K200431 · Surgentec · FMF · May 18, 2020 · General Hospital
Device Facts
Record ID
K200431
Device Name
3D GraftRasp System
Applicant
Surgentec
Product Code
FMF · General Hospital
Decision Date
May 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The 3D GraftRasp System is intended to be used in orthopedic procedures to rasp or decorticate bone and for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site.
Device Story
3D GraftRasp System comprises decortication rasps, graft pushers, bone funnel, and dilators; used in orthopedic procedures. Device enables manual tissue dilation and bone decortication via rasping surface. Following preparation, graft delivery occurs by inserting graft tube into rasp lumen or using bone funnel; compatible with Graftgun Universal Graft Delivery System for graft extrusion. Operated by surgeons in clinical/surgical settings. Output is physical bone preparation and graft placement at surgical site. Benefits include improved access to bony anatomy and efficient graft delivery. System includes reusable and single-use components (e.g., replaceable rasp teeth, flexible pusher tips).
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1:2018), sterilization validation (ISO 17665-1), graft extrusion functionality using various materials, rasp decortication performance on bone blocks, and volume dispensing accuracy verification.
Technological Characteristics
Materials: 316L/304 stainless steel (ASTM F899-12b, passivated per ASTM A967-13), Nitinol (ASTM F2063). Manual operation; no energy source. Sterilization: Steam. Components: Rasps, dilators, pushers, funnel. Connectivity: None.
Indications for Use
Indicated for patients undergoing orthopedic procedures requiring bone decortication and delivery of bone graft materials (allograft, autograft, or synthetic).
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Graftgun Universal Graft Delivery System (K170675)
Bone Rasp No64 7 inches 3mm/4mm Millennium Surgical
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 18, 2020
SurGenTec. LLC. Andrew Shoup, COO 911 Clint Moore Rd Boca Raton, Florida 33847
Re: K200431
Trade/Device Name: 3D GraftRasp System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, HTR Dated: February 21, 2020 Received: February 21, 2020
Dear Andrew Shoup:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200431
Device Name 3D GraftRasp System
Indications for Use (Describe)
The 3D GraftRasp System is intended to be used in orthopedic procedures to rasp or decorticate bone and for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5.0 510(k) Summary
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the 3D GraftRasp System is provided below.
## 5.1 Submitter Information:
| Submitter: | SurGenTec, LLC<br>911 Clint Moore Rd<br>Boca Raton, FL 33487<br>Telephone: (561) 990-7882 |
|----------------------------|-------------------------------------------------------------------------------------------|
| Contact: | Andrew Shoup<br>COO<br>Phone 561-990-7882<br>Email: ashoup@surgentec.com |
| Date Prepared: | May 15, 2020 |
| 5.2 Name of Device | |
| Device Proprietary Name: | 3D GraftRasp System |
| Device Common Name: | Piston Syringe and Bone Rasp |
| Classification Regulation: | 21 CFR 880.5860 Class II |
| Classification name: | Piston Syringe |
| Panel: | General & Plastic Surgery |
| Product Code: | FMF, HTR |
#### Legally Marketed Predicate Device 5.3
Predicate(s): Graftgun Universal Graft Delivery System Surgentec, LLC. Class II Device (K170675)
> Bone Rasp No64 7 inches 3mm/4mm Millennium Surgical Class I Exempt
## 5.4 Device Description
The 3D GraftRasp system contains various decortication rasps, graft pushers, a bone funnel, and dilators. The 3D GraftRasp system is compatible with the Graftgun Universal Graft Delivery System for graft delivery to the surgical site. The decortication rasps can be used to rasp or decorticate bone. After rasping, a 5cc or 7.5cc graft tube can be inserted into the rasp lumen to extrude graft from the tip of the rasp. A set of optional sequential dilators and rasp pushers may also be provided to
{4}------------------------------------------------
aid in the use of the decortication rasp. A bone funnel may optionally be used to extrude graft through the rasps instead of the graftgun.
#### Indication for Use ર્સ રાજ્યના સાથે જિલ્લાના પાકની ખેતી કરવામાં આવે છે. આ ગામનાં મુખ્યત્વે ખેત-ઉત્પત્તમજૂરી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં મુખ્યત્વે આવેલું એક ગામના લોક
The 3D GraftRasp System is intended to be used in orthopedic procedures to rasp or decorticate bone and for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site.
| | 5.6 Technological Characteristics and Substantial Equivalence | | |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Substantial<br>Equivalence<br>Topic | 3D GraftRasp System | Graftgun Universal<br>Graft Delivery System | Millennium Surgical<br>Bone Rasp |
| 510(k) | K200431 | K170675 | N/A – Class I Device |
| Regulation<br>Description | 21 CFR 880.5860 | 21 CFR 880.5860 | 21 CFR 878.4800 |
| Device Name | 3D GraftRasp System | Graftgun Universal Graft<br>Delivery System | Bone Rasp No64 7 inches<br>3mm/4mm |
| Product Code | FMF, HTR | FMF | HTR |
| Classification | Class II | Class II | Class I |
| Indications for<br>Use | The 3D GraftRasp System is<br>intended to be used in<br>orthopedic procedures to rasp<br>or decorticate bone and for<br>the delivery of hydrated<br>allograft, autograft, or<br>synthetic bone graft material<br>to an orthopedic surgical site. | The Graftgun Universal<br>Graft Delivery System is<br>intended to be used for the<br>delivery of hydrated<br>allograft, autograft, or<br>synthetic bone graft<br>material to an orthopedic<br>surgical site. | Millennium Surgical<br>Orthopedic Rasps are<br>coarse files used in<br>orthopedic procedures to<br>sculpt bone. |
| Single Use | Yes / No* | Yes | No |
| Sterility | Steam Sterilization | Gamma irradiation to a<br>SAL of 10-6 | Steam Sterilization |
| Patient Contact<br>Material | Medical Grade:<br>Stainless Steel –<br>316L, 304 per<br>ASTM F899-12b,<br>passivated per<br>ASTM A967-13Nitinol Superelastic<br>per ASTM F2063,<br>passivated per<br>ASTM A967-13 | Medical Grade:<br>PolycarbonatePolypropyleneStainless Steel –<br>316L, 316F,<br>304H, 304HC per<br>ASTM F899-12b,<br>passivated per<br>ASTM A967-13<br>ABS | Medical Grade:<br>Surgical Grade<br>Stainless Steel –<br>420 per ASTM<br>F899-12b<br>passivated per<br>ASTM A967-13 |
| Packaging | Packaged in sterilization<br>tray, sterilized and cleaned<br>prior to each use via<br>autoclave.<br>Replacement parts shipped in<br>nonsterile packaging<br>including labeling, IFU, and<br>cleaning instructions. Need<br>to be assembled, cleaned and<br>sterilized prior to use. | Device tray containing the<br>required components to<br>action the device | Packaged in sterilization<br>tray, sterilized and cleaned<br>prior to each use via<br>autoclave. |
| Operating<br>Principle | The user can dilate tissue for<br>easier access to the surgical<br>site. The decortication rasp is<br>then used to remove cortical<br>bone by manually scraping<br>the bony anatomy. Once the<br>bony anatomy is prepared.<br>graft is applied using the<br>Graftgun Universal Graft<br>Delivery Device. It's inserted<br>into the rasp and delivers<br>graft through the rasp lumen<br>directly to the prepared<br>surgical site. The shape of<br>the rasp allows to access to<br>adjacent vertebral bodies or<br>other bony anatomy during<br>decortication and graft<br>delivery. | The user can apply a wide<br>variety of autograft,<br>allograft, and synthetic<br>bone grafts directly to a<br>surgical site using the<br>GraftGun using a<br>percutaneous approach.<br>The device comprises a<br>trigger which is manually<br>squeezed to dispense bone<br>graft. | The device is manually<br>maneuvered and scraped<br>across the bone to shape<br>and decorticate bony<br>anatomy. |
| Components | Orthopedic Rasp, Dilators,<br>Pushers, Graft Funnel | Graftgun, Graft Tubes,<br>Loading Device, Funnel<br>and Pusher, Plunger Rods | Orthopedic Rasp |
Technological Characteristics and Substantial Equivalence 5 6
{5}------------------------------------------------
SurGenTec. LLC
*Some components of the GraftRasp System are designed to be reusable while others are single use such as the replaceable rasp teeth and the flexible tip of the rasp pusher.
## 5.7 Performance Data
The following non-clinical performance data were provided to demonstrate substantial equivalence of the subject device to the predicate.
- Biocompatibility per ISO 10993-1:2018 ●
- Sterilization validation per ISO 17665-1:2006/(R) 2013 ●
- Bench Testing ●
- o Functionality Common Material Test
- Various graft materials were tested to ensure the Graftgun can ■ successfully extrude them through the various rasps of the 3D GraftRasp System. For each rasp, the Graftgun was able to successfully extrude the graft materials using an acceptable hand force.
- o Rasp Functionality Test
- The rasping surface of the rasp was tested on a sample bone ■ block to ensure the worst case scenario rasp of the 3D GraftRasp System is capable of bone decortication without failure. The rasp was able to successfully decorticate a portion of the bone block without failure of either the teeth of the rasping surface or of the main rasp body.
- Volume Dispensing Verification Test o
- The Graftgun Universal Graft Delivery system was tested to ensure an accurate amount of bone graft was dispensed with every squeeze of the trigger. It was justified that using the Graftgun along with either rasp and the accompanying graft pusher of the 3D GraftRasp System was adequate to maintain
{6}------------------------------------------------
SurGenTec, LLC
the acceptance criteria of accuracy.
# 5.8 Conclusion
The design characteristics of the 3D GraftRasp System do not raise different questions of safety and effectiveness. Non-clinical study data supports that the device is safe and effective. This data supports that the 3D GraftRasp System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.