K200461 · Detax GmbH & Co. KG · EBI · Nov 16, 2020 · Dental
Device Facts
Record ID
K200461
Device Name
Freeprint denture
Applicant
Detax GmbH & Co. KG
Product Code
EBI · Dental
Decision Date
Nov 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Indications for Use
FREEPRINT denture is a light-cured resin indicated for the fabrication of all types of denture bases, for instance full and partial removable dentures. FREEPRINT denture is intended for continuous use in the oral environment, exclusively for professional dental work. FREEPRINT denture can be used in combination with a stereolithographic 3D printer using a 385nm light source. A 3D-printer is not part of the device.
Device Story
FREEPRINT denture is a liquid, one-component, dimethacrylate-based photo-curable resin; supplied in HDPE bottles. Used by dental professionals to fabricate denture bases via stereolithographic 3D printing (DLP). The resin is polymerized using a 385nm light source; post-cured with a xenon flashlight in a nitrogen atmosphere. The device is non-sterile. It provides the physical material for denture base construction; clinical decision-making relies on the dental professional's design and the printer's output. Benefits include the creation of custom-fit, durable denture bases for patients requiring tooth replacement.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated against ISO 20795-1 for compressive/tensile strength, hardness, water sorption, and solubility. Biocompatibility testing performed per ISO 10993-3, 5, 10, 11, 17, and ISO 7405.
Technological Characteristics
Dimethacrylate-based resin; one-component system. Polymerization via 385nm light source. Processed via stereolithographic 3D printing (DLP). Complies with ISO 20795-1 for mechanical properties and ISO 10993 series for biocompatibility. Provided non-sterile; 2-year shelf life.
Indications for Use
Indicated for the fabrication of full and partial removable denture bases for professional dental use.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
November 16, 2020
Detax GmbH & Co. KG % Jan-Paul Van Loon Consultant Qserve Consultancy BV. Utrechtseweg 310 - Bldg B42 Arnhem, Gelderland 6812 AR Netherlands
Re: K200461
Trade/Device Name: Freeprint denture Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: October 19, 2020 Received: October 20, 2020
Dear Jan-Paul Van Loon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph. D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200461
Device Name FREEPRINT denture
#### Indications for Use (Describe)
FREEPRINT denture is a light-cured resin indicated for the fabrication of all types of denture bases, for instance full and partial removable dentures.
FREEPRINT denture is intended for continuous use in the oral environment, exclusively for professional dental work. FREEPRINT denture can be used in combination with a stereolithographic 3D printer using a 3Dprinter is not part of the device.
Type of Use (Select one or both, as applicable):
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K200461 - 510(k) Summary
| I SUBMITTER | |
|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Name: | DETAX GmbH & Co.KG. |
| Submitter Address: | Carl-Zeiss-Stasse 4<br>D-76275 Ettingen Germany |
| Phone Number: | +49 (0)7243/510-138 |
| Contact Person: | M. (Markus) Stratmann |
| Date Prepared: | February 6, 2020 |
| II. DEVICE | |
| Device Trade Name: | FREEPRINT denture |
| Common Name | Denture relining, repairing, or rebasing resin |
| Classification | 21 CFR 872.3760 |
| Product Code: | EBI |
| Review Panel: | Dental |
| III. PREDICATE DEVICE | K162572, NextDent™ Denture/E-Denture, Vertex-Dental BV., |
| IV. DEVICE DESCRIPTION | |
| Description | FREEPRINT denture is a light-cured one-component<br>materials for the Digital Light Processing (DLP)-printing of<br>denture base products. It is stored in 500 and 1000ml HDPE<br>bottles. This Product is a liquid photo-curable material that is<br>polymerized, by the photo-initiator contained in the resin. This<br>device can be manufactured with 3D stereolithographic<br>printer with 385nm wavelength light source. Curing in a 3D<br>printer is related to the conditions of the printer equipment<br>and is typically 100µm in thickness. FREEPRINT denture<br>products are post cured with a xenon flashlight, in a nitrogen<br>atmosphere. |
| Physical Description: | FREEPRINT denture: dimethacrylate based resin<br>1-component system.<br>Processed (3D printed) using 385nm light source. |
| V. INTENDED USE /<br>INDICATION FOR USE<br>STATEMENT | FREEPRINT denture is a light-cured resin indicated for the<br>fabrication of all types of denture bases, for instance full and<br>partial removable dentures.<br>FREEPRINT denture is intended for continuous use in the<br>oral environment, exclusively for professional dental work.<br>FREEPRINT denture can be used in combination with a<br>stereolithographic 3D printer using a 385nm light source. A<br>3D-printer is not part of the device. |
| VI SUMMARY OF | |
| PREDICATE DEVICES<br>Indications for Use<br>and<br>Technological<br>Characteristics: | FREEPRINT denture has a similar Indications for Use<br>statement as the predicate device.<br>The FREEPRINT denture has the following similarities in the<br>technological characteristics to the predicate device,<br>NextDent™ Denture/E-Denture, Vertex-Dental BV<br>(K162572):<br>- Same type of materials (chemical composition) that<br>form the basis for the device used<br>- Same technology, however different wavelength used<br>for polymerization (Curing) method<br>- Similar mechanical and physical properties of<br>processed device. |
| Summary of Performance<br>Data and Design<br>Controls: | Bench testing was carried out on the following<br>characteristics:<br>• Printers validation<br>Four printers are validated for processing (3D printing) using<br>FREEPRINT denture. Additional printers post-510(k)-<br>clearance will be added to the labelling by means of the<br>Quality Systems and within the validation plan presented in<br>this 510(k). |
| | Referenced Standards and Performance Testing:<br>Performance testing confirmed the FREEPRINT denture<br>demonstrated performance to the acceptance criteria referred<br>to ISO 20795-1.<br>The composition of FREEPRINT denture exhibits sufficient<br>strengths and performances in all intraoral conditions and will<br>sufficiently resist compressive & tensile loads, hardness,<br>water sorption and solubility. |
| | Biocompatibility testing:<br>FREEPRINT denture was tested with respect to<br>biocompatibility according to ISO 10993-3, ISO 10993-5, ISO<br>10993-10, ISO 10993-11 and ISO 10993-17, taking ISO 7405<br>into account. The results showed that the insolubility is in<br>compliance with the requirements of the standard. There is<br>no evidence that hazardous effects will arise by leachable<br>ingredients/contaminants. |
| | Clinical Tests<br>No clinical tests were performed with FREEPRINT denture |
| Sterility and Shelf-Life<br>Testing: | Sterilization<br>FREEPRINT denture is provided non-sterile. |
| | From the Shelf-life testing, FREEPRINT denture has a shelf<br>life of two years. |
| VIII. CONCLUSION OF<br>SUBSTANTIAL | FREEPRINT denture and the predicate have the same<br>intended use and similar technological characteristics. |
### VI. SUMMARY OF TECHNOLOGICAL CHRACHTERISTICS OF DEVICE COMPARED TO
{4}------------------------------------------------
{5}------------------------------------------------
The results of the performed tests show that FREEPRINT EQUIVALENCE denture meets the requirements mentioned in the applicable standards, and confirm that the device performs similarly to the predicate device.
It is therefore concluded that FREEPRINT denture is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.