← Product Code [JOW](/productcode/JOW) · K200564

# Aria System (K200564)

_Inova Labs · JOW · Apr 3, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200564

## Device Facts

- **Applicant:** Inova Labs
- **Product Code:** [JOW](/productcode/JOW.md)
- **Decision Date:** Apr 3, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5800
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Aria sequential circulator is a programmable sequential, pneumatic compression device intended for use by medical professionals and patients at home, for the treatment of the following conditions: - Chronic edema - Lymphedema - Venous insufficiency - Wound healing

## Device Story

Aria System is a programmable sequential pneumatic compression device for home or clinical use. It consists of a flow generator and a limb garment with seven air chambers. The flow generator, powered by a brushless low-voltage DC motor, inflates garment chambers to provide 0-45 mmHg compression. This compress-and-release action stimulates lymphatic vessels to encourage fluid clearance. The device features a simple ON/OFF interface and Bluetooth connectivity for data export to a paired app. It delivers one pre-programmed therapy mode over a 60-minute cycle. Healthcare providers prescribe the device to manage edema, lymphedema, venous insufficiency, and wound healing. Patients operate the device at home to receive therapy, which aims to reduce fluid accumulation and improve patient outcomes through consistent compression treatment.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including pressure stability, sleeve burst/leakage/integrity, pressure accuracy, chamber filling cycle time, and total therapy time. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-6, and ISO 10993-1 standards was verified.

## Technological Characteristics

Microcontroller-based flow generator; brushless low-voltage DC motor; 7-chamber pneumatic garment; 0-45 mmHg pressure range; Bluetooth classic connectivity. Power: 100-240V, 50-60Hz. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), IEC 60601-1-11 (home use), ISO 10993-1 (biocompatibility).

## Regulatory Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

## Predicate Devices

- Tactile Systems Technology Inc, Entre Model PD08-U ([K143185](/device/K143185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.

April 3, 2020

Inova Labs % Rafael Aguila Responsible Third-Party Official Accelerated Device Approval Services, LLC 6800 S.W. 40th Street. Ste. 444 Ludlum, Florida 33155-3708

Re: K200564

Trade/Device Name: Aria System Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible limb sleeve Regulatory Class: Class II Product Code: JOW Dated: March 3, 2020 Received: March 4, 2020

Dear Rafael Aguila:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K200564

Device Name

Aria™ System

Indications for Use (Describe)

The Aria sequential circulator is a programmable sequential, pneumatic compression device intended for use by medical professionals and patients at home, for the treatment of the following conditions:

- Chronic edema
- Lymphedema
- Venous insufficiency

Type of Use (Select one or both, as applicable)

- Wound healing

| <span style="font-size:12px;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:12px;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

K200564

# InovaLabs

## 510(k) Summary Inova Labs Aria System

| Date Prepared            | April 3, 2020                                                             |
|--------------------------|---------------------------------------------------------------------------|
| Submitter                | Inova Labs<br>7581 Convoy Court, San Diego CA 92111                       |
| Official Contact         | Reuben Lawson<br>Regulatory Affairs Consultant<br>reuben.lawson@gmail.com |
| Product Codes            | JOW                                                                       |
| Class                    | Class II                                                                  |
| Classification Reference | 21 CFR 870.5800                                                           |
| Common/Usual Name        | Sleeve, Limb, Compressible                                                |
| Proprietary Name         | Aria                                                                      |
| Predicate Device(s)      | Tactile Systems Technology Inc, Entre Model PD08-U (K143185)              |
| Reason for submission    | New Device                                                                |

{4}------------------------------------------------

#### Indication for Use

The Aria sequential circulator is a programmable sequential, pneumatic compression device intended for use by medical professionals and patients at home, for the treatment of the following conditions:

- · Chronic edema
- · Lymphedema
- · Venous insufficiency
- · Wound healing

#### Device Description

The Aria system consists of two main components: A flow generator and a garment. The garment is to be wrapped around the limb, providing a comfortable fit. The garment has seven (7) chambers that are filled with air by the flow generator to provide compression on the extremity. The Aria system uses a compressand-release massage action, similar to the predicate, in order to stimulate lymphatic vessels in the treated area and encourage fluid clearance.

The Aria system retains similar hardware and performance features of the predicate device. Key features include flow generator, valves, A/C plug pack, lower limb garment, tubing, no-LCD User Interface and ON/OFF button. The Aria System contains a microprocessor-controlled flow generator/blower system that generates pressure from 0-45 mmHG to provide for effective treatment of the conditions described in the IFU.

The Aria flow generator has no control settings and delivers one pre-programmed therapy mode.

#### Characteristics between predicate and new device

| Characteristic                     | Predicate Device Entre Model PD08-U (K143185)                                                                                                                                                                                                         | New device: Aria System                                                                                                                                                                                                                                                                              |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for use                 | The entré System is intended for use by medical<br>professionals and patients who are under<br>medical supervision, for the<br>treatment of the following conditions:<br>• Chronic edema<br>• Lymphedema<br>• Venous insufficiency<br>• Wound healing | The Aria sequential circulator is a programmable<br>sequential, pneumatic compression device<br>intended for use by medical professionals and<br>patients at home, for the treatment of the<br>following conditions:<br>• Chronic edema<br>• Lymphedema<br>• Venous insufficiency<br>• Wound healing |
| Pressure range                     | 0-45 mmHG (at 'moderate' setting)                                                                                                                                                                                                                     | 0-45 mmHG                                                                                                                                                                                                                                                                                            |
| Cycle Time                         | 65 seconds                                                                                                                                                                                                                                            | 56 seconds                                                                                                                                                                                                                                                                                           |
| Total Therapy Time                 | Approx. 60 minutes                                                                                                                                                                                                                                    | Approx. 60 minutes                                                                                                                                                                                                                                                                                   |
| Modes of operation                 | Sequential gradient compression therapy                                                                                                                                                                                                               | Sequential gradient compression therapy                                                                                                                                                                                                                                                              |
| System Components                  | • Flow generator                                                                                                                                                                                                                                      | • Flow generator                                                                                                                                                                                                                                                                                     |
|                                    | • Valves                                                                                                                                                                                                                                              | • Valves                                                                                                                                                                                                                                                                                             |
|                                    | • Tubing                                                                                                                                                                                                                                              | • Tubing                                                                                                                                                                                                                                                                                             |
|                                    | • Garment                                                                                                                                                                                                                                             | • Garment                                                                                                                                                                                                                                                                                            |
| Flow Generator<br>operating system | Microcontroller                                                                                                                                                                                                                                       | Microcontroller                                                                                                                                                                                                                                                                                      |
| Garment Air<br>Chambers            | 8                                                                                                                                                                                                                                                     | 7                                                                                                                                                                                                                                                                                                    |

{5}------------------------------------------------

| Characteristic                     | Predicate Device Entre Model PD08-U (K143185)                        | New device Aria System                                               |
|------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|
| Tubing                             | Entre Tubing (2m length)                                             | Aria tubing (1.8m length)                                            |
| User Interface                     | On/Off Button<br>Pressure Low/Med/High Button<br>Start Therapy/Pause | On/Off (Start/Stop Therapy) Button                                   |
| Connectivity                       | None                                                                 | Bluetooth classic to allow export of system data<br>to a paired app. |
| Motor type                         | Compressor                                                           | Brush-less low voltage DC                                            |
| Power supply                       | 100-240V, 50-60Hz                                                    | 100-240V, 50-60Hz                                                    |
| Weight                             | 2.5lb                                                                | 0.65lb                                                               |
| Dimensions H x W x D<br>(inches)   | 11 x 6 x 8                                                           | Flow generator unit:<br>2 x 3.3 x 5.3                                |
| IEC 60601-1 (electrical<br>safety) | Yes                                                                  | Yes                                                                  |
| IEC 60601-1-2 (EMC)                | Yes                                                                  | Yes                                                                  |
| IEC 60601-1-6<br>(Usability)       | Yes                                                                  | Yes                                                                  |
| IEC 60601-1-11<br>(Home medical)   | Yes                                                                  | Yes                                                                  |
| ISO 10993-1                        | Not referenced in 510(k) summary                                     | Yes                                                                  |

# InovaLabs

### Substantial Equivalence

The new device has the following similarities to the previously cleared predicate devices.

- Same intended use A
- >> Similar operating principle
- > Similar technology

As a result of the Risk Analysis review and design input requirements. non-clinical verification activities were performed on the Aria System. The results demonstrate that the Aria system raises no new safety or effectiveness concerns and is substantially equivalent to the predicate device. The Aria System complies with the applicable requirements referenced in the FDA guidance documents:

- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices A (May 11, 2005)
- FDA Guidance for Industry and FDA Staff Design Considerations for Devices Intended for Home A Use- Document Issued on: Nov 24, 2014
- > FDA Guidance for Industry and FDA Staff Radio Frequency Wireless Technology in Medical Devices - Document Issued on: August 14, 2013

### Non-Clinical Testing:

Side-by-Side bench testing was performed to verify that the Aria System met the Aria System Specification when compared to the predicate device.

{6}------------------------------------------------

# InovaLabs

The bench testing included performance comparisons to the predicate device and covered:

- Pressure stability A
- Sleeve burst test A
- Sleeve leakage test A
- Sleeve integrity test A
- A Pressure accuracy test
- A Chamber filling cycle time testing
- Total therapy time »

The Aria system has been tested to appropriate standards and other applicable requirements:

- > IEC 60601-1:2005 A1:2015 Ed. 3.1, Medical electrical equipment Part 1: General requirements for safety Medical electrical equipment - General requirements for basic safety and essential performance
- > IEC 60601-1-2:2013 Ed. 4.0 Medical Electrical Equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- > IEC 60601-1-11:2015 Ed 2.0 Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- > FDA guidance on software development for medical devices
- > ISO 10993-1 Biological evaluation of medical devices
- > IEC 60601-1-6 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

### Clinical Testing:

As with the predicate system, clinical testing was not conducted for Aria system. The technology used in JOW devices is well known and understood. Bench testing against recognized consensus standards, is sufficient to demonstrate no new risk.

### Conclusion

The Aria System is substantially equivalent the predicate device Entre PD08-U (K143185) and is as safe and as effective as the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K200564](https://fda-staging.innolitics.com/device/K200564)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
