← Product Code [PGY](/productcode/PGY) · K200485

# RadiForce RX1270, RadiForce RX1270-AR (K200485)

_Eizo Corporation · PGY · Mar 23, 2020 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200485

## Device Facts

- **Applicant:** Eizo Corporation
- **Product Code:** [PGY](/productcode/PGY.md)
- **Decision Date:** Mar 23, 2020
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

This product is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

## Device Story

RadiForce RX1270/RX1270-AR is a 30.9-inch color LCD monitor for diagnostic radiology. It accepts digital video inputs (DisplayPort, HDMI) and displays high-resolution (12MP) medical images. The device uses IPS panel technology and LED backlighting. It features factory-calibrated lookup tables to ensure DICOM GSDF compliance. The RX1270-AR variant includes an anti-reflection coating, while the RX1270 uses anti-glare treatment. Used in clinical settings by radiologists and medical practitioners, the monitor provides high-fidelity visualization of mammography and tomosynthesis scans. It integrates with RadiCS software for quality control, luminance calibration, and performance testing. By providing a large, high-resolution display, it replaces traditional dual-monitor setups, potentially improving diagnostic workflow efficiency and image review accuracy for the clinician.

## Clinical Evidence

Bench testing only. No clinical or animal data provided. Testing followed FDA guidance 'Display Devices for Diagnostic Radiology' (2017) and AAPM TG18 guidelines, covering MTF, luminance, DICOM GSDF conformance, temporal response, uniformity, and reflection measurements.

## Technological Characteristics

TFT Color LCD panel (IPS) with LED backlight. 30.9" screen, 4200 x 2800 resolution (12MP), 0.1554 mm pixel pitch. 1200 cd/m² max luminance. Connectivity: DisplayPort, HDMI. Integrated optical sensor for calibration. RadiCS software for QC. Anti-glare (RX1270) or anti-reflection (RX1270-AR) surface treatment.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- RadiForce RX560, RX560-AR ([K172738](/device/K172738.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the agency's name on the right. The symbol on the left is a stylized depiction of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "FOOD & DRUG" are larger than the other words.

March 23, 2020

EIZO Corporation % Mr. Hiroaki Hashimoto Senior Manager of Regulatory Compliance and Safety Department 153 Shimokashiwano, Hakusan Ishikawa 924-8566 JAPAN

Re: K200485

Trade/Device Name: RadiForce RX1270, RadiForce RX1270-AR Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: PGY Dated: February 25, 2020 Received: February 27, 2020

Dear Mr. Hashimoto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200485

Device Name RadiForce RX1270, RadiForce RX1270-AR

Indications for Use (Describe)

This product is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

| Type of Use (Select one or both, as applicable)                                 |                                                                                |
|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span>    ☑ Prescription Use (Part 21 CFR 801 Subpart D)   </span> </div> | <div> <span>    ☐ Over-The-Counter Use (21 CFR 801 Subpart C)   </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for EIZO. The logo consists of a stylized square made up of smaller squares, next to the company name "EIZO" in bold, sans-serif font. A registered trademark symbol is located to the right of the company name. The logo is in black and white.

EIZO Corporation, 153 Shimokashiwano, Hakusan, Ishikawa 071-2566 Isnan

U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center - WO66-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002

Name Department Hiroaki Hashimoto Regulatory Compliance and Safety

Telephone Fax E-Mail

+81 (76) 274-2468 +81 (76) 274-2484 hiroaki.hashimoto@eizo.com

#### 510(k) Summary K200485

#### 1. Submitter

EIZO Corporation 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan Phone: +81 (76) 274-2468 Fax: +81 (76) 274-2484

Contact Person: Hiroaki Hashimoto Date of Prepared: February 25th, 2020

#### 2. Device

| • Name of Device:       | RadiForce RX1270,<br>RadiForce RX1270-AR                         |
|-------------------------|------------------------------------------------------------------|
| • Common or Usual Name: | 78.4 cm (30.9 inch) class Color LCD Monitor                      |
| • Classification Name:  | Picture archiving and communications system<br>(21 CFR 892.2050) |
| • Regulatory Class:     | II                                                               |
| • Product Code:         | PGY                                                              |

#### 3. Predicate Device

| • | 510(k) Holder: | EIZO Corporation |
|---|----------------|------------------|
|---|----------------|------------------|

- 510(k) Number: K172738
- Name of Device: RadiForce RX560, RX560-AR ●

II

- Common or Usual Name: 54.1 cm (21.3 inch) class Color LCD Monitor ●
- Classification Name: Picture archiving and communications system ● (21 CFR 892.2050)
- Regulatory Class:
- PGY ● Product Code:

{4}------------------------------------------------

# 4. Device Description

RadiForce RX1270 is a color LCD monitor for viewing medical images including those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 4.200 x 2.800 pixels (12MP) with a pixel pitch of 0.1554 mm. With the matrix size (or resolution) of 4.200 x 2.800 pixels (12MP), the RX1270 is an optimal replacement for traditional dual head 2,048 x 2,560 pixels (5MP) display installations.

Image /page/4/Picture/2 description: The image shows two medical monitors displaying different types of medical scans. The monitor on the left displays a series of CT scans of the chest and lungs, including cross-sectional views and 3D reconstructions. The monitor on the right shows a mammogram on the left screen and ultrasound images of the breast on the right screen. The images suggest the use of advanced medical imaging technologies for diagnostic purposes.

RX1270 (12MP)

RX560 (5MP) x 2

Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.

There are two model variations, RX1270 and RX1270-AR. The difference of the two variations is the surface treatment of the display screens; the surface treatment of the RX1270 is Anti-Glare (AG) treatment and that of the RX1270-AR is Anti-Reflection (AR) coating.

RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce RX1270 based on the OC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce RX1270.

{5}------------------------------------------------

#### 5. Indications for use

The stated intended use of the two devices is substantially the same:

# ● RadiForce RX1270

This product is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.

### ● RadiForce RX560

This product is intended to be used in displaying and viewing digital images, including standard and multiframe digital mammography, for review, analysis and diagnosis by trained medical practitioners. It is specially designed for breast tomosynthesis applications.

{6}------------------------------------------------

# 6. Comparison of Technological Characteristics with the predicate device

The comparison table below enumerates information derived from the product brochure and measured values of the each device and different technological characteristics are discussed in it:

| Attributes                            | Proposed Device:<br>RadiForce RX1270                                                                                                                                                      | Predicate Device:<br>RadiForce RX560                                                    |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Display Technology                    | TFT Color<br>LCD Panel (IPS)                                                                                                                                                              | TFT Color<br>LCD Panel (IPS)                                                            |
| Screen size                           | 78.4cm / 30.9"                                                                                                                                                                            | 54.1cm / 21.3"                                                                          |
| Resolution                            | 4200 x 2800 (3:2 aspect ratio)                                                                                                                                                            | 2048 x 2560 (4:5 aspect ratio)                                                          |
| Pixel pitch                           | 0.1554 x 0.1554 mm                                                                                                                                                                        | 0.165 x 0.165 mm                                                                        |
| Backlight type                        | LED                                                                                                                                                                                       | LED                                                                                     |
| Maximum luminance                     | 1200 cd/m²                                                                                                                                                                                | 1100 cd/m²                                                                              |
| DICOM calibrated luminance            | 500 cd/m²                                                                                                                                                                                 | 500 cd/m²                                                                               |
| Contrast Ratio (typical)              | 1500:1                                                                                                                                                                                    | 1500:1                                                                                  |
| Response Time (typical)               | 12 ms (black-white-black)                                                                                                                                                                 | 12 ms (black-white-black)                                                               |
| Frame rate and refresh rate           |                                                                                                                                                                                           |                                                                                         |
| Digital Scanning Frequency<br>(H / V) | HDMI: 31 - 160 kHz / 59 - 61 Hz<br>(VGA Text: 69 - 71 Hz)<br>DisplayPort: 31 - 175 kHz / 29 -<br>61 Hz<br>(VGA Text: 69 - 71 Hz)<br>Frame synchronous mode:<br>29.5 - 30.5 Hz, 59 - 61 Hz | 31 - 135 kHz / 23 - 61 Hz<br>Frame synchronous mode: 23.5 -<br>25.5 Hz, 47 - 51 Hz      |
| Analog Scanning Frequency<br>(H / V)  | HDMI: 31 - 160 kHz / 59 - 61 Hz<br>(VGA Text: 69 - 71 Hz)<br>DisplayPort: 31 - 175 kHz / 29 -<br>61 Hz<br>(VGA Text: 69 - 71 Hz)<br>Frame synchronous mode:<br>29.5 - 30.5 Hz, 59 - 61 Hz | 31 - 135 kHz / 23 - 61 Hz<br>Frame synchronous mode: 23.5 -<br>25.5 Hz, 47 - 51 Hz      |
| Input video signals                   | DisplayPort x 2,<br>HDMI                                                                                                                                                                  | DVI-D (dual link) x 1,<br>DisplayPort x 1                                               |
| Output video signals                  | -                                                                                                                                                                                         | DisplayPort x 1 (daisy chain)                                                           |
| Video bandwidth                       |                                                                                                                                                                                           |                                                                                         |
|                                       | HDMI: 260MHz<br>DisplayPort: 765MHz                                                                                                                                                       | DVI: 290MHz<br>DisplayPort: 290MHz                                                      |
| Ambient light sensing                 |                                                                                                                                                                                           |                                                                                         |
| Ambient light sensor                  | Yes                                                                                                                                                                                       | Yes                                                                                     |
| Luminance calibration tools           |                                                                                                                                                                                           |                                                                                         |
|                                       | Integrated optical sensor<br>External optical sensor<br>Calibration software:<br>RadiCS                                                                                                   | Integrated optical sensor<br>External optical sensor<br>Calibration software:<br>RadiCS |
| Quality-control procedures            |                                                                                                                                                                                           |                                                                                         |
|                                       | Software: RadiCS                                                                                                                                                                          | Software: RadiCS                                                                        |

{7}------------------------------------------------

It is clear that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the RX1270.

{8}------------------------------------------------

# 7. Performance Testing

The bench tests below were performed on the RadiForce RX1270 following the instructions in "Guidance for Industry and FDA Staff: Display Devices for Diagnostic Radiology" issued on October 2, 2017:

- . Measurement of spatial resolution expressed as modulation transfer function (MTF)
- The maximum number allowed for each type of pixel defects/faults ●
- . Visual check of presence or absence of miscellaneous artifacts on the display screen as specified in TG18 guideline
- Measurement of temporal response
- Measurement of Luminance ●
- Verification of the conformance to DICOM GSDF as specified in Assessment of . Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)
- . Measurement of the angular dependency of luminance response in horizontal, vertical and diagonal directions
- . Measurement of the luminance non-uniformity characteristics of the display screen as specified in TG18 guideline
- . Measurement of the chromaticity non-uniformity characteristics of the display screen as specified in TG18 guideline
- . Performance data on luminance stability
- Measurement of noise expressed as noise power spectrum (NPS)
- Measurement of display reflections including specular, diffuse and haze components ●
- . Measurement of small-spot contrast ratio
- Measurement of pixel aperture ratio ●
- Measurement of Color tracking and Gray tracking

The test results showed that the RadiForce RX1270 has display characteristics equivalent to those of the predicate device, RadiForce RX560.

Besides, the display characteristics of the RadiForce RX1270 meet the pre-defined criteria when criteria are set.

No animal or clinical testing was performed on the RadiForce RX1270.

{9}------------------------------------------------

## 8. Conclusion

The RadiForce RX1270 was determined to be substantially equivalent to the predicate device due to the following reasons:

- . The stated intended use is substantially the same as that of the predicate device.
- . It was confirmed that the technological characteristics differences from those of the predicate device do not affect the safety or the effectiveness.
- The bench tests demonstrated that the display characteristics are equivalent to those of . the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K200485](https://fda-staging.innolitics.com/device/K200485)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).