← Product Code [GKT](/productcode/GKT) · K200530

# AMICUS Separator System (K200530)

_Fresenius Kabi AG · GKT · Sep 11, 2020 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200530

## Device Facts

- **Applicant:** Fresenius Kabi AG
- **Product Code:** [GKT](/productcode/GKT.md)
- **Decision Date:** Sep 11, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.9245
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The AMICUS Separator System is an automated blood component separator intended for use in the collection of blood components and mononuclear cells. The AMICUS Separator System is an automated blood cell separator intended for use in therapeutic apheresis applications and may be used to perform Therapeutic Plasma Exchange (TPE). The AMICUS Separator System is an automated blood component separator intended for use in therapeutic apheresis applications and may be used to perform red blood cell exchange, depletion, and depletion/exchange (RBCX) procedures.

## Device Story

AMICUS Separator System is a continuous-flow, centrifugal blood cell separator. Device draws whole blood from donor/patient; separates components via density-based centrifugation; collects target components; returns remainder to donor/patient. System utilizes pumps, clamps, valves, and single-use sterile fluid path kits. Operator monitors procedure via touch screen interface with audible/visual alerts. Used in clinical settings for therapeutic apheresis (TPE, RBCX) or blood component collection. Output includes collected blood products (e.g., platelets, plasma, mononuclear cells). Device facilitates therapeutic management of sickle cell disease and blood component banking. Modifications include updated controller circuit boards, QNX operating system, and added connectivity (USB, Ethernet, Wi-Fi).

## Clinical Evidence

Bench testing only. Software and systems verification testing, including regression testing, confirmed performance equivalent to the predicate. Electrical safety and EMC testing performed per ANSI/AAMI ES60601-1 and IEC 60601-1-2.

## Technological Characteristics

Continuous-flow centrifugal separator. Components: touch screen (1024x768), pumps, clamps, sensors, centrifuge rotor. Connectivity: USB, Ethernet, Wi-Fi. Software: QNX OS. Safety: air detector, pressure sensors, centrifuge leak sensor, audible/visual alerts. Sterilization: single-use sterile fluid path kits.

## Regulatory Identification

An automated blood cell separator is a device that uses a centrifugal or filtration separation principle to automatically withdraw whole blood from a donor, separate the whole blood into blood components, collect one or more of the blood components, and return to the donor the remainder of the whole blood and blood components. The automated blood cell separator device is intended for routine collection of blood and blood components for transfusion or further manufacturing use.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is a guidance for industry and FDA staff entitled “Class II Special Controls Guidance Document: Automated Blood Cell Separator Device Operating by Centrifugal or Filtration Separation Principle.”

## Predicate Devices

- AMICUS Separator System ([K192150](/device/K192150.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 11, 2020

Fresenius Kabi AG Kim Forch Manager, Regulatory Affairs 3 Corporate Drive Lake Zurich, IL 60047

Re: K200530 Trade/Device Name: AMICUS Separator System Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: LKN, GKT Dated: August 11, 2020 Received: August 12, 2020

Dear Kim Forch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K200530

Device Name

AMICUS Separator System

#### Indications for Use (Describe)

Depending on the AMICUS Separator System disposable used in the therapeutic apheresis procedure, the AMICUS Separator System has been cleared for the following:

The AMICUS Separator System is an automated blood cell separator indicated to perform therapeutic plasma exchange (TPE). (K111702)

The AMICUS Exchange Kit is indicated for use in therapeutic plasma exchange (TPE). The kit is for use with the AMICUS separator. (K111702)

The AMICUS Separator System is an automated blood component separator indicated to perform red blood cell exchange (RBCX), including exchange and depletion/exchange procedures, for the transfusion management of sickle cell disease in adults and children. (K180615)

The AMICUS Exchange Kit – Therapeutics is indicated for use in therapeutic plasma exchange (TPE) and red blood cell exchange (RBCX). The kit is for use with the AMICUS separator. (K111702, K180615)

The waste transfer set is indicated for use in red blood cell exchange (RBCX). The set is for use with the AMICUS separator. (K180615)

The Blood Component Filter Set with Vented Spike and Luer Adapter is indicated for the administration of blood and blood components during a therapeutic plasma exchange (TPE) or red blood cell exchange (RBCX) therapeutic apheresis procedure. The set is for use with the AMICUS separator. (K111702, K180615)

The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells.

Depending on the AMICUS Separator System apheresis kit used in the collection of products, the AMICUS Separator System has been cleared to collect:

- · Platelets Pheresis, Leukocytes Reduced (single, double, or triple units)(BK960005, BK990009)
- · Platelets Pheresis, Leukocytes Reduced, Platelet Additive Solution (InterSol)(single, double or triple units) (BK090065)
- · Red Blood Cells, Leukocytes Reduced (by apheresis) (BK000039)
- Mononuclear Cells (BK000047)
- Plasma (BK960005, BK120041)
	- o Fresh Frozen Plasma
		- · Must be prepared and placed in a freezer at -18° C or colder within eight hours after phlebotomy.

o Plasma Frozen Within 24 Hours After Phlebotomy (PF24)

- Must be stored at 1-6°C within eight hours after phlebotomy and placed in a freezer at -18° C or colder within 24

{3}------------------------------------------------

hours after phlebotomy.

- · Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.
- o Plasma Frozen Within 24 Hours After Phlebotomy (PF24) Held at Room Temperature Up to 24 Hours After Phlebotomy (PF24RT24)
	- · Can be stored at room temperature for up to 24 hours after phlebotomy. Product must be placed in a freezer at -18° C or colder within 24 hours after phlebotomy.
	- · Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.
- o Source Plasma

The device is designed to collect products while maintaining an extracorporeal volume at or below 10.5ml/kg and a donor post-platelet count greater than or equal to 100,000 platelets/microliter.

Platelet Pheresis (single, double, or triple units) may be manufactured from products that do not meet leukocyte reduction product standards. This does not apply to Platelet Additive Solution (InterSol) (single, double, or triple units).

The AMICUS platelet storage container is cleared to store Platelets Pheresis, Leukocytes Reduced in 100% plasma for up to seven days. Additionally, for platelet units stored past five days and through seven days, every product must be tested with a bacterial device cleared by FDA and labeled as a "safety measure."

| Type of Use (Select one or both, as applicable)                                   |                                                                                 |
|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>       Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) SUMMARY

#### Date Prepared

April 21, 2020

## Owner/Operator

Fresenius Kabi AG 61346 Bad Homburg Germany

#### Contact Person (include full address, phone and fax numbers)

Kim Forch Manager, Regulatory Affairs Fresenius Kabi Three Corporate Drive Lake Zurich, IL 60047 USA T: 847-550-7962 F: 847-550-2960

#### Device Trade Name

AMICUS Separator System

#### Common Name/Usual Name:

Automated Separator, Blood Cell and Plasma, Therapeutic Automated Blood Cell Separator (Centrifugal Separation Principle)

#### Classification Name

Automated separators, used for separation of blood cells and plasma for therapeutic purposes, have not been classified under a regulation by the Center for Devices and Radiological Health due to pre-amendment status.

21 CFR 864.9245 Automated Blood Cell Separator

Automated blood cell separators which are based on centrifugation type technology have been classified by the Center for Biologics Evaluation and Research as Class II devices with Special Controls (Docket 2005N-0017, Final Rule, 30-Nov-07, updated March 28, 2011 OMB Control No: 0910-0594).

## Product Code and Classification Panel

LKN (Gastroenterology/Urology panel) - Unclassified (due to pre-amendment status) GKT (Hematology panel) - Separator, Automated, Apheresis

{5}------------------------------------------------

## Legally Marketed Device Under Which Substantial Equivalence is Being Claimed

Fresenius Kabi is claiming substantial equivalence with the currently cleared version of the AMICUS Separator System. The AMICUS Separator System is an automated blood cell separator intended for use in therapeutic apheresis applications including Therapeutic Plasma Exchange (TPE), Red Blood Cell Exchange and Red Blood Cell Depletion (RBCX procedures). The AMICUS Separator System was most recently cleared by CDRH under K192150 on Nov. 13, 2019. This includes all operating protocols and changes previously cleared for the AMICUS Separator System.

No reference devices were used in this submission.

#### Device Description

The AMICUS Separator System is comprised of the AMICUS separator instrument and a disposable apheresis kit specific to the procedure being performed. The instrument is a continuous-flow, centrifugal device that draws whole blood from a donor/patient, separates the blood into its components, collects one or more of the blood components, and returns the remainder of the blood components to the donor/patient. The instrument operates using pumps, clamps and valves that move donor/patient blood through a singleuse, sterile fluid path disposable kit. The cells are centrifugally separated within the kit by density differences.

The operator is responsible for connecting and monitoring the donor/patient and operating and monitoring the AMICUS separator during the procedure. The operator controls the separator through a touch screen. When necessary, the operator is warned of problems with messages on the screen and corresponding audible alerts.

Once the cell separation is complete, the operator disconnects the donor/patient, dismantles the kit, and disposes of the kit in a safe manner. The kit is packaged in a recyclable plastic tray.

#### Statement of Intended Use

The AMICUS Separator System is an automated blood component separator intended for use in the collection of blood components and mononuclear cells.

The AMICUS Separator System is an automated blood cell separator intended for use in therapeutic apheresis applications and may be used to perform Therapeutic Plasma Exchange (TPE).

The AMICUS Separator System is an automated blood component separator intended for use in therapeutic apheresis applications and may be used to perform red blood cell exchange, depletion, and depletion/exchange (RBCX) procedures.

{6}------------------------------------------------

#### Indications for Use

Depending on the AMICUS Separator System disposable used in the therapeutic apheresis procedure, the AMICUS Separator System has been cleared for the following:

The AMICUS Separator System is an automated blood cell separator indicated to perform therapeutic plasma exchange (TPE). (K111702)

The AMICUS Exchange Kit is indicated for use in therapeutic plasma exchange (TPE). The kit is for use with the AMICUS separator. (K111702)

The AMICUS Separator System is an automated blood component separator indicated to perform red blood cell exchange (RBCX), including exchange and depletion/exchange procedures, for the transfusion management of sickle cell disease in adults and children. (K180615)

The AMICUS Exchange Kit – Therapeutics is indicated for use in therapeutic plasma exchange (TPE) and red blood cell exchange (RBCX). The kit is for use with the AMICUS separator. (K111702, K180615)

The waste transfer set is indicated for use in red blood cell exchange (RBCX). The set is for use with the AMICUS separator. (K180615)

The Blood Component Filter Set with Vented Spike and Luer Adapter is indicated for the administration of blood and blood components during a therapeutic plasma exchange (TPE) or red blood cell exchange (RBCX) therapeutic apheresis procedure. The set is for use with the AMICUS separator. (K111702, K180615)

The AMICUS Separator System is an automated blood cell separator indicated for the collection of blood components and mononuclear cells.

Depending on the AMICUS Separator System apheresis kit used in the collection of products, the AMICUS Separator System has been cleared to collect:

- Platelets Pheresis, Leukocytes Reduced (single, double, or triple units) (BK960005, . BK990009)
- Platelets Pheresis, Leukocytes Reduced, Platelet Additive Solution (InterSol) (single, ● double or triple units) (BK090065)
- Red Blood Cells, Leukocytes Reduced (by apheresis) (BK000039)
- Mononuclear Cells (BK000047) ●
- Plasma (BK960005, BK120041)
	- Fresh Frozen Plasma o

{7}------------------------------------------------

- . Must be prepared and placed in a freezer at -18° C or colder within eight hours after phlebotomy.
- Plasma Frozen Within 24 Hours After Phlebotomy (PF24) O
	- . Must be stored at 1-6℃ within eight hours after phlebotomy and placed in a freezer at -18° C or colder within 24 hours after phlebotomy.
	- י Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.
- Plasma Frozen Within 24 Hours After Phlebotomy (PF24) Held at Room о Temperature Up to 24 Hours After Phlebotomy (PF24RT24)
	- Can be stored at room temperature for up to 24 hours after phlebotomy. I Product must be placed in a freezer at -18° C or colder within 24 hours after phlebotomy.
	- י Indicated for replacement of non-labile clotting factors. This product is not equivalent to Fresh Frozen Plasma.
- o Source Plasma

The device is designed to collect products while maintaining an extracorporeal volume at or below 10.5ml/kg and a donor post-platelet count greater than or equal to 100,000 platelets/microliter. (BK960005 and BK990009)

Platelet Pheresis (single, double, or triple units) may be manufactured from products that do not meet leukocyte reduction product standards. This does not apply to Platelet Pheresis, Platelet Additive Solution (InterSol) (single, double, or triple units). (BK990009 and BK090065)

The AMICUS platelet storage container is cleared to store Platelets Pheresis, Leukocytes Reduced in 100% plasma for up to seven days. Additionally, for platelet units stored past five days and through seven days, every product must be tested with a bacterial device cleared by FDA and labeled as a "safety measure." (BK050038 and BK150242)

#### Technological Comparison as Compared to the Predicate Device

The technological characteristics of the AMICUS separator are standardized across procedures and remain the same as the predicate AMICUS device. This includes the centrifuge system, separation technology, fluid control system, safety management system (including safety sensors and alerts) and anticoagulant management system which monitors and controls the delivery of anticoagulant solution based on operator parameters. There is no change in the fundamental scientific technology or principles of operation.

{8}------------------------------------------------

The physical design of the AMICUS instrument includes three main areas of operation (touch screen, top panel and centrifuge) that have the same technological characteristics as the predicate AMICUS device. There is no change in physical dimensions, manufacturing facility or processes. The performance specifications and data management capabilities remain the same as the cleared AMICUS device. The AMICUS apheresis kits are unaffected and remain the same as the currently cleared kits including design, labeling, materials and manufacturing methods.

| Element                       | Predicate Device                                                                                                                                              | Proposed Device                                                                                                 |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Technological Characteristics |                                                                                                                                                               |                                                                                                                 |
| Touch Screen                  | LCD display                                                                                                                                                   | SAME<br>Note: The resolution of the LCD<br>display on the touch screen was<br>changed from 800x600 to 1024x768. |
| Top Panel                     | Disposable Control (e.g., pumps, clamps and sensors)                                                                                                          | SAME                                                                                                            |
| Centrifuge System             | Continuous Flow Centrifugation<br>Spool/Spool Holder<br>Interface Detector<br>Centrifuge Rotor Speed<br>Centrifugal Force                                     | SAME                                                                                                            |
| Fluid Control                 | Container Weight Scales and Pumps                                                                                                                             | SAME                                                                                                            |
| Safety System                 | Air Detector<br>Air Traps<br>Pressure Sensors<br>Return line filter<br>Centrifuge Leak Sensor<br>Audible Alerts with Visual Display<br>Power Failure Recovery | SAME                                                                                                            |
| Anticoagulant<br>Management   | Parameters for ACD Ratio and<br>Citrate Infusion Rate                                                                                                         | SAME                                                                                                            |

A comparison of the technological characteristics between the predicate device and the subject device is provided in the table below.

## Modification to the Existing Device

The primary changes to the AMICUS device in support of this submission are the replacement of obsolete controller circuit boards and associated redesign of other boards within the interior portion of the instrument. The replacement boards have resulted in new proposed product codes (6R4590, 6R4590TH, 6R4590TH-M and 6R4590TH-T). There was a modification made to the rear door design of the instrument's exterior interface to accommodate two new USB ports, one new ethernet port, one serial port (reduced from four) and one Wi-Fi antenna.

An update to the cleared AMICUS 6.0 software, including a change in the operating system from pSOS to QNX, was developed to run on the proposed device with the

{9}------------------------------------------------

replacement boards. The software changes do not involve changes to the core blood component collection or therapeutic procedure functionality. The software updates were made to enable use with the new hardware as well as some updates to diagnostics mode to enhance serviceability. Other non-procedural updates include embedded Wi-Fi and ethernet connectivity and the ability to export procedural data and network configuration settings via USB. The barcode module in the AMICUS software was updated for compatibility with the new USB-powered bar code scanner.

A new off-the-shelf bar code scanner with USB interface has been qualified for use with the USB port in the new I/O board. The USB bar code scanner will function the same as the serial-based bar code scanner used with the predicate device.

The cleared SW 6.0 AMICUS Operator's Manual, 6.0 Data Management Supplement and EMC Addendum have been updated in support of the proposed device and updated software. A new 6.0 Cyber Security Supplement has been created. New product labeling has been created for the new product codes.

The AMICUS disposable apheresis kits remain the same as the currently cleared kits, including design, materials and manufacturing methods. The labeling for the AMICUS apheresis disposable kits are not affected by the proposed AMICUS device and remain unchanged. The data management capabilities remain the same as the cleared AMICUS device.

#### Performance Data

Software and systems verification testing of impacted requirements were performed with the replacement hardware components and updated 6.0 software. Regression testing was performed including paired testing with the predicate device on key product quality and performance characteristics. Results from the software and systems verification testing indicate that the subject device is as safe and performs as well as the predicate device.

Electrical safety and EMC testing were also conducted on the subject device. The device passed all applicable tests in compliance with:

ANSI/AAMI ES60601-1:2005/(R)2012 and A1: 2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Edition 3.1) "Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance."

IEC 60601-1-2:2014 (Edition 4) "Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests."

#### Conclusion

Based on the verification activities performed, the AMICUS Separator System modified with the replacement hardware components and updated 6.0 software provides a device system that is substantially equivalent to the currently marketed AMICUS Separator System.

---

**Source:** [https://fda-staging.innolitics.com/device/K200530](https://fda-staging.innolitics.com/device/K200530)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
