K200428 · Nvision Biomedical Technologies, Inc. · MBI · Nov 10, 2020 · Orthopedic
Device Facts
Record ID
K200428
Device Name
Multi-Drive Interference Screw System
Applicant
Nvision Biomedical Technologies, Inc.
Product Code
MBI · Orthopedic
Decision Date
Nov 10, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Multi-Drive Interference Screw System is intended for soft tissue reattachment, i.e. fixation of ligament and tendon graft tissue and tendon transfers in surgeries of the shoulder, elbow, knee, foot/ankle, and hand /wrist. Specifically: Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Sepair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus Transfer for Achilles Tendon Reconstruction, and Flexor Digitorum Longus Transfer for Posterior Tibial Tendon Reconstruction Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction Hand/Wrist: Scapholunate Ligament Reconstruction, Unar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs. Ligament Reconstruction and Tendon Interposition
Device Story
Multi-Drive Interference Screw System consists of threaded metallic or HA-enhanced PEEK screws used for soft tissue reattachment and fixation in orthopedic surgeries. Screws feature a cannulated design and a multi-drive mating interface for insertion/removal. HA-enhanced PEEK versions include tantalum pins for radiographic visibility. Used by surgeons in clinical settings to secure ligament and tendon grafts to bone. Device provides mechanical fixation to facilitate tissue healing and reconstruction. Multi-drive interface offers surgeons flexibility in tool compatibility during implantation.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F543 (insertion, torque, static axial pullout). Pyrogenicity testing performed per ANSI/AAMI ST72:2011/R2016, confirming bacterial endotoxins < 20 EU/device.
Technological Characteristics
Materials: HA-enhanced PEEK, titanium alloy. Features: Threaded shank, cannulation hole, multi-drive mating interface, tantalum pins (in PEEK versions). Mechanical testing per ASTM F543. Sterilization: Not specified.
Indications for Use
Indicated for soft tissue reattachment, including ligament/tendon graft fixation and tendon transfers in shoulder, elbow, knee, foot/ankle, and hand/wrist surgeries.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Responsive Arthroscopy's Interference Screw System (K180573)
Parcus Medical's SLIK Fix Interference Screw (K170877)
Reference Devices
Nvision Biomedical's Vector Hammertoe Correction System (K183055)
Nvision Biomedical's Trigon™ Ti Stand-Alone Wedge Fixation System (K192645)
Nvision Biomedical's Healix Compression Screw System (K182949)
Submission Summary (Full Text)
{0}------------------------------------------------
November 10, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it.
Nvision Biomedical Technologies, Inc % Jeffrey Brittan Vice President of Product Realization Watershed Idea Foundry 1815 Aston Avenue, Suite 106 Carlsbad, California 92008
## Re: K200428
Trade/Device Name: Multi-Drive Interference Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: October 14, 2020 Received: October 15, 2020
## Dear Mr. Brittan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K200428
Device Name Multi-Drive Interference Screw System
#### Indications for Use (Describe)
The Multi-Drive Interference Screw System is intended for soft tissue reattachment, i.e. fixation of ligament and tendon graft tissue and tendon transfers in surgeries of the shoulder, elbow, knee, foot/ankle, and hand /wrist. Specifically:
#### Shoulder:
Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Sepair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
#### Foot/Ankle:
Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus Transfer for Achilles Tendon Reconstruction, and Flexor Digitorum Longus Transfer for Posterior Tibial Tendon Reconstruction
#### Knee:
Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis
#### Elbow:
Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
#### Hand/Wrist:
Scapholunate Ligament Reconstruction, Unar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs. Ligament Reconstruction and Tendon Interposition
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Nvision Biomedical Technologies. The logo features a stylized letter N with a series of short, parallel lines extending from the left side of the N. The text "VISION" is written in a bold, sans-serif font to the right of the N. Below the main logo, the words "biomedical technologies" are written in a smaller, elegant font.
510(k) Number: K200428 Date Received: 02/24/2020
K200428 Page 1 of 3
# 510(k) Summary
## DATE PREPARED
October 22, 2020
## MANUFACTURER AND 510(k) OWNER
Nvision Biomedical Technologies, Inc. 4590 Lockhill Selma San Antonio, TX 78249, USA Telephone: (210) 545-3713 Fax: (866) 764-1139 Official Contact: Diana Langham, Director of Regulatory and Corporate Compliance
## REPRESENTATIVE/CONSULTANT
Jeffrey Brittan Vice President of Product Realization Watershed Ideas Foundry Telephone: (714) 287-6780 Email: jeffbrittan@watershedideas.com
## PROPRIETARY NAME OF SUBJECT DEVICE
Multi-Drive Interference Screw System
### COMMON NAME Interference Screw (fastener, fixation, nondegradable, soft tissue)
## DEVICE CLASSIFICATION
Smooth or threaded metallic bone fixation fastener (Classification Regulations: 21 CFR 888.3040, Product Code: MBI, Class: II)
## PREMARKET REVIEW
Orthopedic Devices
## INDICATIONS FOR USE
The Multi-Drive Interference Screw System is intended for soft tissue reattachment, i.e. fixation of ligament and tendon graft tissue and tendon transfers in surgeries of the shoulder, elbow, knee, foot/ankle, and hand /wrist. Specifically:
#### Shoulder:
Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
## Foot/Ankle:
Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Nvision biomedical technologies. The logo features a stylized letter N with a series of horizontal lines emanating from it. The word "vision" is written in a modern, sans-serif font to the right of the N. Below the word "vision" is the phrase "biomedical technologies" in a smaller, serif font.
Transfer for Achilles Tendon Reconstruction, and Flexor Digitorum Longus Transfer for Posterior Tibial Tendon Reconstruction
### Knee:
Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis
### Elbow:
Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction
### Hand/Wrist:
Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs. Ligament Reconstruction and Tendon Interposition
## DEVICE DESCRIPTION
The Multi-Drive Interference Screw System is a family of interference screws for the reattachment and fixation of tissue in surgeries of the shoulder, elbow, knee, foot/ankle, and hand /wrist. The interference screws are manufactured from HA Enhanced PEEK or titanium alloy in multiple lengths and diameters with key features including a threaded shank, cannulation hole, and multiple drive mating interface. The HA PEEK versions also incorporate tantalum pins for imaging visibility.
## PREDICATE DEVICE IDENTIFICATION
The Multi-Drive Interference Screw System is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name & Manufacturer | Primary<br>Predicate |
|---------------|----------------------------------------------------|----------------------|
| K183690 | Paragon 28's Tenodesis Screw System | ✓ |
| K180573 | Responsive Arthroscopy's Interference Screw System | |
| K170877 | Parcus Medical's SLIK Fix Interference Screw | |
The following reference devices are also cited in this submission:
- Nvision Biomedical's Vector Hammertoe Correction System (K183055)
- Nvision Biomedical's Trigon™ Ti Stand-Alone Wedge Fixation System (K192645) ●
- Nvision Biomedical's Healix Compression Screw System (K182949) ●
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for NVision biomedical technologies. The logo has a stylized "N" with horizontal lines in gold and black. The word "VISION" is in a sans-serif font in black, and the words "biomedical technologies" are in a smaller font in gold below the main logo.
## SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the Multi-Drive Interference Screw System. The following testing was performed:
- . Mechanical testing in accordance with ASTM F543 including insertion testing, insertion/removal torque, and static axial pullout
- Engineering analysis comparison of mechanical strength
- . Pyrogenicity testing (LAL method), with results demonstrating bacterial endotoxins less than 20 EU/device in accordance with ANSI/AAMI ST72:2011/R2016
The results of these tests indicate that the Multi-Drive Interference Screw System is substantially equivalent to the predicate devices.
## EQUIVALENCE TO PREDICATE DEVICES
Nvision believes that the Multi-Drive Interference Screw System is substantially equivalent to the identified predicate devices. The subject device has similar dimensions, and uses similar or identical materials. The subject device has the same indications for use and intended use, as well as similar technological characteristics (threaded screws in a range of diameters and lengths with a mating interface and central cannulation hole). The subject screw incorporates a multi-drive mating interface while the predicates utilize a single drive, and although this multi-drive offers additional options for insertion/removal, it utilizes the equivalent operating principle and function. These technological characteristics have undergone testing/analysis to ensure the subject device is equivalent to the predicates.
## CONCLUSION
Based on the testing performed, including engineering analysis of mechanical strength, it can be concluded that the subject device does not raise new issues of safety or efficacy compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed Multi-Drive Interference Screw System are assessed to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.