K200425 · Medsafety Solutions, LLC · FPA · Oct 29, 2020 · General Hospital
Device Facts
Record ID
K200425
Device Name
Female Luer Lock Cap
Applicant
Medsafety Solutions, LLC
Product Code
FPA · General Hospital
Decision Date
Oct 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Female Luer Lock Cap is indicated for use as a cap for male Luer fittings on medical devices such as manifolds, stopcocks or IV sets.
Device Story
The Female Luer Lock Cap is a sterile, single-use, disposable injection-molded polyethylene component designed to cover male Luer fittings on medical devices like manifolds, stopcocks, and IV sets. It functions by screwing onto the threads of a male Luer fitting to provide a secure closure. Used primarily in clinical settings and pharmacy compounding by pharmacists, technicians, and healthcare professionals, the device serves to prevent touch contamination and medication leakage. It is provided terminally gamma-sterilized. The device does not involve electronic processing or software.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), sterility (ISO 11137), package integrity (ASTM F1929, ASTM F88), and functional performance testing per ISO 594 and ISO 80369-7.
Technological Characteristics
Material: Polyethylene. Sensing/Actuation: Mechanical Luer lock interface. Form factor: Single-use cap. Sterilization: Gamma radiation (SAL 10^-6). Standards: ISO 594 (conical fittings), ISO 80369-7 (small-bore connectors), ISO 10993 (biocompatibility), ANSI/AAMI/ISO 11137 (sterilization), ASTM F1929/F88 (packaging).
Indications for Use
Indicated for use as a cap for male Luer fittings on medical devices (e.g., manifolds, stopcocks, IV sets) by pharmacy technicians, pharmacists, and healthcare professionals to reduce touch contamination and leakage.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 29, 2020
MedSafety Solutions LLC % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K200425
Trade/Device Name: Female Luer Lock Cap Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: FPA Dated: September 14, 2020 Received: September 16, 2020
# Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for
combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Payal Patel for Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200425
Device Name Female Luer Lock Cap
Indications for Use (Describe)
The Female Luer Lock Cap is indicated for use as a cap for male Luer fittings on medical devices such as manifolds, stopcocks or IV sets.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 5: 510(K) SUMMARY
#### 510(K) SUMMARY: K200425
#### DATE PREPARED: 10/19/2020
#### OWNER:
MedSafety Solutions LLC 7012 S. Revere Pkwy, Suite #120 Centennial, Colorado 80112
#### CONTACT PERSON:
Richard Weiskopf Sr. Director Quality & Regulatory Phone: 720-500-2500 ext. 210 Fax: 720-500-2571 Email: rich@medsafetysolutions.com
# IDENTIFICATION OF THE DEVICE:
| Trade Name: | Female Luer Lock Cap |
|--------------------------|-----------------------|
| Common Name: | Female Luer Lock Cap |
| Classification Name: | IV Administration Set |
| Classification Panel: | 80 General Hospital |
| Classification: | 21 CRF 880.5440 |
| Class: | Class II |
| Product Code: | FPA |
| Subject Device Model No: | 75309 |
# PREDICATE DEVICE(S):
| Device: | Dual Luer Lock Cap |
|-----------------|-------------------------------|
| Company: | Baxter Healthcare Corporation |
| 510(k) Number: | K101385 |
| Clearance Date: | June 22, 2010 |
# DESCRIPTION OF DEVICE:
The Female Luer Lock Cap is an individually packaged, sterile, single use disposable product used to cover male Luer fittings. Luer fittings are commonly used on IV administration products and accessories to provide universal compatibility and are well established in the market. Luer lock fittings are securely joined by means of a tabbed hub on the female fitting which screws into threads in a sleeve on the male fitting. The Female Luer Lock Cap is an injection molded, polyethylene component. Polyethylene material is durable, biologically inert, gamma stable, and used extensively in the medical device industry. The product is terminally gamma sterilized to a Sterility Assurance Level (SAL) of 10-6.
The Female Luer Lock Cap is not associated with any specific medical condition. The user group consists of those involved in pharmacy compounding, mainly pharmacy
{4}------------------------------------------------
technicians and pharmacists, as well as health care professionals. Covering Luer fittings helps reduce the risk of touch contamination and medication leakage.
The basis for the premarket notification is a new device that has not been marketed previously by MedSafety Solutions. The Indications for Use, the basic design and function for the subject device are equivalent to the legally marketed predicate device.
# INDICATIONS FOR USE:
The Female Luer Lock Cap is indicated for use as a cap for male Luer fittings on medical devices such as manifolds, stopcocks or IV sets.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
The proposed device is substantially equivalent to Baxter Healthcare's current legally marketed device cleared under 510(k) premarket notification K101385 cleared June 22, 2010 given that the subject device has the same Indications for Use, basic design, and function as the predicate device.
The primary difference between the predicate device and the subject device is that the predicate has both male and female Luer connections whereas the subject device only has a male Luer lock connection.
The change in material from the predicate device from Polypropylene to Polyethylene does not raise new questions of safety or efficacy. Polypropylene and Polyethylene are both used extensively for medical products. Polyethylene was selected for the subject device because of its low moisture absorption, is biologically inert, gamma stable, is mechanically stable and the material's sustainability (i.e. recyclable). Performance testing of the subject device for the design and functional performance and biocompatibility assessment have demonstrated the subject device is substantially equivalent to the predicate device.
| Feature /<br>Specification | Predicate Device<br>Dual Luer Lock<br>Cap<br>K101385 | Subject Device<br>Female Luer Lock<br>Cap | Comparison |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>Classification | Class II -FPA<br>IV Administration Set | Class II -FPA<br>IV Administration Set | Identical |
| Indications for<br>Use | The Dual Luer Lock<br>Cap is indicated for<br>use as a cap for<br>male or female Luer<br>ports on medical<br>devices such as<br>manifolds, stopcocks<br>or sets. | The Female Luer<br>Lock Cap is<br>indicated for use as<br>a cap for male Luer<br>fittings on medical<br>devices such as<br>manifolds, stopcocks<br>or IV sets. | Substantially<br>Equivalent -<br>difference is the<br>predicate has both<br>male and female<br>Luer connections<br>whereas the subject<br>device only has a |
| Feature /<br>Specification | Predicate Device<br>Dual Luer Lock<br>Cap<br>K101385 | Subject Device<br>Female Luer Lock<br>Cap | Comparison |
| | | | male Luer lock<br>connection |
| Sterile | Yes | Yes | Identical |
| Sterility<br>Assurance Level | SAL of 10-6 | SAL of 10-6 | Identical |
| Single Use | Yes | Yes | Identical |
| Non-Pyrogenic | Yes | Yes | Identical |
| Materials | Injection molded,<br>Polypropylene<br>component<br>Biocompatibility<br>ISO 10993<br>• Cytotoxicity<br>• Sensitization<br>• Intracutaneous<br>Reactivity/<br>Irritation<br>• Systemic<br>Toxicity<br>• Hemo-<br>compatibility<br>• Material-<br>Mediated<br>Pyrogen<br>• Subacute<br>toxicity<br>• Subchronic<br>toxicity<br>Material shelf-life<br>ISO 594<br>• Ease of<br>assembly<br>ISO 80369-7<br>• Positive Pressure<br>Liquid Leakage<br>• Sub-atmospheric<br>Pressure Air<br>Leakage<br>• Stress Cracking<br>• Resistance to | Injection molded,<br>Polypropylene<br>component<br>Biocompatibility<br>ISO 10993<br>• Cytotoxicity<br>• Sensitization<br>• Intracutaneous<br>Reactivity/<br>Irritation<br>• Systemic<br>Toxicity<br>• Hemo-<br>compatibility<br>• Material-<br>Mediated<br>Pyrogen<br>• Subacute<br>toxicity<br>• Subchronic<br>toxicity<br>Material shelf-life<br>ISO 594<br>• Ease of<br>assembly<br>ISO 80369-7<br>• Positive Pressure<br>Liquid Leakage<br>• Sub-atmospheric<br>Pressure Air<br>Leakage<br>• Stress Cracking<br>• Resistance to | Substantially<br>Equivalent - material<br>difference does not<br>raise new questions<br>of safety or<br>effectiveness |
| Feature /<br>Specification | Predicate Device<br>Dual Luer Lock<br>Cap<br>K101385 | Subject Device<br>Female Luer Lock<br>Cap | Comparison |
| | Axial Load<br>• Resistance to<br>Separation from<br>Unscrewing<br>• Resistance to<br>Overriding | Axial Load<br>• Resistance to<br>Separation from<br>Unscrewing<br>• Resistance to<br>Overriding | |
# Table 5-1: Comparison of Technological Characteristics
{5}------------------------------------------------
{6}------------------------------------------------
# PERFORMANCE DATA:
Bench testing was performed to demonstrate that the subject device is substantially equivalent to the predicate device, and that the basic design and function of the Female Luer Lock Cap complies with International Standards that specify dimensions and requirements for the design and functional performance of small-bore connectors intended to be used for intravascular applications of medical devices and accessories.
Shelf life testing includes both simulated and real time aged product for both package integrity and product performance testing. Product and packaging underwent simulated distribution and conditioning prior to shelf life testing. Product and packaging testing are summarized in Table 5-2; testing is performed at multiple time points to support a shelf life of three years.
A product risk analysis was conducted according to ISO 14971 Medical devices -Application of risk management to medical devices and there were no new issues of safety and effectiveness.
Performance bench testing results meet acceptance criteria and verified the Female Luer Lock Cap meets applicable Industry Standard requirements for the design and functional performance, biocompatibility, sterilization and packaging, and demonstrates substantial equivalently to the predicate device.
| | Table 5-2: Performance Summary | |
|--|--------------------------------|--|
|--|--------------------------------|--|
| Performance Test | Testing Standard |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Luer Connector<br>Standard | ISO 594 Conical fittings with 6% (Luer) taper for syringes,<br>needles, and certain other medical equipment Part 1: General &<br>Part 2: Lock fittings (FDA FR 6-129)<br>• Ease of assembly |
| Performance Test | Testing Standard |
| Luer Connector<br>Standard | ISO 80369-7 Small-bore connectors for liquids and gases in<br>healthcare applications - Part 7: Connectors with 6% (Luer)<br>taper for intravascular or hypodermic applications (FDA FR 5-<br>115) |
| | • Positive Pressure Liquid Leakage |
| | • Sub-atmospheric Pressure Air Leakage |
| | • Stress Cracking |
| | • Resistance to Separation from Axial Load |
| | • Resistance to Separation from Unscrewing<br>• Resistance to Overriding |
| Biocompatibility | ISO 10993-1 Biological Evaluation of Medical Devices - Part 1:<br>Evaluation and Testing (FDA FR 2-258) |
| | • Cytotoxicity |
| | • Sensitization |
| | • Intracutaneous Reactivity / Irritation |
| | • Systemic Toxicity |
| | • Hemocompatibility |
| | • Material-Mediated Pyrogen |
| | • Bacterial Endotoxins Test<br>• Subacute Toxicity<br>• Subchronic Toxicity |
| Sterilization | ANSI/AAMI/ISO11137-1 Sterilization of health care products.<br>Radiation - Part 1: Requirements for development, validation<br>and routine control of a sterilization process for medical devices<br>(FDA FR 14-528) and ANSI/AAMI/ISO 11137-2 Sterilization of<br>health care products - Radiation - Part 2: Establishing the<br>sterilization dose (FDA FR 14-409) |
| | • VDmax15<br>• Sterility Assurance Level of 10-6 |
| Package Integrity | ASTM F1929-15 Standard Test Method for Detecting Seal Leaks<br>in Porous Medical Packaging by Dye Penetration (FDA FR 14-<br>484)<br>ASTM F88/F88M-15 Standard Test Method for Seal Strength of<br>Flexible Barrier Materials (FDA FR 14-482) |
{7}------------------------------------------------
{8}------------------------------------------------
# ASSESSMENT OF CLINICAL DATA:
There were no clinical studies conducted in support of a substantial equivalence determination.
# CONCLUSION:
Review of the bench performance test data as well as comparison of the device classification, indications for use, operating principle, technological characteristics, sterility, and biocompatibility demonstrate that the subject device, Female Luer Lock Cap is substantially equivalent to the predicate device, the Baxter Dual Luer Lock Cap (K101385). Any differences between the subject and the predicate device do not raise any issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.