K200595 · CellaVision AB · JOY · Oct 16, 2020 · Hematology
Device Facts
Record ID
K200595
Device Name
CellaVision DC-1, CellaVision DC-1 PPA
Applicant
CellaVision AB
Product Code
JOY · Hematology
Decision Date
Oct 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5260
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K200595 · Oct 16, 2020
CellaVision DC-1, CellaVision DC-1 PPA
CellaVision AB
Leftover clinical samples from laboratory daily workflow
Leftover clinical samples were used to evaluate the precision, reproducibility, and method comparison (against the predicate device and manual microscopy) of the CellaVision DC-1 system for WBC differential counts, RBC morphology characterization, and platelet estimation.
WBC, RBC, and PLT Precision and Reproducibility Studies; Retrospective analysis of leftover clinical samples; Follow-up/Duration: 20 days (repeatability); 5 days (reproducibility)
Patients with normal and abnormal hematological characteristics; Sample Size: Not explicitly stated (multiple samples per site across 3 sites); Number of Sites: 3
Not applicable for this study
Repeatability and reproducibility of WBC pre-classification, RBC pre-characterization, and PLT estimation
Method Comparison with Predicate Device; Retrospective method comparison
598 samples (WBC/PLT) and 586 samples (RBC) from laboratory daily workload; Sample Size: 598 (WBC/PLT); 586 (RBC); Number of Sites: 7
CellaVision DM1200 (predicate device) and manual light microscope
Agreement (PPA, NPA, OA) between DC-1 and predicate device/manual method
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
White blood cell differential count
Convolutional neural network
—
OA 86.8% (95% CI: 83.8%, 89.3%)
—
—
Method comparison study: 598 samples
>1 (skilled operators)
Red blood cell morphology characterization
Convolutional neural network
—
Size OA 88.2% (95% CI: 85.3%, 90.6%); Shape OA 85.2% (95% CI: 82.0%, 87.8%)
—
—
Method comparison study: 586 slides
>1 (skilled operators)
Platelet estimation
—
—
Level 2 OA 96.3%; Level 3 OA 96.2%; Level 4 OA 95.5%
—
—
Method comparison study: 598 slides
>1 (skilled operators)
Indications for Use
CellaVision® DC-1 is an automated cell-locating device intended for in-vitro diagnostic use in clinical laboratories. CellaVision® DC-1 is intended to be used by operators, trained in the use of the device. The Peripheral Blood Application is intended for differential count of white blood cells (WBC), characterization of red blood cell (RBC) morphology and platelet estimation. The CellaVision® DC-1 with the Peripheral Blood Application automatically locates blood cells on peripheral blood (PB) smears. The application presents images of the blood cells for review. A skilled operator trained in recognition of blood cells, identifies and verifies the suggested classification of each cell according to type.
Device Story
Device is an automated cell-locating system for peripheral blood smears; used in clinical laboratories by trained operators. Input: stained peripheral blood smear on glass slide. Operation: device scans slide in battlement pattern using motorized microscope, LED illumination, and digital camera; identifies WBCs and RBC monolayer. Processing: Convolutional Neural Networks (CNN) pre-classify WBCs and pre-characterize RBC morphology; overview image used for platelet estimation. Output: digital images of cells presented on screen for operator review. Clinical workflow: operator verifies or modifies pre-classifications/characterizations; results stored in database; system integrates with LIS for data transfer. Benefit: assists skilled operators in performing differential counts and morphology assessment, improving efficiency in identifying abnormal blood samples.
Clinical Evidence
Clinical and analytical performance evaluated across three sites. Precision (repeatability/reproducibility) assessed per CLSI EP05-A3. Method comparison (N=598 samples) against DM1200 showed WBC distribution OA 89.8%, morphology OA 91.3%. RBC morphology agreement (OA) ranged 79.9%-88.2%. PLT estimation agreement evaluated via Cohen's Kappa (0.894). Results demonstrate no systematic difference between DC-1 and predicate.
Technological Characteristics
System includes motorized microscope, LED illumination, digital color camera, XY stage, and integrated PC with SSD. Materials: glass slides, Romanowsky stain. Connectivity: LIS integration. Software: CellaVision DM Software (CDMS) v7.0 with Peripheral Blood Application (PBA). Algorithm: Convolutional neural network for WBC pre-classification. Calibration: Annual service engineer requirement. Sterilization: Not applicable (in-vitro diagnostic).
Indications for Use
Indicated for patients whose blood samples have been flagged as abnormal by an automated cell counter. Used by trained operators in clinical laboratories for differential count of WBCs, characterization of RBC morphology, and platelet estimation.
Regulatory Classification
Identification
An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K200595
B Applicant
CellaVision AB
C Proprietary and Established Names
CellaVision® DC-1, CellaVision® DC-1 PPA
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JOY | Class II | 21 CFR 864.5260 - Automated Cell-Locating Device | HE - Hematology |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
The measurand for CellaVision DC-1 include white blood cells, red blood cells, and platelet.
C Type of Test:
The CellaVision DC-1 facilitates the differential count of white blood cells (WBC), characterization of red blood cell (RBC) morphology and platelet estimation, by capturing and pre-classifying images of WBCs and RBCs on a stained peripheral blood smear, as well as capturing and presenting an overview image of the blood smear for platelet analysis.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Intended use of CellaVision DC-1
CellaVision DC-1 is an automated cell-locating device intended for in-vitro diagnostic use in clinical laboratories. CellaVision DC-1 is intended to be used by operators, trained in the use of the device.
Intended use of the Peripheral Blood Application
The CellaVision Peripheral Blood Application is intended for differential count of white blood cells (WBC), characterization of red blood cell (RBC) morphology and platelet estimation. The CellaVision DC-1 with the Peripheral Blood Application automatically locates blood cells on peripheral blood (PB) smears. The application presents images of the blood cells for review. A skilled operator trained in recognition of blood cells, identifies and verifies the suggested classification of each cell according to type.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
None
IV Device/System Characteristics:
A Device Description:
The CellaVision DC-1 is a cell-locating device. The CellaVision DC-1 consists of a built-in PC with a Solid-State Disk (SSD) containing CellaVision DM Software (CDMS), a high-power magnification focusing nosepiece microscope with a LED illumination, an XY stage, a proprietary camera with firmware, built-in motor- and illumination LED controller, a casing and an external power supply.
The CDMS is the software that controls the collection of cell images, stores the images of the located cells and the results, and displays the images in an organized manner to the operator.
The Peripheral Blood Application (PBA) is a software module included as part of the CDMS. The PBA performs pre-classification of WBCs, pre-characterization of RBCs, and facilitates platelet estimation by presenting an overview image captured of a monolayer area.
K200595 - Page 2 of 17
{2}
B Principle of Operation:
The typical workflow with respect to the CellaVision DC-1 is as follows.
- From a peripheral blood sample collected in K₂EDTA or K₃EDTA tube, typically flagged by a cell-counter indicating an abnormal morphology, a thin blood film is wedged on a glass slide (a blood smear). The blood smear is then stained with Romanowsky stain. The system uses stained microscope slides made either manually or by using an automatic slide maker-stainer.
- The operator enters the order ID for the slide, either manually or using an optional barcode reader. If a Laboratory Information System (LIS) is used, the device automatically fetches order data for the sample from the LIS.
- The operator places the slide in the loading tray of the device and closes the input hatch. The device automatically moves the slide under the microscope. Starting at 33 mm from the edge of the slide, the device looks for a monolayer in the smear. Once a monolayer is found, the device scans the monolayer in a battlement pattern. While doing this, the device locates any WBCs and stores high quality images of each located WBC. The device also locates and stores an overview image of a part of the RBC monolayer.
- The device software, PBA, pre-classifies each located WBC. It also pre-characterizes the RBC morphology based on the overview image. By using the overview image, the operator can calculate or estimate the level of PLTs.
- During the review/verification stage, a skilled operator, trained in the use of the software and in recognition of blood cells, then opens the order to review and verify (or modify) the preliminary results. The review can be done either at the device or using the CellaVision Remote Review Software (CRRS).
Instrument Description Information:
1. Instrument Name: CellaVision DC-1
2. Specimen Identification: All slides must be labeled clearly with patient or order information. To use the optional handheld barcode reader, slides must be labeled with a barcode label.
3. Specimen Sampling and Handling: Slides are processed one at a time. To load slide, manually open the input hatch, place the slide, with the blood smear facing up, in the loading tray, and close the hatch.
4. Calibration: CellaVision recommends that calibration is performed once a year by a service engineer.
5. Quality Control:
K200595 - Page 3 of 17
{3}
Self-Test - The CellaVision DC-1 analyzer performs self-tests during startup of the software, and at certain points during the operation of the analyzer. When the software starts, the analyzer is checked before the user can start analyzing. During this phase, both the hardware and the software components are tested for anomalies, as well as various requirements for the operation of the analyzer.
Cell Location Test - A cell location test shall be run at least once a day and after any changes in the staining procedure or the staining solutions for quality control. Cell location slides are prepared and processed by the customer from a freshly stained slide from a blood sample with a WBC count within the normal range. The cell location test verifies that the quality of the slide preparation process is good enough to allow the analyzer to locate the cells needed for the analysis. It also verifies the analyzer's ability to locate cells.
# V Substantial Equivalence Information:
A Predicate Device Name(s):
Cellavision DM1200 Automated Hematology Analyzer, Model Xu-10127
B Predicate 510(k) Number(s):
K092868
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K200595 | K092868 |
| --- | --- | --- |
| Device Trade Name | CellaVision DC-1 | CellaVision DM1200 |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | CellaVision DC-1 is an automated cell-locating device intended for in-vitro diagnostic use in clinical laboratories. CellaVision DC-1 is intended to be used by operators, trained in the use of the device. Peripheral Blood Application: The CellaVision Peripheral Blood Application is intended for differential count of white blood cells (WBC), characterization of red blood cell (RBC) morphology and platelet estimation. The CellaVision DC-1 with the Peripheral Blood Application | CellaVision DM1200 is an automated cell-locating device. CellaVision DM1200 automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type. CellaVision DM1200 is intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells. |
K200595 - Page 4 of 17
{4}
K200595 - Page 5 of 17
| Device & Predicate Device(s): | K200595 | K092868 |
| --- | --- | --- |
| Device Trade Name | CellaVision DC-1 | CellaVision DM1200 |
| | automatically locates blood cells on peripheral blood (PB) smears. The application presents images of the blood cells for review. A skilled operator trained in recognition of blood cells, identifies and verifies the suggested classification of each cell according to type. | |
| Intended use population | The intended use population is patients whose blood samples have been flagged as abnormal by an automated cell counter. | Same |
| Analytes | Automated cell-locating device for cell-location and identification of RBC, WBC or platelets for in-vitro use.
Verification of results by human operator. | Same |
| Light source | LED (Light Emitting Diode) | Same |
| Classification software | CellaVision DM Software including Peripheral blood application (version 7.0) | CellaVision DM Software including Peripheral blood application (version 6.0) |
| Sample type | Stained blood film on glass slides of peripheral whole blood. | Same |
| Sample preparation | Romanowsky stain | Same |
| Analysis technique | • White blood cells:
Cells are located/counted by moving according to the battlement track pattern. Cell images are analyzed using artificial intelligence trained to distinguish between classes of white blood cells.
The cell images are preclassified and the operator verifies the suggested classification by accepting or reclassifying the white blood cells.
• Red blood cells:
The device presents an overview image. | Same |
{5}
K200595 - Page 6 of 17
| Device & Predicate Device(s): | K200595 | K092868 |
| --- | --- | --- |
| Device Trade Name | CellaVision DC-1 | CellaVision DM1200 |
| | The cell images are precharacterized and the operator verifies or re-characterizes red blood cell morphology from the image.
• Platelets:
The device presents an overview image (corresponding to eight high power fields).
The operator manually counts and estimates the platelet concentration from the overview image according to a standardized procedure. | |
| Optical means for magnifying images of white blood cells for observation and interpretation | Individual white blood cells are magnified and imaged on a camera sensor by a microscope. The camera sensor produces images on a screen for view and interpretation (cell class verification). | Same |
| Viewing of white blood cell images | Individual located white blood cells are presented in an organized manner and observed on a screen. | Same |
| Classification of white blood cells | White blood cells are pre-classified and presented to the operator. To complete the differential, the operator needs to verify that all located WBCs are correctly classified. All cells must be classified and verified before the order can be signed. | Same |
| Characterization of the red blood cell morphology | An overview image is collected and presented from which the operator can characterize the red blood cells morphology. | Same |
| Estimating the platelet level | From an overview image corresponding to eight high power fields the platelet level is estimated. | Same |
| Image interpretation requirements | A skilled operator is required to differentiate and finally modify and/or confirm the preclassification/characterization of | Same |
{6}
K200595 - Page 7 of 17
| Device & Predicate Device(s): | K200595 | K092868 |
| --- | --- | --- |
| Device Trade Name | CellaVision DC-1 | CellaVision DM1200 |
| | the located blood cells. | |
| Information transfer from instrument to Printer or network | The system can interact with a laboratory information system (LIS). The system will retrieve order data from the LIS and send results back to the LIS. | Same |
| Result format | The differential proportional count is normally based on 100 white blood cells. The number of WBCs can be modified if required. The result can be presented as an absolute number or as % of total number of WBCs. The result of RBC characterization is presented as a grading for each morphology. | Same |
| Technological characteristics | The system locates white blood cells, stores digital images of the cells and suggests a cell class for each white blood cell to aid operators in performing the white blood cell differential procedure. The system captures an overview image of the RBC monolayer for RBC characterization. The operator estimates the platelet concentration using the overview image. | Same |
| Operators competence | The operator is trained in the recognition of blood cells and in the use of the device. | Same |
| Decision support | The device includes white blood cell reference cells. The operator can add his/her own reference cells. | Same |
| Calibration | Recommended calibration once a year by a service engineer. | Same |
| General Device Characteristic Differences | | |
| Major parts of the system | • Computer module (integrated)
• Motorized microscope
• Digital color camera | • System computer (stand-alone)
• Motorized microscope
• Digital color camera |
{7}
| Device & Predicate Device(s): | K200595 | K092868 |
| --- | --- | --- |
| Device Trade Name | CellaVision DC-1 | CellaVision DM1200 |
| | • XY stage
• Control unit (integrated in camera)
• Casing
• Database | • Robot gripper unit
• Control unit
• Casing
• Database
• Immersion oil unit
• Barcode reader |
| Neural network | Neural network of convolutional type. | Artificial neural network of multiple perception type. |
| Loading capacity | 1 slide | 12 slides |
| Immersion oil application | Manual application | Automatic application |
VI Standards/Guidance Documents Referenced:
Guidance Documents
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, October 2014
- Guidance for Industry – Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software, January 2005
- Guidance for Industry, FDA Reviewer and Compliance on Off-The-Shelf Use in Medical Devices, September 1999
- General Principles of Software Validation; Final Guidance for industry and FDA Staff, January 2002
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 2005
- Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, and Food and Drug Administration Staff, April 2006
- Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests - Guidance for Industry and FDA Staff, March 2007
FDA-recognized consensus standards
- ASTM D4169-16 - Standard Practice for Performance Testing of Shipping Containers and Systems (Recognition Number: 14-499)
- AUTO11-A2 - Information Technology Security of In Vitro Diagnostic Instruments and Software Systems; Approved Standard - Second Edition (Recognition Number: 13-85)
- CLSI EP05-A3 - Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition, (Recognition Number: 7-251)
- CLSI EP12-A2 - User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition (Recognition Number: 7-152)
K200595 - Page 8 of 17
{8}
- CLSI H20-A2 - Reference Leukocyte (WBC) Differential Count (Proportional) and Evaluation of Instrumental Methods; Approved Standard - Second Edition, (Recognition Number: 7-165)
- IEC 62366-1:2015 - Medical devices - Part 1: Application of usability engineering to medical devices [Including Corrigendum 1 (2016)], (Recognition Number: 5-114)
- ISO 14155:2011 - Clinical investigations of medical devices for human subject- Good clinical practice, (Recognition Number: 2-205)
- ISO 14971:2007 - Medical devices - Application of risk management to medical devices, (Recognition Number: 5-40)
- ISO 15223-1:2016 - Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements, (Recognition Number: 5-117)
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
#### 1. Precision/Reproducibility:
Three repeatability studies were performed for white blood cell (WBC), red blood cell (RBC), and platelet (PLT), respectively to demonstrate the repeatable performance of CellaVision DC-1 across multiple days. Three reproducibility studies were also conducted for WBC, RBC, and PLT to demonstrate that the results generated by the CellaVision DC-1 are reproducible across sites. Each study was conducted at one U.S. site and two sites outside the U.S. The study designs and representative results are summarized below.
#### i. WBC Pre-classification Repeatability
The repeatability of WBC pre-classification results generated by CellaVision DC-1 was evaluated across multiple days. At each site, the study included five samples, one was normal sample and four displayed various clinical pathological characteristics (i.e. abnormal samples). The study samples were selected from leftover samples from the laboratory's daily workflow. All samples were initially analyzed on an automated cell counter available at the site. Each sample was made into two slides to represent two replicates (i.e. both slides were analyzed once in each run to represent two replicate measurements). Each sample was run twice per day for 20 days, resulting in $20 \times 2 \times 2$ measurements. The proportional cell count in percent for each cell class was used to estimate total variance and variance components for repeatability, including within-run, between-run, between-day, and within-laboratory, by an ANOVA. The WBC results were evaluated separately for each sample and site.
#### ii. RBC Pre-characterization Repeatability
The repeatability of RBC pre-characterization results generated by CellaVision DC-1 was evaluated across multiple days. At each site, the study included five samples, one was normal sample and four displayed abnormal levels of Polychromatic cells,
K200595 - Page 9 of 17
{9}
Hypochromatic cells, Anisocytosis, Microcytes, Macrocytes, Poikilocytosis (i.e. abnormal samples). The study samples were selected from leftover samples from the laboratory's daily workflow. All five samples were initially analyzed on an automated cell counter available at the site. Each sample was made into two slides to represent two replicates (i.e. both slides were analyzed once in each run to represent two replicate measurements). Each sample was run twice per day for 20 days, resulting in 20 × 2 × 2 measurements. The repeatability in terms of grade (0, 1, 2 or 3) for each morphological characteristic was evaluated. The most prevalent grade reported was evaluated, and the agreement (percentage of runs reporting the grade) was calculated for each morphology. The agreement was evaluated for each slide and run, as well as for each sample. The RBC results were evaluated separately for each site.
iii. PLT Estimation Repeatability
The repeatability of PLT estimation using the overview image captured by the CellaVision DC-1 was evaluated across multiple days. The procedure for estimating PLT concentration with the overview image provided by CellaVision DC-1 includes: 1) First, the overview image is divided into multiple grids and the operator counts the number of platelet in each grid and enters the counted values for all corresponding grids to estimate the concentration of PLT in the sample; 2) Based on the estimated concentration of PLT, the device reports a grade (1 = significant decreased, 2 = decreased, 3 = normal, and 4 = increased) corresponding to the predefined ranges. At each site, the study included four samples, one was normal sample (~250 × 10⁹/L) and three abnormal samples, including one significantly decreased (~30 × 10⁹/L), one decreased (~100 × 10⁹/L), and one increased (~600 × 10⁹/L) samples. The study samples were selected from leftover samples from the laboratory's daily workflow. All four samples were initially analyzed on an automated cell counter available at the site. Each sample was made into two slides to represent two replicates (i.e. both slides were analyzed once in each run to represent two replicate measurements). Each sample was run twice per day for 20 days, resulting in 20 × 2 × 2 measurements. The agreement (i.e. percentage of runs reporting the actual PLT level) was evaluated for each PLT level. The PLT results were evaluated separately for each site.
iv. WBC Reproducibility
A three-site reproducibility study was conducted to demonstrate that the WBC results obtained from using the CellaVision DC-1 were reproducible across multiple sites (devices). The study included five samples, one was normal sample and four displayed various clinical pathological characteristics (i.e. abnormal samples). The study samples were selected from leftover samples from the laboratories' daily workflow. All five samples were initially analyzed on an automated cell counter. Each sample was made into five slides to represent five replicates (i.e. five slides were analyzed once in each run to represent five replicate measurements) that were run once per day for five days per site, resulting in 3 × 5 × 5 measurements in total. In the primary analysis, the operator-verified results, including the proportional cell count in percent for each cell class, were used to estimate total variation and variance. In addition to analyzing the operator-verified
K200595 - Page 10 of 17
{10}
results, a secondary analysis was performed on the preclassification outputs generated by the CellaVision DC-1 (i.e. without operator verification).
# v. RBC Reproducibility
A three-site reproducibility study was conducted to demonstrate that the RBC results obtained from using the CellaVision DC-1 were reproducible. The study included four samples, one was normal sample and three displayed abnormal levels of Polychromatic cells, Hypochromatic cells, Anisocytosis, Microcytes, Macrocytes, Poikilocytosis (i.e. abnormal samples). The study samples were selected from leftover samples from the laboratory's daily workflow. All four samples were initially analyzed on an automated cell counter available at the site. Each sample was made into five slides to represent five replicates (i.e. five slides were analyzed once in each run to represent five replicate measurements). The slides were run once per day for five days per site, resulting in $3 \times 5$ x 5 measurements in total. Reproducibility in terms of grade (0, 1, 2 or 3) for each morphological characteristic was evaluated. The most prevalent grade reported was evaluated, and the agreement (percentage of runs reporting the grade) was calculated.
# vi. PLT Reproducibility
A three-site reproducibility study was conducted to demonstrate that the PLT results obtained from using the overview image provided by CellaVision DC-1 were reproducible. The study included four samples, one was normal sample ( $\sim 250 \times 10^{9} / \mathrm{L}$ ) and three abnormal samples, including one significantly decreased ( $\sim 30 \times 10^{9} / \mathrm{L}$ ), one decreased ( $\sim 100 \times 10^{9} / \mathrm{L}$ ), and one increased ( $\sim 600 \times 10^{9} / \mathrm{L}$ ) samples. The study samples were leftover samples selected from different laboratories' daily workflow. All four samples were initially analyzed on an automated cell counter. Each sample was made into five slides to represent five replicates (i.e. five slides were analyzed once in each run to represent five replicate measurements) that were run once per day for five days per site, resulting in $3 \times 5 \times 5$ measurements in total. For each PLT level (significantly decreased, decreased, normal or increased) the agreement (i.e. percentage of runs reporting the actual PLT level) was evaluated. The agreement was evaluated for each sample and site and representative results by sample are presented below.
Agreement $(\%)$ per sample
| Sample | Level 1 | Level 2 | Level 3 | Level 4 | Total number of runs (n) | Agreement % |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | 0 | 0 | 1 | 74 | 75 | 98.6 |
| 2 | 0 | 75 | 0 | 0 | 75 | 100 |
| 3 | 75 | 0 | 0 | 0 | 75 | 100 |
| 4 | 0 | 2 | 73 | 0 | 75 | 97.3 |
# 2. Linearity:
Not applicable.
K200595 - Page 11 of 17
{11}
3. Analytical Specificity/Interference:
Not applicable.
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Not applicable.
6. Detection Limit:
Not applicable.
7. Assay Cut-Off:
Not applicable.
8. Accuracy (Instrument):
Not applicable.
9. Carry-Over:
Not applicable.
## B Comparison Studies:
1. Method Comparison with Predicate Device:
Three multi-site method comparison studies were conducted for WBC, RBC, and PLT, respectively to demonstrate that the CellaVision DC-1 (DC-1) performs equivalently to the predicate device, CellaVision DM1200 (DM1200). Each study involved three sites, one of which was a U.S. site. A total of seven sites participated in the three studies.
i. WBC Comparison
A total of 598 samples were analyzed on both the DM1200 and on the DC-1. The samples were leftover samples from the laboratory's daily workload. Each site collected approximately 200 samples, 100 of which were normal (non-diseased) and 100 were samples with various clinical pathological characteristics (abnormal samples). Normal and abnormal samples were defined in accordance with CLSI H20-A2. The samples covered a wide range of ages and both genders. Three slides (A, B, and Spare) were prepared from each sample and stained using Romanowsky stain as per standard clinical practice. The operator-verified WBC results from the DC-1 were compared to the
K200595 - Page 12 of 17
{12}
corresponding results from the DM1200. The results of one slide (slide A) from the DC-1 were compared to the mean results (slide A and slide B) from the DM1200, also known as the single slide evaluation.
## WBC Evaluation
In this evaluation, the results from one slide per sample (slide A) from the DC-1 were compared to the mean results (slides A and B) from the DM1200. This evaluation was referred to as the duplicate slide evaluation.
DC-1 slide A versus DM1200 mean: Total Distribution Agreement
| | DM1200 Positive | DM1200 Negative | Total |
| --- | --- | --- | --- |
| DC-1 Positive | 259 | 32 | 291 |
| DC-1 Negative | 47 | 260 | 307 |
| Total | 306 | 292 | 598 |
| | Result | 95% CI | |
| --- | --- | --- | --- |
| | | LL | UL |
| OA | 86.8% | 83.8% | 89.3% |
| PPA | 84.6% | 80.2% | 88.2% |
| NPA | 89.0% | 84.9% | 92.1% |
DC-1 slide A versus DM1200 mean: Total Morphology Agreement
| | DM1200 Positive | DM1200 Negative | Total |
| --- | --- | --- | --- |
| DC-1 Positive | 162 | 48 | 210 |
| DC-1 Negative | 22 | 366 | 388 |
| Total | 184 | 414 | 598 |
| | Result | 95% CI | |
| --- | --- | --- | --- |
| | | LL | UL |
| OA | 88.3% | 85.5% | 90.6% |
| PPA | 88.0% | 82.6% | 92.0% |
| NPA | 88.4% | 85.0% | 91.1% |
Regression analysis was performed for WBC classes that have normal range of 5% or more, comparing slide A results from DC-1 to mean results from DM1200. Segmented Neutrophils, Lymphocytes, Eosinophils, and Monocytes were included in this analysis. The fit results are summarized below.
| Cell Type | Regression Fit | R² |
| --- | --- | --- |
| Segmented Neutrophils | Y = 0.9926X + 0.0027 | 0.969 |
| Lymphocytes | Y = 0.9799X + 0.0017 | 0.946 |
| Eosinophils | Y = 0.9884X + 0.0014 | 0.928 |
| Monocytes | Y = 0.9838X + 0.0025 | 0.897 |
ii. RBC Comparison
K200595 - Page 13 of 17
{13}
A total of 586 slides (one slide per sample) were analyzed on both the DC-1 and the DM1200 across three sites. The samples were leftover samples from the laboratory's daily workload. Each site collected approximately 200 samples, with 50 samples for each group: normal, color (Polychromasia, Hypochromasia), size (Anisocytosis, Microcytosis, Macrocytosis) and shape (Poikilocytosis). The samples covered a wide range of ages and both genders. The slides were divided into four sets, one for each operator (e.g. 50 slides in each). Each of the operators verified their own allocated slides on both the DM1200 and the DC-1.
The RBCs were characterized for each morphology according to a predefined 4-graded scale (normal (0), slight (1), moderate (2) or marked (3)). The operator-verified RBC results from the DC-1 were compared to the corresponding results from the DM1200. PPA, NPA and OA with two-sided 95% confidence intervals, comparing the DC-1 results with the DM1200 results, were calculated. The results were assumed to be normally distributed and calculations of 95% confidence intervals were performed as described in CLSI EP12-A2. The overall agreement results for each group are presented below.
Morphology reference limits
| Group | RBC characteristics | Abnormal grade |
| --- | --- | --- |
| Color | Polychromatic cells | ≥1 |
| | Hypochromatic cells | ≥2 |
| Size | Anisocytosis | ≥1 |
| | Microcytes | ≥1 |
| | Macrocytes | ≥2 |
| Shape | Poikilocytosis | ≥1 |
Color: DC-1 versus DM1200
| | DM1200 Positive | DM1200 Negative | Total |
| --- | --- | --- | --- |
| DC-1 Positive | 159 | 96 | 255 |
| DC-1 Negative | 22 | 309 | 331 |
| Total | 181 | 405 | 586 |
| | Result | 95% CI | |
| --- | --- | --- | --- |
| | | LL | UL |
| OA | 79.9% | 76.4% | 82.9% |
| PPA | 87.8% | 82.3% | 91.8% |
| NPA | 76.3% | 71.9% | 80.2% |
Size: DC-1 versus DM1200
| | DM1200 Positive | DM1200 Negative | Total |
| --- | --- | --- | --- |
| DC-1 Positive | 354 | 29 | 383 |
| DC-1 Negative | 40 | 162 | 202 |
| Total | 394 | 191 | 585* |
*One sample was excluded due to missing grading information for Microcytes.
K200595 - Page 14 of 17
{14}
| | Result | 95% CI | |
| --- | --- | --- | --- |
| | | LL | UL |
| OA | 88.2% | 85.3% | 90.6% |
| PPA | 89.8% | 86.3% | 92.2% |
| NPA | 84.8% | 79.0% | 89.2% |
Shape: DC-1 versus DM1200
| | DM1200 Positive | DM1200 Negative | Total |
| --- | --- | --- | --- |
| DC-1 Positive | 199 | 58 | 257 |
| DC-1 Negative | 29 | 300 | 329 |
| Total | 228 | 358 | 586 |
| | Result | 95% CI | |
| --- | --- | --- | --- |
| | | LL | UL |
| OA | 85.2% | 82.0% | 87.8% |
| PPA | 87.3% | 82.3% | 91.0% |
| NPA | 83.8% | 79.6% | 87.3% |
RBC Supplemental Study: DC-1 versus manual light microscope
To supplement the results for the Color group, the sponsor conducted a study in which results for the group Color from the DC-1 were compared to the results from manual light microscope. An experienced operator analyzed 70 slides made up of normal and abnormal (polychromatic and hypochromatic) samples on both the DC-1 and the manual light microscope with a wash-out period between the two devices. The results are summarized below.
Color: DC-1 versus Manual Light Microscope
| | MLM Positive | MLM Negative | Total |
| --- | --- | --- | --- |
| DC-1 Positive | 19 | 10 | 29 |
| DC-1 Negative | 0 | 41 | 41 |
| Total | 19 | 51 | 70 |
| | Result | 95% CI | |
| --- | --- | --- | --- |
| | | LL | UL |
| OA | 85.7% | 75.7% | 92.0% |
| PPA | 100% | 83.2% | 100% |
| NPA | 80.4% | 67.5% | 89.0% |
iii. PLT Comparison
A total of 598 slides were analyzed on both the DC-1 and the DM1200 across three sites. The samples were leftover samples from the laboratory's daily workload. Each site
K200595 - Page 15 of 17
{15}
collected 200 samples (50 from each PLT level based on cell counter results). The samples covered a wide range of patient demographics. All samples were collected based on the results from the automated cell counters. Each level was assigned a number: 1= Significant Decreased, 2= Decreased, 3=Normal, 4=Increased. PPA, NPA and OA for each PLT level were calculated by using 2x2 contingency tables. The 95% confidence intervals were calculated according to CLSI EP12-A2.
PLT analysis between the DC-1 device and the DM1200
| Site | Agreement | Level 2
(95% CI) | Level 3
(95% CI) | Level 4
(95% CI) |
| --- | --- | --- | --- | --- |
| All combined
n=598 | OA | 96.3%
(94.5%, 97.6%) | 96.2%
(94.3%, 97.4%) | 95.5%
(93.5%, 96.9%) |
| | PPA | 95.9%
(91.3%, 98.1%) | 96.7%
(94.1%, 98.2%) | 96.7%
(94.7%, 97.9%) |
| | NPA | 96.5%
(94.3%, 97.8%) | 95.5%
(92.5%, 97.4%) | 88.6%
(80.3%, 93.7%) |
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
F Other Supportive Instrument Performance Characteristics Data:
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
K200595 - Page 16 of 17
{16}
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K200595 - Page 17 of 17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.