The Inbody blood pressure monitor is designed to measure blood pressure (diastolic and systolic) and pulse rate in adult patients with arm circumference range between 17cm - 42cm.
Device Story
InBody Blood Pressure Monitors (BP170 series, BP160 series, BPBIO250) are automatic oscillometric devices for measuring systolic/diastolic blood pressure and pulse rate. Devices consist of a measurement unit, pump, and external cuff (17-42cm circumference). Operation involves wrapping the cuff around the upper arm; the device automatically inflates/deflates to detect pressure oscillations. Results are displayed on an LCD. Models are portable, battery-operated (or AC), and designed for clinical or home use. Data is processed by an internal 32-bit microcontroller using software reused from the predicate. Future Bluetooth implementation is intended for MDDS (Medical Device Data System) functionality. The device provides objective blood pressure readings to assist healthcare providers in patient monitoring and diagnosis.
Clinical Evidence
Clinical testing performed per ANSI/AAMI SP10:2002/A1:2003/(R)2008. BP170 tested on 106 subjects; BPBIO250 tested on 104 subjects (312 observations). Primary endpoints were mean error of measurement (max 5mmHg) and experimental standard deviation (max 8mmHg). Results confirmed substantial equivalence to manual mercury manometer measurements.
Technological Characteristics
Oscillometric measurement principle. Components: LCD, external cuff, pump, 32-bit MCU. Power: 4-AA alkaline or rechargeable lithium with AC adapter. Dimensions vary by model. Connectivity: Future Bluetooth (MDDS). Standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11, IEC 80601-2-30, IEC 60601-1-2, IEC 62304.
Indications for Use
Indicated for adult patients with arm circumference 17cm-42cm for measurement of systolic/diastolic blood pressure and pulse rate. BP170, BP160, BP160B are for prescription and OTC use; BPBIO250 is for prescription use only.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 4, 2020
InBody Co. Ltd. % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114
Re: K200442
Trade/Device Name: InBody Blood Pressure Monitors: BP170B; BP160; BP160B; BPBIO250 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 3, 2020 Received: November 4, 2020
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200442
Device Name
InBody Blood Pressure Monitors: BP170; BP170B; BP160; BP160B; BPBIO250
#### Indications for Use (Describe)
The Inbody blood pressure monitor is designed to measure blood pressure (diastolic and systolic) and pulse rate in adult patients with arm circumference range between 17cm - 42cm.
The models of BP170, BP160 and BP160B are designated as both Prescription use and Over-the-counter use.
The model of BPBIO250 is designated as Prescription use only.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K200442
# 510(k) Summary K200442
Image /page/3/Picture/2 description: The image shows the word "InBody" in a bold, sans-serif font. The color of the text is a dark reddish-brown. The letters are closely spaced together, creating a solid block of text. The background is plain white, which makes the text stand out.
InBody Co., Ltd. InBody Bldg., 625, Eonju-ro, Gangnam-gu Seoul KOREA 06106 TEL. +82 02-501-3939 Date prepared: December 2, 2020 Contact: Kichul Cha, CEO
- 1. Identification of the Devices: Proprietary-Trade Names: BP170; BP170B; BP160; BP160B; BPBIO250. Common/Usual Name: Blood pressure monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Meter Regulatory Class: Class II Product Code: DXN
- 2. Equivalent legally marketed device: K131064, Biospace Corporation Limited (Former company name) Proprietary-Trade Names: BPBIO320/BPBIO320n Common/Usual Name: Blood pressure monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Non-Invasive Blood Pressure Meter Regulatory Class: Class II Product Code: DXN
- 3. Indications for Use (intended use) The Inbody blood pressure monitor is designed to measure blood pressure (diastolic and systolic) and pulse rate in adult patients with arm circumference range between 17cm - 42cm.
- 4. Descriptions of the different models: All models have this common feature: Automatic blood pressure measurement using the oscillometric technique. This is the same method as our predicate. We have employed essentially the same technology as we used in our previous model (the predicate) but have packaged the unit differently in order to make it more portable. Instead of having a built in blood pressure cuff, we have moved that component outside of the unit so that now it plugs in to the measurement unit in the same way a large number of similar devices are designed. Instead of AC line operation only, the units can run on primary or on rechargeable batteries. The predicate device used an 8 bit microcontroller with 128Kbytes of in-system programmable flash memory. The BP170 series and the BPBIO250 use 32 bit microcontrollers with 128Kbytes of programmable flash. Most of the measurement software has been re-used from our predicate device and is virtually the same between the two new sets of models.
{4}------------------------------------------------
## K200442
The BP170 is a fully automatic portable battery operated blood pressure monitor using the oscillometric technique which comes with batteries, a standard cuff, a storage pouch, and a user's manual. Model variations:
| | Options | | |
|------------|-------------------|----------------|------------|
| Model name | One-touch<br>Cuff | Normal<br>Cuff | Bluetooth* |
| BP170 | Yes | No | No |
| BP170B | Yes | No | Yes |
| BP160 | No | Yes | No |
| BP160B | No | Yes | Yes |
*Bluetooth is not currently supported. For future use. When implemented, it will operate as an MDDS, a medical device data system, product code OUG.
Image /page/4/Picture/4 description: The image shows a blood pressure monitor displaying a reading of 126/80 with a pulse rate of 75 bpm. The monitor also indicates the date as 2/15 and the time as 9:00 AM. The device has several buttons, including M1, M2, up and down arrows, and a start/stop button. The brand name "InBody" is visible at the bottom of the device.
Photo of BP170
BPBIO250 This product is a professional automatic blood pressure monitor used by medical professionals. 5 measurement modes can be selected according to various medical environments. Rechargeable batteries can be used without replacing batteries. The User can measure blood pressure even in dark environments. - This blood pressure monitor has cuffs of various sizes. So, User can measure the blood pressure and measure pulse rate to children and adult patients whose arm circumferences are between 17 cm and 42 cm. Bluetooth, when implemented (future option) it will operate as an MDDS, a medical device data system, product code OUG.
# Photo of BPBIO250
Image /page/4/Picture/8 description: The image shows a blood pressure monitor with a cuff. The monitor has a large digital display that shows the systolic blood pressure as 128, the diastolic blood pressure as 64, and the pulse rate as 72. The monitor is white with a black trim around the display screen.
- 5. Safety and Effectiveness, comparison to predicate device. The testing results and specification comparisons indicate that the new models are as safe and effective as the predicate device. A comparison table is presented below.
{5}------------------------------------------------
| K200442 | | | |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Item | Predicate Device<br>BPBIO320/BPBIO320n K131064,<br>Biospace Corporation Limited (Our<br>former company name) | Proposed Models: BP170; BP170B; BP160;<br>BP160B; BPBIO250 | Comparison<br>Result |
| Intended Use | The Biospace blood pressure<br>monitor is designed to measure<br>blood pressure (diastolic and<br>systolic) and pulse rate in adult<br>patients with arm circumference<br>range between 17cm - 42cm. | The InBody blood pressure monitor is<br>designed to measure blood pressure<br>(diastolic and systolic) and pulse rate in<br>adult patients with arm circumference<br>range between 17cm - 42cm. | SAME |
| Patient<br>Population | Adult | Adult | Same |
| Measurement<br>location | Upper Arm | Upper Arm | Same |
| Measurement<br>Principle | Oscillometric | Oscillometric | Same |
| Measured: | Blood Pressure and Pulse Rage | SAME | Same |
| Blood<br>Pressure Cuff | Internal to the device | External to the device | Makes the<br>new device<br>more<br>portable |
| Components | LCD, Cuff, MCU, Pump | LCD, Cuff, MCU, Pump | Same |
| Power Source | AC Line | BP170; 4-AA Size Alkaline (or AC Line)<br>BPBIO250; 4-AA Size Alkaline (or AC Line)<br>Rechargeable Lithium with AC Line<br>adapter | All of the<br>new models<br>can work on<br>batteries. |
| Dimensions | 489(W) x 409(L) x 284(H)mm | BP170; BP170B; BP160; BP160B:<br>99(W) X 191(L) X 26(H): mm<br>BPBIO250: 122(W) x 150(L) x 195(H):<br>mm | The new<br>units are<br>smaller |
| Measurement<br>range | Blood pressure:<br>40-300mmHg,<br>Pulse rate: 30-240bpm | BP170: 0-300 mmHg,<br>Pulse: 30-240 bpm<br>BPBIO250; 0-300mmHg,<br>Pulse: 30-240 bpm | Ranges are<br>comparable. |
| Accuracy | Pressure: ±3mmHg<br>Pulse rate: ±2% | BP170: Pressure: ±3 mmHg,<br>Pulse: Within ±3%<br>BPBIO250: Pressure: Within ±3 mmHg,<br>Pulse: Within ±3%<br>Pressure: ± 2mmHg Pulse: Within ± 1.5% | Accuracies are<br>not<br>significantly<br>different. The<br>new offers<br>the best<br>accuracy |
| Range of arm<br>circumference | 17cm - 42cm. | SAME 17cm-42cm | SAME |
| K200442 | | | |
| Item | Predicate Device<br>BPBIO320/BPBIO320n K131064,<br>Biospace Corporation Limited (Our<br>former company name) | Proposed Models: BP170; BP170B; BP160;<br>BP160B; BPBIO250 | Comparison<br>Result |
| Photos | Image: BPBIO350 | Image: New Models BP170, BPBIO250 | The new<br>models take<br>up less table<br>space and use<br>a traditional<br>cuff. |
| Bluetooth | Not available | BP170B<br>BP160B<br>BPBIO250 | Future option<br>as MDDS |
{6}------------------------------------------------
- 6. Summary of technological characteristics of the device compared to the predicate device. These blood pressure/pulse rate meters are intended to be used in measuring human systolic, diastolic and pulse rate by oscillometric (or manual) method. Performance characteristics are in accordance with ANSI/AAMI SP10:2002/ (R) 2008. The substantial equivalence between these new meters and the predicate BPBIO320n can be evaluated from several aspects as listed in above table. The following FDA guidance was consulted in the design and testing of the device: Non-Invasive Blood Pressure (NIBP) Monitor Guidance MARCH 1997, Final
- 7. Non-clinical testing: The proposed new models were tested and found to conform to the following international standards:
IEC 60601-1 :2005/AMD1:2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance FDA recognition 19-4
IEC 60601-1-6:2010, +AMD1:2013, General requirements for safety - Collateral Standard: Usability FDA recognition 5-89
IEC 60601-1-11:2015, MEDICAL ELECTRICAL EQUIPMENT -Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment FDA recognition 19-14 (Applies only to models BP170; BP170B; BP160; BP160B)
IEC 80601-2-30:2009, +AMD1:2013 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers FDA recognition 3-152
IEC 60601-1-2 [2014] Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic Compatibility Requirements and tests FDA recognition 19-8
{7}------------------------------------------------
### K200442
IEC 62304 [2006] Medical device software life-cycle processes FDA recognition 13-79
- 8. Clinical Testing: Successful testing was performed according to ANSI/AAMI SP10 : 2002 & ANSI/AAMI SP10 : 202/A1 : 2003/ (R) 2008 and the FDA Guidance Document. The SP10 standard calls for a minimum of 85 subjects using a minimum of 255 paired observations, comparing to a manual blood pressure measurement. Separate clinical investigations were performed for the BP170 Series and the BPBIO250. For the BP170, 106 subjects were tested. For the BPBIO250, 104 subjects with 312 observations were tested. The criteria for the overall system accuracy follow the SP10 requirements:
a) maximum mean error of measurement: 5mmHg
- b) maximum experimental standard deviation: 8mmHg
InBody has compared the accuracy between mercury manometer using the stethoscope and the non-invasive automatic sphygmomanometers. We found and concluded that they are substantially equal to each other.
- 9. Conclusion, Comparison to the predicate device. Proposed Models: BP170; BP160B; BPBIO250 are substantially equivalent to the BPBIO320/BPBIO320n whose 510(k) number is K131064. The devices are identical in the intended use, and very similar in the design principles, the performance and the applicable standards. Only their appearance, use of an external blood pressure cuff, and the user interfaces are different.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.