K200601 · GreenMark Biomedical, Inc. · LFC · Mar 29, 2021 · Dental
Device Facts
Record ID
K200601
Device Name
LumiCare Caries Diagnostic Rinse
Applicant
GreenMark Biomedical, Inc.
Product Code
LFC · Dental
Decision Date
Mar 29, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1740
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
LumiCare™ Caries Detection Rinse is indicated to aid the dental professional in visualization of carious lesions.
Device Story
LumiCare is an oral rinse containing fluorescein-tagged submicron starch particles. Used in a dental operatory by a clinician; the patient swishes the rinse, followed by a water rinse and air drying of teeth. The clinician then exposes the teeth to a blue light source (450-470 nm, 450 mW/cm2) and observes the teeth through light orange filter accessories (including protective eyewear). Areas of tooth decay temporarily stain, fluorescing under the blue light. This visual indicator assists the clinician in identifying carious lesions during examination, potentially improving the accuracy of caries detection and guiding clinical decision-making regarding treatment.
Clinical Evidence
No clinical data. Bench testing on extracted teeth compared the device to a histologic reference standard, demonstrating reproducibility and performance comparable to the predicate. Simulated use testing in a clinical environment using a manikin model confirmed ease of use. Biocompatibility testing (cytotoxicity, sensitization, irritation) was performed per ISO 10993-5 and 10993-10.
Technological Characteristics
Liquid oral rinse containing fluorescein-tagged submicron starch particles in a propylene glycol/water carrier. Requires external blue light source (450-470 nm, 450 mW/cm2) and orange filter accessories for visualization. Biocompatible per ISO 10993-5 and 10993-10. Shelf life of 2 years supported by accelerated testing.
Indications for Use
Indicated for dental professionals to aid in the visualization of carious lesions in adults and children.
Regulatory Classification
Identification
The caries detection device is a device intended to show the existence of decay in a patient's tooth by use of electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 29, 2021
GreenMark Biomedical Inc. % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K200601
Trade/Device Name: LumiCare Caries Diagnostic Rinse Regulation Number: 21 CFR 872.1740 Regulation Name: Caries detection device Regulatory Class: Class II Product Code: LFC, EAQ Dated: March 24, 2021 Received: March 26, 2021
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200601
Device Name LumiCare^TM Caries Detection Rinse
Indications for Use (Describe) To aid the dental professional in visualization of carious lesions
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Expiration Date: 06/30/2023 See PRA Statement below.
Form Approved: OMB No. 0910-0120
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for GreenMark Biomedical Inc. The word "GREENMARK" is written in a combination of green and black letters, with "GREEN" in green and "MARK" in black. A trademark symbol is located to the right of the word "MARK". Below the word "GREENMARK" is the text "BIOMEDICAL INC." in black. To the right of the text is a green leaf design.
K200601
### 510(k) Summary
### I. SUBMITTER
GreenMark Biomedical Inc.
Mailing Address: 325 E. Grand River, Suite 314 East Lansing, MI 48823
Offices and lab at: 1600 Huron Parkway, Ann Arbor, MI 48109
Phone: 517-896-3665 Contact Person: Steven Bloembergen: info@greenmark.bio Date Prepared: March 24, 2021
II. DEVICE
Name of Device: LumiCare™ Caries Detection Rinse Common or Usual Name: Caries detector Classification Name: Caries Detection Device (21 CFR 872.1740) Regulatory Class: II Product Code: LFC Review Panel: Dental
### III. PREDICATE DEVICE
Pulpdent SnoopTM Caries Detecting Dye, K964430
### IV. DEVICE DESCRIPTION
The LumiCare™ device is an oral rinse solution containing starch-based particles tagged with fluorescein. A specific volume is swished in the mouth, followed by a water rinse and air drying of teeth prior to exposure of the teeth to a blue light source with an irradiance of 450 mW/cm2 and wavelength of 450-470 nm. The clinician completes the clinical examination by observing fluorescence through light orange filter accessories, including protective eyewear, to aid the visualization of dental caries.
Page 1 of 4
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for GreenMark Biomedical Inc. The logo has the word "GREENMARK" in green and black lettering, with the "GREEN" portion in green and the "MARK" portion in black. To the right of the word is a green leaf design. Below the word "GREENMARK" is the phrase "BIOMEDICAL INC." in black lettering.
## V. INDICATIONS FOR USE / INTENDED USE
LumiCare™ Caries Detection Rinse is indicated to aid the dental professional in visualization of carious lesions.
The Indications for Use statement for the LumiCare™ device is identical to that of the predicate device, with minor exception. This difference does not alter the intended use of the device nor does it affect the safety and effectiveness of the device relative to the predicate.
Both the subject and predicate device have the same intended use of detection of dental caries.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Characteristic | Subject Device | Predicate Device |
|--------------------------------------|-----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Device Name | LumiCare™ Caries<br>Detection Rinse | Pulpdent Snoop™<br>(K964430) |
| Regulation Number | 872.1740 | 872.1740 |
| Identification | Caries Detection Device | Caries Detection Device |
| Intended Use | Detection of dental caries | Detection of dental caries |
| Indication for disease | Dental caries | Dental caries |
| Indication for patient<br>population | Adults and children | Not specified |
| Indication for Use<br>statement | To aid the dental<br>professional in<br>visualization of carious<br>lesions | To assist the dental<br>professional in the<br>detection of carious dentin |
| Anatomic Sites | Directly visible tooth<br>surfaces | Directly visible tooth<br>surfaces |
| Use Environment | Dental Operatory | Dental Operatory |
| Prescription/OTC | Prescription only | Prescription only |
| Product Form | Liquid | Liquid |
| Primary Carrier Fluids | Propylene Glycol/Water | Propylene Glycol/Water |
| Contains Staining<br>Agent | Yes | Yes |
| Use | Apply product to teeth | Apply product to teeth |
| Detection Method | Areas of tooth decay<br>temporarily stain | Areas of tooth decay<br>temporarily stain |
| Shelf Life | 2 years | 2 years |
| Staining Agent | Fluorescein | Dark blue dye |
## Table 1. Comparison to Predicate Device
Page 2 of 4
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for GreenMark Biomedical Inc. The word "GREENMARK" is written in a combination of green and black letters, with "GREEN" in green and "MARK" in black. To the right of the text is a green leaf design. Below the text, "BIOMEDICAL INC." is written in smaller, black letters.
| Characteristic | Subject Device | Predicate Device |
|----------------------------|-------------------------------------|---------------------------------------------------------------------|
| Carrier for staining agent | Submicron Starch Particle | None |
| Viewing Conditions | Blue light/orange filter | Ambient light/white light with the naked eye or under magnification |
| Stimulation Wavelength | Blue light (450-470 nm, peak value) | Visible light |
The subject and the predicate device are both dye-containing liquids which bind to carious lesions on all (occlusal, facial, lingual, proximal) directly visible surfaces of the teeth, providing a visible indicator of the presence of caries. The subject and predicate devices have similar technological elements: liquid formulation, application directly to teeth and the liquid contains carriers, preservatives, and a staining agent.
Technological differences between the subject and predicate device include:
- 1. Viewing Conditions: The predicate device's stain is visible under visible light whereas the subject device's stain is visible under blue light conditions when viewed through an orange filter. This difference does not impact safety or performance.
- 2. Staining agent and carrier: The subject device contains submicron starch particles as a carrier for the staining agent, fluorescein, whereas the predicate contains blue dye with no carrier. This difference does not impact safety or effectiveness. The fluoresceinated starch particle was demonstrated to be non-cytotoxic, non-irritating and non-sensitizing.
## VII. PERFORMANCE DATA
- A. Non-Clinical Performance Testing
## Bench Testing
Sensitivity and Specificity testing was completed on extracted teeth using histologic reference standard for the LumiCare™ rinse and the predicate, demonstrating a high degree of reproducibility for the LumiCare™ rinse, and performance comparable to the predicate device. No FDA-recognized standards were available to inform the testing.
## Simulated Use Testing
Simulated use testing in a clinical environment was performed by clinicians using a manikin model with extracted teeth placed in a typodont, demonstrating ease and effectiveness of use within the clinical workflow. No FDA-recognized standards were available to inform the testing.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for GreenMark Biomedical Inc. The word "GREENMARK" is written in a combination of green and black letters, with "GREEN" in green and "MARK" in black. To the right of the word "MARK" is a green leaf design. Below the word "GREENMARK" is the phrase "BIOMEDICAL INC." in black letters.
## Biocompatibility
A biocompatibility risk assessment was performed (per 2016 FDA Biocompatibility Guidance), along with cytotoxicity, sensitization, and irritation testing (per ISO 10993-5 and 10993-10), indicating the LumiCare™ rinse is biocompatible for use as intended.
## Shelf Life Testing
Based on accelerated testing, a shelf life of two years is supported. Real time aging is being performed. No FDA-recognized standards were available to inform the testing.
- B. Clinical Performance Testing
Clinical performance testing was not conducted.
## VIII. CONCLUSION
Based upon the indications for use and comparison of technology, together with the results from non-clinical performance testing, we find the LumiCare™ rinse used together with a blue light source, having an irradiance of 450 mW/cm2 and wavelength of 450-470 nm, is substantially equivalent to the predicate device Pulpdent Snoop™.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.