← Product Code [FED](/productcode/FED) · K200419

# Kalera Vacuum Aspiration Catheter (K-VAC) (K200419)

_Kalera Medical, Inc. · FED · May 22, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200419

## Device Facts

- **Applicant:** Kalera Medical, Inc.
- **Product Code:** [FED](/productcode/FED.md)
- **Decision Date:** May 22, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Kalera Vacuum Aspiration Catheter™ (K-VAC™) is used to establish a conduit during endoscopic urological procedures for the treatment of urinary stones.

## Device Story

K-VAC is a sterile, single-use, dual-lumen aspiration-irrigation catheter designed for stone fragment removal during ureteroscopy. The device features an outer irrigation lumen and a larger inner aspiration lumen (2.5 mm ID). The distal 5 cm is steerable via a handle-mounted control lever and embedded wires, allowing 180° articulation to access renal calyces. A radiopaque introducer with a soft tapered tip is included for guidewire-compatible placement. Used by physicians in clinical settings under fluoroscopic guidance, the device is inserted through a standard 12/14F ureteral access sheath. Irrigation fluid is delivered to fluidize stones, while a standard hospital vacuum source (150-200 mm Hg) is applied to the aspiration lumen to remove debris. The steerable tip improves access to the renal pelvis and calyces compared to non-steerable sheaths, potentially increasing stone clearance efficiency and reducing procedure time.

## Clinical Evidence

No human clinical data. Evidence includes bench testing (dimensional conformance, flow, cycling, tortuous path, actuation, torque, bond testing) and a GLP porcine animal study. The animal study compared K-VAC to the predicate, showing equivalent pathology, histopathology, fluoroscopy time, and procedure time. Usability assessment during the animal study confirmed successful task completion. Biocompatibility testing (ISO 10993-1) for limited duration, tissue-contacting use passed all metrics (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity). Sterilization validated to SAL 10^-6 per ISO 11135:2014.

## Technological Characteristics

Dual-lumen catheter; outer irrigation lumen, inner aspiration lumen (2.5 mm ID). Steerable distal tip (180° articulation) via mechanical handle lever and embedded wires. Radiopaque construction. Compatible with 12/14F ureteral access sheaths and 0.035"/0.038" guidewires. Passive open-system irrigation. Sterile, single-use, EtO sterilized (ISO 11135:2014). Biocompatible materials per ISO 10993-1. Mechanical device; no electrical components or software.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- ClearPetra Suction-Evacuation Sheath ([K161110](/device/K161110.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 22, 2020

Kalera Medical, Inc. % Diane Horwitz, Ph.D., RAC Regulatory Consultant Mandell Horwitz Consultants, LLC 5 Lake Como Ct. Greenville, SC 29609

Re: K200419

> Trade/Device Name: Kalera Vacuum Aspiration Catheter™ (K-VACTM) Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FED Dated: April 27, 2020 Received: April 28, 2020

Dear Diane Horwitz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Martha W. Betz, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K200419

Device Name

Kalera Vacuum Aspiration Catheter™ (K-VACTM)

Indications for Use (Describe)

The Kalera Vacuum Aspiration Catheter (K-VAC™) is used to establish a conduit during endoscopic urological procedures for the treatment of urinary stones.

| Type of Use (Select one or both, as applicable)                                               |                                                                                              |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">Over-The-Counter Use</span> (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

#### 1. GENERAL INFORMATION

#### Submitter and 510(k) Owner 1.1

Kalera Medical, Inc. 4473 Willow Road, Suite 100 Pleasanton, CA 94588

#### 1.2 Official Correspondent

Diane Horwitz, Ph.D. Regulatory Consultant 5 Lake Como Ct. Greenville, SC 29609 Telephone: 703.307.2921 Email: dmh@mandellhorwitzconsulting.com

#### Date of Preparation 1.3

May 21, 2020

#### NAME OF THE DEVICE 2.

#### 2.1.1 Trade/Proprietary Name

Kalera Vacuum Aspiration Catheter™ (K-VACTM)

#### 2.1.2 Common/Usual Name

Endoscopic Access Overtube, Gastroenterology-Urology

#### 2.1.3 Classification Information

| Classification Name:       | Endoscope and Accessories                                 |
|----------------------------|-----------------------------------------------------------|
| Classification Regulation: | 21 CFR 876.1500                                           |
| Class:                     | II                                                        |
| Product Code:              | FED, Endoscopic Access Overtube, Gastroenterology-Urology |
| Panel:                     | Gastroenterology / Urology                                |

#### 3. PREDICATE DEVICE

The predicate device is ClearPetra Suction-Evacuation Sheath (Well Lead Medical Co., Ltd., K161110).

{4}------------------------------------------------

#### 4. DESCRIPTION OF THE DEVICE

K-VAC is a sterile, single-use aspiration-irrigation catheter that is designed to assist in the removal of stone fragments during standard ureteroscopy. K-VAC is a dual-lumen catheter with two concentric lumens. The outer lumen is the irrigation lumen that accommodates delivery of irrigation/contrast solution. The larger, inner lumen is the aspiration lumen through which the stone fragments are removed. The inner diameter of the aspiration lumen is 2.5 mm. The distal 5 cm of the K-VAC tip is soft with a steering mechanism on the catheter handle and embedded wires to actuate the tip. In this way, stones and debris in the renal calyces can be fluidized with irrigation and aspirated.

An additional component to the device is a radiopaque introducer, which is 0.035" and 0.038" guidewire compatible. The tip of the introducer has a soft tapered tip.

The design of K-VAC allows the trained operator to maneuver and position K-VAC via a standard 12/14F ureteral access sheath under fluoroscopic guidance, followed by controlled irrigation and aspiration of stones and debris from the renal pelvis and calyces.

K-VAC is composed of well-characterized biocompatible materials commonly used in urinary catheters. The catheter is mechanical and contains no electrical components.

#### ട. INTENDED USE

The intended use / indications for use for the Kalera Vacuum Aspiration Catheter (K-VAC) as follows:

The Kalera Vacuum Aspiration Catheter™ (K-VAC™) is used to establish a conduit during endoscopic urological procedures for the treatment of urinary stones.

#### INTENDED USE COMPARED TO THE PREDICATE 6.

The intended use of the Kalera Vacuum Aspiration Catheter (K-VAC) is the same as the predicate device. Both devices are used to establish a conduit for irrigation and aspiration of kidney stones during standard ureteroscopy, are composed of biocompatible materials, and are sterile single-use devices.

#### TECHNOLOGY CHARACTERISTICS COMPARED TO THE PREDICATE 7.

There are similarities and differences between the technological characteristics of K-VAC and the predicate device. The devices are similar in their use and purpose, as both devices are used to introduce irrigation to fluidize kidney stones and both devices remove stones using aspiration. A few features of the devices are different. Table 1 on the following page highlights the similarities and differences.

{5}------------------------------------------------

|                                             | Subject Device<br>K-VAC<br>(Kalera Medical Inc.)                                                                                                                                                                                | Predicate<br>ClearPetra K161110                                                                                                                                                    | Similar / Different<br>Impact Safety and<br>Effectiveness                                                                                                                                                                                                                                                                                                                       |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Models and<br>Sizes                         | One model, two sizes<br>SKU KVAC60: 12F OD, 60<br>cm<br>SKU KVAC70: 12F OD, 70<br>cm                                                                                                                                            | ClearPetra Ureteral Access<br>Sheath with 30 SKUs, ranging<br>from 10F to 14F ID, varying<br>lengths                                                                               | Similar sizes, no impact on<br>safety or effectiveness                                                                                                                                                                                                                                                                                                                          |
| Design                                      | Dual lumen catheter with<br>Aspiration Connector,<br>Irrigation Port and articulation<br>Control Lever. Radiopaque<br>introducer with a soft, tapered<br>tip. Inserted with introducer<br>over a guidewire under<br>fluoroscopy | Single lumen access sheath<br>with aspiration connector.<br>Irrigation via ureteroscope<br>Radiopaque introducer<br>Inserted with introducer over a<br>guidewire under fluoroscopy | Both devices are similar in the<br>use of luer adapters and<br>connectors, guidewires and<br>introducers, irrigation and<br>aspiration for stone removal<br>and use of fluoroscopy<br>The devices are different in<br>their steerability<br>This difference does not raise<br>any new issues of safety or<br>effectiveness as shown through<br>product testing                  |
| Lumen<br>Dimension(s)                       | Dual lumen device, compatible<br>with 12/14F standard ureteral<br>access sheaths<br>Outer lumen diameter (OD): 12F (4 mm) Inner lumen diameter (ID): 7.5F (2.5 mm)                                                              | Single lumen device<br>Sizes ranging from 10F (3<br>mm) to 14F (4.7 mm)                                                                                                            | Same: K-VAC within size<br>range of predicate                                                                                                                                                                                                                                                                                                                                   |
| Radiopacity                                 | K-VAC and introducer are<br>radiopaque                                                                                                                                                                                          | ClearPetra and introducer are<br>radiopaque                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                            |
| Irrigation                                  | Via the Irrigation Port, a luer<br>fitting, using hospital supplies<br>Passive open system design<br>allowing continuous outflow<br>of irrigation fluid                                                                         | Via the irrigation port of the<br>ureteroscope using hospital<br>supplies<br>Passive open system design<br>allowing continuous outflow<br>of irrigation fluid                      | Similar method of introducing<br>irrigation with the same goal<br>of fluidizing the kidney stones<br>and fragments<br>Same passive open system<br>design                                                                                                                                                                                                                        |
| Aspiration                                  | May be used with standard<br>hospital vacuum sources at 150<br>to 200 mm Hg                                                                                                                                                     | May be used with standard<br>hospital vacuum sources at 150<br>to 200 mm Hg                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                            |
| Steerability<br>via actuating<br>distal end | Control Lever can actuate the<br>distal 5 cm of the catheter 180°<br>to each side and can be rotated<br>to access areas of the renal<br>pelvis and calyces                                                                      | Not steerable.<br>The ureteroscope inside the<br>sheath is steerable and actuates<br>similarly to the subject device                                                               | Differences between K-VAC<br>and the predicate devices<br>include:<br>Steerable tip A handle lever that<br>controls the actuation of<br>the tip Differences between subject<br>and predicate device were<br>shown not to impact safety or<br>effectiveness of the device.<br>Evidence comes from bench<br>design verification and<br>validation studies and GLP<br>animal study |

{6}------------------------------------------------

#### 8. PERFORMANCE TESTING

The 510(k) submission provided performance data to establish the substantial equivalence of K-VAC (the subject device) to the predicate device. A summary of these performance tests is provided below.

Performance Test Bench (Laboratory Testing): Bench performance testing was performed to evaluate dimensional conformance, flow, product cycling, tortuous path testing, actuation testing, torque, and bond testing. Results of verification and validation testing confirm that the aspiration and irrigation and other device characteristics meet product specifications and the device meets design input requirements.

Performance Testing Animal: Testing with K-VAC and the predicate device in a GLP Animal Study evaluated device performance under fluoroscopy, irrigation and aspiration using a porcine model. Equivalent pathology and histopathology findings were observed between K-VAC and the predicate device; some observations in both groups were a result of the difficulty in establishing access in a tortuous porcine renal model and were not attributed to either device. Average overall fluoroscopy time and procedure time were substantially equivalent between K-VAC and the predicate device.

Usability Assessment: A usability assessment was conducted as a part of the GLP Animal Study, and confirmed that all participants were successful in all tasks during use of K-VAC.

Biocompatibility Data: Biocompatibility testing was conducted for K-VAC as an external communicating device, tissue/bone/dentin contacting, limited duration exposure. Cytotoxicity (MEM Elution Method), Sensitization (Guinea Pig Maximization Test, Magnusson-Kligman Method), Irritation (Intracutaneous test), Acute Systemic Toxicity, Material Mediated Pyrogen Test (LAL Test) testing all resulted in passing test results. This leads to the conclusion that K-VAC is biocompatible and meets the requirements of ISO 10993-1.

Sterilization: Sterilization validation studies using ethylene oxide (EtO) confirmed sterilization to a sterility assurance level (SAL) of 106. The sterilization process was validated in accordance with ANSI AAMI ISO 11135: 2014, Over-Kill Approach.

Packaging and Shelf Life: Packaging validation testing, including transportation and distribution testing demonstrates the soundness of the product packaging to protect the device and maintain product sterility, integrity and functionality and supports a 12-month shelf life for K-VAC.

#### 9. CONCLUSIONS

This 510(k) submission demonstrates that K-VAC is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K200419](https://fda-staging.innolitics.com/device/K200419)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).