K200397 · Olympus Medical Systems Corp. · EOQ · Apr 24, 2020 · Ear, Nose, Throat
Device Facts
Record ID
K200397
Device Name
Single Use Biopsy Forceps FB-456D
Applicant
Olympus Medical Systems Corp.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Apr 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
The biopsy forceps has been designed specifically to collect tissue endoscopically for examination in conjunction with a flexible Olympus bronchoscope.
Device Story
Single-use biopsy forceps for endoscopic tissue collection; used with flexible Olympus bronchoscope. Device consists of handle (body/slider) and insertion portion (distal cups, operation wire, coil sheath). Physician inserts device into bronchoscope channel; advances/retracts slider to open/close distal cups to bite lesion surface; withdraws device to retrieve biopsy sample. Sterilized by ethylene oxide. Provides mechanical tissue sampling capability for diagnostic examination.
Clinical Evidence
Bench testing only. Performance verified via compatibility with endoscope, cup operation, repeated operation testing, insertion diameter measurement, visual inspection, and mechanical strength testing of junctions and cups. Biocompatibility, sterilization (ISO 11135), and shelf-life (ASTM F1980-16) validated.
Technological Characteristics
Single-use, ethylene oxide sterilized biopsy forceps. Components: handle (body/slider), insertion portion (cups, operation wire, coil sheath). Standards: ISO 14971 (risk management), ISO 10993-1/5/7/10/11 (biocompatibility), ISO 11135 (sterilization), ISO 11607-1/2 (packaging), ASTM F1980-16 (accelerated aging). Mechanical actuation via slider-wire mechanism.
Indications for Use
Indicated for endoscopic tissue collection for examination using a flexible Olympus bronchoscope.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
April 24, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Olympus Medical Systems Corp. % Jonathan Gilbert Regulatory Affairs Consultant to OCA Olympus Corporation of the Americas 3500 Corporation Parkway PO Box 610 Center Valley, Pennsylvania 18034-0610
Re: K200397
Trade/Device Name: Single Use Biopsy Forceps FB-456D Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: February 15, 2020 Received: February 18, 2020
Dear Jonathan Gilbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael J. Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K200397
Device Name
Single Use Biopsy Forceps FB-456D
Indications for Use (Describe)
The biopsy forceps has been designed specifically to collect tissue endoscopically for examination with a flexible Olympus bronchoscope.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly shadowed, giving them a three-dimensional appearance. A thin, light-yellow line is visible beneath the word, adding a subtle highlight.
# K200397 510(k) Summary Single Use Biopsy Forceps FB-456D
April 23, 2020
# 1. General Information
| ■ 510(k) submitter: | OLYMPUS MEDICAL SYSTEMS CORP. |
|--------------------------|---------------------------------------------------|
| | 2951 Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, |
| | Japan |
| | Establishment Registration No: 8010047 |
| ■ Contact Person: | Jon Gilbert fbo Sheri L. Musgnung |
| | Olympus Corporation of the Americas |
| | 3500 Corporate Parkway PO Box 610 |
| | Center Valley, PA 18034-0610, USA |
| | Phone: 484-896-5691 |
| | FAX: 484-896-7128 |
| | Email: sheri.musgnung@olympus.com |
| ■ Manufacturing site: | Aomori Olympus Co., Ltd. |
| | 2-248-1 Okkonoki, Kuroishi-shi, Aomori, 036-0357 |
| | Japan |
| | Establishment Registration No.: 9614691 |
| 2. Device Identification | |
| ■ Device Trade Name: | Single Use Biopsy Forceps FB-456D |
| ■ Common Name: | Single Use Biopsy Forceps |
| ■ Regulation Number: | 874.4680 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced. A registered trademark symbol is present to the right of the word.
- Bronchoscope (flexible or rigid) and accessories ■ Regulation Name:
- Regulatory Class: II
- Ear Nose & Throat ■ Classification Panel:
- Product Code: EOQ
### 3. Predicate Device Information
Predicate device
- Device Name: Single Use Biopsy Forceps FB-433D ■ Common Name: Biopsy Forceps ■ Applicant OLYMPUS MEDICAL SYSTEMS CORP. ■ 510(k) No. K172726
{5}------------------------------------------------
## 4. Device Description
The biopsy forceps has been designed specifically to collect tissue endoscopically for examination in conjunction with a flexible Olympus bronchoscope.
The subject device is single-use biopsy forceps sterilized by Ethylene Oxide.
The subject device is inserted into the channel of a bronchoscope to collect tissue by biting the surface of lesions with the pair of cups affixed to the distal end of the subject device. Users withdraw the subject device from the bronchoscope channel to collect the biopsy sample.
The subject device consists of the Handle and the Insertion portion. The Handle consists of the Body and the Slider. The Insertion portion is divided into the Distal end including the Cups as well as other connection parts, and the Sheath portion including the Operation wire, the Coil sheath and other joint pipes. The Handle Slider is designed to connect to the Operation wire in the Insertion and the Cups, actuating the Cups to open/close by advancing and retracting the Slider.
### 5. Indications for Use
The biopsy forceps has been designed specifically to collect tissue endoscopically for examination in conjunction with a flexible Olympus bronchoscope.
### 6. Comparison of Technological Characteristics
In comparison to the predicate device, the subject device, Single Use Biopsy Forceps FB-456D, has similar technological characteristics except for the following differences.
- Cup shape -
- Maximum insertion portion diameter -
- -Bronchoscope compatibility
- Configuration of handle -
- Patient-contact material -
The differences above have been validated and demonstrate that these technological features do not raise new issues in terms of the claims of substantial equivalence to the subject device.
{6}------------------------------------------------
### 7. Summary of non-clinical testing
The technological differences between the predicate device and the subject device have been verified and validated by means of the following tests and standards to endorse the claims of substantial equivalence to the predicate device.
Risk analysis was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
Biocompatibility testing for Single Use Biopsy Forceps FB-456D were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
Sterilization and shelf-life testing for the Single Use Biopsy Forceps FB-456D was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Sterile". Sterilization validation was carried out with the method of Half-cycle approach in accordance with ISO 11135:2014. The shelf-life for three years has been validated by accelerated testing according to ASTM F1980-16. Three-year aging analyses will be performed to support the results of the accelerated aging test.
The requirements on packaging for terminally sterilized medical device per AAMI/ ANSI/ISO 11607-1/2 have also been validated. Performance testing as below was carried out to demonstrate the safety and the effectiveness of the subject device.
- 1. Compatibility with the endoscope
- 2. Open and close of the cups
- 3. Performance after repeated operation
- 4. Maximum insertion portion diameter
- 5. Visual inspection of the insertion portion
- 6. Strength of the junction and cups
- 7. Strength when opening/closing the cups
- 8. Package inspection
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly blurred, and there is a registered trademark symbol to the right of the letter "S". There is a blurred yellow line underneath the word.
The following standards have been applied to the subject device.
| Standard No. | Standard Title |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO11135 Second Edition<br>2014 | Sterilization of Health-Care Products, Ethylene Oxide -<br>Requirements for the Development, Validation and<br>Routine Control of a Sterilization Process for Medical<br>Devices |
| ISO 10993-7 Second<br>edition 2008-10-15 | Biological evaluation of medical devices - part 7: ethylene<br>oxide sterilization residuals |
| ISO 11607-1 First edition<br>2006-04-15 | Packaging for terminally sterilized medical devices - part<br>1: requirements for materials, sterile barrier systems and<br>packaging systems |
| ISO 11607-2 First edition<br>2006-04-15 | Packaging for terminally sterilized medical devices - part<br>2: validation requirements for forming, sealing and<br>assembly processes |
| ASTM F1980-16 | Standard Guide For Accelerated Aging Of Sterile Barrier<br>Systems For Medical Devices |
| ISO 10993-1 Fourth<br>edition 2009-10-15 | Biological evaluation of medical devices - part 1:<br>evaluation and testing within a risk management process |
| ISO 10993-5 Third edition<br>2009-06-01 | Biological evaluation of medical devices - part 5: tests for<br>in vitro cytotoxicity |
| ISO 10993-10 Third<br>Edition 2010-08-01 | Biological evaluation of medical devices - part 10: tests<br>for irritation and skin sensitization |
| ISO 10993-11 Third<br>edition 2017-09 | Biological evaluation of medical devices - part 11: tests<br>for systemic toxicity |
| ISO 14971 Second edition<br>2007-03-01 | Medical devices - application of risk management to<br>medical devices |
# 8. Conclusion
In comparison to the predicate device, the subject Single Use Biopsy Forceps FB-456D device does not raise different questions of safety and effectiveness and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.