K200631 · Deroyal Industries, Inc. · FLL · Dec 23, 2020 · General Hospital
Device Facts
Record ID
K200631
Device Name
DeRoyal Temperature Monitoring Probe
Applicant
Deroyal Industries, Inc.
Product Code
FLL · General Hospital
Decision Date
Dec 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The DeRoyal Temperature Monitoring Probe is used for routine monitoring of the patient's core body or skin surface temperature.
Device Story
Single-use temperature monitoring probe; consists of wire set with proximal blue connector and distal thermistor chip. Thermistor passively modifies electrical current based on temperature. Interfaces with DeRoyal-branded cables connected to YSI 400/700 series compatible monitors (e.g., Mindray, Philips, Siemens/Draeger, GE). Used in hospital settings by licensed healthcare practitioners. Three configurations: General Purpose (nasopharyngeal, esophageal, rectal), Adult Skin (adhesive cover), and Tympanic (foam ear plug). Output displayed on independent patient monitor; assists clinicians in routine core or skin temperature tracking. Benefits patient through continuous, accurate temperature monitoring.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (cytotoxicity, skin sensitization, irritation), electrical safety, EMC, leakage current, and temperature accuracy/response time per ISO 80601-2-56. Results confirmed the encapsulation modification does not impact safety or efficacy.
Technological Characteristics
Passive thermistor-based temperature sensing. Materials: PVC tubing/insulation, copper wire, ceramic thermistor, brass connector, UV-cured adhesive. Dimensions/form factor: wire set with specific distal interfaces (tube, adhesive cover, or foam plug). Connectivity: wired interface to YSI 400/700 compatible monitors. Sterilization: Ethylene Oxide (General Purpose/Skin) or non-sterile (Tympanic).
Indications for Use
Indicated for routine monitoring of core body or skin surface temperature in adult and pediatric patients. General Purpose probes are for nasopharyngeal, esophageal, or rectal use. Adult Skin sensors are for skin surface application. Tympanic probes are for aural canal use. For use by licensed healthcare practitioners only.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Exac-Temp & Clini-Temp Probe w/ Temperature Sensor (K925791)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The text reads "U.S. FOOD & DRUG ADMINISTRATION".
December 23, 2020
DeRoyal Industries, Inc. Sarah Bennett Senior Regulatory Affairs Specialist 200 DeBusk Lane Powell, Tennessee 37849
Re: K200631
Trade/Device Name: DeRoyal Temperature Monitoring Probe Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: November 23, 2020 Received: November 24, 2020
Dear Sarah Bennett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan M. Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K200631
#### Device Name
DeRoyal Temperature Monitoring Probe
Indications for Use (Describe)
The DeRoyal Temperature Monitoring Probe is used for routine of the patient's core body or skin surface temperature.
The probe is offered in the following three configurations :
- General Purpose Temperature Probe for routine monitoring of the core body temperature in adult and pediatric patients by insertion into the nasopharyngeal, esophageal, or rectal cavities.
- Adult Skin Temperature Sensor for routine monitoring of skin temperature by application of the probe's adhesive cover to an adult patient's skin surface.
- Tympanic Temperature Probe for routine monitoring of the core body temperature in adult and pediatric patients by insertion of the ear piece into the auralcanal.
The device is single use and for use by licensed healthcare practitioners only. The probes are designed to interface with DeRoyal- branded cables for connection with YSI 400 or 700 series compatible monitors, including the following patient monitors and equivalent models: Mindray Passport. Philips IntelliVue, Siemens/Draeger Infinity, and GE Datex-Ohmeda brands.
Type of Use (Select one or both, as applicable)
| X Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------|
|------------------------------------------------|
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperw ork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperw ork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."
{3}------------------------------------------------
# 510(k) Summary
Image /page/3/Picture/1 description: The image features the logo for "DeRoyal®". The logo consists of a stylized blue swoosh that curves over the text "DeRoyal". The "®" symbol is present to the right of the word "DeRoyal", indicating that it is a registered trademark.
Corporate 200 DeBusk Lane Powell, TN 37849 USA 865.938.7828 800.251.9864 www.deroyal.com
| Date prepared: | November 23, 2020 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | DeRoyal Industries, Inc.<br>200 DeBusk Lane<br>Powell, TN 37849<br>Owner/Operator #1044833 |
| 510(k) Contact: | Sarah Bennett<br>Senior Regulatory Affairs Specialist<br>P: 865-362-6112<br>F: 865-362-3741<br>sabennett@deroyal.com |
| Contract Manufacturer: | DeRoyal Cientifica de Latinoamerica<br>S.R.L.<br>Global Park<br>Box 180-3006, 602 Parkway<br>La Aurora, Heredia Heredia, Costa<br>Rica 146<br><br>DeRoyal Industries, Inc.<br>185 Richardson Way<br>Maynardville, TN 37807 |
| Trade Name: | DeRoyal Temperature Monitoring<br>Probe |
| Common Name: | Temperature Monitoring Probe |
| Classification Name: | Thermometer, Electronic, Clinical |
| Device Product Code: | FLL |
| Regulatory Class: | Class II |
| Classification Panel: | General Hospital |
| Regulation Number: | 21 CFR 880.2910 |
| Predicate Devices: | Exac-Temp & Clini-Temp Probe w/<br>Temperature Sensor [K925791] –<br>Primary Predicate<br><br>Exac-Temp & Clini-Temp<br>Tympanic Probe [K925792] |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image features the logo for "DeRoyal". The logo consists of a stylized, curved blue shape resembling a checkmark or a swoosh, positioned to the left of the company name. The text "DeRoyal" is written in a blue, sans-serif font, with the registered trademark symbol appearing next to the word "Royal".
Corporate 200 DeBusk Lane Powell, TN 37849 USA 865.938.7828 800.251.9864 www.deroyal.com
Reference Devices:
Skin Surface Temperature Sensor [K925006]
Skin Sensor Probe, General Purpose Probe, Tympanic Probe [K101244]
Esophageal/Rectal Temperature Probe [K140134]
Disposable Temperature Probe [K181967]
## Device Description
The DeRoyal Temperature Monitoring Probe is used for routine patient temperature monitoring. Each probe is single use and contains a wire set with a blue connector at the proximal end and a thermistor chip at the distal end. The thermistor passively modifies the electrical current traveling through the probe. The connector interfaces with a cable that is connected to an independent temperature monitor used to display the temperature readings.
There are three probe configurations included in this submission:
- 1. An adult skin temperature sensor that consists of the wire set with an adhesive probe cover that is applied to the patient's intact skin for monitoring of skin surface temperature.
- 2. A tympanic temperature probe that consists of the wire set with a foam ear plug that is inserted into the aural canal for monitoring of core body temperature.
- 3. A general purpose temperature probe that consists of the wire set enclosed in a tube that may be inserted into the esophageal, rectal, or nasopharyngeal cavities for monitoring of core body temperature.
All probes are individually packaged. The tympanic configuration is non-sterile. The other two configurations are sold sterile. The probes are designed to interface with DeRoyal-branded cables for connection with YSI 400 or 700 series compatible monitors, including the following patient monitors and equivalent models: Mindray Passport, Philips IntelliVue, Siemens/Draeger Infinity, and GE Datex-Ohmeda brands.
## Intended Use
The DeRoyal Temperature Monitoring Probe is used for routine monitoring of the patient's core body or skin surface temperature.
{5}------------------------------------------------
### Summary of Technological Characteristics
| Characteristic | Configuration | Proposed Device | Predicate 510(k) | Predicate Device | Differences |
|---------------------|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | General Purpose | The DeRoyal<br>Temperature<br>Monitoring Probe is<br>used for routine<br>monitoring of the<br>patient's core body or<br>skin surface<br>temperature. | K925791 – Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | The Exac-Temp<br>General Purpose<br>Temperature Probe is to<br>be used for routine<br>monitoring of<br>anesthetized patients by<br>inserting the probe into<br>the esophagus or<br>rectum. | The Indications for Use<br>have been updated for<br>the proposed device to<br>include the reference<br>body site and<br>measuring site as<br>defined by ISO 80601-<br>2-56 and the intended<br>treatment population<br>and compatible<br>monitors as requested<br>by the FDA. |
| | Adult Skin Sensor | The probe is offered in<br>the following three<br>configurations:<br>- General Purpose<br>Temperature Probe for<br>routine monitoring of<br>the core body<br>temperature in adult<br>and pediatric patients<br>by insertion into the<br>nasopharyngeal,<br>esophageal, or rectal<br>cavities;<br>- Adult Skin<br>Temperature sensor for<br>routine monitoring of<br>skin temperature by<br>application of the<br>probe's adhesive cover<br>to an adult patient's<br>skin surface.<br>- Tympanic<br>Temperature Probe for<br>routine monitoring of<br>the core body<br>temperature in adult<br>and pediatric patients | K925006 – Exac-Temp<br>Skin Surface<br>Temperature Sensor | The Exac-Temp Skin<br>Surface Temperature<br>Sensor is to be used for<br>routine monitoring of a<br>patient's skin surface. | |
| | Tympanic Probe | | K925792 – Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | The Exac-Temp<br>Tympanic Temperature<br>Probe is to be used for<br>routine monitoring of<br>the patient temperature<br>using the aural canal<br>(ear canal or the<br>external auditory<br>meatus). | |
{6}------------------------------------------------
| Prescription Only | General Purpose | by insertion of the ear piece into the aural canal.<br>The device is single use and for use by licensed healthcare practitioners only. The probes are designed to interface with DeRoyal-branded cables for connection with YSI 400 or 700 series compatible monitors, including the following patient monitors and equivalent models: Mindray Passport, Philips IntelliVue, Siemens/ Draeger Infinity, and GE Datex-Ohmeda brands. | Yes | K925791 - Exac-Temp & Clini-Temp Probe w/ Temp. Sensor | Yes | Same |
|-------------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------------------------------------------------|-------------|------|
| | Adult Skin Sensor | | | K925006 - Exac-Temp Skin Surface<br>Temperature Sensor | | |
| | Tympanic Probe | | | K925792 - Exac-Temp & Clini-Temp<br>Tympanic Probe | | |
| Mode of Operation | General Purpose | | Direct Mode | K925791 - Exac-Temp & Clini-Temp Probe w/ Temp. Sensor | Direct Mode | Same |
| | Adult Skin Sensor | | | K925006 - Exac-Temp Skin Surface<br>Temperature Sensor | | |
| | Tympanic Probe | | | K925792 - Exac-Temp & Clini-Temp<br>Tympanic Probe | | |
{7}------------------------------------------------
| Measuring Site | General Purpose | Rectum, Esophagus,<br>Nasopharynx | K925791 – Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | Esophagus, Rectum | DeRoyal intends to add<br>the nasopharynx to the<br>General Purpose Probe<br>as a measuring site. ISO<br>80601-2-56 defines<br>measuring site as the<br>"part of the patient<br>where the temperature<br>is measured." The FDA<br>has cleared three<br>510(k)s for temperature<br>probes that utilize the<br>nasopharynx as a<br>measuring site in<br>addition to the<br>esophagus and rectum:<br>K101244, K140134,<br>and K181967. |
|---------------------|-------------------|-----------------------------------|--------------------------------------------------------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Adult Skin Sensor | Skin Surface | K925006 – Exac-Temp<br>Skin Surface<br>Temperature Sensor | Skin Surface | Same |
| | Tympanic Probe | Auditory Canal | K925792 - Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | Auditory Canal | Same |
| Reference Body Site | General Purpose | Core Body | K925791 - Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | Core Body | Same |
| | Adult Skin Sensor | Skin Surface | K925006 - Exac-Temp<br>Skin Surface<br>Temperature Sensor | Skin Surface | Same |
| | Tympanic Probe | Core Body | K925792 – Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | Core Body | Same |
| Use Environment | General Purpose | Hospital | K925791 – Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | Hospital | Same |
| | Adult Skin Sensor | | K925006 – Exac-Temp<br>Skin Surface<br>Temperature Sensor | | |
{8}------------------------------------------------
| | Tympanic Probe | | K925792 - Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | | |
|----------------------|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Rated Output Range | General Purpose | 25°C to 45°C | K925791 - Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | 25°C to 45°C | Same |
| | Adult Skin Sensor | | K925006 - Exac-Temp<br>Skin Surface<br>Temperature Sensor | | |
| | Tympanic Probe | | K925792 - Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | | |
| Accuracy | General Purpose | $\u00b10.2°C$ | K925791 - Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | $\u00b10.2°C$ | Same |
| | Adult Skin Sensor | | K925006 - Exac-Temp<br>Skin Surface<br>Temperature Sensor | | |
| | Tympanic Probe | | K925792 - Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | | |
| Operating Conditions | General Purpose | 25°C to 45°C | K925791 - Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | 25°C to 45°C | Same |
| | Adult Skin Sensor | | K925006 - Exac-Temp<br>Skin Surface<br>Temperature Sensor | | |
| | Tympanic Probe | | K925792 - Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | | |
| Design | General Purpose | Wire set with a<br>thermistor chip at the<br>distal end and a blue<br>connector at the<br>proximal end. The wire<br>set is enclosed in a tube<br>that may be inserted<br>into the application site. | K925791 - Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | Wire set with a<br>thermistor chip at the<br>distal end and a blue<br>connector at the<br>proximal end. The wire<br>set is enclosed in a tube<br>that may be inserted<br>into the application site. | Same |
| | Adult Skin Sensor | Wire set with a | K925006 - Exac-Temp | Wire set with a | Same |
{9}------------------------------------------------
| | | thermistor chip at the distal end and a blue<br>connector at the proximal end. An<br>adhesive probe cover<br>applies the device to the<br>patients' skin. | Skin Surface<br>Temperature Sensor | thermistor chip at the distal end and a blue<br>connector at the proximal end. An<br>adhesive probe cover<br>applies the device to the<br>patients' skin. | |
|--|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Tympanic Probe | Wire set with a<br>thermistor chip at the distal end and a blue<br>connector at the proximal end. A foam<br>ear plug is used to<br>insert the device into<br>the patient's aural<br>canal. | K925792 – Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | Wire set with a<br>thermistor chip at the distal end and a blue<br>connector at the proximal end. A foam<br>ear plug is used to<br>insert the device into<br>the patient's aural<br>canal. | Same |
| | Materials | General Purpose | Tube: PVC<br>Wire: Copper with PVC<br>insulation<br>Thermistor: Ceramic<br>Connector: PVC-<br>molded brass<br>Strain Relief: PVC<br>Cap: UV-cured<br>adhesive | K925791 – Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | Tube: PVC<br>Wire: Copper with PVC<br>insulation<br>Thermistor: Ceramic<br>Connector: PVC-<br>molded brass<br>Strain Relief: PVC<br>Cap: PVC and epoxy<br>glue |
| | | Adult Skin Sensor | Cover: Adhesive foam<br>Wire: Copper with PVC<br>insulation<br>Thermistor: Ceramic<br>Connector: PVC-<br>molded brass<br>Strain Relief: PVC<br>Cap: UV-cured<br>adhesive | K925006 - Exac-Temp<br>Skin Surface<br>Temperature Sensor | Cover: Adhesive foam<br>Wire: Copper with PVC<br>insulation<br>Thermistor: Ceramic<br>Connector: PVC-<br>molded brass<br>Strain Relief: PVC<br>Cap: PVC and epoxy<br>glue |
| | | Tympanic Probe | Ear Plug: Foam and/or<br>cotton ball<br>Wire: Copper with PVC<br>insulation<br>Thermistor: Ceramic | K925792 – Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | Ear Plug: Foam and/or<br>cotton ball<br>Wire: Copper with PVC<br>insulation<br>Thermistor: Ceramic |
| | | | | | |
{10}------------------------------------------------
| | | Connector: PVC-<br>molded brass<br>Strain Relief: PVC<br>Cap: UV-cured<br>adhesive | | Connector: PVC-<br>molded brass<br>Strain Relief: PVC<br>Cap: PVC and epoxy<br>glue | Safety and EMC testing<br>also were performed on<br>the proposed device and<br>are contained within<br>Section 17. The results<br>of this testing support<br>the proposed device's<br>safety and<br>effectiveness. |
|------------------------|-------------------|------------------------------------------------------------------------------------|--------------------------------------------------------------|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Use of a Probe Cover | General Purpose | No | K925791 – Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | No | Same |
| | Adult Skin Sensor | Yes | K925006 - Exac-Temp<br>Skin Surface<br>Temperature Sensor | Yes | Same |
| | Tympanic Probe | No | K925792 – Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | No | Same |
| Nature of Body Contact | General Purpose | Mucosal Membrane | K925791 – Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | Mucosal Membrane | Same |
| | Adult Skin Sensor | Intact Skin | K925006 - Exac-Temp<br>Skin Surface<br>Temperature Sensor | Intact Skin | Same |
| | Tympanic Probe | Intact Skin | K925792 — Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | Intact Skin | Same |
| Duration of Contact | General Purpose | Limited (≤ 24 hours) | K925791 – Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | Limited (≤ 24 hours) | Same |
| | Adult Skin Sensor | Prolonged (≥ 24 hours) | K925006 - Exac-Temp<br>Skin Surface<br>Temperature Sensor | Prolonged (≥ 24 hours) | Same |
| | Tympanic Probe | Prolonged (≥ 24 hours) | K925792 – Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | Prolonged (≥ 24 hours) | Same |
| Sterilization | General Purpose | Sterilized with Ethylene<br>Oxide | K925791 - Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | Sterilized with Ethylene<br>Oxide | Same |
{11}------------------------------------------------
| Adult Skin Sensor | Sterilized with Ethylene Oxide | K925006 – Exac-Temp<br>Skin Surface<br>Temperature Sensor | Sterilized with Ethylene Oxide | Same | |
|------------------------|--------------------------------|-----------------------------------------------------------|--------------------------------------------------------------|----------------|------|
| Tympanic Probe | Non-sterile | K925792 – Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | Non-sterile | Same | |
| Storage Conditions | General Purpose | -25°C to +55°C | K925791 – Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | -25°C to +55°C | Same |
| Adult Skin Sensor | | K925006 – Exac-Temp<br>Skin Surface<br>Temperature Sensor | | | |
| Tympanic Probe | | K925792 – Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | | | |
| Reusable or Disposable | General Purpose | Disposable | K925791 – Exac-Temp<br>& Clini-Temp Probe w/<br>Temp. Sensor | Disposable | Same |
| Adult Skin Sensor | | K925006 – Exac-Temp<br>Skin Surface<br>Temperature Sensor | | | |
| Tympanic Probe | | K925792 – Exac-Temp<br>& Clini-Temp<br>Tympanic Probe | | | |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for DeRoyal. The logo features a stylized blue swoosh above the company name, also in blue. The word "DeRoyal" is written in a serif font, with a registered trademark symbol next to the word "Royal".
Corporate 200 DeBusk Lane Powell, TN 37849 USA 865.938.7828 800.251.9864 www.deroyal.com
### Summary of Performance Tests
The proposed device has been tested and/or evaluated according to the following standards: ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 80601-2-56, IEC 60601-1, and IEC 60601-1-2.
The following biocompatibility tests were performed on final, finished products manufactured with the proposed change: cytotoxicity, skin sensitization, and irritation. This testing was performed on each of the three configurations contained within this submission.
Testing according to IEC 60601-1 and IEC 60601-1-2 was performed to ensure the change in the encapsulation did not affect the electrical safety of the device. Accuracy and time response testing according to ISO 80601-2-56 was performed to ensure the proposed device completed its essential performance safely and effectively. A leakage current test also was performed after submerging the device in solution to ensure the encapsulation method is effective.
All testing was performed on final, finished products manufactured with the proposed modification. The test results met the requirements of the aforementioned standards and demonstrate that the proposed modification does not impact the safety or efficacy of the device.
### Conclusion
The results of performance testing demonstrate the DeRoyal Temperature Monitoring Probe is substantially equivalent to the predicates. With the exception of the proposed change to the encapsulation method, the proposed device is identical to the predicate devices, which have been on the market since 1994. Therefore, the proposed device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.