The Halyard Black-Fire Powder-Free Nitrile Exam Glove tested for use with Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The Halyard Purple Nitrile Powder-Free Exam Glove tested for use with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile, ambidextrous nitrile examination gloves; textured fingertips; chlorinated donning side. Used by healthcare professionals to prevent cross-contamination between patient and examiner. Subject devices tested for permeation resistance against Fentanyl Citrate (both) and various chemotherapy drugs (Purple Nitrile only) per ASTM D6978-05. Output is a physical barrier; healthcare providers use gloves during patient examinations or chemotherapy handling to protect against chemical exposure and biological contaminants. Benefits include reduced risk of exposure to hazardous drugs and patient-to-examiner contamination.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ASTM D5151 (leakage), D6319 (physical properties), D6124 (residual powder), and D6978 (chemotherapy drug permeation). Biocompatibility confirmed via ISO 10993-10 (irritation/sensitization) and ISO 10993-11 (systemic toxicity).
Indicated for use as a disposable patient examination glove to prevent contamination between patient and examiner. The Purple Nitrile variant is indicated for use with specific chemotherapy drugs and Fentanyl Citrate; the Black-Fire variant is indicated for use with Fentanyl Citrate. Contraindicated for use with Carmustine and ThioTEPA.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Kimberly-Clark PURPLE NITRILE Powder-Free Exam Glove with Tested for Use with Chemotherapy Drugs (K101596)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.
October 11, 2020
0 & M Halyard, Inc Steven Dowdley Associate Director of Regulatory Affairs 5405 Windward Parkway Alpharetta, Georgia 30004
Re: K200633
Trade/Device Name: Halyard Black-Fire Powder-Free Nitrile Exam Glove tested for use with Fentanyl Citrate; Halyard Purple Nitrile Powder-Free Exam Glove tested for use with Chemotherapy Drugs and Fentanyl Citrate
Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZA, QDO, LZC Dated: September 8, 2020 Received: September 10, 2020
Dear Steven Dowdley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200633
Device Name
Halyard Black Fire Powder-Free Nitrile Exam Glove tested with Fentanyl Citrate
Indications for Use (Describe)
The Halyard Black Fire Powder-Free Nitrile Exam Glove tested with Fentany] Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
The following drugs showed not breakthrough at 240 minutes:
Fentanyl Citrate, 100 mcg/2ml
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|----------------------------------------------|--------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200633
Device Name
Halyard Purple Nitrile Powder-Free Exam Glove tested with Chemotherapy Drugs and Fentanyl Citrate
Indications for Use (Describe)
The Halyard Purple Nitrile Powder-Free Exam Glove tested with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
The following drugs showed not breakthrough at 240 minutes: Bleomycin (15.0 mg/ml) Busulfan (6.0 mg/ml) Carboplatin (10.0 mg/ml) Cisplatin (1.0 mg/ml) Cyclophosphamide (20.0 mg/ml) Cytarabine (100.0 mg/ml) Dacarbazine (DTIC) (10.0 mg/ml) Daunorubicin (5.0 mg/ml) Docetaxel (10.0 mg/ml) Doxorubicin Hydrochloride (2.0 mg/ml) Ellence (2.0 mg/ml) Etoposide (Toposar) (20.0 mg/ml) Fludarabine (25.0 mg/ml) Fluorouracil (50.0 mg/ml) Gemcitabine (Gemzar) (38.0 mg/ml) Idarubicin (1.0 mg/ml) Ifosfamide (50.0 mg/ml) Irinotecan (20.0 mg/ml) Mechlorethamine HC1 (1.0 mg/ml) Melphalan (5.0 mg/ml) Methotrexate (25.0 mg/ml) Mitomycin C (0.5 mg/ml) Mitoxantrone (2.0 mg/ml) Paclitaxel (Taxol) (6.0 mg/ml) Paraplatin (10.0 mg/ml) Rituximab (10.0 mg/ml) Trisonex (1.0 mg/ml) Vincrinstine (1mg/ml)
Fentanyl Citrate, 100 mcg/2ml
The following drugs showed breakthrough detected in less than 30 minutes: Carmustine (3.3mg/ml) 1.8 minutes and ThioTEPA (10.0mg/ml): 1.7 minutes.
Warning- Not for use with Carmustine and ThioTEPA.
{4}------------------------------------------------
Prescription Use (Part 21 CFR 801 Subpart D)
- X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
### K200633
| Date Summary was Prepared: | October 10, 2020 |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter: | Steven Dowdley<br>Associate Director, Regulatory Affairs<br>505 Windward Parkway<br>Alpharetta, GA 30004<br>Email: steven.dowdley@hyh.com<br>Phone: 678-451-8062 |
| Primary Contact for this 510(k): | Same as above |
| Device1 Trade Name: | Halyard Black-Fire Powder-Free Nitrile Exam Glove tested for use with<br>Fentanyl Citrate |
| Device2 Trade Name: | Halyard Purple Nitrile Powder-Free Exam Glove tested for use with<br>Chemotherapy Drugs and Fentanyl Citrate |
| Common Name: | Powder Free Nitrile Patient Examination Glove |
| Device1 Product Code: | LZA, QDO |
| Device2 Product Code: | LZC, QDO |
| Classification: | Class I |
| Regulation: | 21 CFR §880.6250 |
| Predicate Device1: | K153708<br>Halyard Black-Orange Powder-Free Nitrile Exam Glove |
| Predicate Device2: | K101596<br>Kimberly-Clark PURPLE NITRILE Powder-Free Exam Glove with Tested for<br>Use with Chemotherapy Drugs |
#### DEVICE DESCRIPTION
The Halyard Black-Fire Powder-Free Nitrile Exam Glove tested for use with Fentanyl Citrate is a disposable, blackcolored on the grip side and orange colored on the donning side, nitrile, powder- free, textured fingertip, ambidextrous, non-sterile patient examination glove that are chlorinated on the donning side and are packed in a cardboard dispenser box.
The Halyard Purple Nitrile Powder-Free Exam Glove tested for use with Chemotherapy Drugs and Fentanyl Citrate is a disposable, nitrile, powder-free, textured fingertip, ambidextrous, non-sterile patient examination gloves that are chlorinated on the donning side and are packed in a cardboard dispenser box.
#### INTENDED USE
The Halvard Black-Fire Powder-Free Nitrile Exam Glove tested for use with Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
The Halyard Purple Nitrile Powder-Free Exam Glove tested for use with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
{6}------------------------------------------------
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
| | Subject Device1 K200633 | Predicate Device1 K153708 | Remarks |
|-------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code | LZA, QDO | LZA | Different |
| Classification | Class 1 | Class 1 | Same |
| Trade Name | Halyard Black-Fire Powder-Free Nitrile | Halyard Black-Orange Powder-Free Nitrile | Similar |
| | Exam Glove tested for use with | Exam Glove | |
| | Fentanyl Citrate | | |
| Indications for Use | The Halyard Black-Fire Powder-Free<br>Nitrile Exam Glove tested for use with<br>Fentanyl Citrate is a disposable device<br>intended for medical purposes that is<br>worn on the examiner's hand to<br>prevent contamination between<br>patient and examiner.<br>The following drugs showed no<br>breakthrough at 240 minutes:<br>Fentanyl Citrate, 100 cg/2ml | The Halyard® Black-Orange Powder-Free<br>Nitrile Exam Glove is a disposable device<br>intended for medical purposes that is<br>worn on the examiner's hand to prevent<br>contamination between patient and<br>examiner. | Similar |
| Technological<br>Characteristics | Black on the grip side, orange on the<br>donning side, nitrile, powder-<br>free, textured fingertip, ambidextrous,<br>non- sterile patient examination glove. | Black on the grip side, orange on the<br>donning side, nitrile, powder-<br>free, textured fingertip, ambidextrous,<br>non- sterile patient examination glove. | Same |
| Sterilization | Gloves are supplied non-sterile | Gloves are supplied non-sterile | Same |
| ISO 10993-10<br>Biological evaluation of<br>medical devices -<br>Tests for Irritation | Acceptance criteria: No<br>erythema/edema up to 72 hours post<br>exposure. Result: Erythema/edema<br>was negligible. Meets acceptance<br>criteria. PASS | Acceptance criteria: No<br>erythema/edema up to 72 hours post<br>exposure. Result: Erythema/edema<br>was negligible. Meets acceptance<br>criteria. PASS | Same |
| ISO 10993-10<br>Biological evaluation of<br>medical devices -Tests<br>for Skin<br>Sensitization | Acceptance criteria: No evidence of<br>delayed dermal contact sensitivity at 24<br>and 48 hours post injection. Result: Not<br>a sensitizer under conditions of the<br>study. Meets acceptance criteria. PASS | Acceptance criteria: No evidence of<br>delayed dermal contact sensitivity at 24<br>and 48 hours post injection. Result: Not<br>a sensitizer under conditions of the<br>study. Meets acceptance criteria. PASS | Same |
| ISO 10993 Biological<br>evaluation of medical<br>devices - Tests for<br>Systemic Toxicity | Acceptance criteria: No signs of<br>systemic toxicity up to 72 hours post<br>injection. Result: No systemic toxicity<br>observed. Meets acceptance criteria.<br>PASS | Acceptance criteria: No signs of systemic<br>toxicity up to 72 hours post injection.<br>Result: No systemic toxicity observed.<br>Meets acceptance criteria. PASS | Same |
# Table 1 comparison between the predicate Device 1 and subject Device 1
# Table 2 comparison between the predicate Device2 and subject Device2
| | Subject Device2<br>K200633 | Predicate Device2<br>K101596 | Remarks |
|-------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code | LZC, QDO | LZC | Different |
| Classification | Class 1 | Class 1 | Same |
| Trade Name | Halyard Purple Nitrile Powder-Free<br>Exam Glove tested for use with<br>Chemotherapy Drugs and Fentanyl<br>Citrate | Kimberly-Clark PURPLE Nitrile Powder-Free<br>Exam Glove with Tested for Use with<br>Chemotherapy Drugs Labeling Claim | Similar |
| Indications for Use | Halyard Purple Nitrile Powder-Free<br>Exam Glove tested for use with | A powder-free patient examination glove<br>is a disposable device intended for | Similar |
| | Chemotherapy Drugs and Fentanyl<br>Citrate is a disposable device intended<br>for medical purposes that is worn on<br>the examiner's hand to prevent<br>contamination between patient and<br>examiner. The following drugs showed<br>no breakthrough at 240 minutes:<br><br>Fentanyl Citrate, 100 cg/2ml<br>Bleomycin (15.0 mg/ml)<br>Busulfan (6.0 mg/ml)<br>Carboplatin (10.0 mg/ml)<br>Cisplatin (1.0 mg/ml)<br>Cyclophosphamide (20.0 mg/ml)<br>Cytarabine (100.0 mg/ml)<br>Dacarbazine (DTIC) (10.0 mg/ml)<br>Daunorubicin (5.0 mg/ml)<br>Docetaxel (10.0 mg/ml)<br>Doxorubicin Hydrochloride (2.0<br>mg/ml)<br>Ellence (2.0 mg/ml)<br>Etoposide (Toposar) (20.0 mg/ml)<br>Fludarabine (25.0 mg/ml)<br>Fluorouracil (50.0 mg/ml)<br>Gemcitabine (Gemzar) (38.0mg/ml)<br>Idarubicin (1.0 mg/ml)<br>Ifosfamide (50.0 mg/ml)<br>Irinotecan (20.0 mg/ml)<br>Mechlorethamine HCl (1.0 mg/ml)<br>Melphalan (5.0 mg/ml)<br>Methotrexate (25.0mg/ml)<br>Mitomycin C (0.5 mg/ml)<br>Mitoxantrone (2.0mg/ml)<br>Paclitaxel (Taxol) (6.0 mg/ml)<br>Paraplatin (10.0 mg/ml)<br>Rituximab (10.0 mg/ml)<br>Trisonex (1.0 mg/ml)<br>Vincrinstine (1mg/ml)<br><br>The following drugs showed<br>breakthrough detected in less than 30<br>minutes:<br>Carmustine (3.3mg/ml): 1.8 minutes<br>ThioTEPA(10.0mg/ml) 1.7 minutes | medical purposes that is worn on the<br>examiners hand or finger to prevent<br>contamination between patient and<br>examiner.<br>In addition, these chemotherapy gloves<br>were tested for use with the following<br>drug concentrations per ASTMV D6978-<br>05:<br>The following drugs had no breakthrough<br>detected up to 240 minutes:<br>Bleomycin (15.0 mg/ml)<br>Busulfan (6.0 mg/ml)<br>Carboplatin (10.0 mg/ml)<br>Cisplatin (1.0 mg/ml)<br>Cyclophosphamide (20.0 mg/ml)<br>Cytarabine (100.0 mg/ml)<br>Dacarbazine (DTIC) (10.0 mg/ml)<br>Daunorubicin (5.0 mg/ml)<br>Docetaxel (10.0 mg/ml)<br>Doxorubicin Hydrochloride (2.0 mg/ml)<br>Ellence (2.0 mg/ml)<br>Etoposide (Toposar) (20.0 mg/ml)<br>Fludarabine (25.0 mg/ml)<br>Fluorouracil (50.0 mg/ml)<br>Gemcitabine (Gemzar) (38.0mg/ml)<br>Idarubicin (1.0 mg/ml)<br>Ifosfamide (50.0 mg/ml)<br>Irinotecan (20.0 mg/ml)<br>Mechlorethamine HCl (1.0 mg/ml)<br>Melphalan (5.0 mg/ml)<br>Methotrexate (25.0mg/ml)<br>Mitomycin C (0.5 mg/ml)<br>Mitoxantrone (2.0mg/ml)<br>Paclitaxel (Taxol) (6.0 mg/ml)<br>Paraplatin (10.0 mg/ml)<br>Rituximab (10.0 mg/ml)<br>Trisonex (1.0 mg/ml)<br>Vincrinstine (1mg/ml)<br><br>The following drugs showed<br>breakthrough detected in less than 30<br>minutes:<br>Carmustine (3.3mg/ml): 1.8 minutes<br>ThioTEPA(10.0mg/ml) 1.7 minutes | |
| Technological<br>Characteristics | Purple nitrile, powder-free, textured<br>fingertip, ambidextrous, non-sterile<br>patient examination glove. | Purple nitrile, powder- free, textured<br>fingertip, ambidextrous, non-sterile<br>patient examination glove. | Similar |
| Sterilization | Gloves are supplied non- sterile | Gloves are supplied non- sterile | Same |
| ISO 10993-10<br>Biological evaluation of<br>medical devices -<br>Tests for Irritation | Acceptance criteria: No<br>erythema/edema up to 72 hours post<br>exposure. Result: Erythema/edema<br>was negligible. Meets acceptance<br>criteria PASS | Acceptance criteria: No<br>erythema/edema up to 72 hours post<br>exposure. Result: Erythema/edema was<br>negligible. Meets acceptance criteria.<br>PASS | Same |
| ISO 10993-10<br>Biological evaluation of<br>medical devices -Tests<br>for Skin<br>Sensitization | Acceptance criteria: No evidence of<br>delayed dermal contact sensitivity at<br>24 and 48 hours post injection. Result:<br>Not a sensitizer under conditions of<br>the study. Meets acceptance criteria.<br>PASS | Acceptance criteria: No evidence of<br>delayed dermal contact sensitivity at 24<br>and 48 hours post injection. Result: Not a<br>sensitizer under conditions of the study.<br>Meets acceptance criteria. PASS | Same |
| ISO 10993 Biological<br>evaluation of medical<br>devices - Tests for<br>Systemic Toxicity | Acceptance criteria: No signs of<br>systemic toxicity up to 72 hours post<br>injection. Result: No systemic toxicity<br>observed. Meets acceptance criteria.<br>PASS | Acceptance criteria: No signs of systemic<br>toxicity up to 72 hours post injection.<br>Result: No systemic toxicity observed.<br>Meets acceptance criteria. PASS | Same |
{7}------------------------------------------------
{8}------------------------------------------------
# SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
Subject Device1, Halyard Black-Fire Powder-Free Nitrile Exam Glove tested for use with Fentanyl Citrate and subject Device2, Halyard Purple Nitrile Powder-Free Exam Glove tested for use with Chemotherapy Drugs and Fentanyl Citrate were both tested for conformance to the applicable sections of the following standards:
- ASTM D5151-06 Standard Test Method for Detection of Holes in Medical Gloves ●
- ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical Applications
- ASTM D6124-06 Standard Test Method for Residual Powder on Medical Gloves
- ASTM D 6978-05 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs
- . ISO 10993-10 Biological evaluation of medical devices -Tests for Irritation and Skin Sensitization
- . ISO 10993-11 Biological evaluation of medical devices - Tests for Systemic Toxicity.
| Performance | K200633 | K153708 | Remarks |
|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| | Subject Device1 The Halyard Black-Fire<br>Powder-Free Nitrile Exam Glove tested<br>for use with Fentanyl Citrate | Predicate Device1 Halyard Black-Orange<br>Powder-Free Nitrile Exam Glove | |
| ASTM D5151-06<br>Standard Test<br>Method for<br>Detection of<br>Holes in Medical<br>Gloves | device shows it meets the 2.5 AQL<br>requirement in the standards for<br>leakage. The device meets the<br>acceptance criteria of the standard.<br>PASS | device shows it meets the 2.5 AQL<br>requirement in the standards for<br>leakage. The device meets the<br>acceptance criteria of the standard.<br>PASS | Same |
| ASTM D6319<br>Standard<br>Specification for<br>Nitrile Examination<br>Gloves for Medical<br>Applications | Properties | Test results | ASTM SPEC |
| | Holes | AQL 1.0% | AQL 2.5% |
| | Length | 230-258mm | >230mm |
| | Width | 85-105mm | 85-105mm |
| | Palm | .05-07mm | >.050mm |
| | Unaged<br>Tensile | >16MPa | >14MPa |
| | Unaged<br>Elongation | > 500% | > 500% |
| | Aged<br>Tensile | >15MPa | >14MPa |
| | Aged<br>Elongation | >450% | >400% |
| | Powder | <2mg/glove | <2mg/glove |
| | Properties | Test results | ASTM SPEC |
| | Holes | AQL 1.0% | AQL 2.5% |
| | Length | 230-258mm | >230mm |
| | Width | 85-105mm | 85-105mm |
| | Palm | .05-.07mm | >.050mm |
| | Unaged<br>Tensile | >16MPa | >14MPa |
| | Unaged<br>Elongation | > 500% | > 500% |
| | Aged<br>Tensile | >15MPa | >14MPa |
| | Aged<br>Elongation | >450% | >400% |
| | Powder | <2mg/glove | <2mg/glove |
| | Same | | |
| | | | Same |
| ASTM D 6978<br>Standard Practice for<br>Assessment of<br>Resistance of<br>Medical Gloves to | The following drugs showed no<br>breakthrough at 240 minutes:<br><br>Fentanyl Citrate, 100 mcg/2ml | Not applicable | Different,<br>adding<br>Fentanyl<br>Citrate |
### Table 3. Non-clinical performance testing for the subject Device1 compared to Predicate device 1
{9}------------------------------------------------
| Permeation by | | | |
|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Chemotherapy Drugs | | | |
| ASTM D6124<br>Standard Test<br>Method for<br>Residual Powder on<br>Medical Gloves | Residual powder on the subject device is<br>within the powder-free limit of < 2 mg<br>maximum powder per glove and meets<br>the acceptance criteria for powder-free.<br>PASS | Residual powder on the subject device is<br>within the powder-free limit of < 2 mg<br>maximum powder per glove and meets<br>the acceptance criteria for powder-free.<br>PASS | Same |
# Table 4. Non-clinical Performance testing for the subject Device2 compared to Predicate device 2
| Performance | K200633 | K101596 | Remarks |
|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| | Subject Device2 Halyard Purple Nitrile<br>Powder-Free Exam Glove tested for use<br>with Chemotherapy Drugs and Fentanyl<br>Citrate | Predicate Device2 Kimberly-Clark PURPLE<br>Nitrile Powder-Free Exam Glove with<br>Tested for Use with Chemotherapy Drugs<br>Labeling Claim | |
| ASTM D5151-06<br>Standard Test<br>Method for<br>Detection of<br>Holes in Medical<br>Gloves | device shows it meets the 2.5 AQL<br>requirement in the standards for<br>leakage. The device meets the<br>acceptance criteria of the standard.<br>PASS | device shows it meets the 2.5 AQL<br>requirement in the standards for<br>leakage. The device meets the acceptance<br>criteria of the standard.<br>PASS | Same |
| ASTM D6319<br>Standard<br>Specification for<br>Nitrile Examination<br>Gloves for Medical<br>Applications | Properties | Test results | ASTM SPEC |
| | Holes | AQL 1.0% | AQL 2.5% |
| | Length | 230-258mm | >230mm |
| | Width | 85-105mm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.