcode/GEX) · K200659

# Dermalux Tri-Wave MD (K200659)

_Aesthetic Technology, Ltd. · GEX · May 9, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200659

## Device Facts

- **Applicant:** Aesthetic Technology, Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** May 9, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Derrmalux® Tri-Wave MD 's use of the red, blue and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions. The blue light (415nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris. The red light (633nm wavelength) is generally indicated for treatment of superficial, benign, vascular and pigmented lesions. The near-infrared light (830nm wavelength) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

## Device Story

Dermalux Tri-Wave MD is a floor-standing, prescription-use phototherapy device. It utilizes 1440 LEDs across 4 adjustable panels to emit narrow-band light at 415nm (blue), 633nm (red), and 830nm (near-infrared). Operated by clinicians in a clinical setting, the device is controlled via a touch screen interface to deliver specific energy doses (Joules/cm2) to the face or body. The device transforms electrical power into therapeutic light energy to treat dermatological conditions, manage pain, and increase local circulation. Clinicians use the device to provide targeted phototherapy; the output directly affects tissue through light-based energy delivery, potentially benefiting patients by reducing acne, treating skin lesions, and alleviating musculoskeletal discomfort.

## Clinical Evidence

Bench testing only. No clinical or animal studies were performed. Performance was verified through compliance with international and FDA-recognized consensus standards, including EN/IEC 60601-1 (safety), EN/IEC 60601-1-2 (EMC), EN/IEC 62304 (software), EN/IEC 62471 (photobiological safety), IEC 60601-2-57 (non-laser light source safety), IEC 62366-1 (usability), and EN/IEC 60601-1-6 (usability).

## Technological Characteristics

Floor-standing unit with 4 adjustable LED panels (1440 total LEDs). Wavelengths: 415nm±5nm (blue), 633nm±5nm (red), 830nm±5nm (NIR). Power: 100-240Vac. Interface: Touch screen. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62471, IEC 60601-2-57, IEC 62366-1, IEC 60601-1-6. Software life cycle per IEC 62304.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Phototherapy System ([K190938](/device/K190938.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".

Aesthetic Technology Ltd. % Richard Hamer US Agent Richard Hamer Associates LLC 705 Spring Lakes Blvd Bradenton, Florida 34210

Re: K200659

Trade/Device Name: Dermalux Tri-Wave MD Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 9, 2020 Received: March 12, 2020

Dear Richard Hamer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin Kejing Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K200659

Device Name

Derrmalux® Tri-Wave MD

#### Indications for Use (Describe)

The Derrmalux® Tri-Wave MD 's use of the red, blue and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions.

The blue light (415nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.

The red light (633nm wavelength) is generally indicated for treatment of superficial, benign, vascular and pigmented lesions.

The near-infrared light (830nm wavelength) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

| Type of Use (Select one or both, as applicable)                                               |                                                                                                                                           |
|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 16px;"> <span style="font-family: DejaVu Sans, sans-serif;"> <span style="color: black;">☑</span> </span> </span> |
| Over-The-Counter Use (21 CFR 801 Subpart C)                                                   | <span style="font-size: 16px;"> <span style="font-family: DejaVu Sans, sans-serif;"> <span style="color: black;">☐</span> </span> </span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summarv

### I. ADMINISTRATIVE

#### Submitter:

Aesthetic Technology Ltd. 211 Europa Blvd. Warrington, Cheshire WA5 7TN United Kingdom +44 (0) 845 689 1789

Contact Person: Dale Needham

Date of Preparation: May 6, 2020

### II. DEVICE NAME

Proprietarv Name: Dermalux® Tri-Wave MD

Common Name: Powered Laser Surgical Instrument

Classification Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology

Regulation Number: 21 CFR §878.4810

Regulatory Class: Class II

Product Code: GEX

### III. PREDICATE DEVICE

Phototherapy System; K190938; Shanghai Apolo Medical Technology Co, Ltd.

### IV. DEVICE DECRIPTION

The Dermalux® Tri-Wave MD is a floor standing Class II medical device which emits specific wavelengths of low level, narrow band light for the treatment of dermatological conditions. The wavelengths used in the Tri-Wave MD system are Blue 415nm, Red 633nm and Near Infrared 830nm.

The device consists of a main body that contains the power supply, power switch, touch screen control panel, and a 4 panel adjustable treatment head that contains Light Emitting Diodes ena

{4}------------------------------------------------

bling treatment of the face and the body. The panels can be easily adjusted for focused intensity treatments and application to larger body areas.

## V. INDICATIONS FOR USE

The Dermalux® Tri-Wave MD's use of the red, blue and infrared regions of the spectrum is intended to emit energy to treat dermatological conditions.

The blue light (415nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.

The red light (633nm wavelength) is generally indicated for treatment of superficial, benign, vascular and pigmented lesions.

The near infrared light (830nm wavelength) is generally indicated for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.

| Parameter                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Device (K190938)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product name                   | Dermalux® Tri-Wave MD                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Photodynamic Therapy System (HS-770)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product code                   | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Regulation No.                 | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Class                          | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Intended use                   | Use of the red, blue and near-infrared regions of the spectrum to emit energy to treat dermatological conditions                                                                                                                                                                                                                                                                                                                                                                            | Use of the red, blue and infrared regions of the spectrum to emit energy to treat dermatological conditions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for use            | Blue light (415nm wavelength): Treatment of dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.<br>Red light (633nm wavelength: Treatment of superficial, benign, vascular and pigmented lesions.<br>NIR (830nm wavelength): Temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied. | The blue light (415nm wavelength) is generally indicated to treat dermatological conditions and specifically indicated to treat moderate inflammatory acne vulgaris.<br>The red light (630nm wavelength) is generally indicated for treatment of superficial, benign vascular and pigmented lesions.<br>The infrared light (835nm wavelength) is generally used for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation where applied. |
| Parameter                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Predicate Device (K190938)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Power supply                   | 100-240Vac, 50/60Hz, 4.6 -1.85A, 460-<br>430W.                                                                                                                                                                                                                                                                                                                                                                                                                                              | AC 100-240V 50/60Hz 10A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Wavelength                     | Red light: 633nm±5nm<br>Blue light: 415nm±5nm<br>NIR light: 830nm±5nm                                                                                                                                                                                                                                                                                                                                                                                                                       | Red light: 630nm±15nm<br>Blue light: 415nm±15nm<br>IR light: 835nm±15nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Panels Type                    | 4 Panels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 3 Panel: 180 EA LEDs 4 Panel: 240 EA LEDs              The panels may emit the three lights (red, blue, infrared) individually or in combination                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Light frequency                | N/A as DC Power                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 200 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Output Power                   | Red - 633nm (105mW/cm2)<br>Blue - 415nm (40mW/cm2)<br>NIR - 830nm (55mW/cm2)                                                                                                                                                                                                                                                                                                                                                                                                                | Each LED lamp bead has 4 diodes that emit different colors; the energy power of a diode is 3W.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Maximum power<br>density in mW | Red: 105mW/cm2<br>Blue: 40mW/cm2<br>NIR: 55mW/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                           | Red light: 115mW/cm2<br>Blue light: 120mW/cm2<br>IR: 70mW/cm2<br>Red/IR: 120mW/cm2<br>Blue/IR: 150mW/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Standard dose in<br>Joules     | Red: 126J/cm2<br>Blue: 48J/cm2<br>NIR: 66J/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                              | Red: 138J/cm2<br>Blue: 144J/cm2<br>IR: 84J/cm2<br>Red/IR: 144J/cm2<br>Blue/IR: 180J/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Adjustable dose<br>range       | Red: 1-126J/cm2<br>Blue: 1-48J/cm2<br>NIR: 1-66J/cm2<br>Red/IR: 1-192J/cm2<br>Blue/IR: 1-114J/cm2<br>Blue/Red: 1-174J/cm2                                                                                                                                                                                                                                                                                                                                                                   | Red: 1-242J/cm2<br>Blue: 1-180J/cm2<br>IR:1-147J/cm2<br>Red/IR: 1-144J/cm2<br>Blue/IR:1-180J/cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Treatment area                 | 792cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 756cm2 and 1008cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Treatment time                 | Up to 20 minutes                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 20 minutes (recommended treatment time)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Numbers of LEDs                | 120 LED's per color, per panel. Total of<br>480 LED's per color per unit. 1440<br>LED's in total.                                                                                                                                                                                                                                                                                                                                                                                           | 3 panels: 180EA<br>4 panels: 240EA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Working distance               | 2.5cm minimum distance from source                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 10~15cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Operation interface            | Display Screen                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Display Screen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Dimension                      | 1400mm[H]×500mm[W]×700mm[D]                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 500mm[H]×500mm[W]×1350mm[D]                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Safety<br>classification       | Class I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Class I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Software                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

## VI. COMPARISON TO PREDICATE DEVICE

{5}------------------------------------------------

## VII. PERFORMANCE TESTING

Bench: Performance testing of the Dermalux® Tri-Wave MD was conducted to verify that the device met all design specifications. The test results demonstrate that the Dermalux® Tri-Wave MD complies with all requirements, including international and FDA-recognized consensus standards:

{6}------------------------------------------------

EN/IEC 60601-1 (FDA #19-4) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance: 2005

EN/IEC 60601-1-2 (FDA #19-8) Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests: 2018

EN/IEC 62304 (FDA #13-79) Medical device software - Software life cycle processes: 2006

EN/IEC 62471 (FDA #12-249) Photobiological safety of lamps and lamp systems: 2008

IEC 60601-2-57 (FDA #12-242) Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use: 2011

IEC 62366-1 (FDA #5-113) Medical devices - Part 1: Application of usability engineering to medical devices: 2015

EN/IEC 60601-1-6 (FDA #5-89) Medical electrical equipment -Part 1-6: General requirements for basic safety and essential performance - Collateral Standard: Usability: 2013

Photometric Testing of LED Light Panel

Animal: No studies were performed.

Clinical: No studies were performed.

### VIII. CONCLUSION

Based on design considerations and testing of product attributes, we conclude that the Dermalux® Tri-Wave MD performs at least as well as the predicate device. The Dermalux® Tri-Wave MD is therefore considered to be substantially equivalent to the above-mentioned predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K200659](https://fda-staging.innolitics.com/device/K200659)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
