VitaVitro® Fertilization Medium is intended for preparation and handling of human gametes and for in vitro fertilization. VitaVitro® Gamete Buffer Medium is intended for human gamete and embryo short-term handling procedures outside the incubator, including washing and intracytoplasmic sperm injection (ICSI). VitaVitro® Flushing Buffer Medium is intended for use during ovarian follicle flushing and oocyte collection procedures for use in in vitro fertilization procedures.
Device Story
VitaVitro media products (Fertilization Medium, Gamete Buffer Medium, Flushing Buffer Medium) are specialized solutions for assisted reproductive technology (ART). Fertilization Medium supports gamete preparation and fertilization; Gamete Buffer Medium facilitates short-term handling outside incubators (e.g., washing, ICSI); Flushing Buffer Medium assists in oocyte retrieval during follicle flushing. Used in clinical IVF laboratories by embryologists and clinicians. Products are aseptically filled in PETG bottles, single-use, and stored at 2-8°C. These media provide the necessary chemical environment (pH, osmolality, nutrients) to maintain gamete/embryo viability during laboratory manipulation, directly supporting successful fertilization and embryo development outcomes.
Clinical Evidence
Bench testing only. Performance validated via pH testing (7.2-7.6), osmolality (260-290 mOsm/kg), endotoxin (<0.25 EU/ml), and Mouse Embryo Assay (MEA) showing ≥80% expanded blastocyst development. Sterility confirmed per USP <71>. Biocompatibility for Flushing Buffer Medium (cytotoxicity, sensitization, irritation) performed per ISO 10993. Shelf-life validated via accelerated aging.
Technological Characteristics
Aqueous-based reproductive media containing glucose, salts, amino acids, taurine, EDTA, phenol red, gentamicin sulfate, and HSA. Packaged in PETG bottles. Aseptically processed and sterile-filtered. pH 7.2-7.6; osmolality 260-290 mOsm/kg. Biocompatibility testing per ISO 10993-1, 10993-5, and 10993-10. Transportation testing per ASTM D4169-2016. Sterilization validation per ISO 13408-1 and 13408-2.
Indications for Use
Indicated for use in human assisted reproduction procedures, specifically for the preparation, handling, and washing of gametes and embryos, and for oocyte collection and follicle flushing.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of human figures, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue font.
October 16, 2020
Shenzhen VitaVitro Biotech Co., Ltd. Donghai Pan International Regulatory Specialist R601, Building B, Hai Ke Xing Tech Park, Baoshan Road No. 16 Shenzhen, Guangdong 518118 China
Re: K200408 Trade/Device Name: VitaVitro® Fertilization Medium, VitaVitro® Gamete Buffer Medium, VitaVitro® Flushing Buffer Medium Regulation Number: 21 CFR § 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MQL Dated: September 4, 2020 Received: September 14, 2020
Dear Donghai Pan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
# 510(k) Number (if known) K200408
# Device Name
VitaVitro® Fertilization Medium, VitaVitro® Gamete Buffer Medium, VitaVitro® Flushing Buffer Medium
# Indications for Use (Describe)
VitaVitro® Fertilization Medium is intended for preparation and handling of human gametes and for in vitro fertilization.
VitaVitro® Gamete Buffer Medium is intended for human gamete and embryo short-term handling procedures outside the incubator, including washing and intracytoplasmic sperm injection (ICSI).
VitaVitro® Flushing Buffer Medium is intended for use during ovarian follicle flushing and oocyte collection procedures for use in in vitro fertilization procedures.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div><span style="font-family: Arial, sans-serif;"> <span style="text-decoration: underline;"> <svg height="15" width="15"> <rect height="15" style="fill:lightgrey;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> <path d="M 2 2 L 13 13 M 13 2 L 2 13" stroke="black" stroke-width="2"></path> </svg> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-family: Arial, sans-serif;"> <span style="text-decoration: underline;"> <svg height="15" width="15"> <rect height="15" style="fill:lightgrey;stroke-width:1;stroke:rgb(0,0,0)" width="15"></rect> </svg> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | |
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K200408
#### 1. Submission Sponsor
Shenzhen VitaVitro Biotech Co., Ltd. R601, Building B, Hai Ke Xing Tech Park, Baoshan Road No.16 Shenzhen, Guangdong, 518118, P. R. China Contract Person: Jenny Lin Title: General Manager Tel: 86-755-84511813 Fax: 86-755-85235226 Email: jenny@vitavitro.com
#### Date Prepared 2.
October 15, 2020
### Device Identification 3.
| Trade/Device Name: | VitaVitro® Fertilization Medium, VitaVitro® Gamete Buffer Medium, VitaVitro®<br>Flushing Buffer Medium |
|------------------------|--------------------------------------------------------------------------------------------------------|
| Common Name: | Assisted Reproduction Medium |
| Regulation Number: | 21 CFR 884.6180 |
| Regulation Name: | Reproductive Media and Supplements |
| Device Classification: | Class II |
| Product Code: | MQL (Media, Reproductive) |
#### 4. Predicate Device
| Subject Device | VitaVitro® Fertilization Medium | VitaVitro® Gamete Buffer Medium | VitaVitro® Flushing Buffer Medium | |
|---------------------|--------------------------------------------|----------------------------------|---------------------------------------|--------------------------------------|
| Name | Global® Total® for<br>Fertilization w/ HSA | Quinn's HTF Medium<br>with HEPES | Sydney IVF Follicle Flush<br>Buffer | |
| Predicate<br>Device | Manufacturer | Lifeglobal Group, LLC | Advanced Reproductive<br>Technologies | William A. Cook<br>Australia Pty Ltd |
| | 510(k) No. | K142991 | K991395 | K153290 |
The predicate devices have not been subjected to any design related recalls.
### 5. DEVICE DESCRIPTION
This submission includes three media products:
- . VitaVitro® Fertilization Medium is intended for preparation and handling of human gametes and for in vitro fertilization. Media provided in volumes of 30 ml and 60 ml.
- . VitaVitro® Gamete Buffer Medium intended for human gamete and embryo short-term handling procedures outside the incubator, including washing and intracytoplasmic sperm injection (ICSI). Media provided in a
{4}------------------------------------------------
single volume of 60 ml.
- . VitaVitro® Flushing Buffer Medium intended for use during in vitro fertilization procedures for follicle flushing and oocyte collection. Media provided in volumes of 60 ml and 125 ml.
All devices are aseptically filled into radiation-sterilized PETG bottles, are single-use only, and have a shelf-life of six-months when stored as recommended at 2-8°C. Additional information on the formulation of the subject devices is provided in Section 7 of this summary.
### INDICATIONS FOR USE: 6.
The subject device Indications for Use statements (IFU) are shown below:
| No. | Device Name | INDICATIONS FOR USE |
|-----|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | VitaVitro®<br>Fertilization Medium | VitaVitro® Fertilization Medium is intended for preparation and handling of<br>human gametes and for in vitro fertilization. |
| 2 | VitaVitro® Gamete<br>Buffer Medium | VitaVitro® Gamete Buffer Medium is intended for human gamete and<br>embryo short-term handling procedures outside the incubator, including<br>washing and intracytoplasmic sperm injection (ICSI). |
| 3 | VitaVitro® Flushing<br>Buffer Medium | VitaVitro® Flushing Buffer Medium is intended for use during ovarian<br>follicle flushing and oocyte collection procedures for use in in vitro<br>fertilization procedures. |
A comparison of the subject and predicate device intended use is shown below:
| Subject Device IFU statement | Predicate Device IFU Statement | Intended Use Comparison |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| VitaVitro® Fertilization Medium<br>is intended for preparation and<br>handling of human gametes and<br>for in vitro fertilization. | Global Total Fertilization with<br>HSA:<br><br>Oocyte culture and fertilization | The IFU statement wording is<br>different but the intended use is the<br>same i.e. (handling of human<br>gametes and in vitro fertilization). |
| VitaVitro® Gamete Buffer<br>Medium is intended for human<br>gamete and embryo short-term<br>handling procedures outside the<br>incubator, including washing and<br>intracytoplasmic sperm injection<br>(ICSI). | Quinn's HTF medium with HEPES<br>was developed for in vitro<br>procedures involving<br>manipulations of gametes and<br>embryos not requiring the use of a<br>CO2 incubator. Such procedures<br>include oocyte recovery, gamete<br>washing, micromanipulation,<br>embryo transfer and<br>cryopreservation. | The intended use of the subject and<br>predicate device is the same<br>regarding short-term handling<br>procedures of human gametes and<br>embryo outside of an incubator.<br>However, the predicate device has<br>additional uses beyond that of the<br>subject device (embryo transfer and<br>cryopreservation). These<br>differences do not represent a new<br>intended use, but rather a more<br>limited intended use for the subject<br>device. |
| VitaVitro® Flushing Buffer<br>Medium is intended for use<br>during ovarian follicle flushing<br>and oocyte collection procedures<br>for use in in vitro fertilization | Sydney IVF Follicle Flush Buffer<br>is intended for use during in vitro<br>fertilization procedures for follicle<br>flushing and oocyte collection. | The intended use of the predicate<br>and subject device is the same<br>(follicle flushing and oocyte<br>collection). |
{5}------------------------------------------------
procedures.
## 7. Substantial Equivalence Discussion
A comparison of the technological features of the subject and predicate device are described in the table below:
| Device &<br>Predicate<br>Device(s): | K200408<br>Vita Vitro® Fertilization<br>Medium, Vita VitroⓇ<br>Gamete Buffer<br>Medium, and<br>Vita Vitro® Flushing<br>Buffer Medium | K142991<br>Global Total<br>Fertilization<br>with HSA<br>(predicate for<br>Vita Vitro®<br>Fertilization<br>Medium) | K991395<br>Quinn's HTF<br>Medium with<br>HEPES<br>(predicate for<br>Vita VitroⓇ<br>Gamete Buffer<br>Medium) | K153290<br>Sydney IVF<br>follicle flush<br>buffer (predicate<br>for Vita Vitro®<br>Flushing Buffer<br>Medium) | Comments |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Formulation | Glucose<br>Sodium Lactate<br>Sodium Pyruvate<br>HEPES*<br>HEPES Sodium*<br>Sodium Bicarbonate<br>Sodium Chloride<br>Potassium Chloride<br>Potassium Phosphate<br>Magnesium Sulphate<br>Calcium Chloride<br>Amino acids<br>Taurine<br>Sodium Citrate<br>EDTA<br>Phenol red<br>Gentamicin sulfate<br>HSA** | Sodium Chloride<br>Calcium chloride<br>Potassium<br>Phosphorus<br>Magnesium<br>Sulfate<br>Sodium<br>Bicarbonate<br>Glucose<br>Sodium Lactate<br>Sodium Pyruvate<br>Amino acids<br>Glycine<br>glutamine<br>EDTA<br>Phenol red<br>Gentamicin<br>sulfate<br>HSA<br>Water | Not known | D-Glucose<br>Gentamicin<br>Calcium -Lactate<br>Glutamine<br>Glycine<br>L-alanine<br>L-asparagine<br>L-aspartic acid<br>L-glutamic acid<br>L-proline<br>L-serine<br>Magnesium<br>sulphate<br>Potassium chloride<br>Potassium<br>phosphate<br>Purified water<br>Sodium<br>bicarbonate<br>Sodium chloride<br>Sodium pyruvate<br>Taurine | Different: The materials<br>in the formulations of<br>the subject and their<br>respective predicate<br>devices are not identical<br>or is not known (i.e.,<br>predicate for VitaVitro®<br>Gamete Buffer<br>Medium). These<br>differences do not raise<br>different questions of<br>Safety and Effectiveness<br>(S&E). |
| | | | | | |
| Packaging | 30 and 60 ml PETG<br>square bottle –<br>Fertilization Medium<br>60 ml PETG square<br>bottle – Gamete Buffer<br>Medium<br>60 and 125 ml PETG<br>square bottle – Flushing<br>Buffer Medium | 50 and 100 ml | Not known | 20, 50, and 100 ml<br>borosilicate Type 1<br>glass vial with<br>Fluorotec coated<br>rubber stopper | Different: Bottle sizes<br>between the subject<br>devices and their<br>respective predicates are<br>not the same or are not<br>known (i.e., predicate<br>for VitaVitro® Gamete<br>Buffer Medium).<br>Differences in<br>packaging do not raise<br>different questions of<br>S&E. |
| Shelf-life | 6 months | 10 weeks | Not known | 20 weeks | Different: Shelf-life of<br>the subject devices and<br>their respective<br>predicates are not the<br>same or is not known<br>(i.e., predicate for<br>VitaVitro® Gamete<br>Buffer Medium).<br>Differences in shelf-life<br>do not raise different<br>questions of S&E. |
| pH | 7.2-7.6 | 7.2-7.4 | Not known | 7.37.50 | Different. The subject<br>devices pH range is<br>greater than their<br>respective predicate<br>devices (i.e., 7.6) or is<br>not known (i.e.,<br>predicate for VitaVitro®<br>Gamete Buffer<br>Medium). Differences<br>in pH do not raise<br>different questions of<br>S&E. |
| Osmolality | 260-290 | 260-270 | Not known | 285-295 | Different. The subject<br>devices osmolality<br>specifications are<br>different than their<br>respective predicate<br>devices or is not known |
| | | | | | |
| | | | | | (i.e., predicate for<br>VitaVitro® Gamete<br>Buffer Medium).<br>Differences in<br>osmolality do not raise<br>different questions of<br>S&E. |
| Endotoxin | <0.25 | <0.5 EU/ml | Not Known | <0.40 EU/Ml | Different: The<br>endotoxin specifications<br>for the subject devices is<br>lower than their<br>respective predicates or<br>is not known (i.e.,<br>predicate for VitaVitro®<br>Gamete Buffer<br>Medium). Differences<br>in endotoxin<br>specifications do not<br>raise different questions<br>of S & E. |
| MEA | 1-cell MEA: ≥80%<br>expanded blastocyst at 96<br>hours after a 24-hour<br>exposure to media<br>(Fertilization media).<br>1-cell MEA: ≥80%<br>expanded blastocyst at 96<br>hours after a 2-hour<br>exposure to media<br>(Gamete Buffer Medium<br>and Flushing Buffer<br>Medium). | 1-cell MEA≥80%<br>blastocysts at 96<br>h of culture | Not Known | >80% 2-Cell MEA | Different: The MEA<br>specifications for the<br>subject devices (e.g.,<br>embryo type, exposure<br>to test media, etc.) is<br>different than their<br>respective predicates or<br>is not known (i.e.,<br>predicate for VitaVitro®<br>Gamete Buffer<br>Medium). Differences in<br>MEA specifications do<br>not raise different<br>questions of S & E. |
| Sterility | No growth | SAL 10-3 | Not Known | Aseptically filtered | Different: The sterility<br>specification is different<br>than its respective<br>predicate devices or is<br>not known ((i.e.,<br>predicate for VitaVitro®<br>Gamete Buffer<br>Medium). This<br>difference does not raise<br>different questions of |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| | | | | | S&E. |
|--|--|--|--|--|------|
|--|--|--|--|--|------|
*not included in Fertilization Medium
**not included in Flushing Buffer Medium
The technological characteristics of the subject devices and their respective predicate devices are not the same as stated in the table above (e.g., formulation, packaging, shelf-life, device specifications, etc.). However, the differences in technological characteristics do not raise different questions of safety and effectiveness as compared to their respective predicate devices.
#### Summary of Non-Clinical Performance 8.
VitaVitro® Fertilization Medium, VitaVitro® Gamete Buffer Medium, and VitaVitro® Flushing Buffer Medium underwent the following testing to assess their performance and demonstrate substantial equivalence to their respective predicate devices. The testing and specifications below, unless otherwise stated, apply to each of the subject media products:
- . pH testing: 7.20 - 7.60
- Osmolality testing: 260 290 mOsm/kg .
- Endotoxin (per USP<85>): <0.25 EU/ml ●
- . MEA;
- I 1-cell MEA: ≥80% expanded blastocyst at 96 hours after a 24-hour exposure to media (Fertilization media).
- ı 1-cell MEA: ≥80% expanded blastocyst at 96 hours after a 2-hour exposure to media (Gamete Buffer Medium and Flushing Buffer Medium).
- . Sterility Testing (per USP <71>): No growth
- . Shelf-life testing was conducted at time 0 and after accelerated aging (equivalent to six months of real-time aging) to ensure the product specifications listed above were met.
- . Transportation testing per ASTM D4169-2016 (Standard Practice for Performance Testing of Shipping Containers and Systems)
- . Sterilization validation was conducted in accordance with ISO 13408-1: 2015 (Aseptic processing of health care products - Part 1: General requirements) and ISO 13408-2: 2018 (Aseptic processing of health care products -Part 2: Sterilizing filtration).
- . Biocompatibility testing was conducted on VitaVitro® Flushing Buffer Medium as this device will have direct contact with the patient during use. Testing was conducted in accordance with the 2020 FDA guidance document "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process." Testing included:
- I Cytotoxicity per ISO 10993-5: 2009
- . Sensitization per ISO 10993-10: 2010
- I Irritation per ISO 10993-10: 2010
The testing demonstrated that VitaVitro® Flushing Buffer Medium is non-sensitizing, and nonirritating.
#### 9. Conclusion
The results of the performance testing described above demonstrate that the subject media products are as safe and effective as their respective predicate devices and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.