← Product Code [DXN](/productcode/DXN) · K200560

# InBody (K200560)

_Inbody Co., Ltd. · DXN · Jan 6, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200560

## Device Facts

- **Applicant:** Inbody Co., Ltd.
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** Jan 6, 2021
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The subject device is a manual blood pressure (BP) monitor, and requires the user to listen for Korotkoff sounds to determine systolic and diastolic pressure. Further, patients whose Korotkoff sounds are not clear or absent, such as children younger than 3 years of age, will not be suitable for the subject device. Pulse rate is also measured.

## Device Story

Manual blood pressure monitor; requires user-supplied stethoscope. Operates via Korotkoff sound auscultation method; user manually inflates/deflates cuff. Device displays pressure on LCD screen; includes backlight for low-light use. Features 'Mark' function to store up to 5 measurements. Includes automatic rapid exhaust for overpressure (>320mmHg) or emergency. Intended for use by medical professionals in clinical settings. Provides systolic/diastolic pressure readings and incidental pulse rate. Benefits include mercury-free digital visualization of pressure values while maintaining traditional auscultatory measurement principles.

## Clinical Evidence

Bench testing only. Accuracy verified against mercury manometer standard in 102 subjects per EN 1060-4. Results: Systolic (Mean=0mmHg, SD=1.206mmHg), Diastolic (Mean=0mmHg, SD=1.424mmHg). No clinical trials required.

## Technological Characteristics

Non-invasive sphygmomanometer; manual inflation/deflation. Components: Gauge, cuff, pump. Power: 4x AA alkaline batteries. Display: Vertical LCD with backlight. Safety: Automatic quick exhaust valve (>320mmHg). Standards: IEC 60601-1, IEC 60601-1-2, EN 1060-4.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- A&D Medical UTM-101 Digital Blood Pressure Monitor ([K061456](/device/K061456.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 6, 2021

InBody Co. Ltd. % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114

Re: K200560

Trade/Device Name: InBody Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, DXQ Dated: December 5, 2020 Received: December 9, 2020

Dear Daniel Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200560

Device Name BPBIO210/220/210T/220T

#### Indications for Use (Describe)

The subject device is a manual blood pressure (BP) monitor, and requires the user to listen for Korotkoff sounds to determine systolic and diastolic pressure. Further, patients whose Korotkoff sounds are not clear or absent, such as children younger than 3 years of age, will not be suitable for the subject device. Pulse rate is also measured.

Type of Use (Select one or both, as applicable)

| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>             |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K200560 510(k) Summary K200560

Image /page/3/Picture/1 description: The image shows the logo for InBody. The logo is in a bold, sans-serif font and is a dark reddish-brown color. The letters are close together, giving the logo a compact appearance. The background is plain white.

InBody Co., Ltd. InBody Bldg., 625, Eonju-ro, Gangnam-gu Seoul KOREA 06106 TEL. +82 02-501-3939 Date prepared: January 5, 2021 Contact: Kichul Cha, CEO

- 1. Identification of the Devices: Proprietary-Trade Names: InBody Blood Pressure Monitor, BPBIO210/220/210T/220T. Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN, DXQ
- 2. Equivalent legally marketed device: K061456, A&D Engineering, Inc. Proprietary-Trade Names: A&D Medical UTM-101 Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN
- 3. Indications for Use: The subject device is a manual blood pressure (BP) monitor, and requires the user to listen for Korotkoff sounds to determine systolic and diastolic pressure. Further, patients whose Korotkoff sounds are not clear or absent, such as children younger than 3 years of age, will not be suitable for the subject device. Pulse rate is also measured.
- 4. Product Description: This device can be used to measure blood pressure and pulse rate non-persistently using the Korotkoff sound method of measuring blood pressure in vitro. The user can measure the blood pressure using this device and cuff, stethoscope. Basically, it has the same function as a conventional mercury sphygmomanometer. However, instead of using harmful mercury to the human body, it displays the pressure on the LCD screen, so that the blood pressure can be measured harmlessly to the human body. In addition, using the 'Mark' function can conveniently display the systolic and diastolic pressures. And using its backlight, the user can measure blood pressure in the dark place. If air pressure exceeds 320mmHg, it will automatically exhaust quickly. If an emergency occurs during the measurement, pressing On/Off button will turn off the equipment and allow rapid exhaust. The device is for use by medical professional. The testing person supplies their own stethoscope, which is required for making the measurements. The measurement method is as follows.

1. BPBIO220/BPBIO210: Place the device on a sturdy table or the cart. For BPBIO220T/BPBIO210T: Assemble the stand and the device.

- 2. Sit properly in chair and place patient's arm at heart height.
- 3. Wind the cuff around the patient's arm.
- 4. Pressurize manually.
- 5. Depressurize manually and listen to the Korotkoff sound.

{4}------------------------------------------------

## K200560

- 6. Determine the systolic blood pressure and the diastolic blood pressure by Korotkoff sound.
7. When the measurement is complete, the heart rate appears on the heart rate display is an incidental result of the Korotkoff sound identification and is not intended to be diagnostic or act as a heart rate monitor or alarm.

- 5. Safety and Effectiveness, comparison to predicate device. The testing results and specification comparisons indicate that the new models are as safe and effective as the predicate device. A comparison table is presented below.

| ltem                     | K061456, A&D Engineering, Inc.<br>UM-101 SPHYGMOMANOMETER                                                                                                   | InBody Blood Pressure Monitor,<br>BPBIO210/220/210T/220T.                                                                                                                                                                                                                                                                                                                         | Comparison<br>Result              |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Intended Use             | Measure blood pressure (systolic and<br>diastolic) and pulse rate.                                                                                          | The subject device is a manual blood<br>pressure (BP) monitor, and requires the<br>user to listen for Korotkoff sounds to<br>determine systolic and diastolic pressure.<br>Further, patients whose Korotkoff sounds<br>are not clear or absent, such as children<br>younger than 3 years of age, will not be<br>suitable for the subject device.  Pulse rate<br>is also measured. | Same                              |
| Patient<br>Population    | Adult                                                                                                                                                       | Adult                                                                                                                                                                                                                                                                                                                                                                             | Same                              |
| Measurement<br>location  | Upper Arm                                                                                                                                                   | Upper Arm                                                                                                                                                                                                                                                                                                                                                                         | Same                              |
| Measurement<br>Principle | Korotkoff sounds.                                                                                                                                           | Korotkoff sounds.                                                                                                                                                                                                                                                                                                                                                                 | Same                              |
| Measured:                | Blood Pressure Pulse Rate                                                                                                                                   | Blood Pressure and Pulse Rate                                                                                                                                                                                                                                                                                                                                                     | Same                              |
| Blood Pressure<br>Cuff   | External                                                                                                                                                    | External                                                                                                                                                                                                                                                                                                                                                                          | Same                              |
| Inflation<br>system      | Manual                                                                                                                                                      | Manual                                                                                                                                                                                                                                                                                                                                                                            | Same                              |
| Components               | Gauge, Cuff, Pump                                                                                                                                           | Gauge, Cuff, Pump                                                                                                                                                                                                                                                                                                                                                                 | Same                              |
| Power Source             | 2 x 1.5V alkaline batteries (LR6 or AA)                                                                                                                     | 4 x alkaline batteries (AA)                                                                                                                                                                                                                                                                                                                                                       | Similar,<br>battery<br>operation  |
| Dimensions               | 12,7" Gauge, Rectangular                                                                                                                                    | 12.6" Gauge, Rectangular                                                                                                                                                                                                                                                                                                                                                          | Nearly<br>identical.              |
| Overpressure<br>safety   | An automatic quick exhaust valve is installed<br>in the device to prevent over<br>pressurization of 320mmHg or higher,<br>therefore protecting the patient. | If air pressure exceeds 320mmHg, it will<br>automatically exhaust quickly. If an emergency<br>occurs during the measurement, pressing<br>On/Off button will turn off the equipment and<br>allow rapid exhaust                                                                                                                                                                     | Same                              |
| Measurement<br>range     | Blood Pressure 0-300 mmHg,<br>Pulse: 30-240 bpm                                                                                                             | Blood Pressure  0-300 mmHg,<br>Pulse: 30-240 bpm                                                                                                                                                                                                                                                                                                                                  | Same                              |
| Measurement<br>storage   | Up to 5 measurements may be stored via a<br>marking on the vertical LCD scale                                                                               | Up to 5 measurements may be stored via a<br>marking on the vertical LCD scale                                                                                                                                                                                                                                                                                                     | Same.                             |
| Accuracy                 | Pressure: ±3mmHg<br>Pulse: ±5 %                                                                                                                             | Pressure: ±3 mmHg,<br>Pulse: ±3%                                                                                                                                                                                                                                                                                                                                                  | Essentially<br>the same           |
| Item                     | K061456, A&D Engineering, Inc.<br>UM-101 SPHYGMOMANOMETER                                                                                                   | InBody Blood Pressure Monitor,<br>BPBIO210/220/210T/220T.                                                                                                                                                                                                                                                                                                                         | Comparison<br>Result              |
| Range of cuff<br>sizes   | Large cuff 33 cm to 45 cm<br>Medium cuff 23 cm to 33 cm<br>Small cuff 16 cm to 23 cm                                                                        | Large cuff 32 cm to 42 cm<br>Medium cuff 22 cm to 32 cm<br>Small cuff 17 cm to 22 cm.                                                                                                                                                                                                                                                                                             | Similar range<br>of sizes         |
| Photos                   | Image: A&D Engineering UM-101 SPHYGMOMANOMETER                                                                                                              | Image: InBody Blood Pressure Monitor BPBIO210/220/210T/220T                                                                                                                                                                                                                                                                                                                       | Very similar<br>in<br>appearance. |

{5}------------------------------------------------

- 6. Summary of technological characteristics of the device compared to the predicate device. These blood pressure meters are intended to be used in measuring human systolic blood pressure using Korotkoff sounds. In addition to being able to manually measure (and store up to 5 measurements) this electronic version of the classical blood pressure meter can display pulse rate. The following FDA guidance was consulted in the design and testing of the device: Non-Invasive Blood Pressure (NIBP) Monitor Guidance MARCH 1997, Final.
- 7. Non-clinical testing: The proposed new models were tested and found to conform to the following international standards:

IEC 60601-1 :2005/AMD1:2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance FDA recognition 19-4

IEC 60601-1-2 [2014] Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic Compatibility Requirements and tests FDA recognition 19-8

In additional to standards testing, we did bench testing to compare the proposed device results to the regular manual method of measuring blood pressure. InBody has implemented the bench testing procedure to determine the overall system accuracy of non-invasive sphygmomanometers with the intent to verity the overall accuracy of BPBIO210/220/210T/220T. 102 subjects were tested. The comparison results are as follows: InBody has compared the accuracy between mercury manometer using the stethoscope and the non-invasive sphygmomanometer, BPBIO210/220/210T/220T. We found and concluded that they are substantially equal to each other. The accuracy of the BPBIO210/220T which acquired from the clinical investigation implemented by InBody is respectively the systolic (Mean = 0mmHg, Std. Dev = 1.206mmHg), and the diastolic (Mean = 0mmHg, Std. Dev = 1.424mmHg). This testing was done in accordance with EN 1060-4. Non-invasive sphygmomanometers. Test procedures to determine the overall system accuracy of automated non-invasive sphygmomanometers. This is an international version of SP-10 which is referenced in the FDA Guidance.

{6}------------------------------------------------

### K200560

- 8. Clinical Testing: Not required for this type of device. Blood pressure measurement is manually done by the testing person.
- 9. Conclusion, Comparison to the predicate device. Proposed Models: BPBIO210/2207. are substantially equivalent to the predicate whose 510(k) number is K061456. The devices are identical in the intended use, and similar in the design principles, the performance and the applicable standards. Only their appearance and the user interfaces are different.

---

**Source:** [https://fda-staging.innolitics.com/device/K200560](https://fda-staging.innolitics.com/device/K200560)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).