The FiberOptix® IAB Catheter is utilized for intra-aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation during systole increases blood supply to the heart muscle and decreases work of the left ventricle.
Device Story
FiberOptix IAB Catheter consists of inflatable balloon, inner/outer lumens, and embedded fiber optic pressure sensor at distal tip. Used with Intra-Aortic Balloon Pump (IABP) control system. Device placed in aorta; IABP regulates inflation/deflation cycles. Inflation during diastole increases myocardial blood supply; deflation during systole reduces left ventricular workload. Used in clinical settings (e.g., OR, ICU) by trained healthcare providers. Fiber optic sensor provides real-time pressure monitoring to guide pump timing. Modified sensor configuration improves mechanical robustness compared to predicate. Benefits include improved cardiovascular function via counterpulsation therapy.
Clinical Evidence
Bench testing only. Verification testing included insertion, durability, and static pressure tests, including simulated misuse scenarios. Results met all acceptance criteria and demonstrated performance equivalent to the predicate device.
Technological Characteristics
Catheter includes inflatable balloon, inner/outer lumens, and embedded fiber optic pressure sensor. Operates via connection to external IABP control system. Modified design features fiber optic sensor configuration along distal length for increased robustness. Sterilization and specific material standards not disclosed.
Indications for Use
Indicated for patients with Acute Coronary Syndrome, those undergoing Cardiac or Non-Cardiac Surgery, or those experiencing complications of Heart Failure.
Regulatory Classification
Identification
An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
April 9, 2020
Arrow International Inc. Apurva Gokhale Senior Regulatory Affairs Specialist 16 Elizabeth Drive Chelmsford, Massachusetts 01824
Re: K200634
Trade/Device Name: FiberOptix Intra-Aortic Balloon Catheter Kit Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-Aortic Balloon And Control System Regulatory Class: Class II Product Code: DSP Dated: March 6, 2020 Received: March 10, 2020
Dear Apurva Gokhale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200634
#### Device Name
FiberOptix® Intra-Aortic Balloon Catheter Kit
Indications for Use (Describe)
The FIBEROPTIX IAB Catheter with the Intra-Aortic Balloon Pump as a control system is clinically indicated for use in any of the following conditions:
1. Acute Coronary Syndrome
2. Cardiac and Non-Cardiac Surgery
3. Complications of Heart Failure
Type of Use (Select one or both, as applicable)| | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 2. 510(k) Summary
[As required by 21 CFR 807.92]
Date Prepared: 6 March 2020
| Submitter | Arrow International, Inc. (Subsidiary of Teleflex, Inc.)<br>16 Elizabeth Drive<br>Chelmsford, MA 01824<br>Establishment Registration: 3010532612 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| | Arrow International, Inc. (Subsidiary of Teleflex, Inc.)<br>Reading, PA 19605<br>Owner/ Operator: 2518433 |
| Company Contact | Apurva N. Gokhale<br>Sr. Regulatory Affairs Specialist<br>Arrow International, Inc.<br>16 Elizabeth Drive<br>Chelmsford, MA 01824 |
Apurva.gokhale@teleflex.com
| Regulatory Information | |
|------------------------|--------------------------------------------------|
| Proprietary Name: | FiberOptix® Intra-Aortic Balloon Catheter Kit |
| Common Name: | System, Balloon, Intra-Aortic and Control System |
| Product Code: | DSP |
| Regulation #: | 870.3535 |
| Description: | Intra-aortic balloon & control system |
| Regulatory Class: | Class II |
Phone: 978-250-5100 extension 722202
- Leqally Marketed ● K190117 Arrow FiberOptix IAB (Arrow International, Inc. - Cleared June 13, Predicate Device 2019)- Primary Predicate
The FiberOptix® Intra-Aortic Balloon (IAB) Catheter consists of an inflatable balloon, Device Description which is placed in the aorta to improve cardiovascular functioning. A computerized control system, also known as Intra-Aortic Balloon Pump (IABP) is utilized to regulate the inflation and deflation of the balloon.
> The FiberOptix® IAB Catheter consists of an inner lumen, an outer lumen, and an inflatable balloon. The outer lumen is comprised of an inflatable balloon connected to the distal tip of the catheter shaft and to the IAB catheter tip outer surface. The inner lumen is comprised of a luer adapter connected to the proximal end of the inner lumen and to the IAB catheter tip inner surface.
> The FiberOptix® IAB Catheter has a fiber optic pressure sensor which acts as a pressure transducer embedded in the catheter tip.
{4}------------------------------------------------
The FiberOptix® IAB Catheter is utilized for intra-aortic balloon counterpulsation Intended Use therapy in the aorta, whereby balloon inflation, during diastole and deflation during systole increases blood supply to the heart muscle and decreases work of the left ventricle.
The FiberOptix® IAB Catheter with the Intra-Aortic Balloon Pump as a control Indications for Use system is clinically indicated for use in any of the following conditions:
- 1. Acute Coronary Syndrome
- 2. Cardiac and Non-Cardiac Surgery
- 3. Complications of Heart Failure
The subject FiberOptix® IAB catheter is similar in design and identical in Technological indications for use to the predicate devices. Arrow Intra-Aortic Balloon Fiber Optic Characteristics Comparison Sensor/ Fiber Optic Measurement System and Arrow FiberOptix IAB Catheter. Compared to the predicate devices, the subject FiberOptix® IAB has a modified configuration of the fiber optic sensor along the distal length of the IAB catheter to improve its robustness.
> The technological differences between the subject and the predicate devices have been evaluated through bench tests to provide evidence that the FiberOptix® IAB Catheter is substantially equivalent to the predicate devices. The modified design of the FiberOptix® IAB Catheter has been verified through the following tests:
- Insertion, Durability Test & (FOS) Static Pressure Test ●
- . Simulated Misuse Test & (FOS) Static Pressure Test
The results of the verification test met the specified acceptance criteria and performed similar to the predicate device. The testing demonstrates that the subject FiberOptix® IAB catheter is substantially equivalent to the predicate device.
The subject FiberOptix® IAB Catheter is substantially equivalent to the Substantial specified predicate device, FiberOptix® IAB Catheter (K021462 and K190117) Equivalence Conclusion based on a comparison of the device functionality, materials, technological characteristics, and indications for use. The device modifications and results of design verification tests do not raise new or different questions of safety or effectiveness; therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.