K200574 · Ekso Bionics, Inc. · PHL · Jun 19, 2020 · Physical Medicine
Device Facts
Record ID
K200574
Device Name
EksoNR
Applicant
Ekso Bionics, Inc.
Product Code
PHL · Physical Medicine
Decision Date
Jun 19, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3480
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EksoNR™ is intended to perform ambulatory functions in rehabilitation institutions under the supervision of a trained physical therapist for the following populations: • Individuals with acquired brain injury, including traumatic brain injury and stroke (upper extremity motor function of at least 4/5 in at least one arm). • Individuals with spinal cord injuries at levels T4 to L5 (upper extremity motor function of at least 4/5 in both arms). • Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D with upper extremity motor function of at least 4/5 in both arms). The therapist must complete a training program prior to use of the device. The devices are not intended for sports or stair climbing.
Device Story
EksoNR is a powered motorized orthosis; consists of fitted metal brace supporting legs, feet, and torso; worn via straps. Battery-powered motors drive hip and knee joints. Operated by physical therapist using handheld interface to initiate steps; patient weight shift also triggers movement. Used in rehabilitation clinics with assistive devices like walkers, crutches, or canes. Provides visual/auditory feedback to therapist. Facilitates gait ambulation for patients with acquired brain injury or spinal cord injury. Failsafe mechanism locks knees and frees hips during power loss. Benefits include assisted mobility and gait training in clinical settings.
Clinical Evidence
Two prospective, open-label, non-comparative studies (Kessler ABI, n=24; Kessler TBI, n=25). Kessler ABI study showed improvement in Motor FIM (20.4 to 54.2) and Cognitive FIM (12.2 to 23.0) scores. Kessler TBI study showed stable 10MWT (17 to 18s), 2MWT (126 to 128m), and TUG (19.3 to 21.7s) scores. No falls or adverse events reported in either study. Data supports safety and efficacy for ABI/TBI populations.
Indicated for individuals with acquired brain injury (including stroke and TBI) with upper extremity motor function ≥4/5 in at least one arm, or spinal cord injuries (T4-L5 with ≥4/5 motor function in both arms; C7-T3 ASIA D with ≥4/5 motor function in both arms). Requires supervision by trained physical therapist.
Regulatory Classification
Identification
A powered lower extremity exoskeleton is a prescription device that is composed of an external, powered, motorized orthosis that is placed over a person's paralyzed or weakened limbs for medical purposes.
Special Controls
In combination with the general controls of the FD&C Act, the Powered Exoskeleton is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Elements of the device materials that may contact the patient must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility/interference (EMC/EMI), electrical safety, thermal safety, mechanical safety, battery performance and safety, and wireless performance, if applicable.
(3) Appropriate software verification, validation, and hazard analysis must be performed.
(4) Design characteristics must ensure geometry and materials composition are consistent with intended use.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must include:
(i) Mechanical bench testing (including durability testing) to demonstrate that the device will withstand forces, conditions, and environments encountered during use;
(ii) Simulated use testing (
*i.e.,* cyclic loading testing) to demonstrate performance of device commands and safeguard under worst case conditions and after durability testing;(iii) Verification and validation of manual override controls are necessary, if present;
(iv) The accuracy of device features and safeguards; and
(v) Device functionality in terms of flame retardant materials, liquid/particle ingress prevention, sensor and actuator performance, and motor performance.
(6) Clinical testing must demonstrate a reasonable assurance of safe and effective use and capture any adverse events observed during clinical use when used under the proposed conditions of use, which must include considerations for:
(i) Level of supervision necessary, and
(ii) Environment of use (
*e.g.,* indoors and/or outdoors) including obstacles and terrain representative of the intended use environment.(7) A training program must be included with sufficient educational elements so that upon completion of training program, the clinician, user, and companion can:
(i) Identify the safe environments for device use,
(ii) Use all safety features of device, and
(iii) Operate the device in simulated or actual use environments representative of indicated environments and use.
(8) Labeling for the Physician and User must include the following:
(i) Appropriate instructions, warning, cautions, limitations, and information related to the necessary safeguards of the device, including warning against activities and environments that may put the user at greater risk.
(ii) Specific instructions and the clinical training needed for the safe use of the device, which includes:
(A) Instructions on assembling the device in all available configurations;
(B) Instructions on fitting the patient;
(C) Instructions and explanations of all available programs and how to program the device;
(D) Instructions and explanation of all controls, input, and outputs;
(E) Instructions on all available modes or states of the device;
(F) Instructions on all safety features of the device; and
(G) Instructions for properly maintaining the device.
(iii) Information on the patient population for which the device has been demonstrated to have a reasonable assurance of safety and effectiveness.
(iv) Pertinent non-clinical testing information (
*e.g.,* EMC, battery longevity).(v) A detailed summary of the clinical testing including:
(A) Adverse events encountered under use conditions,
(B) Summary of study outcomes and endpoints, and
(C) Information pertinent to use of the device including the conditions under which the device was studied (
*e.g.,* level of supervision or assistance, and environment of use (*e.g.,* indoors and/or outdoors) including obstacles and terrain).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 15, 2020
Ekso Bionics, Inc. Jack Peurach CEO 1414 Harbour Way South, Suite 1201 Richmond, CA 94804
Re: K200574
Trade/Device Name: EksoNRTM Regulation Number: 21 CFR 890.3480 Regulation Name: Powered Lower Extremity Exoskeleton Regulatory Class: Class II Product Code: PHL Dated: February 27, 2020 Received: March 5, 2020
Dear Mr. Peurach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Acting Assistant Director, Acute Injury Devices DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
## 510(k) Number (if known) K200574 Device Name EksoNR™M
Indications for Use (Describe)
The EksoNR™ is intended to perform ambulatory functions in rehabilitations under the supervision of a trained physical therapist for the following populations:
- · Individuals with acquired brain injury, including traumatic brain injury and stroke (upper extremity motor function of at least 4/5 in at least one arm).
- · Individuals with spinal cord injuries at levels T4 to L5 (upper extremity motor function of at least 4/5 in both arms).
- · Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D with upper extremity motor function of at least 4/5 in both arms).
The therapist must complete a training program prior to use of the devices are not intended for sports or star climbing.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary as required by 21 CFR 807.92(c)
| Device name | Ekso |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitters name &<br>contact info | Ekso Bionics® Inc.<br>1414 Harbour Way South<br>Suite 1201<br>Richmond, CA 94804 |
| | <b>Contact Details:</b><br>Jack Peurach<br>CEO |
| | Tel: +1 510-260-7610<br>Email: jack@eksobionics.com |
| | Ekso Bionics Tel: +1 (510) 984-1761<br>Ekso Bionics Fax: +1 (510) 927-2647 |
| Preparation Date | February 27, 2020 |
| Device Name &<br>Classification | Trade Name: EksoNR™<br>Common Name: Exoskeleton<br>Classification Name: Powered Exoskeleton<br>Device Classification: Class II, 21 CFR 890.3480<br>Product Code: PHL |
| Legally Marketed | K161443, Ekso, Ekso Bionics, Inc. |
| Device Description | The Ekso is a powered motorized orthosis. It consists of a fitted metal brace that<br>supports the legs, feet, and torso. It is worn via straps on the body, legs, and feet.<br>Battery powered motors drive knee and hip joints. It has an integrated solid torso<br>containing the computer and power supply. It has a hand-held user interface to<br>specify settings and initiate steps. The Ekso is used with a cane, crutch, or walker. |
| Indication for Use<br>Statement | The EksoNR™ is intended to perform ambulatory functions in rehabilitation<br>institutions under the supervision of a trained physical therapist for the following<br>populations:<br>• Individuals with acquired brain injury, including traumatic brain injury and<br>stroke (upper extremity motor function of at least 4/5 in at least one arm).<br>• Individuals with spinal cord injuries at levels T4 to L5 (upper extremity<br>motor function of at least 4/5 in both arms).<br>• Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D with upper<br>extremity motor function of at least 4/5 in both arms).<br>The therapist must complete a training program prior to use of the device. The<br>devices are not intended for sports or stair climbing. |
| Substantial Equivalence<br>Discussion | This device and the previously cleared (predicate) device (K161443) are essentially the same products. |
| | Differences in Indications for Use |
| | The purpose of this 510(k) is to update the indications for use relative to the predicate device. The indications for use are identical to that of the predicate device, other than the following (changed wording is <b>bolded</b> ): |
| | Individuals with hemiplegia due to stroke (upper extremity motor function of at least 4/5 in at least one arm) |
| | has been updated to: |
| | Individuals with an acquired brain injury, including traumatic brain injury and stroke (upper extremity motor function of at least 4/5 in at least one arm) |
| | The change noted above expands the existing stroke indication to the broader acquired brain injury (ABI) population. The intended use of the device is unchanged. For the purposes of this intended use (ambulatory rehabilitation), ABI patients present similarly, are screened for device safety requirements including sufficient cognitive ability to follow instructions and safely use the device, and are treated in the same way as stroke patients (a subset of ABI). As such, gait ambulation effectiveness of the device is no different when used on the general ABI population. When used as instructed, the device is as safe to use with the general ABI population as the already cleared stroke population. |
{4}------------------------------------------------
#### Technical Characteristics
The device is essentially unchanged from the current (predicate) device. All changes since the previous submission fall below the threshold requiring a 510(k) per the FDA's October 25, 2017, "Guidance for Industry and FDA Staff: Deciding When to Submit a 510(k) for a Change to an Existing Device". Even though there is no change, for completeness, the technical characteristics of the device are compared with the current device below.
| Comparison of Technical Characteristics | | | |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|------------------------|
| | | Current Device | |
| Manufacturer | Ekso Bionics®, Inc. | Ekso Bionics®, Inc. | |
| Trade Name | EksoNR™ | Ekso GT® and EksoNR™ | Differences |
| 510(k) Number | | K161443 | N/A |
| Product Code | PHL | PHL | Same |
| Regulation Number | 890.3480 | 890.3480 | Same |
| Regulation Name | Powered Exoskeleton | Powered Exoskeleton | Same |
| Body Coverage | Worn over legs and upper<br>body with rigid torso | Worn over legs and upper<br>body with rigid torso | Same |
| Size of Components | Adjustable upper leg, lower<br>leg, and hip width; control unit<br>integrated into the torso | Adjustable upper leg, lower<br>leg, and hip width; control unit<br>integrated into the torso | Same |
| Mobility Aid | Walker, Crutches, Cane | Walker, Crutches, Cane | Same |
| Comparison of Technical Characteristics | | | |
| | | Current Device | |
| Manufacturer | Ekso Bionics®, Inc. | Ekso Bionics®, Inc. | |
| Trade Name | EksoNR™ | Ekso GT® and EksoNR™ | Differences |
| Ability of User Mobility | Sit, stand, walk, and<br>turn | Sit, stand, walk, and<br>Turn | Same |
| Walking Speed | ~2 km/hr | ~2 km/hr | Same |
| Grade of Inclination | 1.15 deg | 1.15 deg | Same |
| Type of Surface | Smooth, cement, carpet | Smooth, cement, carpet | Same |
| Height of Patient | ~62" to 74" (1.58 m to 1.88 m) | ~62" to 74" (1.58 m to 1.88 m) | Same |
| Weight of Patient | Up to 220 lbs (100kg) | Up to 220 lbs (100kg) | Same |
| Control Method | Handheld interface for PT;<br>weight shift to initiate steps | Handheld interface for PT;<br>weight shift to initiate steps | Same |
| Range of Motion | Hips: 135° flexion to 20°<br>extension<br>Knees: 130° flexion to 0°<br>extension<br>Ankles: 10° flexion to 10°<br>extension | Hips: 135° flexion to 20°<br>extension<br>Knees: 130° flexion to 0°<br>extension<br>Ankles: 10° flexion to 10°<br>extension | Same |
| Device Weight | 50 lbs (23 kg) | 50 lbs (23 kg) | Same |
| Rechargeable Battery | Rechargeable lithium ion<br>batteries 48.1V, 30A peak<br>current, 1 hour of continuous<br>usage per charge | Rechargeable lithium ion<br>batteries 48.1V, 30A peak<br>current, 1 hour of continuous<br>usage per charge | Same |
| Battery Charge Time | 1 hour | 1 hour | Same |
| Expected Useable Life | 4 years | 4 years | Same |
| Training Program | Yes | Yes | Same |
| Certification Program | Yes | Yes | Same |
| User Feedback | Provides visual feedback on the<br>handheld controller and<br>auditory feedback | Provides visual feedback on the<br>handheld controller and<br>auditory feedback | Same |
| Fall Detection and<br>Mitigation | None | None | Same |
| Failsafe Feature | In event of power failure-<br>knees become locked and hips<br>free (similar to typical passive<br>leg braces) | In event of power failure-<br>knees become locked and hips<br>free (similar to typical passive<br>leg braces) | Same |
| Operating Temperature | 10° to 95°F (-12° to -35° C) | 10° to 95°F (-12° to -35° C) | Same |
| Operating Humidity | Not available | Not available | Same |
| Electrical Safety Testing | IEC 60601-1:2005 with US<br>deviations | IEC 60601-1:2005 with US<br>deviations | Same |
| Electromagnetic<br>Compatibility Testing | Passed IEC 60601-1-2: 2014 | Passed IEC 60601-1-2: 2007 | Same, newer<br>version |
#### Comparison of Technical Characteristics
{5}------------------------------------------------
## Comparison of Technical Characteristics
{6}------------------------------------------------
#### Non-clinical Performance Summary
| Technical Area | Tests Completed |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety and<br>Electromagnetic Compatibility | IEC 60601-1:2005, IEC 60601-1-2:2014, low battery testing, IEC 62133, IEC 61960 parts 7.4 and 7.5, UN 38.3, UN Manual (ST/SG/AC. 10/11/Rev.5/Amend.1), battery life cycle testing |
| Durability | Worst case loading of knee and hip joints beyond service life, worst case loading of structure beyond service life, ankle spring durability, component strength testing |
| Thermal | IEC 60601-1:2005 (ISO 7176 not required; batteries on the Ekso are mounted to the aluminum Torso frame, enclosed in an aluminum case, and not near any flammable material. The testing conducted per IEC 60601 -1 in terms of flame retardant evaluations is sufficient to support the device functionality in terms of flame retardant materials.) |
| Software | Verification, Validation, and hazard analysis |
| Bench Testing | IEC 60601-1:2005 sec. 15.3 (IP22 not required) |
#### Clinical Performance Summary
The current device utilized 5 different studies to demonstrate safety and efficacy of the current indications for use (K161443). This submission builds from those 5 studies with 2 additional clinical studies (summarized below) focusing on the specific populations added by the expanded indication for use in this submission. These studies demonstrate the device effectively enables gait ambulation and there were no reported adverse events indicating the device is safe to use with this expanded population.
| Study | Description |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Kessler ABI | Single center, open-label, non-comparative, non-randomized, prospective study of patients with acquired brain injury (only non-stroke patients included in this submission). |
| | Duration of Intervention |
| | Ambulation during a mean of 5.8 sessions (range 1 to 20) |
| | 24 subjects total used the Ekso |
| | Mean time since injury 92.2 days (range of 10 to 835 days) 6 closed TBI 1 TBI unspecified 3 non-TBI anoxic 2 non-TBI neoplasm NOS 2 non-TBI malignant neoplasm 3 non-TBI benign neoplasm of meninges 1 non-TBI non traumatic hemorrhage |
| | Results |
| | Mean Motor FIM scores pre and post-training were 20.4 and 54.2, respectively |
- Mean Cognitive FIM scores pre and post-training were 12.2 and 23.0, respectively ●
- There were no falls or other adverse events reported ●
{7}------------------------------------------------
Kessler TBI Single center, open-label, non-comparative, non-randomized, prospective study of patients with traumatic brain injury.
Duration of Intervention
- Ambulation during a mean of 4.9 sessions (range 2 to 5) .
25 subjects total used the Ekso
- Mean time since injury 3540 days (range of 603 to 21360 days) ●
- . All were unspecified TBI
Results
- . Mean 10MWT pre and post-training were 17 and 18 seconds, respectively
- Mean 2MWT pre and post-training were 126 and 128 meters, respectively
- Mean TUG scores pre and post-training were 19.3 and 21.7, respectively
- There were no falls or other adverse events reported ●
#### Substantial Equivalence Conclusion
This device is substantially equivalent to the current devices have the same technical characteristics and the same intended uses, to facilitate gait ambulation. The supporting clinical data demonstrating the use of the product with patients with Acquired Brain Injury (ABI), including both Traumatic Brain Injury (TBI) and non-Traumatic Brain Injury (non-TBI), show that the device effectively facilitates gait ambulation in the expanded patient population. The clinical data reported no adverse events demonstrating the device is safe on this patient population when used in accordance with existing labeling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.