K200400 · Tear Film Innovations, Inc. · ORZ · May 21, 2020 · Ophthalmic
Device Facts
Record ID
K200400
Device Name
Systane iLux2
Applicant
Tear Film Innovations, Inc.
Product Code
ORZ · Ophthalmic
Decision Date
May 21, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Systane iLux2 is indicated for the application of localized heat and pressure therapy in adult patients with Melbonian Gland Dysfunction (MGD), which is associated with evaporative dry eye, and to capture/store digital images and video of the meibomian glands.
Device Story
Systane iLux2 is a handheld, battery-powered ophthalmic device for treating Meibomian Gland Dysfunction (MGD) and performing meibography. It uses lime-green and IR LED-based heat (38-44°C) to melt meibum; Eye Care Professionals (ECP) manually apply compression via a control button to express glands. The device features a reusable handheld instrument with an integrated LCD touchscreen for real-time monitoring of eyelid temperature and gland visualization. A sterile, single-use 'Smart Tip' interfaces with the eyelid during treatment, while a reusable 'Photography Tip' facilitates image/video capture of the eyelid margin and glands using white or IR light. The device includes a docking station for battery charging and data transfer (USB/HDMI) to external monitors or PCs. ECPs use the visual feedback to guide treatment, potentially improving patient comfort and dry eye symptoms by restoring gland function.
Clinical Evidence
Bench testing only. Evaluation included biocompatibility, sterility, packaging integrity, transportation, optical radiation safety, electromagnetic compatibility, electrical safety, stability/shelf-life, performance/functionality, software validation, and simulated use (human factors). Results demonstrate the device meets safety and performance requirements for its intended use.
Technological Characteristics
Handheld, battery-powered instrument with OLED touchscreen. Heat source: lime-green and IR LEDs. Patient interface: sterile, single-use Smart Tip (biocompatible) and reusable Photography Tip. Connectivity: USB/HDMI for image transfer. Sterilization: Ethylene Oxide (SAL 10^-6) for Smart Tip. Control: Dedicated processor for power/heat management. Temperature accuracy: +/- 1.0°C. Target range: 38-44°C.
Indications for Use
Indicated for adult patients with Meibomian Gland Dysfunction (MGD) associated with evaporative dry eye; used for localized heat and pressure therapy and meibomian gland imaging.
Regulatory Classification
Identification
An eyelid thermal pulsation system is an electrically-powered device intended for use in the application of localized heat and pressure therapy to the eyelids. The device is used in adult patients with chronic cystic conditions of the eyelids, including meibomian gland dysfunction (MGD), also known as evaporative dry eye or lipid deficiency dry eye. The system consists of a component that is inserted around the eyelids and a component to control the application of heat and pressure to the eyelids.
Special Controls
In addition to the general controls of the Act, the LipiFlow® Thermal Pulsation System is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing should validate electromagnetic compatibility (EMC) and safety of exposure to non-ionizing radiation;
(2) Design, description, and performance data should validate safeguards related to the temperature and pressure aspects of the device, including during fault conditions;
(3) Performance data should demonstrate the sterility of patient-contacting components and the shelf-life of these components;
(4) The device should be demonstrated to be biocompatible; and
(5) Performance data should demonstrate that any technological changes do not adversely effect safety and effectiveness.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The seal features a stylized depiction of a human figure, while the text is in a clear, sans-serif font, with "FDA" in a blue square.
May 21, 2020
Tear Film Innovations, Inc. Neal Hartman Sr. Director, Regulatory Affairs/Quality Assurance 5924 Balfour Court, Suite 100 Carlsbad, CA 92008
Re: K200400
Trade/Device Name: Systane iLux2 Regulation Number: 21 CFR 886.5200 Regulation Name: Eyelid Thermal Pulsation System Regulatory Class: Class II Product Code: ORZ, HKI Dated: April 15, 2020 Received: April 16, 2020
Dear Neal Hartman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products
{1}------------------------------------------------
(see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K200400
Device Name Systane iLux2
Indications for Use (Describe)
The Systane iLux2 is indicated for the application of localized heat and pressure therapy in adult patients with Melbonian Gland Dysfunction (MGD), which is associated with evaporative dry eye, and to capture/store digital images and video of the meibomian glands.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Tearfilm Innovations. The logo features a stylized eye on the left, with the words "tearfilm" in blue and green to the right of the eye. Below "tearfilm" is the word "innovations" in blue.
# 510(K) SUMMA
### Submitter Information
| Company Name: | Tear Film Innovations, Inc. |
|--------------------|-----------------------------------------------------------------------------------------------|
| Company Address: | 5924 Balfour Court, Suite 100<br>Carlsbad, CA 92008 |
| Company Phone: | (760) 444-9555 |
| Company Facsimile: | (442) 244-4696 |
| Contact Person: | Neal Hartman<br>Sr. Director of Regulatory Affairs/Quality Assurance<br>nhartman@tearfilm.com |
| Date: | May 20, 2020 |
## Device Identification
| Device Trade Name: | Systane iLux² | |
|----------------------|------------------------------------|-----------------------------------|
| Common Name: | N/A | |
| Classification Names | Eyelid Thermal Pulsation<br>System | Camera, Ophthalmic,<br>Ac-Powered |
| Regulations | 886.5200 | 886.1120 |
| Device Class | 2 | 2 |
| Product Codes | ORZ | HKI |
| Advisory Panel | Ophthalmic | Ophthalmic |
#### ldentification of Predicate Devices
The System Device is substantially equivalent to the following device:
| Device Name | Classification Regulation | Product Code | 510(K) Number | Clearance Date |
|-------------|--------------------------------------------|--------------|---------------|----------------|
| iLux System | 886.8200 – Eyelid Thermal Pulsation System | ORZ | K172645 | 12/26/2017 |
#### Device Description
The Subject Device is indicated for the application of localized heat and pressure therapy in adult patients with Meibomian Gland Dysfunction (MGD), which is associated with evaporative dry eye. The device can capture images and videos of the eyelid margin and meibomian glands. These images can be use in meibography (i.e., visual morphology of meibomian glands).
{4}------------------------------------------------
In treatment mode, light-based heat is used to warm the eyelid tissue to a target range of 38-44 ℃ to melt the meibum blocking the gland orifices. The Eye Care Professionals (ECP) can apply compression to the eyelid to express the meibum through the orifices. At all times, the amount of the heat and pressure applied is under direct control of the ECP who monitors the response of the glands and the comfort of the patient. The inner eyelid temperature is displayed during the treatment process. The source of heating is lime-green and IR optical radiation produced by LEDs incorporated in the Instrument. Images and video can be captured during treatment.
In meibography mode, infrared (IR) light is used in the visualization of meibomian gland morphology. Images can be captured with IR or white light. Video is captured only with white light.
The Subject Device consists of the following:
- Reusable, Hand-Held Instrument Incorporates hardware, software, and . mechanics to facilitate treatment and imaging.
The Instrument incorporates an LCD touchscreen user interface that is used to view the eyelid margin as well as provides instructions to the ECP during use. Selections can be made through the user interface and images/video can be viewed.
The Instrument is powered by a rechargeable lithium ion battery.
- Docking Station Is used to charge the Instrument's internal battery and transfer . images/video to an external monitor or PC.
- . AC Power Supply - Transfer electricity for the AC electrical outlet to the Docking Station.
- . Photography Tip (Photo Tip) - Is reusable, up to a specified number of attachments, and when coupled with the Instrument it is used to capture images and video. There is electronics in the Photo Tip that interfaces with the Instrument. The Photo Tip is positioned on the cheek bone or eyebrow. The patient contacting material is biocompatible.
- . Smart Tip Patient Interface (Smart Tip) - Is sterile and single-use that when coupled with the Instrument treatment can be provided. There is electronics in the Smart Tip that interfaces with the Instrument. A feature of the Smart Tip is positioned behind the eyelid during the treatment process. The patient contacting material is biocompatible.
- USB & HMDI Cable Attaches Docking Station to external monitor/PC. .
#### Indications for Use
The Systane iLux2 is indicated for the application of localized heat and pressure therapy in adult patients with Meibomian Gland Dysfunction (MGD), which is associated with
{5}------------------------------------------------
evaporative dry eye, and to capture/store digital images and video of the meibomian glands
| Comparison<br>Feature | Subject Device | Predicate Device | Reference Device |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Systane iLux² | iLux System | LipiScan Dynamic<br>Meibomian Imager |
| Manufacturer | Tear Film Innovations, Inc. | Tear Film Innovations, Inc. | TearScience, Inc<br>(Johnson & Johnson) |
| Device<br>classification | 2 | 2 | 2 |
| Product Code | ORZ | ORZ | HKI |
| Indications for<br>Use | The Systane iLux² is indicated<br>for the application of localized<br>heat and pressure therapy in<br>adult patients with Meibomian<br>Gland Dysfunction (MGD),<br>which is associated with<br>evaporative dry eye, and to<br>capture/store digital images and<br>video of the meibomian glands | The iLux System is indicated<br>for the application of localized<br>heat and pressure therapy in<br>adult patients with chronic<br>disease of the eyelids,<br>including meibomian gland<br>dysfunction (MGD), also known<br>as evaporative dry eye. | LipiScan Dynamic<br>Meibomian Imager is an<br>ophthalmic imaging device<br>intended for use by a<br>physician in adult patients to<br>capture, archive, manipulate<br>and store digital images of<br>the meibomian glands. |
| System<br>components | • Reusable Handheld<br>Instrument<br>• Smart Tip Patient Interface<br>• Photography Tip<br>• Charging/Image Transfer<br>Station | • Reusable Handheld<br>Instrument<br>• Smart Tip Patient Interface<br>• Charging Station | • Computer system<br>• Electronics<br>• Chin and forehead rest<br>• Camera and attached<br>lens<br>• Illuminator<br>• Touchscreen display<br>• Handheld near-infrared<br>(IR) lid everter<br>• Power supply |
| Patient contact<br>system<br>component | • Smart Tip Patient Interface<br>• Photography Tip | • Smart Tip Patient Interface | • Chin and forehead rest<br>• Handheld near-infrared<br>(IR) lid everter |
| Single use | • Smart Tip Patient Interface | • Smart Tip Patient Interface | N/A |
| Sterile | • Smart Tip Patient Interface | • Smart Tip Patient Interface | N/A |
| Method of<br>sterilization,<br>SAL | Ethylene Oxide, SAL 10-6 | Ethylene Oxide, SAL 10-6 | N/A |
| Packaging,<br>Sterile Barrier | Thermoform Tray/Tyvek Lidding<br>Stock | Tyvek Pouch | N/A |
| System control<br>component | Reusable Handheld Instrument | Reusable Handheld Instrument | Window based PC |
| Software | Yes | Yes | Yes |
| Operating<br>orientation | Handheld | Handheld | Stationary |
| User Interface | • Touchscreen OLED Display<br>• Control Button<br>• Smart Tip Patient Interface | • Control Button<br>• Smart Tip Patient Interface | • Chin and forehead rest<br>• Touchscreen display |
| Comparison<br>Feature | Subject Device | Predicate Device | Reference Device |
| Device name | Systane iLux2 | iLux System | LipiScan Dynamic<br>Meibomian Imager |
| | • Photography Tip | | • Handheld near-infrared<br>(IR) lid everter |
| System's<br>component for<br>viewing glands | OLED Display | Magnifying Lens | Integrated Monitor |
| Tip Attachment<br>to Handheld<br>Instrument | Mechanical (identical to<br>Predicate) | Mechanical | N/A |
| Patient interface<br>during treatment | Smart Tip Patient Interface<br>(identical to Predicate) | Smart Tip Patient Interface | N/A |
| Treatment of<br>upper and lower<br>eyelids | Sequential | Sequential | N/A |
| Treatment Time | Typically, 8-12 minutes | Typically, 8-12 minutes | N/A |
| Safety-Related<br>Power/ Heat<br>Management | Dedicated Processor | Dedicated FPGA | N/A |
| Heat Source | LEDs (lime-green and IR<br>wavelengths) with reflective<br>shroud (identical to Predicate) | LEDs (lime-green and IR<br>wavelengths) with reflective<br>shroud | N/A |
| Rate of heating<br>(time to target<br>temperature) | 15 to 50 seconds | 15 to 50 seconds | N/A |
| Temperature,<br>accuracy | +/- 1.0° | +/- 1.0° | N/A |
| Target<br>temperature<br>range | 38 to 44 °C | 38 to 44 °C | N/A |
| Maximum Inner<br>Eyelid Surface<br>temperature<br>limit | 44 °C | 44 °C | N/A |
| Maximum Outer<br>Eyelid Surface<br>temperature<br>limit | 45 °C | 45 °C | N/A |
| Compression<br>Mechanism | Mechanical (identical to<br>Predicate) | Mechanical | N/A |
| Pressure<br>Control | Manual via control button<br>(identical to Predicate): Eye<br>Care Professional determines<br>(based on patient feedback and<br>direct viewing of glands) | Manual via control button: Eye<br>Care Professional determines<br>(based on patient feedback<br>and direct viewing of glands) | N/A |
| Pressure type | Compression, repeated as<br>necessary | Compression, repeated as<br>necessary | N/A |
| Patient Interface<br>during Imaging | Photography Tip | N/A | • Chin and forehead rest<br>• Handheld near-infrared<br>(IR) lid everter |
| Imaging<br>Controls | Touchscreen display | N/A | Touchscreen display |
| Comparison<br>Feature | Subject Device | Predicate Device | Reference Device |
| Device name | Systane iLux2 | iLux System | LipiScan Dynamic<br>Meibomian Imager |
| Illumination,<br>Imaging | White light or infrared (IR) | No | Infrared (IR) |
| Ability to<br>capture images | Yes, Photography Tip use to<br>maintain focal point | No | Yes |
| Meibography<br>imaging | Yes, Photography Tip use to<br>maintain focal point | No | Yes |
| Ability to<br>capture video | Yes, Photography Tip use to<br>maintain focal point | No | No |
| Resolution,<br>Onboard | 240 x 170 (Portrait)<br>320 x 240 (Landscape) | N/A | Unknown |
| Resolution,<br>External | 1080p (Video)<br>2304 x 1536 (Images) | N/A | Unknown |
| Image Transfer<br>Method | Docking station-to-HDMI cable-<br>to-external monitor<br>Docking station-to-USB cable-<br>to-PC<br>SD card-to-PC | N/A | HDMI cable to monitor<br>USB 2.0 |
| Power Source/<br>Charging | Batteries, DC/ Charging/Image<br>Transfer station via AC power<br>adapter | Batteries, DC/ charging station<br>via AC power adapter | AC |
| Installation<br>required | No | No | No |
Comparison of Technological Characteristics with Predicate and Reference Devices
{6}------------------------------------------------
{7}------------------------------------------------
## Summary of Testing Performed
A program of design verification and validation testing was performed that includes the following:
- Biocompatibility ●
- Sterility and EO Residual
- Packaging Integrity (i.e., Sterile Barrier)
- Transportation
- Optical Radiation Safety ●
- Electromagnetic Compatibility and Electrical Safety
- Stability/Shelf-Life
- Performance/Functionality/Safety ●
- Software .
- Simulated Use (Human Factors Evaluation) ●
Results of the evaluations demonstrate that the Subject Device met the safety and performance requirements as it relates to its indication for use.
#### Conclusions Drawn from Nonclinical Evaluation
The results of the evaluation demonstrate that the Subject Device is substantially equivalent to the Predicate and Reference Devices as it pertains to the indications for use and device performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.