The VisualEyes system provides information to assist in the nystagmographic evaluation, diagnosis and documentation of vestibular disorders. Nystagmus of the eye is recorded by use of a goggle mounted with cameras. These images are measured, recorded, displayed and stored in the software. This information then can be used by a trained medical professional to assist in diagnosing vestibular disorders. The target population for VisualEyes system is 5 years of age and above.
Device Story
VisualEyes 505/515/525 is a software-based vestibular analysis system; operates on Windows PC; utilizes goggle-mounted cameras to record eye movements during standard VNG protocols. System transforms video input into measured, displayed, and stored data; includes torsional eye tracking algorithms (Johns Hopkins libraries). Used by audiologists, ENT surgeons, and hearing healthcare professionals in clinical settings. Output assists clinicians in diagnosing vestibular disorders by providing quantitative metrics (latency, accuracy, velocity) for tests including caloric, saccade, and vHIT. Benefits include standardized, objective documentation of vestibular function to support clinical decision-making.
Clinical Evidence
Bench testing and external beta testing conducted. Sequential testing of subjects on VisualEyes and predicate devices (MMT or IA) performed. Data sets compared by clinical audiologists; statistical analysis showed negligible differences beneath acceptance criteria. No clinical data in the sense of prospective trials, but validation confirmed equivalent clinical findings to predicates.
Technological Characteristics
Software-based system running on Microsoft Windows PC. Integrates with external hardware including video goggles, cameras, rotary chairs, and irrigators. Features include torsional eye tracking algorithms and various vestibular test modules. Connectivity via standard PC interfaces. No specific materials or sterilization methods listed as it is a software-based analysis system.
Indications for Use
Indicated for patients 5 years of age and older requiring nystagmographic evaluation, diagnosis, and documentation of vestibular disorders.
Regulatory Classification
Identification
A nystagmograph is a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 12, 2020
Interacoustics A/S Erik Nielson Director, Regulatory & Compliance Audiometer Alle 1 Middelfart, 5500 Denmark
Re: K200534
Trade/Device Name: VisualEyes Regulation Number: 21 CFR 882.1460 Regulation Name: Nystagmograph Regulatory Class: Class II Product Code: GWN Dated: July 10, 2020 Received: July 13, 2020
Dear Erik Nielson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina Eydelman, M.D. Office Director DHT1C: Division of ENT. Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200534
Device Name VisualEyes 515 / VisualEyes 525 / VisualEyes 505
Indications for Use (Describe)
The VisualEyes system provides information to assist in the nystagmographic evaluation, diagnosis and documentation of vestibular disorders. Nystagmus of the eye is recorded by use of a goggle mounted with cameras. These images are measured, recorded, displayed and stored in the software. This information then can be used by a trained medical professional to assist in diagnosing vestibular disorders. The target population for VisualEyes system is 5 years of age and above
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> <b> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
## VisualEyes 505/515/ 525
### Submitter Information:
Company Name Address
Phone Fax e-mail Contact Person Date Summary Prepared Interacoustics A/S Audiometer Allé 1 5500 Middelfart Denmark +45 6371 3555 +45 6371 3522 erni@demant.com Erik Nielsen, Director, Regulatory & Compliance August 07, 2020
### Device Identification:
Trade Name
Common Name Classification Name Product Code Class Panel Device Class
Predicate Device: Predicate Device Manufacturer 510(k) No. Date Cleared
Reference devices: Reference Device Manufacturer 510(k) No. Date Cleared
Reference Device Manufacturer 510(k) No. Date Cleared
VisualEyes 505 VisualEyes 515 VisualEyes 525 Vestibular analysis device Nystagmograph, apparatus, vestibular analysis GWN Neurology Class II (According to 21 CFR 882.1460)
VisualEyes 505/515/ 525 Interacoustics A/S K163149 04/26/2017
VIDEO EYE TRAKKER (Aka Spectrum) MicroMedical Technologies Inc K964646 07/15/1997
EyeSeeCam VHIT Interacoustics A/S K131681 09/06/2013
{5}------------------------------------------------
#### VisualEyes 505/515/ 525 is a software program that analyzes eye Device movements recorded from a camera mounted to a video goggle. A Description standard Video Nystagmography (VNG) protocol is used for the testing. VisualEyes 505/515/ 525 is and update/change, replacing the existing VisualEyes 515/525 release 1 (510(k) cleared under K152112).
The software is intended to run on a Microsoft Windows PC platform. The "525" system is a full featured system (all vestibular tests as listed below) while the "515" system has a subset of the "525" features. "505" is a simple video recording mode.
The VisualEyes 505/ 515/ 525 software is designed to perform the following vestibular tests:
| Test | Availability |
|----------------------------------------------|--------------|
| Calibration | 515 / 525 |
| Spontaneous nystagmus | 515 / 525 |
| Dix-Hallpike | 515 / 525 |
| Positional | 515 / 525 |
| Caloric | 515 / 525 |
| Gaze | 525 |
| Smooth Pursuit | 525 |
| Random Saccade | 525 |
| Saccadometry | 525 |
| Optokinetic | 525 |
| Sinusoidal Harmonic Acceleration | 515/525 |
| Step Velocity | 515/525 |
| VOR Suppression | 515/525 |
| Visual VOR | 515/525 |
| Subjective Visual Vertical (SVV)- Static | 525 |
| Subjective Visual Vertical (SVV)-<br>Dynamic | 525 |
| Video Frenzel | 525 |
| Ocular Counter Roll | 525 |
| VORTEQ AHR | 525 |
| vHIT for VORTEQ | 525 |
| Dynamic Visual Acuity (DVA) | 515/ 525/505 |
| Dix-Hallpike Advanced | |
| Lateral Head Roll | |
| vHIT for EyeSeeCam | |
| - Lateral | |
| - LARP | 515/525/505 |
| - RALP<br>- SHIMP | |
| Video recording | 505 |
{6}------------------------------------------------
| Indications for<br>Use | The VisualEyes system provides information to assist in the<br>nystagmographic evaluation, diagnosis and documentation of<br>vestibular disorders. Nystagmus of the eye is recorded by use of a<br>goggle mounted with cameras. These images are measured,<br>recorded, displayed and stored in the software. This information<br>then can be used by a trained medical professional to assist in<br>diagnosing vestibular disorders. The target population for<br>VisualEyes system is 5 years of age and above. |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>operator | The VisualEyes VNG system is to be used by trained personnel<br>only, such as audiologists, ENT surgeons, doctor's, hearing<br>healthcare professionals or personnel with a similar level of<br>qualifications. The device should not be used without the necessary<br>knowledge and training to understand its use and how results should<br>be interpreted. |
| Technological<br>characteristics | The VisualEyes is a software system is intended to incorporate<br>various goggle/cameras, rotary chairs, irrigators and other<br>accessories |
### Comparison of technological characteristics with predicate device
We have chosen to compare the VisualEyes 3 software with the VisualEyes 2 as the characteristics of VisualEyes 3 is most similar to VisualEyes 2. The technological principles for VisualEyes 3 is based on refinements from VisualEyes 2 and adoptions of functionality of the two reference devices.
### Comparison tables
VisualEyes 3 to Predicate Device VisualEyes 2
| Functionality Description | Equivalence |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Indications for use:<br>The VisualEyes system provides information to<br>assist in the nystagmographic evaluation, diagnosis<br>and documentation of vestibular disorders.<br>Nystagmus of the eye is recorded by use of a<br>goggle mounted with cameras. These images are<br>measured, recorded, displayed and stored in the<br>software. This information then can be used by a<br>trained medical professional to assist in diagnosing<br>vestibular disorders. The target population for<br>VisualEyes system is 5 years of age and above | Same |
| Spontaneous Nystagmus Test | Same |
| Gaze Test | Same |
| Smooth Pursuit Test | Same |
| Saccade Test | Same |
| Optokinetic Test | Same |
| Dix-Hallpike Test | Same |
| Positional Test | Same |
| Caloric Test | Same |
{7}------------------------------------------------
| SHA Test | Same |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Step Test | Same |
| VOR Suppression Test | Same |
| Saccadometry | This is the same.<br>Saccadometry is a broad term that means saccadic<br>eye movements being measured. The VisualEyes<br>saccade test involves a single stimulus (dot) moving<br>from one position to the other. The patient either<br>looks to or away from the dot. There is no<br>difference in the stimulus presentation between the<br>predicate and new software. The only difference is<br>that for the antisaccade the patient looks away from<br>the target instead of towards the target. The exact<br>same measures of latency, accuracy and velocity<br>are recorded and plotted.<br>Since the stimulus presented and the algorithms<br>have not changed for the saccade tests there are<br>no device safety or effectiveness differences<br>between the two devices. |
| Advanced Dix Hallpike Test | This is the same.<br>For the Advanced Dix Hallpike there are no<br>changes in how the test is performed. The only<br>change was the inclusion of the new torsional eye<br>tracker from Johns Hopkins. This has been verified<br>and demonstrates that there are no device safety or<br>effectiveness differences between the two devices. |
VisualEyes 3 software also adopts functions from the Micromedical Technologies Spectrum software as the Spectrum supports many of the protocols that are being ported into VisualEyes 3.
The VHIT function is adopted into VisualEyes 3 software from EyeSeeCam VHIT
## Functionality adopted from reference device VIDEO EYE TRAKKER
| Functionality Description | Equivalence |
|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ENG | Same |
| Ocular Counter roll | Uses the Johns Hopkins torsional algorithm not the<br>Spectrum tracking algorithm |
| VORTEQ Active Head Rotation (AHR) | Same |
| VORTEQ Dynamic Visual Acuity (DVA) | Same |
| VORTEQ Video Head Impulse Test (vHIT) | Same |
| Dynamic Subjective Visual Vertical (SVV) in<br>Orion AT chair | Same |
| Static Subjective Visual Vertical in the Orion<br>Comprehensive Chair | Same |
| Torsional Tracking Algorithm | This is the same.<br>Improved in release VisualEyes 3 with the addition<br>of libraries from Johns Hopkins. This has been<br>verified and demonstrates that there are no device<br>safety or effectiveness differences between the two<br>devices. |
{8}------------------------------------------------
| Functionality adopted from reference device EyeSeeCam | |
|-------------------------------------------------------|-------------|
| Functionality Description | Equivalence |
| vHIT | Same |
### Verification and validation summary
In order to validate the new software, bench testing was conducted internally, and beta testing was conducted with various groups in different geographical locations externally. The purpose of the testing is to confirm that the results collected on the updated VisualEyes software generated the same clinical findings as the existing predicate devices.
The beta testing was conducted in external sites that had either MMT or IA existing predicate devices. All data was exported from the external beta sites and transferred to licensed internal clinical audiologists who compared the test results for substantial equivalency. The data was collected sequentially, which involved having the same subject tested on VisualEyes and then tested again on either IA or MMT predicate devices. The data sets were compared to determine if there were any statistically significant differences in the data collected from predicate and current devices.
### Summary
There were no differences found in internal bench testing comparisons or the external beta testing statistical comparisons. VisualEyes 505/515/525 performs substantially equivalent to the predicate devices used for the validation.
For the statistical comparison, all data sets showed a negligible statistical difference beneath the specified acceptance criteria.
It is the professional opinion of both clinical reviewers of the validation that all the data between the new VisualEyes software and the data collected and analyzed with both predicate devices are substantially equivalent. There are no differences between any of the products tested.
### Conclusion
We have compared key issues for the VisualEyes 505/515/ 525 and the predicate devices. We have performed a comparison validation between VisualEves 505/515/ 525 and the predicate devices.
Any deviations between VisualEyes 505/515/ 525 and predicate devices are appraised to have no adverse effect on the safety and effectiveness of the device.
We trust that the results of these comparisons demonstrate that the VisualEyes 505/515/ 525 is substantially equivalent to the marketed predicate devices (K163149).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.