K200404 · Boston Scientific Corporation · KNS · Dec 3, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K200404
Device Name
ORISE ProKnife
Applicant
Boston Scientific Corporation
Product Code
KNS · Gastroenterology, Urology
Decision Date
Dec 3, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ORISE™ ProKnife has been designed to be used with endoscopes and electrosurgical units to cut tissue within the gastrointestinal tract using high-frequency current. The electrosurgical knife also has the capability of delivering saline/ submucosal lifting agent into submucosal tissue layers under direct visualization through an endoscope.
Device Story
ORISE ProKnife is a sterile, single-use electrosurgical device for endoluminal surgery, including endoscopic submucosal dissection (ESD). Used with endoscopes and electrosurgical units; device delivers high-frequency monopolar energy for tissue incision. Integrated fluid channel allows delivery of saline or submucosal lifting agents directly into submucosal tissue layers without requiring device exchange. Operated by physicians in clinical settings under direct endoscopic visualization. Output is controlled tissue incision and submucosal cushion maintenance. Benefits include procedural efficiency by enabling cutting and injection via a single device.
Clinical Evidence
No clinical data. Evidence consists of bench-top and ex-vivo testing, electrical safety testing, and biocompatibility evaluation. Bench testing confirmed functional performance, including knife operation with compatible endoscopes, dimensional accuracy, durability, and injection capability. Electrical safety verified per IEC 60601-1, IEC 60601-2-2, and IEC 60601-2-18. Biocompatibility evaluated per ISO 10993-1 for limited contact (<24 hours) with breached/compromised body surfaces. All tests passed.
Technological Characteristics
Sterile, single-use electrosurgical knife. Features integrated fluid delivery channel for submucosal lifting agents. Compatible with standard endoscopes and high-frequency electrosurgical generators. Electrical safety compliant with IEC 60601-1, IEC 60601-2-2, and IEC 60601-2-18. Biocompatibility compliant with ISO 10993-1. Design parameters include specific tip/shaft diameters and power handling capabilities consistent with predicate devices.
Indications for Use
Indicated for use in endoluminal surgery procedures, including endoscopic submucosal dissection (ESD), for cutting gastrointestinal tissue and delivering submucosal lifting agents.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
Single Use Electrosurgical Knife (Dual Knife J) (K171158)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
December 3, 2020
Boston Scientific Corporation Jennifer Edouard Senior Regulatory Affairs Specialist 300 Boston Scientific Way Marlborough, Massachusetts 01752
Re: K200404
Trade/Device Name: ORISE ProKnife Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit And Accessories Regulatory Class: Class II Product Code: KNS Dated: October 9, 2020 Received: October 13, 2020
Dear Jennifer Edouard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200404
Device Name ORISE™ ProKnife
#### Indications for Use (Describe)
The ORISE™ ProKnife has been designed to be used with endoscopes and electrosurgical units to cut tissue within the gastrointestinal tract using high-frequency current. The electrosurgical knife also has the capability of delivering saline/ submucosal lifting agent into submucosal tissue layers under direct visualization through an endoscope.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div><span></span> X Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span></span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K200404 Page 1 of 3
# 510(k) Summary for ORISE™ ProKnife
#### 1. Submitter
Boston Scientific Corporation Endoscopy Division 200 Boston Scientific Way Marlborough, MA 01752
| Contact: | Jennifer Edouard<br>Sr. Regulatory Affairs Specialist<br>Phone: (508) 683-6134<br>E-mail: jennifer.edouard@bsci.com |
|----------|---------------------------------------------------------------------------------------------------------------------|
|----------|---------------------------------------------------------------------------------------------------------------------|
| <b>Secondary Contact:</b> | Karin Desjardins |
|---------------------------|----------------------------------|
| | Sr. Manager Regulatory Affairs |
| | Phone: (508) 683-5748 |
| | Email: karin.desjardins@bsci.com |
Date Prepared: October 9, 2020
## 2. Device
| Trade Name: | ORISE™ ProKnife |
|----------------------------|-------------------------------------------------|
| Classification Name: | Endoscopic Electrosurgical Unit and Accessories |
| Product Code: | KNS |
| Device Class and Panel: | Class II, Gastroenterology/Urology |
| Classification Regulation: | 21 CFR 876.4300 |
## 3. Predicate Devices
| Trade Name: | Single Use Electrosurgical Knife (Dual Knife J) |
|----------------------------|-------------------------------------------------|
| Manufacturer: | Olympus Medical Systems Corp. |
| Clearance Number: | K171158 |
| Classification Name: | Endoscopic Electrosurgical Unit and Accessories |
| Product Code: | KNS |
| Device Class and Panel: | Class II, Gastroenterology/Urology |
| Classification Regulation: | 21 CFR 876.4300 |
| | |
| Trade Name: | ERBE Hybrid Knife |
|----------------------------|------------------------------------------------------------|
| Manufacturer: | ERBE USA Incorporated |
| Clearance Number: | K083608 |
| Classification Name: | Jet Lavage |
| Product Code: | FQH – Lavage, Jet |
| | GEI - Electrosurgical, Cutting & Coagulation & Accessories |
| Device Class and Panel: | Class II, General & Plastic Surgery |
| Classification Regulation: | 21 CFR 880.5475 |
{4}------------------------------------------------
## 4. Proposed Device Description
The ORISE™ ProKnife is an electrosurgical knife intended for use in Endoluminal surgery procedures, including but not limited to Endoscopic Submucosal Dissection (ESD). The ProKnife can deliver high frequency monopolar energy to facilitate incision of tissue. The ORISE™ ProKnife also has the capability of delivering saline/submucosal lifting agent through the electrode to maintain a submucosal fluid cushion without the need for a device exchange.
## 5. Indications for Use
The ORISE™ ProKnife has been designed to be used with endoscopes and electrosurgical units to cut tissue within the gastrointestinal tract using high-frequency current. The electrosurgical knife also has the capability of delivering saline/submucosal lifting agent into submucosal tissue layers under direct visualization through an endoscope.
## 6. Technological Characteristics
The ORISE ProKnife, the Olympus Single Use Electrosurgical Knife, and the ERBE Hybrid Knife are all sterile, single use devices and are identical with regard to lifting agent delivery pathway, electrode length, working length, endoscope compatibility, HF Generator Compatibility, Generator Operating Frequency Range and Voltage Capability and Maximum Power.
The proposed and predicate devices share similar characteristics, including principle of operation, tip outer diameter, shaft outer diameter, generator connection. Voltage Capability and Minimum Power, and materials.
# 7. Substantial Equivalence
A direct comparison of key characteristics demonstrates that the ORISE™ ProKnife is substantially equivalent to the predicate devices in terms of intended use, technological characteristics, and performance characteristics.
### 8. Performance Data
Performance tests, which included bench-top and ex-vivo testing, electrical, and biocompatibility were conducted to demonstrate the performance and usability of the ORISE™ ProKnife in endoscopic procedures
### Bench Testing:
Functional and performance tests were performed on the proposed ORISE™ ProKnife to satisfy all design verification requirements. The ORISE™ ProKnife passed all tests.
The following performance tests were performed to satisfy design verification requirements.
- Knife operation with the compatible endoscopes
- o Dimension of each part of the knife
{5}------------------------------------------------
- . General durability
- . Injection capability
All results were passing and demonstrate that the proposed device meets its pre-defined performance specifications.
#### Electrical Testing:
Electrical safety was evaluated in accordance with the following standards:
- IEC 60601-1: 2005 + A1: 2012, Third Edition Medical Electrical Equipment o General Requirements for Basic Safety and Essential Performance
- . IEC 60601-2-2: 2017, Edition 6.0 – Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- IEC 60601-2-18:2009, Ed. 3.0 Medical electrical equipment Part 2: Particular . requirements for the basic safety and essential performance of endoscopic equipment
- IEC 60601-1-6:2010, AMD1:2013 Medical electrical equipment Part 1-6 ● General requirements for safety - Collateral Standard: Usability
#### Biocompatibility Testing:
Biocompatibility of the ORISE™ ProKnife was evaluated in accordance with ISO 10993-1: 2018 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process and FDA Guidance Use of International Standard ISO 10993-1 (June 16, 2016) for devices categorized as limited (<24 hours) surface, breached/compromised body contact.
The results of non-clinical testing demonstrate that the ORISE™ ProKnife is considered safe and effective for its intended use.
#### 9. Conclusion
All performance testing, which include device bench and ex-vivo tests, electrical, and biocompatibility testing results were acceptable. The data demonstrates that the proposed device sufficiently meets the design specifications and is suitable for the intended use.
Boston Scientific has demonstrated that the proposed ORISE™ ProKnife is substantially equivalent to the currently marketed predicate devices and can be safely and effectively used for its proposed indication.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.